A.1.2 IOD Entity-Relationship Model
This Section of an IOD provides the Entity-Relationship (E-R) Model that depicts the relationships of the components or Information Entities (IE) of the specified IOD. It forms an IOD specific information model. This E-R model provides the complete context of how the Composite Instance information shall be interpreted when a Composite Instance is exchanged between two DICOM Application Entities; in particular, an IOD will specify a single IE at the level below the Series IE.
Even though Composite Instances are encoded as discrete individual components, each Composite Instance IOD E-R Model requires that all Composite Instances that are part of a specific Study shall share the same context. That is, all Composite Instances within a specific Patient Study share the same Patient and Study information; all Composite Instances within the same Series share the same Series information; etc.
Figure A.1-1 is the DICOM Composite Instance IOD Information Model. It applies to all Patient-related Composite Instance IODs defined in Annex A. However, a subset of this model may be specified by each individual Composite Instance IOD to accurately define the context for specific Composite Instance exchange.
The sub-sections of this Section describe the Information Entities (IE) that comprise the Composite Instance IODs defined in this Annex.
The Patient IE defines the characteristics of a Patient who is the subject of one or more medical Studies.
Note
A Patient may be a human or a non-human organism.
The Patient IE is modality independent.
The Study IE defines the characteristics of a medical Study performed on a Patient. A Study is a collection of one or more Series of medical images, presentation states, and/or SR documents that are logically related for the purpose of diagnosing a Patient. Each Study is associated with exactly one Patient.
A Study may include Composite Instances that are created by a single modality, multiple modalities or by multiple devices of the same modality.
The Study IE is modality independent.
The Series IE defines the Attributes that are used to group Composite Instances into distinct logical sets. Each Series is associated with exactly one Study.
The following criteria group Composite Instances into a specific Series:
-
All Composite Instances within a Series must be of the same modality
-
Each Series may be associated with exactly one Frame of Reference IE, and if so associated all Composite Instances within the Series shall be spatially or temporally related to each other
-
All Composite Instances within the Series shall be created by the same equipment; therefore, each Series is associated with exactly one Equipment IE
-
All Composite Instances within a Series shall have the same Series information
Presentation States shall be grouped into one or more Series without Images or Waveforms (i.e., in a different Series from the Series containing the Images or Waveforms to which they refer).
Note
-
The Series containing Grayscale, Color and Pseudo-Color Softcopy Presentation States and the Series containing the Images to which they refer are both contained within the same Study, except for Blending Presentation States, which may refer to images from different Studies.
-
The Series containing the Waveform Presentation State and the Series containing the Waveforms to which they refer are both contained within the same Study.
-
The Series containing the Waveform Presentation State and the Series containing Waveform Annotation SRs to which they refer are both contained in the same Study but in different Series.
Waveforms shall be grouped into Series without Images. A Frame of Reference IE may apply to both Waveform Series and Image Series.
SR Documents shall be grouped into Series without Images. The Frame of Reference IE may apply to SR Document Series, for SR Documents that contain 3D spatial coordinates relative to one or more spatial Frames of Reference, or temporal coordinates that require a temporal Frame of Reference.
The Equipment IE describes the particular device that produced the Series of Composite Instances. A device may produce one or more Series within a Study. The Equipment IE does not describe the data acquisition or image creation Attributes used to generate the Composite Instances within a Series. These Attributes are described in the Composite Instance specific IEs (e.g., the Image IE).
A.1.2.5 Frame of Reference IE
The Frame of Reference IE identifies the coordinate system that conveys spatial and/or temporal information of Composite Instances in a Series.
When present, a Frame of Reference IE may be related to one or more Series. In this case, it provides the ability to spatially or temporally relate multiple Series to each other. In such cases, the Series may share the UID of the Frame of Reference, or alternatively, a Registration SOP Instance may specify the spatial relationship explicitly, as a spatial transformation. A Frame of Reference IE may also spatially register a Frame of Reference to an atlas.
The Image IE defines the Attributes that describe the pixel data of an image. The pixel data may be generated as a direct result of Patient scanning (termed an Original Image) or the pixel data may be derived from the pixel data of one or more other images (termed a Derived Image). An image is defined by its image plane, pixel data characteristics, gray scale and/or color mapping characteristics and modality specific characteristics (acquisition parameters and image creation information).
An image is related to a single Series within a single Study.
The pixel data within an Image IE may be represented as a single Frame of pixels or as multiple Frames of pixel data. The Frames of a Multi-frame Image (a Cine Run or the slices of a volume) are sequentially ordered and share a number of common properties. A few Attributes may vary between Frames (e.g., Time, Angular Displacement, Slice Increment). All common Image IE Attributes refer to the first Frame of a Multi-frame Image.
Overlay, Modality and Value of Interest Lookup Table and Real World Value Mapping data may be included within an Image IE only if this information is directly associated with the image.
Overlay data represents graphics or text in a bit-map format, and is used to indicate such items as region of interest, reference marks and annotations.
A.1.2.6.2 Modality LUT Data
Modality LUT data describes the transformation of manufacturer dependent pixel values into pixel values that are manufacturer independent (e.g., Hounsfield units for CT, Optical Density for film digitizers, etc.). The transformation may be linear, described by Rescale Slope (0028,1053) and Rescale Intercept (0028,1052), or non-linear, described by a Lookup Table (LUT).
A.1.2.6.3 Value of Interest LUT Data
The Value of Interest (VOI) LUT data describes the transformation of the modality pixel values into pixel values that are meaningful for print, display, etc. This transformation is applied after any Modality LUT. The transformation may be linear, described by Window Center and Window Width, or non-linear, described by a Lookup Table. A non-linear interpretation of Window Center and Window Width may be defined by VOI LUT Function.
A.1.2.6.4 Real World Value Mapping Data
The Real World Value Mapping data describes the transformation of the image pixel values into Real World Values in defined units. There may be multiple transformations, each scoped by a range of input pixel values. Each transformation may be linear, described by Slope and Intercept, or non-linear, described by a Lookup Table.
A.1.2.7 Overlay IE (Retired)
Retired. See PS3.3-2016a.
Note
Overlays were previously modeled as independent Information Entities; in the current model they are considered Attributes within the Image IE or Presentation State IE. See A.1.2.6.1.
A.1.2.8 Curve IE (Retired)
Retired. See PS3.3-2004.
A.1.2.9 Modality LUT IE (Retired)
Retired. See PS3.3-2016a.
Note
Modality LUTs were previously modeled as independent Information Entities; in the current model they are considered Attributes within the Image IE or Presentation State IE. See A.1.2.6.2.
A.1.2.10 VOI LUT IE (Retired)
Retired. See PS3.3-2016a.
Note
VOI LUTs were previously modeled as independent Information Entities; in the current model they are considered Attributes within the Image IE or Presentation State IE. See A.1.2.6.3.
A.1.2.11 Presentation State IE
The Presentation State IE defines how a referenced image (or images) will be presented (e.g., displayed) in a device independent grayscale space (i.e., in P-Values) or color space (i.e., in PCS-values), and what graphical annotations and spatial and grayscale contrast transformations will be applied to the referenced image pixel data.
Overlay, Modality LUT, and VOI LUT data (see A.1.2.6.1, A.1.2.6.2, and A.1.2.6.3) may be included within a Presentation State IE if this information is to be applied to the referenced image(s).
The Waveform IE represents a multi-channel time-based digitized waveform. The waveform consists of measurements of some physical qualities (e.g., electrical voltage, pressure, gas concentration, or sound), sampled at constant time intervals. The measured qualities may originate, for example, in any of the following sources:
-
the anatomy of the Patient,
-
therapeutic equipment (e.g., a cardiac pacing signal or a radio frequency ablation signal),
-
equipment for diagnostic synchronization (e.g., a clock or timing signal used between distinct devices),
-
the physician's voice (e.g., a dictated report).
The sample data within a Waveform IE may represent one or more acquired channels. Several signal channels acquired at the same sampling rate can be multiplexed (by interleaving samples) in a single multiplex group. (see also Annex C “Waveforms (Informative)” in PS3.17.)
The SR Document IE defines the Attributes that describe the content of an SR Document. These include semantic context as well as Attributes related to document completion, verification and other characteristics. An SR Document SOP Instance is related to a single Series within a single Study.
The Spectroscopy IE defines the Attributes that describe the data of a spectroscopy acquisition created by a magnetic resonance spectroscopy device.
The Raw Data IE defines the Attributes that describe a collection of data that may be used for further processing to produce image data or other data.
Note
For example, raw data may be used with CT and MR systems to reconstruct sets of images or for MR to reconstruct spectroscopic data. The format of the raw data is vendor specific.
A.1.2.16 Encapsulated Document IE
The Encapsulated Document IE defines the Attributes that describe the content of a non-DICOM formatted document that is encapsulated in a DICOM Attribute. These include Attributes related to document origin, title, and other characteristics. An Encapsulated Document SOP Instance is related to a single Series within a single Study.
A.1.2.17 Real World Value Mapping IE
The Real World Value Mapping IE defines the Attributes that describe the mapping of stored pixel data to Real World values (see A.1.2.6.4).
The Surface IE defines the Attributes that describe a surface in a spatial coordinate system. A surface is defined by its shape and can be further defined by normals on that shape. The surface may be reconstructed from either spatial scans (e.g., laser scanners) or based on images. A surface is described by its finite volume and manifold property, gray scale and color mapping characteristics, presentation type, opacity, and modality specific characteristics.
A surface is related to a single Series.
The Measurements IE defines the Attributes that describe the measurements taken by medical instruments.
A.1.2.20 Tractography Results IE
The Tractography Results IE defines the Attributes that describe the results of a tractography application.
Note
The term tractogram is not used because it does not include all types of tractography results.
The Plan IE defines the parameters and instructions to deliver treatment, particularly Radiotherapy, to the Patient. The entity includes the set of machine and positioning parameters to be applied during treatment delivery and instructions guiding the treatment workflow.
A.1.2.22 Content Assessment Result IE
The Content Assessment Result IE contains the results of an assessment of the content of a SOP Instance.
