| DICOM PS3.3 2026c - Information Object Definitions |
|---|
Table A.47-2 specifies the use of the Functional Group Macros used in the Multi-frame Functional Groups Module for the Enhanced XA Image IOD.
Table A.47-2. Enhanced XA Image Functional Group Macros
|
C - Required if the image or Frame has been derived from another SOP Instance. |
||
|
C - Required if the Enhanced Contrast/Bolus Module is present. |
||
|
C - Required if Pixel Intensity Relationship (0028,1040) equals LOG. May be present otherwise. |
||
|
C - Required if C-arm Positioner Tabletop Relationship (0018,9474) is present and equals YES. May be present otherwise. |
||
|
C - Required if Isocenter Reference System Sequence (0018,9462) is present. May be present otherwise. |
||
|
C - Required if X-Ray Receptor Type (0018,9420) is present and equals DIGITAL_DETECTOR. |
||
|
C - Required if C-arm Positioner Tabletop Relationship (0018,9474) is present and equals YES. |
||
|
C - Required if Image Type (0008,0008) Value 1 equals ORIGINAL and C-arm Positioner Tabletop Relationship (0018,9474) is present and equals YES. May be present otherwise. |
||
|
C - Required if Image Type (0008,0008) Value 1 equals ORIGINAL and C-arm Positioner Tabletop Relationship (0018,9474) is present and equals YES. May be present otherwise. |
||
|
C - Required if Image Type (0008,0008) Value 1 equals ORIGINAL. May be present otherwise. |
||
|
U - May not be used if C-arm Positioner Tabletop Relationship (0018,9474) is not present or equals NO. |
||
|
C - Required if Projection Pixel Calibration Sequence (0018,9401) is present. May be present otherwise. |
For Anatomic Region Sequence (0008,2218) DCID 4042 “XA/XRF Anatomy Imaged” shall be used.
| DICOM PS3.3 2026c - Information Object Definitions |
|---|