| DICOM PS3.3 2026d - Information Object Definitions |
|---|
Table C.7-18 specifies the Attributes of the Device Module, which identifies and describes devices or calibration objects (e.g., catheters, markers, baskets) or other quality control materials that are associated with a Study and/or image.
Table C.7-18. Device Module Attributes
|
A Sequence of Items describing devices used that may be visible in the image. |
|||
|
The date the device was originally manufactured or re-manufactured (as opposed to refurbished). |
|||
|
Length in mm of device. See Section C.7.6.12.1.1. |
|||
|
Unit diameter of device. See Section C.7.6.12.1.1. |
|||
|
Volume of device in ml. See Section C.7.6.12.1.1. |
|||
|
Distance in mm between markers on calibrated device. See Section C.7.6.12.1.1. |
|||
|
Further description in free form text describing the device. |
|||
Depending on the type of device specified by Code Value (0008,0100) in an Item of Device Sequence (0050,0010), various device size Attributes (e.g., Device Length (0050,0014), Device Diameter (0050,0016), Device Volume (0050,0018), Inter Marker Distance (0050,0019)) may be required to fully characterize the device.
| DICOM PS3.3 2026d - Information Object Definitions |
|---|