An assessment is part of a process within a clinical workflow, conducted by users or devices, which have the role of assessing the validity and suitability of the content in question, based on subjective or objective criteria. The specific nature of such a process is outside of the scope of this Standard.
A.1.2.23 Spatial Fiducials IE
The Spatial Fiducials IE identifies one or more geometric locations or shapes within a Frame of Reference or image pixel/voxel space that may be correlated with similar locations or shapes within different Frames of Reference or image pixel/voxel spaces.
The Dose IE describes dose distributions calculated by radiotherapy treatment planning systems. These distributions may be represented as 2D or 3D grids, as isodose curves, or as named or unnamed dose points scattered throughout a volume.
A.1.2.25 Structure Set IE
The Structure Set IE describes Regions of Interest (ROI) within a referenced 2D (image) or 3D (volumetric) space. These ROIs may be represented as geometric contours.
A.1.2.26 Treatment Record IE
The Treatment Record IE describes treatments, particularly radiotherapy, for a Patient.
A.1.2.27 Stereometric Relationship IE
The Stereometric Relationship IE defines how referenced images are related as stereometric pairs.
A.1.2.28 Procedure Protocol IE
The Procedure Protocol IE defines the Attributes that describe a Protocol. This IE may encode a Defined Procedure Protocol or a Performed Procedure Protocol.
The Acquisition IE defines the Attributes that describe a single continuous gathering of data.
An Acquisition may result in more than one Series, and a Series may contain Instances from more than one Acquisition.
A.1.2.30 Multi-Resolution Pyramid IE
The Multi-Resolution Pyramid IE describes a set of Images that encode the same image data at different spatial resolutions, i.e., a base (highest resolution) layer that is successively smoothed and down-sampled to create additional lower resolution layers (a multi-resolution decomposition).
Note
No specific method of filtering or down-sampling is specified by the Standard, nor is there a requirement for any specific down-sampling factor between layers, nor that an integer factor be used.
Each layer is encoded as a separate DICOM Image, each of which has a uniform resolution (same Value for Pixel Spacing (0028,0030)). Layers may be tiled, with each tile encoded as a Frame of a Multi-frame Image.
All DICOM Image SOP Instances that constitute a single Multi-Resolution Pyramid shall share the same Frame of Reference, and shall be contained in the same Series.
Only one Multi-Resolution Pyramid shall be contained in a Series (i.e., each such Multi-Resolution Pyramid will be in a different Series).
Note
-
In historical usage, there is no Multi-Resolution Pyramid IE and thus a Series is not constrained to contain only a single conceptual pyramid. However, any instantiation of the Multi-Resolution Pyramid in a Series constrains the Series to one Multi-Resolution Pyramid.
-
Each Multi-Resolution Pyramid may be accompanied in the same Series by LABEL, OVERVIEW and THUMBNAIL images if they share the same Frame of Reference (but not otherwise, per the definition of the Series IE). The THUMBNAIL image rather than a VOLUME image may be the apex (lowest resolution layer) of the Multi-Resolution Pyramid.
-
A unique identifier, Pyramid UID (0008,0019) will be assigned to an instance of a Multi-Resolution Pyramid, and will be shared by all of the layers that constitute that instance of a Multi-Resolution Pyramid, whether or not a particular resolution layer (usually the highest resolution) is deemed to be ORIGINAL, and the lower resolution layers DERIVED (e.g., by some down-sampling image processing operation). By definition, the absence of Pyramid UID (0008,0019) implies the absence of instantiation of the Multi-Resolution Pyramid IE.
That use is distinct from the Pyramid UID (0008,0019) of different Multi-Resolution Pyramids that may be further derived from a Multi-Resolution Pyramid. In otherwords, the Pyramid UID (0008,0019) of a Multi-Resolution Pyramid will not be shared between two pyramids that contain different pixel data (other than differences due to lossless representation of the same pixel data in different Transfer Syntaxes).
A.1.2.31 Waveform Presentation State IE
The Waveform Presentation State IE defines how referenced waveforms will be presented.
The Waveform Presentation State IE comprises text annotations, segments of interest, and montages including filters, colors, gain, and vertical sizes of waveform channels if this information is to be applied to the referenced waveform(s).
It might also contain display information for structured annotations related to the referenced waveform(s).
A.1.4 Overview of the Composite IOD Module Content
The Tables in this Section provide an overview of the Modules used throughout the Composite IODs. This table is for informative purposes only. It is based on the IOD definitions found in the remaining Sections of Annex A that are normative.
Table A.1-1a. Composite Information Object Modules Overview - Images
|
IODs
Modules
|
CR
|
CT
|
Enh CT
|
MR
|
Enh MR
|
Enh MR Col
|
NM
|
US
|
US MF
|
Enh US Vol
|
PA
|
SC
|
SC MF SB
|
SC MF GB
|
SC MF GW
|
SC MF TC
|
PT
|
Enh PT
|
|
Patient
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
|
Clinical Trial Subject
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
|
General Study
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
|
Patient Study
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
|
Clinical Trial Study
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
|
General Series
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
|
Clinical Trial Series
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
|
Enhanced Series
|
|
|
|
|
|
|
|
|
|
|
M
|
|
|
|
|
|
|
|
|
CR Series
|
M
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
CT Series
|
|
|
M
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Single-Frame CT Series
|
|
C
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
NM/PET Patient Orientation
|
|
|
|
|
|
|
M
|
|
|
|
|
|
|
|
|
|
M
|
|
|
Enhanced Patient Orientation
|
|
U
|
U
|
U
|
U
|
U
|
|
|
|
|
|
U
|
U
|
U
|
U
|
U
|
|
U
|
|
PET Series
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
M
|
|
|
PET Isotope
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
M
|
|
|
PET Multi-gated Acquisition
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
C
|
|
|
Enhanced PET Series
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
M
|
|
Enhanced US Series
|
|
|
|
|
|
|
|
|
|
M
|
|
|
|
|
|
|
|
|
|
Frame of Reference
|
|
M
|
M
|
M
|
M
|
M
|
U
|
U
|
U
|
M
|
M
|
|
|
C
|
C
|
