| DICOM PS3.3 2026c - Information Object Definitions |
|---|
The RT Image Frame Context Macro contains information about the context of a Frame constructed for use in, or acquired during, Radiotherapy treatment sessions.
Table C.36.2.4.11-1. RT Image Frame Context Macro Attributes
|
The RT Radiation Instances or the Treatment Position Groups for which the Frame is reconstructed or acquired. |
|||
|
>>Include Table C.36.2.3.3-1 “RT Patient Position Scope With Legacy Support Macro Attributes” |
|||
|
Identification of the RT Radiation Set Deliveryof the referenced RT Radiation Set Instance that this Frame refers to. Required if the Frame is related to a specific Fraction. See Section C.36.20.1.2. |
|||
|
Identification of the RT Treatment Fraction of the referenced RT Radiation Set Instance that this Frame refers to. Required if the Frame is related to a specific Fraction. See Section C.36.20.1.2. |
|||
| DICOM PS3.3 2026c - Information Object Definitions |
|---|