DICOM PS3.3 2026c - Information Object Definitions

A.54.4 X-Ray 3D Craniofacial Image Functional Group Macros

Table A.54-2 specifies the use of the Functional Group Macros used in the Multi-frame Functional Groups Module for the X-Ray 3D Craniofacial Image IOD.

Table A.54-2. X-Ray 3D Craniofacial Image Functional Group Macros

Functional Group Macro

Section

Usage

Pixel Measures

C.7.6.16.2.1

M

Frame Content

C.7.6.16.2.2

M- May not be used as a Shared Functional Group.

Plane Position (Patient)

C.7.6.16.2.3

M

Plane Orientation (Patient)

C.7.6.16.2.4

M

Referenced Image

C.7.6.16.2.5

U

Derivation Image

C.7.6.16.2.6

C - Required if Image Type (0008,0008) Value 1 equals DERIVED.

Frame Anatomy

C.7.6.16.2.8

M

Pixel Value Transformation

C.7.6.16.2.9

U

Frame VOI LUT

C.7.6.16.2.10

M

Real World Value Mapping

C.7.6.16.2.11

U

Contrast/Bolus Usage

C.7.6.16.2.12

C - Required if the Enhanced Contrast/Bolus Module is present.

X-Ray 3D Frame Type

C.8.21.5.1

M


A.54.4.1 X-Ray 3D Craniofacial Image Functional Group Macros Content Constraints

A.54.4.1.1 Frame Anatomy Functional Group Macro

For Anatomic Region Sequence (0008,2218) DCID 4028 “Craniofacial Anatomic Region” or DCID 4016 “Anatomic Region for Intra-oral Radiography” shall be used.

In the case of CID 4016 “Anatomic Region for Intra-oral Radiography”, for Anatomic Region Modifier Sequence (0008,2220), if present, DCID 4017 “Anatomic Region Modifier for Intra-oral Radiography” shall be used, and for Primary Anatomic Structure Sequence (0008,2228), if present, DCID 4026 “Primary Anatomic Structure for Intra-oral and Craniofacial Radiography - Teeth” shall be used.

DICOM PS3.3 2026c - Information Object Definitions