DICOM PS3.3 2026d - Information Object Definitions

A.8.4.5 Multi-frame Grayscale Word Secondary Capture Image Functional Group Macros

Table A.8-4b specifies the use of the Functional Group Macros used in the Multi-frame Functional Groups Module for the Multi-frame Grayscale Word Secondary Capture Image IOD.

Table A.8-4b. Multi-frame Grayscale Word Secondary Capture Image Functional Group Macros

Functional Group Macro

Section

Usage

Pixel Measures

C.7.6.16.2.1

C - Required if Plane Position (Patient) or Plane Orientation (Patient) Macros Present

Plane Position (Patient)

C.7.6.16.2.3

C - Required if Pixel Measures or Plane Orientation (Patient) Macros Present

Plane Orientation (Patient)

C.7.6.16.2.4

C - Required if Pixel Measures or Plane Position (Patient) Macros Present


Note

If the Pixel Measures Macro is present, then the Values of Pixel Spacing (0028,0030) therein are intended to be used for 3D spatial computations, rather than any Values of Nominal Scanned Pixel Spacing (0018,2010) (from the SC Multi-frame Image Module), which may also be present.

DICOM PS3.3 2026d - Information Object Definitions