C
|
M
|
M
|
|
Ultrasound Frame of Reference
|
|
|
|
|
|
|
|
|
|
M
|
M
|
|
|
|
|
|
|
|
|
Synchronization
|
|
C
|
C
|
|
C
|
C
|
C
|
U
|
U
|
M
|
M
|
|
|
U
|
U
|
U
|
C
|
C
|
|
Cardiac Synchronization
|
|
|
C
|
|
C
|
C
|
|
|
|
C
|
|
|
|
|
|
|
|
C
|
|
Respiratory Synchronization
|
|
|
C
|
|
C
|
C
|
|
|
|
C
|
|
|
|
|
|
|
|
C
|
|
Bulk Motion Synchronization
|
|
|
|
|
C
|
C
|
|
|
|
|
|
|
|
|
|
|
|
|
|
General Equipment
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
U
|
U
|
U
|
U
|
U
|
M
|
M
|
|
Enhanced General Equipment
|
|
|
M
|
|
M
|
M
|
|
|
|
M
|
M
|
|
|
|
|
|
|
M
|
|
SC Equipment
|
|
|
|
|
|
|
|
|
|
|
|
M
|
M
|
M
|
M
|
M
|
|
|
|
Acquisition
|
M
|
M
|
|
M
|
|
|
M
|
M
|
M
|
M
|
|
M
|
M
|
M
|
M
|
M
|
M
|
|
|
General Image
|
M
|
M
|
|
M
|
|
|
M
|
M*
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
|
|
General Reference
|
U
|
U
|
|
U
|
|
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
|
|
Image Plane
|
|
M
|
|
M
|
|
|
|
|
|
|
|
|
|
|
|
|
M
|
|
|
Image Pixel
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M*
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
|
NM Image Pixel
|
|
|
|
|
|
|
M
|
|
|
|
|
|
|
|
|
|
|
|
|
Palette Color Lookup Table
|
|
|
|
|
|
|
|
C
|
C
|
|
|
|
|
|
|
|
|
|
|
Supplemental Palette Color Lookup Table
|
|
|
C
|
|
C
|
|
|
|
|
|
C
|
|
|
|
|
|
|
|
|
Enhanced Palette Color Lookup Table
|
|
|
|
|
|
|
|
|
|
U
|
|
|
|
|
|
|
|
|
|
Contrast/Bolus
|
C
|
C
|
|
C
|
|
|
|
C*
|
C
|
|
|
|
|
|
|
|
|
|
|
Enhanced Contrast/Bolus
|
|
|
C
|
|
C
|
C
|
|
|
|
C
|
C
|
|
|
|
|
|
|
|
|
Cine
|
|
|
|
|
|
|
|
|
M
|
|
|
|
C
|
C
|
C
|
C
|
|
|
|
Multi-frame
|
|
|
|
|
|
|
M
|
|
M
|
|
M
|
|
M
|
M
|
M
|
M
|
|
|
|
NM Multi-frame
|
|
|
|
|
|
|
M
|
|
|
|
|
|
|
|
|
|
|
|
|
Frame Pointers
|
|
|
|
|
|
|
|
|
U
|
|
|
|
U
|
U
|
U
|
U
|
|
|
|
Multi-frame Functional Groups
|
|
|
M
|
|
M
|
M
|
|
|
|
M
|
M
|
|
|
U
|
U
|
U
|
|
M
|
|
Multi-frame Dimension
|
|
|
M
|
|
M
|
M
|
|
|
|
M
|
M
|
|
|
U
|
U
|
U
|
|
M
|
|
Excluded Intervals
|
|
|
|
|
|
|
|
|
|
U
|
|
|
|
|
|
|
|
|
|
Display Shutter
|
U
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Device
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
|
|
Intervention
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
U
|
|
Specimen
|
U
|
U
|
U
|
U
|
U
|
C
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
|
CR Image
|
M
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
CT Image
|
|
M
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Enhanced CT Image
|
|
|
M
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
MR Image
|
|
|
|
M
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Enhanced MR Image
|
|
|
|
|
M
|
M
|
|
|
|
|
|
|
|
|
|
|
|
|
|
MR Pulse Sequence
|
|
|
|
|
C
|
C
|
|
|
|
|
|
|
|
|
|
|
|
|
|
NM Image
|
|
|
|
|
|
|
M
|
|
|
|
|
|
|
|
|
|
|
|
|
NM Isotope
|
|
|
|
|
|
|
M
|
|
|
|
|
|
|
|
|
|
|
|
|
NM Detector
|
|
|
|
|
|
|
M
|
|
|
|
|
|
|
|
|
|
|
|
|
NM TOMO Acquisition
|
|
|
|
|
|
|
C
|
|
|
|
|
|
|
|
|
|
|
|
|
NM Multi-Gated Acquisition
|
|
|
|
|
|
|
C
|
|
|
|
|
|
|
|
|
|
|
|
|
NM Phase
|
|
|
|
|
|
|
C
|
|
|
|
|
|
|
|
|
|
|
|
|
NM Reconstruction
|
|
|
|
|
|
|
C
|
|
|
|
|
|
|
|
|
|
|
|
|
US Region Calibration
|
|
|
|
|
|
|
|
U*
|
U
|
|
|
|
|
|
|
|
|
|
|
US Image
|
|
|
|
|
|
|
|
M*
|
M
|
|
|
|
|
|
|
|
|
|
|
Enhanced US Image
|
|
|
|
|
|
|
|
|
|
M
|
|
|
|
|
|
|
|
|
|
IVUS Image
|
|
|
|
|
|
|
|
|
|
C
|
|
|
|
|
|
|
|
|
|
SC Image
|
|
|
|
|
|
|
|
|
|
|
|
M
|
U
|
U
|
U
|
U
|
|
|
|
SC Multi-frame Image
|
|
|
|
|
|
|
|
|
|
|
|
|
M
|
M
|
M
|
M
|
|
|
|
SC Multi-frame Vector
|
|
|
|
|
|
|
|
|
|
|
|
|
C
|
C
|
C
|
C
|
|
|
|
PET Image
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
M
|
|
|
Enhanced PET Isotope
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
M
|
|
Enhanced PET Acquisition
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
M
|
|
Enhanced PET Image
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
M
|
|
Photoacoustic Image
|
|
|
|
|
|
|
|
|
|
|
M
|
|
|
|
|
|
|
|
|
Photoacoustic Acquisition Parameters
|
|
|
|
|
|
|
|
|
|
|
M
|
|
|
|
|
|
|
|
|
Photoacoustic Transducer
|
|
|
|
|
|
|
|
|
|
|
U
|
|
|
|
|
|
|
|
|
Photoacoustic Reconstruction
|
|
|
|
|
|
|
|
|
|
|
U
|
|
|
|
|
|
|
|
|
Overlay Plane
|
U
|
U
|
|
U
|
|
|
U
|
U*
|
U
|
|
|
U
|
|
|
|
|
U
|
|
|
Multi-frame Overlay
|
|
|
|
|
|
|
U
|
|
U
|
|
|
|
|
|
|
|
|
|
|
Modality LUT
|
U
|
|
|
|
|
|
|
|
|
|
|
U
|
|
|
|
|
|
|
|
VOI LUT
|
U
|
U
|
|
U
|
|
|
U
|
U*
|
U
|
|
|
U
|
|
C
|
C
|
|
U
|
|
|
Common Instance Reference
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
|
Acquisition Context
|
|
|
M
|
|
M
|
M
|
U
|
|
|
M
|
M
|
|
|
|
|
|
U
|
M
|
|
ICC Profile
|
|
|
U
|
|
U
|
M
|
U
|
U
|
U
|
U
|
C
|
U
|
|
|
|
U
|
|
|
|
SOP Common
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M*
|
M*
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
|
Frame Extraction
|
|
|
C
|
|
C
|
C
|
C
|
|
C
|
C
|
C
|
|
C
|
C
|
C
|
C
|
|
C
|
* The notation next to M and U indicates a special condition for these Modules. Refer to the corresponding IODs in this Annex for details.
Table A.1-1b. Composite Information Object Modules Overview - More Images
|
IODs
Modules
|
XA
|
Enh XA
|
RF
|
Enh RF
|
3D XA
|
3D CF
|
Br Tomo
|
Br Proj
|
RT IM
|
DX
|
MG
|
IO
|
Par Map
|
|
Patient
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
|
Clinical Trial Subject
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
|
General Study
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
|
Patient Study
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
|
Clinical Trial Study
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
|
General Series
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
|
M
|
M
|
M
|
M
|
|
Clinical Trial Series
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
|
Enhanced Series
|
|
|
|
|
M
|
M
|
|
|
|
|
|
|
|
|
RT Series
|
|
|
|
|
|
|
|
|
M
|
|
|
|
|
|
DX Series
|
|
|
|
|
|
|
|
M
|
|
M
|
M
|
M
|
|
|
Mammo Series
|
|
|
|
|
|
|
|
|
|
|
M
|
|
|
|
Enhanced Mammo Series
|
|
|
|
|
|
|
M
|
M
|
|
|
|
|
|
|
Intra-Oral Series
|
|
|
|
|
|
|
|
|
|
|
|
M
|
|
|
XA/XRF Series
|
|
M
|
|
M
|
|
|
|
|
|
|
|
|
|
|
Parametric Map Series
|
|
|
|
|
|
|
|
|
|
|
|
|
M
|
|
Frame of Reference
|
|
C
|
|
U
|
M
|
M
|
M
|
M
|
U
|
U
|
C
|
U
|
M
|
|
Synchronization
|
U
|
C
|
U
|
C
|
U
|
|
C
|
C
|
|
|
|
|
U
|
|
Cardiac Synchronization
|
|
C
|
|
C
|
C
|
|
|
|
|
|
|
|
U
|
|
Respiratory Synchronization
|
|
C
|
|
C
|
C
|
|
|
|
|
|
|
|
U
|
|
Bulk Motion Synchronization
|
|
|
|
|
|
|
|
|
|
|
|
|
U
|
|
General Equipment
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
|
Enhanced General Equipment
|
|
M
|
|
M
|
M
|
M
|
M
|
M
|
|
|
|
|
M
|
|
Acquisition
|
M
|
|
M
|
|
|
|
|
|
M
|
M
|
M
|
M
|
U
|
|
Multi-Resolution Pyramid
|
|
|
|
|
|
|
|
|
|
|
|
|
U
|
|
General Image
|
M
|
|
