| DICOM PS3.3 2026c - Information Object Definitions |
|---|
Table C.12-9 defines the Attributes that describe the Frames extracted if the SOP Instance was created in response to a Frame-Level retrieve request.
Table C.12-9. Frame Extraction Module Attributes
|
Sequence containing details of how this SOP Instance was extracted from a source multi-frame SOP Instance. If this Instance was created from an Instance that contains a Frame Extraction Sequence, then this Sequence shall contain all of the Items from the parent's Frame Extraction Sequence and a new Item that describes this extraction. |
|||
|
SOP Instance from which the Frames of this Instance are extracted. |
|||
|
A list of Frames that were extracted in the form of a simple list. Required if object extraction is based on a Frame Level Retrieve using the Simple Frame List (0008,1161). See PS3.4 "Instance and Frame Level Retrieve SOP Classes". |
|||
|
A list of Frames that were extracted in the form of one or more triplets Required if object extraction is based on a Frame Level Retrieve using the Calculated Frame List (0008,1162). See PS3.4 "Instance and Frame Level Retrieve SOP Classes". |
|||
|
The start and end times of the Frames that were extracted. Required if object extraction is based on a Frame Level Retrieve using Time Range (0008,1163). See PS3.4 "Instance and Frame Level Retrieve SOP Classes". |
| DICOM PS3.3 2026c - Information Object Definitions |
|---|