| DICOM PS3.3 2026d - Information Object Definitions |
|---|
Table C.4-13 specifies the Attributes of the Performed Procedure Step Relationship Module, which are used to reference other SOP Classes and other Information Entities of the DICOM real-world model as defined in Section 7.3.1.6.
Table C.4-13. Performed Procedure Step Relationship Module Attributes
|
Primary identifier for the Patient. NoteIn the case of imaging a group of small animals simultaneously, the single Value of this identifier corresponds to the identification of the entire group. See also Section C.7.1.4.1.1. |
||
|
Date of birth of the named Patient in the Alternative Calendar (0010,0035). |
||
|
Date of death of the named Patient in the Alternative Calendar (0010,0035). |
||
|
The Alternative Calendar used for Patient's Birth Date in Alternative Calendar (0010,0033) and Patient's Death Date in Alternative Calendar (0010,0034). See Section C.7.1.5 for Defined Terms. Required if either Patient's Birth Date in Alternative Calendar (0010,0033) or Patient's Alternative Death Date in Calendar (0010,0034) is present. |
||
|
See Note 2 and Note 3 in Table C.7-1. |
||
|
An individual's personal sense of being a man, woman, boy, girl, nonbinary, or something else, ascertained by asking them what their identity is. One or more Items are permitted in this Sequence. See Section C.7.2.2.1.5. |
||
|
Only a single Item shall be included in this Sequence. See also Section C.7.2.2.1.4. |
||
|
The date and time at which the content of this Sequence Item begins to be applicable. See Section C.7.2.2.1.5. |
||
|
The date and time at which the content of this Sequence Item ceases to be applicable. See Section C.7.2.2.1.5. |
||
|
Comments on this gender identity, such as the context in which it should be used. |
||
|
Guidance on how to apply settings or reference ranges that are derived from observable information such as an organ inventory, recent hormone lab tests, genetic testing, menstrual status, obstetric history, etc. One or more Items are permitted in this Sequence. See Section C.7.2.2.1.2. |
||
|
The date and time at which the content of this Sequence Item begins to be applicable. See Section C.7.2.2.1.5. |
||
|
The date and time at which the content of this Sequence Item ceases to be applicable. See Section C.7.2.2.1.5. |
||
|
Further description of clinical implications and reasons for the selected code |
||
|
Reference to a resource that explains or extends the Sex Parameters for Clinical Use Category Code |
||
|
The name(s) that should be used when addressing or referencing the person. |
||
|
A name that should be used when addressing or referencing the person. This need not be an official name nor comply with any particular name structure. See Section C.7.2.2.1.3. |
||
|
The date and time at which the content of this Sequence Item begins to be applicable. See Section C.7.2.2.1.5. |
||
|
The date and time at which the content of this Sequence Item ceases to be applicable. See Section C.7.2.2.1.5. |
||
|
Pronoun(s) specified by the patient to use when referring to the patient in speech, in clinical notes, and in written instructions to caregivers. One or more Items are permitted in this Sequence. See Section C.7.2.2.1.4. |
||
|
A single code that specifies the set of third person pronouns to be used in reference to this patient. |
||
|
The date and time at which the content of this Sequence Item begins to be applicable. See Section C.7.2.2.1.5. |
||
|
The date and time at which the content of this Sequence Item ceases to be applicable. See Section C.7.2.2.1.5. |
||
|
>Include Table 10-11 “SOP Instance Reference Macro Attributes” |
||
|
Identification number of the Visit as assigned by the healthcare provider. |
||
|
Identifier of the Assigning Authority that issued the Admission ID. |
||
|
>Include Table 10-17 “HL7v2 Hierarchic Designator Macro Attributes” |
||
|
Identifier of the Service Episode as assigned by the healthcare provider. |
||
|
Identifier of the Assigning Authority that issued the Service Episode ID. |
||
|
>Include Table 10-17 “HL7v2 Hierarchic Designator Macro Attributes” |
||
|
Sequence containing Attributes that are related to the scheduling of the Procedure Step. |
||
|
Uniquely identifies the Study associated with this Scheduled Procedure Step. Only a single Item shall be included in this Sequence. See Section 10.6.1. |
||
|
>>Include Table 10-11 “SOP Instance Reference Macro Attributes” |
||
|
A departmental Information System generated number that identifies the Imaging Service Request. |
||
|
Identifier of the Assigning Authority that issued the Accession Number. |
||
|
>>Include Table 10-17 “HL7v2 Hierarchic Designator Macro Attributes” |
||
|
The order number assigned to the Imaging Service Request by the party placing the order. |
||
|
Identifier of the Assigning Authority that issued the Placer Order Number. |
||
|
>>Include Table 10-17 “HL7v2 Hierarchic Designator Macro Attributes” |
||
|
The order number assigned to the Imaging Service Request by the party filling the order. |
||
|
Identifier of the Assigning Authority that issued the Filler Order Number. |
||
|
>>Include Table 10-17 “HL7v2 Hierarchic Designator Macro Attributes” |
||
|
Institution-generated administrative description or classification of Requested Procedure. |
||
|
A Sequence that conveys the Procedure Type of the requested procedure. |
||
|
Institution-generated description or classification of the Scheduled Procedure Step to be performed. |
||
|
Sequence describing the Scheduled Protocol following a specific Coding Scheme. |
||
|
Sequence that specifies the context for the Scheduled Protocol Code Sequence (0040,0008) Item. One or more Items shall be included in this Sequence. See Section C.4.10.1. |
||
|
>>>Include Table 10-2 “Content Item Macro Attributes” |
||
|
Sequence that specifies modifiers for a Protocol Context Content Item. One or more Items shall be included in this Sequence. See Section C.4.10.1. |
||
|
>>>>Include Table 10-2 “Content Item Macro Attributes” |
||
The Patient information is included in order to relate the Modality Performed Procedure Step SOP Instance to the Study Management SOP Instance and other associated IODs in case the SCU (the modality) is unable to obtain or use the Study Instance UID created by the Information System.
Placer Order Number / Imaging Service Request (0040,2016) and Filler Order Number / Imaging Service Request (0040,2017) are intended to convey the corresponding order numbers as defined in HL7, in the case where interoperability with an HL7 environment is the objective.
Placer Order Number / Imaging Service Request (Retired) (0040,2006) and Filler Order Number / Imaging Service Request (Retired) (0040,2007) were previously defined in DICOM. They have been retired. See PS3.3-1998.
Placer Order Number / Procedure (0040,1006) and Filler Order Number / Procedure (0040,1007) were previously defined in DICOM. They have been retired. See PS3.3-1998.