| DICOM PS3.3 2026c - Information Object Definitions |
|---|
Table C.11.8-1 specifies the Attributes of the Softcopy VOI LUT Module, which describe the Softcopy VOI LUT. These Attributes have the same meaning and behavior as defined in the Section C.11.2 VOI LUT Module.
If a VOI LUT (that applies to a Referenced Image) is not present, it is assumed to be an identity transformation. Any VOI LUT or equivalent values in the Referenced Image will not be used (described in Section N.2 in PS3.4).
Table C.11.8-1. Softcopy VOI LUT Module Attributes
|
Defines a Sequence of VOI LUTs or Window Centers and Widths and to which images and Frames they apply. No more than one VOI LUT Sequence containing a single Item or one pair of Window Center/Width values shall be specified for each image or Frame. |
|||
|
The subset of images and Frames listed in the Presentation State Relationship Module, to which this VOI LUT or Window Center and Width applies. One or more Items shall be included in this Sequence. Required if the VOI LUT transformation in this Item does not apply to all the images and Frames listed in the Presentation State Relationship Module. |
|||
|
>>Include Table 10-3 “Image SOP Instance Reference Macro Attributes” |
|||
| DICOM PS3.3 2026c - Information Object Definitions |
|---|