A.3.3 CT Image IOD Module Table
Table A.3-1 specifies the Modules of the CT Image IOD.
Table A.3-1. CT Image IOD Modules
|
IE
|
Module
|
Reference
|
Usage
|
|
Patient
|
Patient
|
C.7.1.1
|
M
|
|
Clinical Trial Subject
|
C.7.1.3
|
U
|
|
Study
|
General Study
|
C.7.2.1
|
M
|
|
Patient Study
|
C.7.2.2
|
U
|
|
Clinical Trial Study
|
C.7.2.3
|
U
|
|
Series
|
General Series
|
C.7.3.1
|
M
|
|
Clinical Trial Series
|
C.7.3.2
|
U
|
|
Single-Frame CT Series
|
C.8.2.3
|
C - Required if the SOP Class UID of this instance is 1.2.840.10008.5.1.4.1.1.2.3 (CT Image Storage - For Processing). May be present otherwise.
|
|
Frame of Reference
|
Frame of Reference
|
C.7.4.1
|
M
|
|
Synchronization
|
C.7.4.2
|
C - Required if time synchronization was applied.
|
|
Equipment
|
General Equipment
|
C.7.5.1
|
M
|
|
Acquisition
|
General Acquisition
|
C.7.10.1
|
M
|
|
Image
|
General Image
|
C.7.6.1
|
M
|
|
General Reference
|
C.12.4
|
U
|
|
Enhanced Patient Orientation
|
C.7.6.30
|
U
|
|
Image Plane
|
C.7.6.2
|
M
|
|
Image Pixel
|
C.7.6.3
|
M
|
|
Contrast/Bolus
|
C.7.6.4
|
C - Required if contrast media was used in this image
|
|
Device
|
C.7.6.12
|
U
|
|
Specimen
|
C.7.6.22
|
U
|
|
CT Image
|
C.8.2.1
|
M
|
|
Multi-energy CT Image
|
C.8.2.2
|
C - Required if Multi-energy CT Acquisition (0018,9361) is YES.
|
|
Overlay Plane
|
C.9.2
|
U
|
|
VOI LUT
|
C.11.2
|
U
|
|
SOP Common
|
C.12.1
|
M
|
|
Common Instance Reference
|
C.12.2
|
U
|
A.3.3.1 CT Image IOD Content Constraints
If Multi-energy CT Acquisition (0018,9361) is YES the following constraints will apply:
-
The Contrast/Bolus Module shall be present if contrast was administered even if images are processed to remove contrast information from the pixels, e.g. Virtual Non-Contrast images.
-
The Real World Value Mapping Sequence (0040,9096) shall be present in the General Image Module.
-
For Measurement Units Code Sequence (0040,08EA) in the Real World Value Mapping Sequence (0040,9096) DCID 301 “Multi-energy Material Unit” shall be used.