| DICOM PS3.3 2026c - Information Object Definitions |
|---|
Table C.34.8-1 specifies the Attributes of the Patient Positioning Module, which describe details about the positioning of the patient before and during the imaging procedure.
Table C.34.8-1. Patient Positioning Module Attributes
|
Patient position relative to the equipment described by the procedure protocol. See Section C.7.3.1.1.2 for Defined Terms and further explanation. |
|||
|
Instructions for positioning and aligning the patient for the procedure. E.g., aligning an anatomical landmark with laser crosshairs. One or more Items are permitted in this Sequence. The precise correlation between positioning and scan elements is to be described in the Instruction Text if necessary. |
|||
|
Identifies the order in which instruction Sequence Items are presented/performed. The Value shall be an integer, increasing monotonically by 1, starting from 1. |
|||
|
Whether or not this instruction was performed. Required if the Value of SOP Class UID (0008,0016) equals one of the following Values: |
|||
|
Date and time the instruction was performed. Required if Instruction Performed Flag (0018,9918) is present with a Value of YES. |
|||
|
Identifies the method for positioning the patient. Only a single Item shall be included in this Sequence. See Section C.34.8.1. |
|||
|
A coded location identifying the intended landmark used as the basis for positioning. Only a single Item is permitted in this Sequence. See Section C.34.8.1. |
|||
|
>Include Table 10.27-1 “Reference Location Macro Attributes” |
BCID 1000 “CT Transverse Plane Reference Basis” for Reference Basis Code Sequence (0018,9902) BCID 1010 “Reference Geometry - Plane” for Reference Geometry Code Sequence (0018,9903) |
||
|
UID of the Frame of Reference of another data set to which the current procedure is intended to be roughly aligned. Note
|
|||
|
Position Reference Indicator for the Target Frame of Reference UID (0018,991E). See Section C.34.8.1. |
|||
|
Identifies the general anatomic region imaged by the Protocol. See Section C.34.8.2. |
|||
|
Sequence of Items that modifies the anatomic region of interest of this Instance. See Section C.34.8.2. |
|||
|
Identifies the primary anatomic structure(s) of interest in this Protocol. |
|||
| BCID 8134 “Anatomic Structure”. | |||
|
Sequence of Items that modifies the primary anatomic structure of interest in this Instance. |
|||
| DICOM PS3.3 2026c - Information Object Definitions |
|---|