Contrast/Bolus Agent
|
(0018,0010)
|
2
|
Contrast or bolus agent
|
Contrast/Bolus Agent Sequence
|
(0018,0012)
|
3
|
Sequence that identifies the contrast agent.
One or more Items are permitted in this Sequence.
|
>Include Table 8.8-1 “Code Sequence Macro Attributes”
|
BCID 12 “Radiographic Contrast Agent”.
|
Contrast/Bolus Route
|
(0018,1040)
|
3
|
Administration route of contrast agent
|
Contrast/Bolus Administration Route Sequence
|
(0018,0014)
|
3
|
Sequence that identifies the route of administration of contrast agent.
Only a single Item is permitted in this Sequence.
|
>Include Table 8.8-1 “Code Sequence Macro Attributes”
|
BCID 11 “Route of Administration”.
|
>Additional Drug Sequence
|
(0018,002A)
|
3
|
Sequence that identifies any additional drug that is administered with the contrast agent bolus.
One or more Items are permitted in this Sequence.
|
>>Include Table 8.8-1 “Code Sequence Macro Attributes”
|
No Baseline CID is defined.
|
Contrast/Bolus Volume
|
(0018,1041)
|
3
|
Volume injected in milliliters of diluted contrast agent
|
Contrast/Bolus Start Time
|
(0018,1042)
|
3
|
Time of start of injection
|
Contrast/Bolus Stop Time
|
(0018,1043)
|
3
|
Time of end of contrast injection
|
Contrast/Bolus Total Dose
|
(0018,1044)
|
3
|
Total amount in milliliters of the undiluted contrast agent
|
Contrast Flow Rate
|
(0018,1046)
|
3
|
Rate(s) of injection(s) in milliliters/sec
|
Contrast Flow Duration
|
(0018,1047)
|
3
|
Duration(s) of injection(s) in seconds. Each Contrast Flow Duration value shall correspond to a value of Contrast Flow Rate (0018,1046).
|
Contrast/Bolus Ingredient
|
(0018,1048)
|
3
|
Active ingredient of agent.
Defined Terms:
-
IODINE
-
-
GADOLINIUM
-
-
CARBON DIOXIDE
-
-
BARIUM
-
|
Contrast/Bolus Ingredient Concentration
|
(0018,1049)
|
3
|
Milligrams of active ingredient per milliliter of (diluted) agent
|