| DICOM PS3.3 2026c - Information Object Definitions |
|---|
Table A.35.5-1 specifies the Modules of the Mammography CAD SR IOD.
The Specimen Identification Module (Retired) was previously included in this IOD but has been retired. See PS3.3-2008.
The document shall be constructed from TID 4000 “Mammography CAD Report” invoked at the root node.
When a Content Item sub-tree from a prior document is duplicated by-value, its observation context shall be defined by TID 1001 “Observation Context” and its subordinate templates.
All Template and Context Group definitions are located in Annex A “Structured Reporting Templates (Normative)” in PS3.16 and Annex B “DCMR Context Groups (Normative)” in PS3.16, respectively.
Value Type (0040,A040) in Content Sequence (0040,A730) of the SR Document Content Module is constrained to the following Enumerated Values (see Table C.17.3-7 for Value Type definitions):
The Mammography CAD SR IOD makes extensive use of by-reference INFERRED FROM, by-reference HAS PROPERTIES and by-reference SELECTED FROM relationships. Other relationships by-reference are forbidden. Table A.35.5-2 specifies the relationship constraints of this IOD. See Table C.17.3-8 for Relationship Type definitions.
| DICOM PS3.3 2026c - Information Object Definitions |
|---|