M
|
|
|
|
|
|
M
|
M
|
M
|
M
|
M
|
|
General Reference
|
U
|
|
U
|
|
|
|
|
|
U
|
U
|
U
|
U
|
U
|
|
Image Pixel
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
C
|
|
Microscope Slide Layer Tile Organization
|
|
|
|
|
|
|
|
|
|
|
|
|
C
|
|
Floating Point Image Pixel
|
|
|
|
|
|
|
|
|
|
|
|
|
C
|
|
Double Floating Point Image Pixel
|
|
|
|
|
|
|
|
|
|
|
|
|
C
|
|
Palette Color Lookup Table
|
|
|
|
|
|
|
|
|
|
|
|
|
C
|
|
Contrast/Bolus
|
C
|
|
C
|
|
|
|
|
|
C
|
U
|
U
|
U
|
|
|
Enhanced Contrast/Bolus
|
|
C
|
|
C
|
C
|
C
|
C
|
C
|
|
|
|
|
|
|
Cine
|
C
|
|
C
|
|
|
|
|
|
C
|
|
|
|
|
|
Multi-frame
|
C
|
|
C
|
|
|
|
|
|
C
|
|
|
|
|
|
Frame Pointers
|
U
|
|
U
|
|
|
|
|
|
|
|
|
|
|
|
Multi-frame Functional Groups
|
|
M
|
|
M
|
M
|
M
|
M
|
M
|
|
|
|
|
M
|
|
Multi-frame Dimension
|
|
U
|
|
U
|
U
|
U
|
U
|
U
|
|
|
|
|
M
|
|
Mask
|
C
|
U
|
C
|
U
|
|
|
|
|
|
|
|
|
|
|
Display Shutter
|
U
|
|
U
|
|
|
|
|
|
|
U
|
U
|
U
|
|
|
Device
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
|
Intervention
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
|
U
|
U
|
U
|
|
|
Specimen
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
|
U
|
U
|
U
|
U
|
|
Patient Orientation
|
|
|
|
|
U
|
U
|
|
M
|
|
|
|
|
|
|
Image - Equipment Coordinate Relationship
|
|
|
|
|
U
|
U
|
U
|
|
|
|
|
|
|
|
X-Ray Image
|
M
|
|
M
|
|
|
|
|
|
|
|
|
|
|
|
Enhanced XA/XRF Image
|
|
M
|
|
M
|
|
|
|
|
|
|
|
|
|
|
X-Ray Acquisition
|
M
|
|
M
|
|
|
|
|
|
|
|
|
|
|
|
XA/XRF Acquisition
|
|
C
|
|
C
|
|
|
|
|
|
|
|
|
|
|
X-Ray Collimator
|
U
|
|
U
|
|
|
|
|
|
|
U
|
U
|
U
|
|
|
X-Ray Table
|
C
|
|
U
|
|
|
|
|
|
|
|
|
|
|
|
XRF Positioner
|
|
|
U
|
|
|
|
|
|
|
|
|
|
|
|
X-Ray Tomography Acquisition
|
|
|
C
|
U
|
|
|
|
|
|
U
|
U
|
U
|
|
|
X-Ray Acquisition Dose
|
|
|
|
|
|
|
|
|
|
U
|
U
|
U
|
|
|
X-Ray Generation
|
|
|
|
|
|
|
|
|
|
U
|
U
|
U
|
|
|
X-Ray Filtration
|
|
U
|
|
U
|
|
|
|
|
|
U
|
U
|
U
|
|
|
X-Ray Grid
|
|
U
|
|
U
|
|
|
|
|
|
U
|
U
|
U
|
|
|
XA Positioner
|
M
|
|
|
|
|
|
|
|
|
|
|
|
|
|
X-Ray Image Intensifier
|
|
C
|
|
C
|
|
|
|
|
|
|
|
|
|
|
X-Ray Detector
|
|
C
|
|
C
|
|
|
|
|
|
|
|
|
|
|
XA/XRF Multi-frame Presentation
|
|
U
|
|
U
|
|
|
|
|
|
|
|
|
|
|
X-Ray 3D Image
|
|
|
|
|
M
|
M
|
M
|
|
|
|
|
|
|
|
X-Ray 3D Angiographic Image Contributing Sources
|
|
|
|
|
U
|
|
|
|
|
|
|
|
|
|
X-Ray 3D Craniofacial Image Contributing Sources
|
|
|
|
|
|
U
|
|
|
|
|
|
|
|
|
Breast Tomosynthesis Contributing Sources
|
|
|
|
|
|
|
U
|
|
|
|
|
|
|
|
X-Ray 3D Angiographic Acquisition
|
|
|
|
|
U
|
|
|
|
|
|
|
|
|
|
X-Ray 3D Craniofacial Acquisition
|
|
|
|
|
|
U
|
|
|
|
|
|
|
|
|
Breast Tomosynthesis Acquisition
|
|
|
|
|
|
|
U
|
|
|
|
|
|
|
|
X-Ray 3D Reconstruction
|
|
|
|
|
U
|
U
|
U
|
|
|
|
|
|
|
|
Enhanced Mammography Image
|
|
|
|
|
|
|
|
M
|
|
|
|
|
|
|
Breast View
|
|
|
|
|
|
|
M
|
M
|
|
|
|
|
|
|
DX Anatomy Imaged
|
|
|
|
|
|
|
|
|
|
M
|
M
|
M
|
|
|
DX Image
|
|
|
|
|
|
|
|
|
|
M
|
M
|
M
|
|
|
DX Detector
|
U
|
|
U
|
|
|
|
|
|
|
M
|
M
|
M
|
|
|
DX Positioning
|
|
|
|
|
|
|
|
|
|
U
|
U
|
U
|
|
|
Mammo Image
|
|
|
|
|
|
|
|
|
|
|
M
|
|
|
|
Intra-Oral Image
|
|
|
|
|
|
|
|
|
|
|
|
M
|
|
|
RT Image
|
|
|
|
|
|
|
|
|
M
|
|
|
|
|
|
Parametric Map Image
|
|
|
|
|
|
|
|
|
|
|
|
|
M
|
|
Approval
|
|
|
|
|
|
|
|
|
U
|
|
|
|
|
|
Overlay Plane
|
U
|
|
U
|
|
|
|
|
|
|
C
|
C
|
C
|
|
|
Multi-frame Overlay
|
C
|
|
C
|
|
|
|
|
|
|
|
|
|
|
|
Modality LUT
|
C*
|
|
C*
|
|
|
|
|
|
U
|
|
|
|
|
|
VOI LUT
|
U
|
|
U
|
|
|
|
|
|
U
|
C
|
C
|
C
|
|
|
Image Histogram
|
|
|
|
|
|
|
|
|
|
U
|
U
|
U
|
|
|
Common Instance Reference
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
C
|
|
Acquisition Context
|
|
M
|
|
M
|
M
|
M
|
M
|
M
|
|
M
|
M
|
M
|
M
|
|
SOP Common
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
|
Frame Extraction
|
C
|
C
|
C
|
C
|
C
|
C
|
C
|
C
|
C
|
|
|
|
C
|
Table A.1-1c. Composite Information Object Modules Overview - More Images
|
IODs
Modules
|
VL EN
|
VL MC
|
VL SL
|
VL WS
|
VL PH
|
DMS PH
|
CFM
|
CFM TP
|
Vid VL EN
|
Vid VL MC
|
Vid VL PH
|
|
Patient
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
|
Clinical Trial Subject
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
|
General Study
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
|
Patient Study
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
|
Clinical Trial Study
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
|
General Series
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
|
Clinical Trial Series
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
|
Whole Slide Microscopy Series
|
|
|
|
M
|
|
|
|
|
|
|
|
|
Frame of Reference
|
|
|
M
|
M
|
|
U
|
M
|
M
|
|
|
|
|
Synchronization
|
|
|
|
|
|
|
C
|
C
|
|
|
U
|
|
General Equipment
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
|
Enhanced General Equipment
|
|
|
|
M
|
|
M
|
M
|
M
|
|
|
|
|
VL Photographic Equipment
|
|
|
|
|
U
|
U
|
|
|
|
|
|
|
Acquisition
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
|
Multi-Resolution Pyramid
|
|
|
|
U
|
|
|
|
U
|
|
|
|
|
General Image
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
|
General Reference
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
|
Microscope Slide Layer Tile Organization
|
|
|
|
M
|
|
|
|
C
|
|
|
|
|
Image Pixel
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
|
Cine
|
|
|
|
|
|
|
|
|
M
|
M
|
M
|
|
Multi-frame
|
|
|
|
|
|
|
|
|
M
|
M
|
M
|
|
Multi-frame Functional Groups
|
|
|
|
M
|
|
|
M
|
M
|
|
|
|
|
Multi-frame Dimension
|
|
|
|
M
|
|
|
M
|
M
|
|
|
|
|
Device
|
U
|
U
|
U
|
|
U
|
|
|
|
U
|
U
|
U
|
|
Specimen
|
U
|
C
|
M
|
M
|
C
|
|
C
|
C
|
C
|
C
|
C
|
|
VL Image
|
M
|
M
|
M
|
|
M
|
M
|
|
|
M
|
M
|
M
|
|
VL Photographic Acquisition
|
|
|
|
|
U
|
U
|
|
|
|
|
|
|
VL Photographic Geolocation
|
|
|
|
|
U
|
|
|
|
|
|
|
|
Slide Coordinates
|
|
|
M
|
|
|
|
|
|
|
|
|
|
Whole Slide Microscopy Image
|
|
|
|
M
|
|
|
|
|
|
|
|
|
Optical Path
|
|
U
|
U
|
M
|
|
|
M
|
M
|
|
|
|
|
Slide Label
|
|
|
|
C
|
|
|
|
|
|
|
|
|
Dermoscopic Image
|
|
|
|
|
|
M
|
|
|
|
|
|
|
Confocal Microscopy Image
|
|
|
|
|
|
|
M
|
M
|
|
|
|
|
Confocal Microscopy Tiled Pyramidal Image
|
|
|
|
|
|
|
|
M
|
|
|
|
|
Cutaneous Confocal Microscopy Image Acquisition Parameters
|
|
|
|
|
|
|
C
|
C
|
|
|
|
|
Overlay Plane
|
U
|
U
|
U
|
|
U
|
|
|
|
|
|
|
|
Common Instance Reference
|
U
|
U
|
U
|
M
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
|
Acquisition Context
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
|
ICC Profile
|
U
|
U
|
U
|
U
|
U
|
U
|
|
|
U
|
U
|
U
|
|
SOP Common
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
|
Frame Extraction
|
|
|
|
C
|
|
|
C
|
C
|
C
|
C
|
C
|
Table A.1-1d. Composite Information Object Modules Overview - More Images
|
IODs
Modules
|
Oph 8 Bit
|
Oph 16 Bit
|
WF Oph St
