| DICOM PS3.3 2026c - Information Object Definitions |
|---|
Table C.34.12-1 specifies the Attributes of the Performed CT Reconstruction Module, which contain reconstruction parameter values for a performed CT imaging procedure.
This Module contains Attributes that affect machine behavior but not those that are merely descriptive. The latter may be found in the reconstructed images.
Table C.34.12-1. Performed CT Reconstruction Module Attributes
|
Parameter values for each reconstruction protocol element in the Protocol. See Section C.34.11.1. Elements are performed in the order of their Protocol Element Number (0018,9921). |
|||
|
>Include Table 10.28-1 “Protocol Element Identification Macro Attributes” |
|||
|
A value corresponding to the Protocol Element Number (0018,9921) of the Acquisition Protocol Element from this Protocol being reconstructed in this reconstruction protocol element. This may be multi-valued if multiple acquisitions are combined in a single reconstruction protocol element. |
|||
|
One or more values corresponding to the Beam Number (300A,00C0) in the Acquisition Protocol Element (specified in Source Acquisition Protocol Element Number (0018,9938) ) from which data was used in this reconstruction protocol element. |
|||
|
Uniquely identifies the referenced SOP Class. Required if the referenced acquisition protocol element is not in this Instance. Shall have a Value of "1.2.840.10008.5.1.4.1.1.200.2" (CT Performed Procedure Protocol Storage). |
|||
|
The UID of the Instance containing the acquisition protocol element referenced in Source Acquisition Protocol Element Number (0018,9938). Required if the referenced acquisition protocol element is not in this Instance. |
|||
|
Anatomically oriented transverse location where this reconstruction starts. Note
|
|||
|
>>Include Table 10.27-1 “Reference Location Macro Attributes” |
BCID 1000 “CT Transverse Plane Reference Basis” for Reference Basis Code Sequence (0018,9902) BCID 1010 “Reference Geometry - Plane” for Reference Geometry Code Sequence (0018,9903) |
||
|
Anatomically oriented transverse location where this reconstruction ends. Note
|
|||
|
>>Include Table 10.27-1 “Reference Location Macro Attributes” |
BCID 1000 “CT Transverse Plane Reference Basis” for Reference Basis Code Sequence (0018,9902) BCID 1010 “Reference Geometry - Plane” for Reference Geometry Code Sequence (0018,9903) |
||
|
>>Include Table 10-19 “Algorithm Identification Macro Attributes” |
BCID 10033 “CT Reconstruction Algorithm” for Algorithm Family Code Sequence (0066,002F) |
||
|
A label describing the convolution kernel or algorithm used to reconstruct the data. |
|||
|
A label describing the group that the Convolution Kernel (0018,1210) belongs. |
|||
|
The diameter in mm of the region from which data were used in creating the reconstruction of the image. Data may exist outside this region and portions of the patient may exist outside this region. See Section C.8.15.3.6.1. Required if Reconstruction Field of View (0018,9317) is not present. |
|||
|
The field of view width (x-dimension) followed by height (y-dimension) as used for reconstruction in mm. Required if Reconstruction Diameter (0018,1100) is not present. |
|||
|
The x, y, and z coordinates (in the Patient-Based Coordinate System) of the reconstruction center target point as used for reconstruction in mm. See Section C.8.15.3.6.1. |
|||
|
An anatomically based description of a point in the patient used as the reconstruction center target point. |
|||
|
>>Include Table 10.27-1 “Reference Location Macro Attributes” |
BCID 1000 “CT Transverse Plane Reference Basis” for Reference Basis Code Sequence (0018,9902) BCID 1010 “Reference Geometry - Plane” for Reference Geometry Code Sequence (0018,9903) |
||
|
Physical distance in the patient between the center of each reconstructed pixel, specified by a numeric pair - adjacent row spacing (delimiter) adjacent column spacing in mm. See Section 10.7.1.3 for further explanation of the order of the Values. |
|||
|
Angle (in degrees) over which the data from which the Frame was reconstructed was collected. |
|||
|
A label describing the filter applied to the reconstructed image after the original reconstruction has been completed. |
|||
|
A description of the nature or effect of the Image Filter (0018,9320). |
|||
|
Additional processing applied to the reconstructed image after image filter (if any). |
|||
|
Spacing between slices, in mm. The spacing is measured from the center-to-center of each slice. NoteSlice overlap may be constrained implicitly by constraining the Slice Thickness (0018,0050) and Spacing Between Slices (0018,0088). However, since each constraint is considered independently, the overlap may be any value resulting from the combination of allowable values for thickness and spacing. |
|||
|
Preferred value for Window Center (0028,1050) in the Image Instances produced by this reconstruction protocol element. |
|||
|
Preferred value for Window Width (0028,1051) in the Image Instances produced by this reconstruction protocol element. |
|||
|
Requested text to copy into the Series Description (0008,103E) of the images resulting from this reconstruction protocol element. |
|||
|
A code that is intended to be copied into the Series Description Code Sequence (0008,103F) of instances resulting from this Acquisition Protocol Element. |
|||
|
Content Qualification Indicator of Instances resulting from this Reconstruction Protocol Element. See Section C.34.12.1. |
|||
| DICOM PS3.3 2026c - Information Object Definitions |
|---|