| DICOM PS3.3 2026d - Information Object Definitions |
|---|
For the purpose of accommodating large sets of Frames in Multi-frame Image SOP Instances the Real-World Entity Relationship Diagram has been extended to describe the relationships of these instances: Concatenation (see Section 7.5.1) and Dimension Organization (see Section 7.5.2). Figure 7.5-1 depicts the additions to Figure 7-1a.
For implementation specific reasons (such as practical limits on the maximum size of an individual SOP Instance) the content of a Multi-frame Image may need to be split into more than one SOP Instance. These SOP Instances together form a Concatenation, which is a group of SOP Instances within a Series that is uniquely identified by Concatenation UID (0020,9161).
| DICOM PS3.3 2026d - Information Object Definitions |
|---|