|
WF Oph 3DC
|
OPT
|
OCT ENF
|
OCT BSV
|
OPM
|
CM
|
|
Patient
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
|
Clinical Trial Subject
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
|
General Study
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
|
Patient Study
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
|
Clinical Trial Study
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
|
General Series
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
|
Clinical Trial Series
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
|
Ophthalmic Tomography Series
|
|
|
|
|
M
|
|
|
|
|
|
Ophthalmic Tomography En Face Series
|
|
|
|
|
|
M
|
|
|
|
|
Ophthalmic Tomography B-scan Volume Analysis Series
|
|
|
|
|
|
|
M
|
|
|
|
Ophthalmic Thickness Map Series
|
|
|
|
|
|
|
|
M
|
|
|
Corneal Topography Map Series
|
|
|
|
|
|
|
|
|
M
|
|
Frame of Reference
|
|
|
M
|
M
|
C
|
M
|
M
|
|
M
|
|
Synchronization
|
M
|
M
|
M
|
M
|
C
|
|
|
|
|
|
Cardiac Synchronization
|
|
|
|
|
C
|
|
|
|
|
|
General Equipment
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
|
Enhanced General Equipment
|
|
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
|
Acquisition
|
M
|
M
|
M
|
M
|
|
M
|
|
M
|
M
|
|
General Image
|
M
|
M
|
M
|
M
|
|
M
|
|
M
|
M
|
|
General Reference
|
U
|
U
|
U
|
U
|
|
|
|
U
|
U
|
|
Image Pixel
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
|
Palette Color Lookup Table
|
|
|
|
|
|
C
|
|
|
M
|
|
Supplemental Palette Color Lookup Table
|
|
|
|
|
|
|
|
C
|
|
|
Enhanced Contrast/Bolus
|
C
|
C
|
C
|
C
|
C
|
|
|
|
|
|
Cine
|
C
|
C
|
C
|
C
|
|
|
|
|
|
|
Multi-frame
|
M
|
M
|
M
|
M
|
|
|
|
|
|
|
Multi-frame Functional Groups
|
|
|
|
|
M
|
|
M
|
|
|
|
Multi-frame Dimension
|
|
|
|
|
M
|
|
M
|
|
|
|
Bitmap Display Shutter
|
|
|
|
|
|
|
|
C
|
|
|
Device
|
U
|
U
|
|
|
|
|
|
|
|
|
Ophthalmic Photography Image
|
M
|
M
|
M
|
M
|
|
|
|
|
|
|
Wide Field Ophthalmic Photography Stereographic Projection
|
|
|
M
|
|
|
|
|
|
|
|
Wide Field Ophthalmic Photography 3D Coordinates
|
|
|
|
M
|
|
|
|
|
|
|
Wide Field Ophthalmic Photography Quality Rating
|
|
|
C
|
C
|
|
|
|
|
|
|
Ocular Region Imaged
|
M
|
M
|
M
|
M
|
M
|
M
|
|
|
|
|
Ophthalmic Photography Acquisition Parameters
|
M
|
M
|
M
|
M
|
|
|
|
M
|
M
|
|
Ophthalmic Photographic Parameters
|
M
|
M
|
M
|
M
|
|
|
|
|
|
|
Ophthalmic Tomography Image
|
|
|
|
|
M
|
|
|
|
|
|
Ophthalmic Tomography Parameters
|
|
|
|
|
M
|
|
|
|
|
|
Ophthalmic Tomography Acquisition Parameters
|
|
|
|
|
M
|
|
|
|
|
|
Ophthalmic Thickness Map
|
|
|
|
|
|
|
|
M
|
|
|
Ophthalmic Thickness Map Quality Rating
|
|
|
|
|
|
|
|
C
|
|
|
Corneal Topography Map Image
|
|
|
|
|
|
|
|
|
M
|
|
Corneal Topography Map Analysis
|
|
|
|
|
|
|
|
|
M
|
|
Ophthalmic Optical Coherence Tomography En Face Image
|
|
|
|
|
|
M
|
|
|
|
|
Ophthalmic Optical Coherence Tomography B-scan Volume Analysis Image
|
|
|
|
|
|
|
M
|
|
|
|
Ophthalmic Optical Coherence Tomography En Face Image Quality Rating
|
|
|
|
|
|
C
|
|
|
|
|
Common Instance Reference
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
|
|
Acquisition Context
|
U
|
U
|
U
|
U
|
M
|
|
|
M
|
M
|
|
ICC Profile
|
U
|
U
|
C
|
C
|
|
U
|
|
|
U
|
|
SOP Common
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
|
Frame Extraction
|
C
|
C
|
C
|
C
|
C
|
|
C
|
|
|
Table A.1-1e. Composite Information Object Modules Overview - More Images
|
IODs
Modules
|
IVOCT
|
Seg
|
Hgt Map Seg
|
Leg Conv Enh CT
|
Leg Conv Enh MR
|
Leg Conv Enh PET
|
|
Patient
|
M
|
M
|
M
|
M
|
M
|
M
|
|
Clinical Trial Subject
|
U
|
U
|
U
|
U
|
U
|
U
|
|
General Study
|
M
|
M
|
M
|
M
|
M
|
M
|
|
Patient Study
|
U
|
U
|
U
|
U
|
U
|
U
|
|
Clinical Trial Study
|
U
|
U
|
U
|
U
|
U
|
U
|
|
General Series
|
M
|
M
|
M
|
M
|
M
|
M
|
|
Clinical Trial Series
|
U
|
U
|
U
|
U
|
U
|
U
|
|
CT Series
|
|
|
|
M
|
|
|
|
MR Series
|
|
|
|
|
M
|
|
|
Enhanced PET Series
|
|
|
|
|
|
M
|
|
Segmentation Series
|
|
M
|
M
|
|
|
|
|
Intravascular OCT Series
|
M
|
|
|
|
|
|
|
Frame of Reference
|
M
|
C
|
M
|
M
|
M
|
M
|
|
Synchronization
|
M
|
|
|
U
|
U
|
U
|
|
Cardiac Synchronization
|
C
|
|
|
U
|
U
|
U
|
|
Respiratory Synchronization
|
|
|
|
U
|
U
|
U
|
|
Bulk Motion Synchronization
|
|
|
|
|
U
|
|
|
General Equipment
|
M
|
M
|
M
|
M
|
M
|
M
|
|
Enhanced General Equipment
|
M
|
M
|
M
|
U
|
U
|
U
|
|
Acquisition
|
|
U
|
M
|
|
|
|
|
Multi-Resolution Pyramid
|
|
U
|
|
|
|
|
|
General Image
|
|
M
|
M
|
|
|
|
|
General Reference
|
|
U
|
U
|
|
|
|
|
Microscope Slide Layer Tile Organization
|
|
C
|
|
|
|
|
|
Image Pixel
|
M
|
M
|
|
M
|
M
|
M
|
|
Floating Point Image Pixel
|
|
|
M
|
|
|
|
|
Palette Color Lookup Table
|
|
C
|
|
|
|
|
|
Supplemental Palette Color Lookup Table
|
C
|
|
|
|
|
|
|
Contrast/Bolus
|
|
|
|
U
|
U
|
|
|
Enhanced Contrast/Bolus
|
M
|
|
|
U
|
U
|
|
|
Multi-frame Functional Groups
|
M
|
M
|
M
|
M
|
M
|
M
|
|
Multi-frame Dimension
|
M
|
M
|
M
|
U
|
U
|
U
|
|
Device
|
U
|
|
|
U
|
U
|
|
|
Specimen
|
|
U
|
|
U
|
U
|
U
|
|
Intervention
|
|
|
|
|
|
U
|
|
Intravascular OCT Image
|
M
|
|
|
|
|
|
|
Intravascular OCT Acquisition Parameters
|
M
|
|
|
|
|
|
|
Intravascular OCT Processing Parameters
|
C
|
|
|
|
|
|
|
Intravascular Image Acquisition Parameters
|
M
|
|
|
|
|
|
|
Enhanced CT Image
|
|
|
|
M
|
|
|
|
Enhanced MR Image
|
|
|
|
|
M
|
|
|
Enhanced PET Image
|
|
|
|
|
|
M
|
|
Segmentation Image
|
|
M
|
|
|
|
|
|
Height Map Segmentation Image
|
|
|
M
|
|
|
|
|
Common Instance Reference
|
M
|
C
|
M
|
M
|
M
|
M
|
|
ICC Profile
|
|
C
|
M
|
|
|
|
|
Acquisition Context
|
M
|
|
|
M
|
M
|
M
|
|
SOP Common
|
M
|
M
|
M
|
M
|
M
|
M
|
|
Frame Extraction
|
C
|
C
|
C
|
C
|
C
|
C
|
Table A.1-2. Composite Information Object Modules Overview - Non-Images
|
IODs
Modules
|
MR Spectroscopy
|
Raw Data
|
Surface Scan Point Cloud
|
Surface Scan Mesh
|
Content Assessment Result
|
WS Bulk Smpl Ann
|
|
Patient
|
M
|
M
|
M
|
M
|
M
|
M
|
|
Clinical Trial Subject
|
U
|
U
|
U
|
U
|
U
|
U
|
|
General Study
|
M
|
M
|
M
|
M
|
M
|
M
|
|
Patient Study
|
U
|
U
|
U
|
U
|
U
|
U
|
|
Clinical Trial Study
|
U
|
U
|
U
|
U
|
U
|
U
|
|
General Series
|
M
|
M
|
M
|
M
|
M
|
M
|
|
Clinical Trial Series
|
U
|
U
|
U
|
U
|
U
|
U
|
|
Optical Surface Scanner Series
|
|
|
M
|
M
|
|
|
|
Microscopy Bulk Simple Annotations Series
|
|
|
|
|
|
M
|
|
Frame of Reference
|
M
|
U
|
M
|
M
|
|
C
|
|
Sync.
|
C
|
C
|
|
|
|
|
|
Cardiac Sync.
|
C
|
|
|
|
|
|
|
Respiratory Sync.
|
C
|
|
|
|
|
|
|
Bulk Motion Sync.
|
C
|
|
|
|
|
|
|
General Equip.
|
M
|
M
|
M
|
M
|
M
|
M
|
|
Enhanced General Equip.
|
M
|
|
M
|
M
|
M
|
M
|
|
Enhanced Patient Orientation
|
U
|
|
|
|
|
|
|
Multi-frame Functional Groups
|
M
|
|
|
|
|
|
|
Multi-frame Dim.
|
M
|
|
|
|
|
|
|
Specimen
|
U
|
U
|
U
|
U
|
|
|
|
Raw Data
|
|
M
|
|
|
|
|
|
MR Spect.
