The Advanced Blending Presentation State IOD specifies information that may be used to blend two or more sets of images that are referenced from within the IOD for the purpose of presentation (display).
It includes capabilities for specifying:
optional thresholds to restrict contributing areas of an input
optional windowing (VOI LUT) to be applied to grayscale images before pseudo-colorization
optional pseudo-color palette to be applied to grayscale images
definition of blending control values for the different inputs
selection of the area of the output images to display and whether to rotate or flip it
image and display relative annotations, including graphics, text and overlays