| DICOM PS3.3 2026c - Information Object Definitions |
|---|
The Overlay Plane Module and VOI LUT Module shall not be used in a Standard Extended SOP Class of the Ophthalmic Tomography Image.
In order to annotate images, whether during acquisition or subsequently, SOP Instances of the Grayscale Softcopy Presentation State Storage or the Structured Report Storage SOP Classes that reference the image SOP Instance may be used.
The VOI LUT function is provided by a Frame VOI LUT With LUT Functional Group.
Pseudo-color presentation information may be applied through the use of separate Pseudo-Color Softcopy Presentation State SOP Instances.
No standard mechanism is provided for inclusion of annotations within the image SOP Instance itself and implementers are discouraged from using private extensions to circumvent this restriction.
| DICOM PS3.3 2026c - Information Object Definitions |
|---|