DICOM PS3.3 2026c - Information Object Definitions

A.35.10.3 Colon CAD SR IOD Module Table

Table A.35.10-1 specifies the Modules of the Colon CAD SR IOD.

Table A.35.10-1. Colon CAD SR IOD Modules

IE

Module

Reference

Usage

Patient

Patient

C.7.1.1

M

Clinical Trial Subject

C.7.1.3

U

Study

General Study

C.7.2.1

M

Patient Study

C.7.2.2

U

Clinical Trial Study

C.7.2.3

U

Series

SR Document Series

C.17.1

M

Clinical Trial Series

C.7.3.2

U

Equipment

General Equipment

C.7.5.1

M

Enhanced General Equipment

C.7.5.2

M

SR Document

SR Document General

C.17.2

M

SR Document Content

C.17.3

M

SOP Common

C.12.1

M


A.35.10.3.1 Colon CAD SR IOD Content Constraints
A.35.10.3.1.1 Template Constraints

The document shall be constructed from TID 4120 “Colon CAD Report” invoked at the root node.

Note

All Template and Context Group definitions are located in Annex A “Structured Reporting Templates (Normative)” in PS3.16 and Annex B “DCMR Context Groups (Normative)” in PS3.16, respectively.

A.35.10.3.1.2 Value Type

Value Type (0040,A040) in Content Sequence (0040,A730) of the SR Document Content Module is constrained to the following Enumerated Values (see Table C.17.3-7 for Value Type definitions):

Enumerated Values:

TEXT

CODE

NUM

DATE

TIME

UIDREF

PNAME

SCOORD

SCOORD3D

TCOORD

COMPOSITE

IMAGE

CONTAINER

Note

The WAVEFORM Value Type has no practical use for this IOD, so is not included.

A.35.10.3.1.3 Relationship Constraints

The Colon CAD SR IOD makes use of by-reference INFERRED FROM and by-reference HAS ACQ CONTEXT relationships. Other relationships by-reference are forbidden. Table A.35.10-2 specifies the relationship constraints of this IOD. See Table C.17.3-8 for Relationship Type definitions.

Table A.35.10-2. Relationship Content Constraints for Colon CAD SR IOD

Source Value Type

Relationship Type (Enumerated Values)

Target Value Type

CONTAINER

CONTAINS

CODE, NUM, IMAGE, CONTAINER, UIDREF, DATE, TIME

TEXT, CODE, NUM, CONTAINER

HAS OBS CONTEXT

TEXT, CODE, NUM, DATE, TIME, PNAME, UIDREF, COMPOSITE

CONTAINER

HAS OBS CONTEXT

CONTAINER

IMAGE

HAS ACQ CONTEXT

TEXT, CODE, DATE, TIME, NUM, CONTAINER

CONTAINER, CODE, COMPOSITE, NUM

HAS CONCEPT MOD

TEXT, CODE

TEXT, CODE, NUM

HAS PROPERTIES

CONTAINER, TEXT, CODE, NUM, DATE, IMAGE, SCOORD, SCOORD3D, UIDREF

CODE, NUM

INFERRED FROM

CODE, NUM, IMAGE, SCOORD, SCOORD3D, CONTAINER, TEXT

SCOORD

SELECTED FROM

IMAGE

TCOORD

SELECTED FROM

SCOORD, SCOORD3D, IMAGE


Note

The SOP Classes to which an IMAGE or COMPOSITE Value Type may refer are documented in the Conformance Statement for an application (see PS3.2 and PS3.4). The WAVEFORM Value Type is not used in this IOD.

DICOM PS3.3 2026c - Information Object Definitions