Table A.55-1 specifies the Modules of the Breast Tomosynthesis Image IOD.
Table A.55-1. Breast Tomosynthesis Image IOD Modules
The Overlay Plane Module, Modality LUT Module, VOI LUT Module and Softcopy Presentation LUT Module shall not be used in a Standard Extended SOP Class of the Breast Tomosynthesis Image.
For Breast Tomosynthesis Image SOP Instances created from SOP Instances with a SOP Class UID (0008,0016) of "1.2.840.10008.5.1.4.1.1.13.1.4" (Breast Projection X-Ray Image Storage - For Presentation SOP Class) or "1.2.840.10008.5.1.4.1.1.13.1.5" (Breast Projection X-Ray Image Storage - For Processing SOP Class), the isocenter coordinate system shall be used to describe the positioning of the X-Ray Source, Detector and Breast Support (see Section C.8.31.6), and shall use only the Value ISOCENTER for Equipment Coordinate System Identification (0028,9537).