|
M
|
|
|
|
|
|
|
MR Spect. Pulse Sequence
|
C
|
|
|
|
|
|
|
MR Spect. Data
|
M
|
|
|
|
|
|
|
Point Cloud
|
|
|
M
|
|
|
|
|
Surface Mesh
|
|
|
|
M
|
|
|
|
UV Mapping
|
|
|
U
|
U
|
|
|
|
Scan Procedure
|
|
|
M
|
M
|
|
|
|
Content Assessment Results
|
|
|
|
|
M
|
|
|
Microscopy Bulk Simple Annotations
|
|
|
|
|
|
M
|
|
Acquisition Context
|
M
|
M
|
|
|
|
|
|
Common Instance Reference
|
|
|
|
|
M
|
M
|
|
ICC Profile
|
|
|
|
|
|
U
|
|
SOP Common
|
M
|
M
|
M
|
M
|
M
|
M
|
|
Frame Extraction
|
C
|
|
|
|
|
|
Table A.1-3. Composite Information Object Modules Overview - More Non-Images
|
IODs
Modules
|
Ln Mx
|
Aut Ref Mx
|
Ker Mx
|
Sub Ref Mx
|
Vis Acy Mx
|
Op Ax Mx
|
IO Ln Cl
|
OPV
|
Reg
|
Def Reg
|
Fid
|
RWV
|
Str Rl
|
Sfc Seg
|
Bs St Dsp
|
Trc Res
|
|
Patient
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
|
Clinical Trial Subject
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
|
General Study
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
|
Patient Study
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
|
Clinical Trial Study
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
|
General Series
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
|
M
|
M
|
M
|
M
|
|
Clinical Trial Series
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
|
|
U
|
|
U
|
U
|
U
|
U
|
|
Presentation Series
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
M
|
|
|
Spatial Registration Series
|
|
|
|
|
|
|
|
|
M
|
M
|
|
|
|
|
|
|
|
Spatial Fiducials Series
|
|
|
|
|
|
|
|
|
|
|
M
|
|
|
|
|
|
|
Real World Value Mapping Series
|
|
|
|
|
|
|
|
|
|
|
|
M
|
|
|
|
|
|
Stereometric Series
|
|
|
|
|
|
|
|
|
|
|
|
|
M
|
|
|
|
|
Segmentation Series
|
|
|
|
|
|
|
|
|
|
|
|
|
|
M
|
|
|
|
Lensometry Msrmnts. Series
|
M
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Autorefraction Msrmnts. Series
|
|
M
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Keratometry Msrmnts. Series
|
|
|
M
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Subjective Refraction Msrmnts. Series
|
|
|
|
M
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Visual Acuity Msrmnts. Series
|
|
|
|
|
M
|
|
|
|
|
|
|
|
|
|
|
|
|
Ophthalmic Axial Msrmnts. Series
|
|
|
|
|
|
M
|
|
|
|
|
|
|
|
|
|
|
|
Intraocular Lens Calculations Series
|
|
|
|
|
|
|
M
|
|
|
|
|
|
|
|
|
|
|
Visual Field Static Perimetry Msrmnts. Series
|
|
|
|
|
|
|
|
M
|
|
|
|
|
|
|
|
|
|
Tractography Results Series
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
M
|
|
Frame of Reference
|
|
|
|
|
|
|
|
|
M
|
M
|
|
|
|
M
|
|
M
|
|
General Equipment
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
|
Enhanced General Equipment
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
|
M
|
|
|
|
M
|
M
|
M
|
|
Specimen
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
U
|
|
Stereometric Relationship
|
|
|
|
|
|
|
|
|
|
|
|
|
M
|
|
|
|
|
Spatial Registration
|
|
|
|
|
|
|
|
|
M
|
|
|
|
|
|
|
|
|
Deformable Spatial Registration
|
|
|
|
|
|
|
|
|
|
M
|
|
|
|
|
|
|
|
Spatial Fiducials
|
|
|
|
|
|
|
|
|
|
|
M
|
|
|
|
|
|
|
Real World Value Mapping
|
|
|
|
|
|
|
|
|
|
|
|
M
|
|
|
|
|
|
Surface Segmentation
|
|
|
|
|
|
|
|
|
|
|
|
|
|
M
|
|
|
|
Surface Mesh
|
|
|
|
|
|
|
|
|
|
|
|
|
|
M
|
|
|
|
Structured Display
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
M
|
|
|
Structured Display Image Box
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
M
|
|
|
Structured Display Annotation
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
U
|
|
|
General Ophthalmic Refractive Msrmnts.
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
|
|
|
|
|
|
|
|
|
|
Lensometry Msrmnts.
|
M
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Autorefraction Msrmnts.
|
|
M
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Keratometry Msrmnts.
|
|
|
M
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Subjective Refraction Msrmnts.
|
|
|
|
M
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Visual Acuity Msrmnts.
|
|
|
|
|
M
|
|
|
|
|
|
|
|
|
|
|
|
|
Ophthalmic Axial Msrmnts.
|
|
|
|
|
|
M
|
|
|
|
|
|
|
|
|
|
|
|
Intraocular Lens Calculations
|
|
|
|
|
|
|
M
|
|
|
|
|
|
|
|
|
|
|
Visual Field Static Perimetry Test Parameters
|
|
|
|
|
|
|
|
M
|
|
|
|
|
|
|
|
|
|
Visual Field Static Perimetry Test Reliability
|
|
|
|
|
|
|
|
M
|
|
|
|
|
|
|
|
|
|
Visual Field Static Perimetry Test Msrmnts.
|
|
|
|
|
|
|
|
M
|
|
|
|
|
|
|
|
|
|
Visual Field Static Perimetry Test Results
|
|
|
|
|
|
|
|
M
|
|
|
|
|
|
|
|
|
|
Ophthalmic Patient Clinical Information and Test Lens Parameters
|
|
|
|
|
|
|
|
U
|
|
|
|
|
|
|
|
|
|
Common Instance Reference
|
|
|
|
|
|
|
|
|
M
|
M
|
M
|
M
|
M
|
C
|
M
|
M
|
|
General Reference
|
|
|
|
|
|
|
|
|
U
|
U
|
U
|
|
|
U
|
|
|
|
Tractography Results
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
M
|
|
SOP Common
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
Table A.1-4. Composite Information Object Modules Overview - Radiotherapy
|
IODs
Modules
|
RT Dose
|
RT Struc Set
|
RT Plan
|
RT Beam Rec
|
RT Brachy Rec
|
RT Sum
|
RT Ion Plan
|
RT Ion Beams Treat Rec
|
RT Beams Delivery Inst
|
RT Brachy App Setup Delivery Inst
|
|
Patient
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
|
Clinical Trial Subject
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
|
General Study
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
|
Patient Study
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
|
Clinical Trial Study
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
|
General Series
|
|
|
|
|
|
|
|
|
M
|
M
|
|
Clinical Trial Series
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
|
RT Series
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
|
|
|
Frame of Reference
|
M
|
U
|
U
|
|
|
|
M
|
|
|
|
|
General Equipment
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
|
Enhanced General Equipment
|
|
|
|
|
|
|
|
|
|
M
|
|
General Image
|
C
|
|
|
|
|
|
|
|
|
|
|
Image Plane
|
C
|
|
|
|
|
|
|
|
|
|
|
Image Pixel
|
C
|
|
|
|
|
|
|
|
|
|
|
Multi-frame
|
C
|
|
|
|
|
|
|
|
|
|
|
RT Dose
|
M
|
|
|
|
|
|
|
|
|
|
|
RT DVH
|
U
|
|
|
|
|
|
|
|
|
|
|
Structure Set
|
|
M
|
|
|
|
|
|
|
|
|
|
ROI Contour
|
|
M
|
|
|
|
|
|
|
|
|
|
RT ROI Observations
|
|
M
|
|
|
|
|
|
|
|
|
|
RT General Treatment Record
|
|
|
|
M
|
M
|
M
|
|
M
|
|
|
|
RT Treatment Machine Record
|
|
|
|
M
|
M
|
|
|
M
|
|
|
|
Measured Dose Reference Record
|
|
|
|
U
|
U
|
|
|
U
|
|
|
|
Calculated Dose Reference Record
|
|
|
|
U
|
U
|
|
|
U
|
|
|
|
RT Beams Session Record
|
|
|
|
M
|
|
|
|
|
|
|
|
RT Ion Beams Session Record
|
|
|
|
|
|
|
|
M
|
|
|
|
RT Brachy Session Record
|
|
|
|
|
M
|
|
|
|
|
|
|
RT Treatment Summary Record
|
|
|
|
U
|
U
|
M
|
|
U
|
|
|
|
RT General Plan
|
|
|
M
|
|
|
|
M
|
|
|
|
|
RT Prescription
|
|
|
U
|
|
|
|
U
|
|
|
|
|
RT Tolerance Tables
|
|
|
U
|
|
|
|
|
|
|
|
|
RT Ion Tolerance Tables
|
|
|
|
|
|
|
U
|
|
|
|
|
RT Patient Setup
|
|
|
U
|
U
|
U
|
|
U
|
U
|
|
|
|
RT Fraction Scheme
|
|
|
U
|
|
|
|
U
|
|
|
|
|
RT Beams
|
|
|
C
|
|
|
|
|
|
|
|
|
RT Ion Beams
|
|
|
|
|
|
|
M
|
|
|
|
|
RT Brachy Application Setups
|
|
|
C
|
|
|
|
|
|
|
|
|
RT Beams Delivery Instruction
|
|
|
|
|
|
|
|
|
M
|
|
|
RT Brachy Application Setup Delivery Instruction
|
|
|
|
|
|
|
|
|
|
M
|
|
Approval
|
|
U
|
U
|
|
|
|
U
|
|
|
|
|
Common Instance Reference
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
C
|
M
|
|
General Reference
|
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
|
SOP Common
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
|
Frame Extraction
|
C
|
|
|
|
|
|
|
|
|
|
Table A.1-4b. Composite Information Object Modules Overview - Second Generation Radiotherapy
|
IODs
Modules
|
RT Physn Intent
|
RT Seg Annot
|
RT Rad Set
|
C-Arm P-E Rad
|
Tomo Rad
|
Rob Arm Rad
|
RT Rad Rec Set
|
RT Rad Slvg Rec
|
C-Arm P-E Rec
|
Tomo Rec
|
Rob Arm Rec
|
RT Rad Set Del Inst
|
RT Treat Prep
|
|
Patient
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
|
Clinical Trial Subject
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
