DICOM PS3.3 2022e - Information Object Definitions |
---|
Table C.12-8 defines the Attributes that describe the hierarchical relationships of any SOP Instances referenced from other Modules within the Instance in which this Module occurs.
Table C.12-8. Common Instance Reference Module Attributes
Sequence of Items each of which includes the Attributes of one Series. One or more Items shall be included in this Sequence. Required if this Instance references Instances in this Study. |
|||
Unique identifier of the Series containing the referenced Instances. |
|||
Sequence of Items each providing a reference to an Instance that is part of the Series defined by Series Instance UID (0020,000E) in the enclosing Item. |
|||
>>Include Table 10-11 “SOP Instance Reference Macro Attributes” |
|||
Sequence of Items each identifying a Study other than the Study of which this Instance is a part, which Studies contain Instances that are referenced elsewhere in this Instance. One or more Items shall be included in this Sequence. Required if this Instance references Instances in other Studies. |
|||
Unique identifier of the Study containing the referenced Instances. |
|||
>Include Table 10-4 “Series and Instance Reference Macro Attributes” |
DICOM PS3.3 2022e - Information Object Definitions |
---|