| DICOM PS3.3 2026d - Information Object Definitions |
|---|
This Macro defines the opening created by a RT Beam Limiting Device.
Table C.36.2.2.20-1. RT Beam Limiting Device Opening Definition Macro Attributes
|
The Value of Device Index (3010,0039) from the RT Beam Limiting Device Definition Sequence (300A,064D) corresponding to the Beam Limiting Device used in this Item. |
|||
|
The offsets (x,y) in mm of the Parallel RT Beam Delimiter Positions (300A,064A) along the x-axis and the Parallel RT Beam Delimiter Boundaries (300A,0649) along the y-axis of the Beam Modifier Coordinate System after the Beam Modifier Orientation Angle (300A,0645) is applied. See Section C.36.2.2.20.1.1 and Section C.36.2.2.19.1.2. Required if the conditions in Section C.36.2.2.5.1.1 are satisfied. |
|||
|
One-dimensional positions of the tip in mm of beam delimiters. If Device Type Code Sequence (3010,002E) contains Coded Entry Value (130333, DCM, "Single Leaves"), N values shall be provided where N is the Number of Parallel RT Beam Delimiters (300A,0648). If Device Type Code Sequence contains Coded Entry Value (130330, DCM, "Jaw Pair") or (130331, DCM, "Leaf Pairs"), 2N Values shall be provided where N is the Number of Parallel RT Beam Delimiters (300A,0648). The values shall be grouped by the mounting side identified by the Parallel RT Beam Delimiter Leaf Mounting Side (300A,064F) with the values of RT Beam Delimiter Elements on the negative mounting side first. The order of values shall correspond to the order of the Parallel RT Beam Delimiter Boundaries (300A,0649). See Section C.36.2.2.20.1.1, Section C.36.2.2.20.1.2 and Section C.36.2.2.20.1.3. Required if the conditions in Section C.36.2.2.5.1.1 are satisfied and if Device Type Code Sequence contains Coded Entry Value (130330, DCM, "Jaw Pair"), (130331, DCM, "Leaf Pairs") or (130333, DCM, "Single Leaves"), and if the Parallel RT Beam Delimiter Opening Mode (300A,064E) is present and does not have the Value BINARY. |
|||
|
The outline of the Beam Limiting Device opening. Required if the conditions in Section C.36.2.2.5.1.1 are satisfied and if DeviceType Code Sequence (3010,002E) contains Coded Entry Value (130332, DCM, "Variable Circular Collimator"). |
|||
|
>Include Table 10.38-1 “Outline Definition Macro Attributes”. |
The Outline Shape Type (0018,1630) shall be CIRCULAR. The plane is defined in Section C.36.2.2.20.1.1. |
||
All geometric values in Section C.36.2.2.20.1 are defined in the Beam Modifier Definition Plane.
For Device Type Code Sequence (3010,002E) Values of (130330, DCM, "Jaw Pair") or (130331, DCM, "Leaf Pairs"), the order of values are
where N denotes the negative mounting side, P the positive mounting side and the indices increasing corresponding to the order of the Values of Parallel RT Beam Delimiter Boundaries (300A,0649).
The definition of the tip positions in Parallel RT Beam Delimiter Positions (300A,064A) or delimiter outline in the RT Beam Delimiter Geometry Sequence (300A,064C) is as defined by the manufacturer and shall be documented in the Conformance Statement. Typically, this will be the radiological or physical edge.
| DICOM PS3.3 2026d - Information Object Definitions |
|---|