|
General Study
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
|
Patient Study
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
|
Clinical Trial Study
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
|
General Series
|
|
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
|
Clinical Trial Series
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
|
Enhanced RT Series
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
|
General Equipment
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
|
Enhanced General Equipment
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
|
Frame of Reference
|
|
|
M
|
M
|
M
|
M
|
|
U
|
M
|
M
|
M
|
|
|
|
Synchronization
|
|
|
|
|
|
|
|
|
C
|
C
|
C
|
|
|
|
Radiotherapy Common Instance
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
|
RT Physician Intent
|
M
|
|
|
|
|
|
|
|
|
|
|
|
|
|
RT Prescription
|
U
|
|
|
|
|
|
|
|
|
|
|
|
|
|
RT Treatment Phase Intent
|
C
|
|
|
|
|
|
|
|
|
|
|
|
|
|
RT Segment Annotation
|
|
M
|
|
|
|
|
|
|
|
|
|
|
|
|
RT Radiation Set
|
|
|
M
|
|
|
|
|
|
|
|
|
|
|
|
RT Radiation Set Delivery Instruction
|
|
|
|
|
|
|
|
|
|
|
|
M
|
|
|
RT Treatment Preparation
|
|
|
|
|
|
|
|
|
|
|
|
|
M
|
|
RT Dose Contribution
|
|
|
C
|
|
|
|
|
|
|
|
|
|
|
|
RT Radiation Record Set
|
|
|
|
|
|
|
M
|
|
|
|
|
|
|
|
RT Dose Contribution Record
|
|
|
|
|
|
|
C
|
|
|
|
|
|
|
|
RT Delivery Device Common
|
|
|
|
M
|
M
|
M
|
|
M
|
M
|
M
|
M
|
|
|
|
RT Radiation Common
|
|
|
|
M
|
M
|
M
|
|
|
|
|
|
|
|
|
RT Radiation Record Common
|
|
|
|
|
|
|
|
M
|
M
|
M
|
M
|
|
|
|
RT Radiation Salvage Record
|
|
|
|
|
|
|
|
M
|
|
|
|
|
|
|
C-Arm Photon-Electron Delivery Device
|
|
|
|
M
|
|
|
|
|
M
|
|
|
|
|
|
C-Arm Photon-Electron Beam
|
|
|
|
M
|
|
|
|
|
M
|
|
|
|
|
|
Tomotherapeutic Delivery Device
|
|
|
|
|
M
|
|
|
|
|
M
|
|
|
|
|
Tomotherapeutic Beam
|
|
|
|
|
M
|
|
|
|
|
M
|
|
|
|
|
Robotic-Arm Delivery Device
|
|
|
|
|
|
M
|
|
|
|
|
M
|
|
|
|
Robotic-Arm Path
|
|
|
|
|
|
M
|
|
|
|
|
M
|
|
|
|
Segment Reference
|
|
M
|
|
|
|
|
|
|
|
|
|
|
|
|
General Reference
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
|
Common Instance Reference
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
|
SOP Common
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
Table A.1-4c. Composite Information Object Modules Overview - More Second Generation Radiotherapy
|
IODs
Modules
|
Enh RT Img
|
Enh Cont RT Img
|
RT Pat Pos Acq Inst
|
|
Patient
|
M
|
M
|
M
|
|
Clinical Trial Subject
|
U
|
U
|
U
|
|
General Study
|
M
|
M
|
M
|
|
Patient Study
|
U
|
U
|
U
|
|
Clinical Trial Study
|
U
|
U
|
U
|
|
General Series
|
M
|
M
|
M
|
|
Clinical Trial Series
|
U
|
U
|
U
|
|
Enhanced RT Series
|
M
|
M
|
M
|
|
Frame Of Reference
|
M
|
M
|
|
|
Synchronization
|
C
|
C
|
|
|
Cardiac Synchronization
|
C
|
C
|
|
|
Respiratory Synchronization
|
C
|
C
|
|
|
General Equipment
|
M
|
M
|
M
|
|
Enhanced General Equipment
|
M
|
M
|
M
|
|
Image Pixel
|
M
|
M
|
|
|
Enhanced RT Image Device
|
M
|
M
|
|
|
Enhanced RT Image
|
M
|
M
|
|
|
RT Patient Position Acquisition Device
|
|
|
M
|
|
RT Patient Position Acquisition Instruction
|
|
|
M
|
|
Multi-frame Functional Groups
|
M
|
|
|
|
Sparse Multi-frame Functional Groups
|
|
M
|
|
|
Multi-frame Dimension
|
M
|
|
|
|
Enhanced Contrast/Bolus
|
C
|
C
|
|
|
Device
|
U
|
U
|
|
|
General Reference
|
M
|
M
|
M
|
|
Common Instance Reference
|
M
|
M
|
M
|
|
SOP Common
|
M
|
M
|
M
|
|
Radiotherapy Common Instance
|
M
|
M
|
M
|
|
Frame Extraction
|
C
|
|
|
Table A.1-5. Composite Information Object Modules Overview - Implants
|
IODs
Modules
|
Generic Implant Template
|
Implant Assembly Template
|
Implant Template Group
|
|
Generic Implant Template Description
|
M
|
|
|
|
Generic Implant Template 2D Drawings
|
C
|
|
|
|
Generic Implant Template 3D Models
|
C
|
|
|
|
Generic Implant Template Mating Features
|
U
|
|
|
|
Generic Implant Template Planning Landmarks
|
U
|
|
|
|
Implant Assembly Template
|
|
M
|
|
|
Implant Template Group
|
|
|
M
|
|
SOP Common
|
M
|
M
|
M
|
|
Surface Mesh
|
U
|
|
|
Table A.1-6. Composite Information Object Modules Overview - Presentation States
|
IODs
Modules
|
Gray Pres State
|
Col Pres State
|
Pseudo Col Pres State
|
Blend Pres State
|
XA RF Pres State
|
Planar MPR Vol Pres State
|
Vol Rend Vol Pres State
|
Adv Blend Pres State
|
Var Mod LUT Pres State
|
|
Patient
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
|
Clinical Trial Subject
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
|
General Study
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
|
Patient Study
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
|
Clinical Trial Study
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
|
General Series
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
|
Clinical Trial Series
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
|
Presn. Series
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
|
Frame of Reference
|
|
|
|
|
|
M
|
M
|
M
|
|
|
General Equip.
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
|
Enhanced General Equip.
|
|
|
|
|
M
|
M
|
M
|
M
|
M
|
|
Mask
|
C
|
|
C
|
|
|
|
|
|
C
|
|
Display Shutter
|
C
|
C
|
C
|
|
|
|
|
|
C
|
|
Bitmap Display Shutter
|
C
|
C
|
C
|
|
C
|
|
|
|
C
|
|
Palette Color LUT
|
|
|
M
|
M
|
|
|
|
|
C
|
|
Overlay Plane
|
C
|
C
|
C
|
|
C
|
|
|
|
C
|
|
Disp. Area
|
M
|
M
|
M
|
M
|
M
|
|
|
U
|
M
|
|
Overlay Actvn.
|
C
|
C
|
C
|
|
C
|
|
|
|
C
|
|
Graphic Annot.
|
C
|
C
|
C
|
C
|
C
|
U
|
U
|
U
|
C
|
|
Spatial Trans.
|
C
|
C
|
C
|
C
|
C
|
|
|
C
|
C
|
|
Graphic Layer
|
C
|
C
|
C
|
C
|
C
|
C
|
C
|
C
|
C
|
|
Graphic Group
|
U
|
U
|
U
|
U
|
|
U
|
U
|
U
|
U
|
|
Modality LUT
|
C
|
|
C
|
|
|
|
|
|
|
|
Variable Modality LUT
|
|
|
|
|
|
|
|
|
C
|
|
Softcopy VOI LUT
|
C
|
|
C
|
|
C
|
|
|
|
C
|
|
Softcopy Presn. LUT
|
M
|
|
|
|
M
|
|
|
|
C
|
|
Presn. State Ident.
|
M
|
M
|
M
|
M
|
M
|
|
|
M
|
M
|
|
Presn. State Reln.
|
M
|
M
|
M
|
|
M
|
|
|
|
M
|
|
Presn. State Shutter
|
M
|
M
|
M
|
|
M
|
|
|
|
M
|
|
Presn. State Mask
|
M
|
|
M
|
|
|
|
|
|
M
|
|
Presn. State Blending
|
|
|
|
M
|
|
|
|
|
|
|
Adv. Presn. State Blending
|
|
|
|
|
|
|
|
M
|
|
|
Adv. Presn. State Display
|
|
|
|
|
|
|
|
M
|
|
|
XA/XRF Presn. State Mask
|
|
|
|
|
C
|
|
|
|
|
|
XA/XRF Presn. State Shutter
|
|
|
|
|
C
|
|
|
|
|
|
XA/XRF Presn. State Presn.
|
|
|
|
|
C
|
|
|
|
|
|
Volumetric Presentation State Identification
|
|
|
|
|
|
M
|
M
|
|
|
|
Volumetric Presentation State Relationship
|
|
|
|
|
|
M
|
M
|
|
|
|
Volume Cropping
|
|
|
|
|
|
C
|
C
|
|
|
|
Presentation View Description
|
|
|
|
|
|
M
|
M
|
|
|
|
Multi-Planar Reconstruction Geometry
|
|
|
|
|
|
M
|
|
|
|
|
Volume Render Geometry
|
|
|
|
|
|
|
M
|
|
|
|
Render Shading
|
|
|
|
|
|
|
U
|
|
|
|
MPR Volumetric Presentation State Display
|
|
|
|
|
|
M
|
|
|
|
|
Render Display
|
|
|
|
|
|
|
M
|
|
|
|
Volumetric Graphic Annotation
|
|
|
|
|
|
U
|
U
|
|
|
|
Presentation Animation
|
|
|
|
|
|
U
|
U
|
|
|
|
ICC Profile
|
|
M
|
M
|
M
|
|
|
|
M
|
C
|
|
Common Instance Reference
|
|
|
|
|
|
M
|
M
|
M
|
|
|
SOP Common
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
Table A.1-7. Composite Information Object Modules Overview - Structured Reports
|
IODs
Modules
|
Basic Text SR
|
Enhd SR
|
Comp SR
|
Comp 3D SR
|
Extns SR
|
Key Obj Sel
|
Rend Obj Sel
|
Mam CAD
|
Ch CAD
|
Col CAD
|
Impl Plan SR
|
Prc Log
|
|
Patient
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
|
Clinical Trial Subject
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
|
General Study
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
|
Patient Study
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
|
Clinical Trial Study
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
|
Clinical Trial Series
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
|
SR Document Series
|
M
|
M
|
M
|
M
|
M
|
|
|
M
|
M
|
M
|
M
|
M
|
|
Key Object Document Series
|
|
|
|
|
|
M
|
M
|
|
|
|
M
|
|
|
Sync.
|
|
U
|
U
|
U
|
U
|
|
M
|
|
U
|
|
|
M
|
|
General Equip.
