| DICOM PS3.3 2026c - Information Object Definitions |
|---|
This Module contains Attributes describing presentation montages of waveform channels.
Table C.39.6-1. Waveform Presentation Montage Module Attributes
|
Description of the waveform montage(s) in the Waveform Presentation State. |
|||
|
The index of the montage within this Sequence. The Value shall start at 1 and increase monotonically by 1. This index will be used elsewhere to refer to this specific montage Sequence Item. |
|||
|
The channel(s) that comprise this montage. |
|||
|
The recommended time-based waveform data display scale in units of mm/s (see Section C.10.9.1.8). |
|||
|
A color triplet value recommended for rendering the waveform display background on a color display. The units are specified in PCS-Values, and the value is encoded as CIELab. See Section C.10.7.1.1. |
|||
|
Sequence of Items, each Item describing a Presentation Group of one or more waveform channels to be displayed together. |
|||
|
Sequence of Items, each Item describing a channel to be displayed in the Presentation Group. |
|||
|
Number of the montage channel to be displayed in the Presentation Group. This is the ordinal number of the Item in the Montage Channel Sequence (0040,B03C). |
|||
|
The offset in seconds from the beginning of the montage channel waveform data to the first sample to be used for presentation. Value may be negative. |
|||
|
A color triplet value recommended for rendering the channel on a color display. The units are specified in PCS-Values, and the value is encoded as CIELab. See Section C.10.7.1.1. |
|||
|
Position of the channel within the Presentation Group display area (see Section C.10.9.1.9). |
|||
|
Specifies display area shading between the displayed waveform channel and another line. The nature of the shading (e.g., solid, or cross-hatching) is implementation dependent. Enumerated Values:
|
|||
|
Fraction of the Presentation Group vertical display dimension assigned to the unit quantity (least significant bit) of the channel samples (see Section C.10.9.1.10). Required if Absolute Channel Display Scale (003A,0248) is not present, may be present otherwise. |
|||
|
Nominal vertical display height in mm assigned to the unit quantity (least significant bit) of the channel samples (see Section C.10.9.1.10). Required if Fractional Channel Display Scale (003A,0247) is not present, may be present otherwise. |
|||
This Macro consists of Attributes describing a single channel of a waveform montage.
Table C.39.7-1. Montage Channel Macro Attributes
|
Text label of the channel, which may be used for display purposes. |
|||
|
A coded descriptor of the waveform channel source. This identifies a single channel in the recorded waveform in terms of the lead from which it is collected. |
|||
|
DCID 3004 “Arterial Pulse Waveform”. DCID 3005 “Respiration Waveform”. DCID 3031 “Lead Location Near or in Muscle”. |
|||
|
A Sequence that provides reference to a waveform from which this channel was derived. One or more Items shall be included in this Sequence. If there are multiple Items in this Sequence, they shall share the same multiplex group identified by Multiplex Group UID (003A,0310). |
|||
|
>Include Table 10-11 “SOP Instance Reference Macro Attributes” |
|||
|
Identifies the waveform multiplex group (M) and channel (C) within the referenced SOP Instance using a pair of values (M,C). See Section C.10.10.1.1. |
|||
|
A Sequence of Items each representing the source of a channel contributing to this montage. |
|||
|
The relative weight this channel contributes to the montage channel. The weights of all Items in this Sequence shall sum up to 1. |
|||
|
A coded descriptor of the contributing waveform channel source. |
|||
|
DCID 3004 “Arterial Pulse Waveform”. DCID 3005 “Respiration Waveform”. DCID 3031 “Lead Location Near or in Muscle”. |
|||
|
Reference to waveforms from which this channel was derived. One or more Items shall be included in this Sequence. If there are multiple Items in this Sequence, they shall share the same multiplex group identified by Multiplex Group UID (003A,0310). |
|||
|
>>Include Table 10-11 “SOP Instance Reference Macro Attributes” |
|||
|
Identifies the waveform multiplex group (M) and channel (C) within the referenced SOP Instance using a pair of values (M,C). See Section C.10.10.1.1. |
|||
|
Nominal numeric value of unit quantity of sample. See Section C.10.9.1.4.2. Required if samples represent defined (not arbitrary) units. |
|||
|
A coded descriptor of the units of measure for the Channel Sensitivity (003A,0210). See Section C.10.9.1.4.2. |
|||
|
Multiplier to be applied to encoded sample values to match units specified in Channel Sensitivity Units Sequence (003A,0211) (e.g., based on calibration data). See Section C.10.9.1.4.2. |
|||
|
The properties of low frequency (high-pass) filters applied to the waveform montage channel. |
|||
|
>Include Table C.10.12-1 “Waveform Filter Characteristics Macro Attributes” |
|||
|
The properties of high frequency (low-pass) filters applied to the waveform montage channel. |
|||
|
>Include Table C.10.12-1 “Waveform Filter Characteristics Macro Attributes” |
|||
|
The properties of notch filters applied to the waveform montage channel. |
|||
|
>Include Table C.10.12-1 “Waveform Filter Characteristics Macro Attributes” |
|||
| DICOM PS3.3 2026c - Information Object Definitions |
|---|