DICOM PS3.3 2024d - Information Object Definitions |
---|
The Patient Position Acquisition Device Macro contains all parameters that describe a device to acquire artifacts to detect the patient position.
Table C.36.2.2.22-1. Patient Position Acquisition Device Macro Attributes
Number of Acquisition Devices defined in the Acquisition Device Sequence (3002,0117). |
|||
Devices used to generate images. The devices may be physical imaging devices that performed acquisition or virtual imaging devices for which acquisition was simulated. Required if the Number of Acquisition Devices (3002,0116) has a non-zero value. The number of Items included in this Sequence shall equal the value of Number of Acquisition Devices (3002,0116). |
|||
>Include Table C.36.2.2.3-1 “RT Accessory Device Identification Macro Attributes” |
|||
Device Index value that links the device defined by this Sequence Item to the corresponding device in another SOP Radiation Instance. The description of the two devices may or may not be the same. The value is the index of a device in the Acquisition Device Sequence (3002,0117) within the single SOP Instance referenced by a SOP Instance Sequence defined in the Macro invocation. Required if the Instance referenced in the SOP Instance Sequence contains the device that corresponds to the device defined by this Sequence Item. |
DICOM PS3.3 2024d - Information Object Definitions |
---|