| DICOM PS3.3 2026c - Information Object Definitions |
|---|
Table A.32.8-2 specifies the use of the Functional Group Macros used in the Multi-frame Functional Groups Module for the VL Whole Slide Microscopy Image IOD.
Table A.32.8-2. VL Whole Slide Microscopy Image Functional Group Macros
The Plane Position (Slide) and Optical Path Identification Macros are Type C, which allows the Per-Frame Functional Groups Sequence (5200,9230) to be entirely omitted in those cases in which there are no other Per-Frame Functional Group Macros with content (i.e., the Frame Content Macro is empty or absent).
For Purpose of Reference Code Sequence (0040,A170) in the Referenced Image Functional Group DCID 8120 “Whole Slide Microscopy Image Referenced Image Purpose of Reference” shall be used.
Frames shall occupy plane positions in a regular tiling of the Total Pixel Matrix; the Frames may extend beyond the edges of the Total Pixel Matrix. Therefore, the Value of Column Position In Total Pixel Matrix (0048,021E) shall be an integer multiple of the Value of Columns (0028,0011) plus a constant, and the Value of Row Position In Total Pixel Matrix (0048,021F) shall be an integer multiple of the Value of Rows (0028,0010) plus a (possibly different) constant.
Pixels beyond the edge of the Total Pixel Matrix may be encoded with a fixed padding value. For images with MONOCHROME2 Photometric Interpretation, see Pixel Padding Value (0028,0120) in Section C.7.5.1.1.2.
The encoding of tiles may be sparse; i.e., some tiles may not be encoded in Frames. There are no constraints on the ordering of Frames within the pixel data; each Frame specifies its position in the Plane Position (Slide) Functional Group.
Frames associated with different optical paths may coincide in any spatial dimension.
| DICOM PS3.3 2026c - Information Object Definitions |
|---|