| DICOM PS3.3 2026c - Information Object Definitions |
|---|
Table A.74-2 specifies the use of the Functional Group Macros used in the Multi-frame Functional Groups Module for the Breast Projection X-Ray Image IOD.
Table A.74-2. Breast Projection X-Ray Image Functional Group Macros
|
C - Required if the Image Type (0008,0008) Value 1 equals DERIVED or Value 1 is ORIGINAL and Presentation Intent Type equals FOR PRESENTATION. May be present otherwise. |
||
|
C - Required if the Enhanced Contrast/Bolus Module is present. |
||
|
C - Required if X-Ray Source moves relative to the Patient. May be present otherwise. |
||
|
C - Required if X-Ray Detector plane is not normal to the X-Ray beam vector. May be present otherwise. |
||
For Anatomic Region Sequence (0008,2218) DCID 4013 “Anatomic Region for Mammography” shall be used.
| DICOM PS3.3 2026c - Information Object Definitions |
|---|