RT Treatment Phases define the intended phases of treatment and their temporal relationship.
Table C.36.7-1. RT Treatment Phase Intent Module Attributes
|
RT Treatment Phase definitions. RT Treatment Phases define the relationships between RT Radiation Set Instances which contain the treatment parameters for the radiation to be concurrently and/or subsequently delivered. The RT Treatment Phase Index (3010,003A) shall define the temporal sequencing of the phases. |
|||
|
>Include Table C.36.2.1.2-1 “RT Treatment Phase Macro Attributes” |
The RT Treatment Phase Index (3010,003A) shall start at 1 and increase monotonically by 1 for successive Items in this Sequence. |
||
|
Include Table C.36.2.1.3-1 “RT Treatment Phase Interval Macro Attributes” |
|||