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
|
Enhanced General Equip.
|
|
|
|
|
M
|
|
M
|
|
|
M
|
M
|
|
|
SR Document General
|
M
|
M
|
M
|
M
|
M
|
|
|
M
|
M
|
M
|
M
|
M
|
|
SR Document Content
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
|
Key Object Document
|
|
|
|
|
|
M
|
M
|
|
|
|
M
|
|
|
Timezone
|
|
|
|
|
|
|
|
|
|
|
M
|
|
|
SOP Common
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
Table A.1-7b. Composite Information Object Modules Overview - More Structured Reports
|
IODs
Modules
|
RD SR
|
R-RD SR
|
Pat RD SR
|
Enh RD SR
|
Sp Rx Rp
|
Mac Grd Rp
|
Acq Ctx SR
|
Simp Card Echo SR
|
Plan IA Admin SR
|
Perf IA Admin SR
|
WF Ann SR
|
|
Patient
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
|
Clinical Trial Subject
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
|
General Study
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
|
Patient Study
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
|
Clinical Trial Study
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
|
Clinical Trial Series
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
|
SR Document Series
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
|
Key Object Document Series
|
|
|
|
|
|
|
|
|
|
|
|
|
Sync.
|
C
|
C
|
C
|
C
|
|
|
C
|
C
|
|
M
|
C
|
|
General Equip.
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
|
Enhanced General Equip.
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
|
SR Document General
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
|
SR Document Content
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
|
Key Object Document
|
|
|
|
|
|
|
|
|
|
|
|
|
Timezone
|
|
|
|
|
|
|
|
M
|
|
|
|
|
SOP Common
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
Table A.1-8. Composite Information Object Modules Overview - Waveforms
|
IODs
Modules
|
Basic Voice Audio
|
12 Lead ECG
|
General ECG WF
|
General 32-bit ECG WF
|
Ambul ECG WF
|
Hemo WF
|
Basic Cardiac EP WF
|
Arterial Pulse WF
|
Resp WF
|
General Audio WF
|
Scalp EEG WF
|
EMG WF
|
EOG WF
|
Sleep EEG WF
|
Multi Resp WF
|
Body Posn WF
|
|
Patient
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
|
Clinical Trial Subject
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
|
General Study
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
|
Patient Study
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
|
Clinical Trial Study
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
|
General Series
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
|
Clinical Trial Series
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
U
|
|
Synchronization
|
U
|
U
|
U
|
U
|
U
|
C
|
C
|
M
|
M
|
M
|
U
|
U
|
U
|
U
|
U
|
U
|
|
General Equipment
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
|
Enhanced General Equipment
|
|
|
|
M
|
|
|
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
|
Waveform Identification
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
|
Waveform
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
| Waveform Annotation |
U
|
C
|
C
|
C
|
C
|
C
|
C
|
C
|
C
|
C
|
C
|
C
|
C
|
C
|
C
|
C
|
|
Acquisition Context
|
M
|
M
|
M
|
M
|
U
|
M
|
M
|
M
|
M
|
M
|
M
|
U
|
U
|
U
|
U
|
U
|
|
SOP Common
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
M
|
Table A.1-9. Composite Information Object Modules Overview - Protocols
|
IODs
Modules
|
CT Performed Procedure Protocol
|
XA Performed Procedure Protocol
|
CT Defined Procedure Protocol
|
XA Defined Procedure Protocol
|
Protocol Approval
|
|
Patient
|
M
|
M
|
|
|
|
|
Clinical Trial Subject
|
U
|
U
|
|
|
|
|
General Study
|
M
|
M
|
|
|
|
|
Patient Study
|
U
|
U
|
|
|
|
|
Clinical Trial Study
|
U
|
U
|
|
|
|
|
General Series
|
M
|
M
|
|
|
|
|
Clinical Trial Series
|
U
|
U
|
|
|
|
|
Enhanced Series
|
M
|
M
|
|
|
|
|
CT Protocol Series
|
M
|
|
|
|
|
|
XA Protocol Series
|
|
M
|
|
|
|
|
Frame of Reference
|
M
|
M
|
|
|
|
|
General Equipment
|
M
|
M
|
M
|
M
|
M
|
|
Enhanced General Equipment
|
M
|
M
|
M
|
M
|
M
|
|
Protocol Context
|
M
|
M
|
M
|
M
|
|
|
Patient Protocol Context
|
U
|
U
|
|
|
|
|
Clinical Trial Context
|
|
|
U
|
U
|
|
|
Patient Specification
|
|
|
U
|
U
|
|
|
Equipment Specification
|
|
|
M
|
M
|
|
|
Instructions
|
U
|
U
|
U
|
U
|
|
|
Patient Positioning
|
U
|
U
|
U
|
U
|
|
|
General Defined Acquisition
|
|
|
U
|
U
|
|
|
Performed CT Acquisition
|
U
|
|
|
|
|
|
Performed XA Acquisition
|
|
U
|
|
|
|
|
General Defined Reconstruction
|
|
|
U
|
U
|
|
|
Performed CT Reconstruction
|
U
|
|
|
|
|
|
Performed XA Reconstruction
|
|
U
|
|
|
|
|
Defined Storage
|
|
|
U
|
U
|
|
|
Performed Storage
|
U
|
U
|
|
|
|
|
Protocol Approval
|
|
|
|
|
M
|
|
SOP Common
|
M
|
M
|
M
|
M
|
M
|
Table A.1-10. Composite Information Object Modules Overview - Encapsulated Documents
|
IODs
Modules
|
Enc PDF
|
Enc CDA
|
Enc STL
|
Enc OBJ
|
Enc MTL
|
|
Patient
|
M
|
M
|
M
|
M
|
M
|
|
Clinical Trial Subject
|
U
|
U
|
U
|
U
|
U
|
|
General Study
|
M
|
M
|
M
|
M
|
M
|
|
Patient Study
|
U
|
U
|
U
|
U
|
U
|
|
Clinical Trial Study
|
U
|
U
|
U
|
U
|
U
|
|
Clinical Trial Series
|
U
|
U
|
U
|
U
|
U
|
|
Encapsulated Document Series
|
M
|
M
|
M
|
M
|
M
|
|
Frame of Reference
|
|
|
M
|
M
|
|
|
General Equipment
|
M
|
M
|
M
|
M
|
M
|
|
Enhanced General Equipment
|
|
|
M
|
M
|
M
|
|
SC Equipment
|
M
|
M
|
|
|
|
|
Encapsulated Document
|
M
|
M
|
M
|
M
|
M
|
|
Manufacturing 3D Model
|
|
|
M
|
M
|
M
|
|
ICC Profile
|
|
|
U
|
U
|
U
|
|
SOP Common
|
M
|
M
|
|
M
|
M
|
|
Common Instance Reference
|
|
|
C
|
C
|
C
|
Table A.1-11. Real-Time Object Modules Overview - Images
|
IODs
Modules
|
RTV EN
|
RTV PH
|
|
Patient
|
M
|
M
|
|
Clinical Trial Subject
|
U
|
U
|
|
General Study
|
M
|
M
|
|
Patient Study
|
U
|
U
|
|
Clinical Trial Study
|
U
|
U
|
|
General Series
|
M
|
M
|
|
Clinical Trial Series
|
U
|
U
|
|
General Equipment
|
M
|
M
|
|
Enhanced General Equipment
|
M
|
M
|
|
Frame of Reference
|
C
|
C
|
|
Synchronization
|
M
|
M
|
|
General Image
|
M
|
M
|
|
General Reference
|
U
|
U
|
|
Real-Time Bulk Data Flow
|
M
|
M
|
|
Acquisition Context
|
M
|
M
|
|
Device
|
U
|
U
|
|
Specimen
|
C
|
C
|
|
VL Image
|
M
|
M
|
|
ICC Profile
|
M
|
U
|
|
SOP Common
|
M
|
M
|
|
Common Instance Reference
|
M
|
M
|
|
Real-Time Acquisition
|
M
|
M
|
|
Current Frame Functional Groups
|
M
|
M
|
Table A.1-12. Real-Time Object Modules Overview - Waveforms
|
IODs
Modules
|
RTV Basic Voice Audio
|
|
Patient
|
M
|
|
Clinical Trial Subject
|
U
|
|
General Study
|
M
|
|
Patient Study
|
U
|
|
Clinical Trial Study
|
U
|
|
General Series
|
M
|
|
Clinical Trial Series
|
U
|
|
General Equipment
|
M
|
|
Enhanced General Equipment
|
M
|
|
Synchronization
|
M
|
|
Waveform Identification
|
M
|
|
Real-Time Bulk Data Flow
|
M
|
|
Acquisition Context
|
M
|
|
SOP Common
|
M
|
|
Current Frame Functional Groups
|
M
|
Table A.1-13. Composite Information Object Modules Overview – Non-Patient Related Information
|
IODs
Modules
|
Hanging Protocol
|
Color Palette
|
Inventory
|
|
Equipment
|
|
|
M
|
|
Hanging Protocol Definition
|
M
|
|
|
|
Hanging Protocol Environment
|
M
|
|
|
|
Hanging Protocol Display
|
M
|
|
|
|
Palette Color LUT
|
|
M
|
|
|
ICC Profile
|
|
M
|
|
|
Color Palette Definition
|
|
M
|
|
|
Inventory
|
|
|
M
|
|
SOP Common
|
M
|
M
|
M
|
Table A.1-14. Composite Information Object Modules Overview - Waveform Presentation States
|
IODs
Modules
|
Waveform Presentation State
|
Waveform Acquisition Presentation State
|
|
Patient
|
M
|
M
|
|
Clinical Trial Subject
|
U
|
U
|
|
General Study
|
M
|
M
|
|
Patient Study
|
U
|
U
|
|
Clinical Trial Study
|
U
|
U
|
|
General Series
|
M
|
M
|
|
Clinical Trial Series
|
U
|
U
|
|
Presentation Series
|
M
|
M
|
|
Synchronization
|
C
|
C
|
|
General Equipment
|
M
|
M
|
|
Enhanced General Equipment
|
M
|
M
|
|
Presentation State Identification
|
M
|
M
|
|
Waveform Presentation State Relationship
|
M
|
M
|
|
Structured Waveform Annotation
|
U
|
U
|
|
Textual Waveform Annotation
|
U
|
U
|
|
Displayed Waveform Segment
|
U
|
U
|
|
Montage Activation
|
U
|
U
|
|
Waveform Presentation Montage
|
C
|
C
|
|
SOP Common
|
M
|
M
|