DICOM PS3.3 2026d - Information Object Definitions
C.36.2.4.6 KV Radiation Image Acquisition Parameters Macro

The KV Radiation Image Acquisition Parameters Macro contains parameters specifying details of X-Ray generation for image acquisition using kV-level radiation.

Table C.36.2.4.6-1. KV Radiation Image Acquisition Parameters Macro Attributes

Attribute Name

Tag

Type

Attribute Description

Energy Derivation Code Sequence

(3002,0133)

1C

How the value of imaging energy is derived from the device configuration.

Required if KVP (0018,0060) is not present.

Only a single Item shall be included in this Sequence.

See Section C.36.2.4.6.1.1.

>Include Table 8.8-1 “Code Sequence Macro Attributes”

BCID 9262 “Energy Derivation Types”.

KVP

(0018,0060)

2C

Peak kilo voltage of the X-Ray generator used to acquire the image.

Required if Imaging Energy Derivation Code Sequence (3002,0133) is not present.

See Section C.36.2.4.6.1.1.

X-Ray Tube Current in µA

(0018,8151)

3

X-Ray Tube Current in µA.

Exposure Time in µS

(0018,8150)

3

Duration of X-Ray exposure in µsec.

Average Pulse Width

(0018,1154)

3

Average width of X-Ray pulse in msec.

Radiation Mode

(0018,115A)

3

Specifies X-Ray radiation mode.

Enumerated Values:

CONTINUOUS

PULSED

X-Ray Filter Sequence

(0018,9556)

3

Image filter inserted into the X-Ray beam.

One or more Items are permitted in this Sequence.

>Include Table 10.36-1 “Device Identification Macro Attributes”

BCID 10007 “X-Ray Filter Type”.


C.36.2.4.6.1 KV Radiation Image Acquisition Parameters Macro Attribute Descriptions
C.36.2.4.6.1.1 Imaging Energy Category Code Sequence

The Imaging Energy can be described in either Energy Derivation Code Sequence (3002,0133) or KVP (0018,0060).

When KVP (0018,0060) is present and has no Value, no Imaging Energy is described.

DICOM PS3.3 2026d - Information Object Definitions