| DICOM PS3.3 2026c - Information Object Definitions |
|---|
Table C.11.23-1 specifies the Attributes of the Volumetric Presentation State Relationship Module, which describe sets of inputs to a Presentation State and how each input is to be displayed in the presentation.
Table C.11.23-1. Volumetric Presentation State Relationship Module Attributes
|
Input sets specified for use by this Volumetric Presentation State. |
|||
|
Unique identifier of this input specification. See Section C.11.23.4. |
|||
|
See Section C.11.23.1. |
|||
|
The set of images comprising this input volume. One or more Items shall be included in this Sequence. See Section C.11.23.1 for constraints on objects referenced by this Sequence. Required if Presentation Input Type (0070,1202) has a Value of VOLUME. |
|||
|
>>Include Table 10-3 “Image SOP Instance Reference Macro Attributes” |
|||
|
A reference to a Spatial Registration Instance that is used to register the referenced inputs. Only one Item shall be included in this Sequence. All images referenced by the Referenced Image Sequence (0008,1140) of this Item of the Volumetric Presentation State Input Sequence (0070,1201) shall be referenced by the Spatial Registration Instance. See Section C.11.23.3. Required if the Frame of Reference UID (0020,0052) Value of the Images referenced by the Referenced Image Sequence (0008,1140) of this Item of the Volumetric Presentation State Input Sequence (0070,1201) does not match the Frame of Reference UID (0020,0052) Value of this Presentation State Instance. May be present otherwise. |
|||
|
>>Include Table 10-11 “SOP Instance Reference Macro Attributes” |
|||
|
Inputs to the Presentation State. |
|||
|
Identification number of the input. Values shall be ordinal numbers starting from 1 and monotonically increasing by 1 within the Volumetric Presentation State Instance. |
|||
|
Unique identifier of this input in Volumetric Presentation Input Set Sequence (0070,120A). The UID may be shared among multiple Volumetric Presentation State Input Sequence (0070,1201) Items. See Section C.11.23.4. |
|||
|
Position of this input data within a set of sequential inputs. Multiple inputs may share the same value. Distinct values shall be ordinal numbers starting from 1 and monotonically increasing by 1 within the Volumetric Presentation State Instance. See Section C.11.23.2. Required if Presentation Animation Style (0070,1A01) is present with a Value of INPUT_SEQ. |
|||
|
Specifies whether to apply per-input cropping to this input. See Section C.11.23.5. |
|||
|
Values of Cropping Specification Number (0070,1309) of the Item in the Volume Cropping Sequence (0070,1301) specifying the cropping methods to be applied to this input. |
|||
|
The rendering method for this input. Enumerated Values:
Required if MPR Thickness Type (0070,1502) is present with a Value of SLAB. |
|||
|
Specifies whether to apply the cropping specified by Global Cropping Specification Index (0070,120C) to every input of the Volumetric Presentation State. See Section C.11.23.5. |
|||
|
Values of Cropping Specification Number (0070,1309) of one or more Items in the Volume Cropping Sequence (0070,1301) specifying the cropping methods to be applied to every input. Volume Cropping Method (0070,1302) shall include only BOUNDING_BOX or OBLIQUE_PLANE values. |
|||
SOP Instances with Presentation Input Type (0070,1202) Value of VOLUME referenced by one Item in Volumetric Presentation Input Set Sequence (0070,1209) are a collection of Image Instances or Frames within Image Instances that conforms to the following requirements:
All Instances or Frames shall have the following Attributes of the same value:
All Instances or Frames have orthogonal row and column vectors
No two Instances or Frames share the same Image Position (Patient) (0020,0032) or Image Position (Volume) (0020,9301). Two Instances or Frames should be treated as sharing the same position if the differences between their positions is sufficiently small.
All Instances or Frames shall be parallel. Instances or Frames may be treated as parallel if the differences in their Image Orientation (Patient) (0020,0037) or Image Orientation (Volume) (0020,9302) are sufficiently small.
All Instances or Frames shall be sufficiently aligned such that a single ray can pass through the upper-left hand corner of all Instances or Frames. That ray shall be orthogonal to the plane of the Frame (i.e., normal to the input Frames). Instances or Frames may be treated as aligned if the degree of misalignment between Frames or relative to the normal is sufficiently small.
| DICOM PS3.3 2026c - Information Object Definitions |
|---|