N.5.2.8.2 SCU of the Basic Color Print Management Meta SOP Class
The Basic Color Print Management Meta SOP Class is composed of the mandatory SOP Classes listed in Table N.5-29:
Table N.5-29. Basic Color Print Management Meta SOP Classes
|
SOP Class Name
|
SOP Class UID
|
|
Basic Film Session
|
1.2.840.10008.5.1.1.1
|
|
Basic Film Box
|
1.2.840.10008.5.1.1.2
|
|
Basic Color Image Box
|
1.2.840.10008.5.1.1.4.1
|
|
Printer
|
1.2.840.10008.5.1.1.16
|
N.5.2.8.2.1 Basic Film Session SOP Class
[If your system also supports the Basic Grayscale Print Management Meta SOP Class and the Film Session parameters are identical for color,
see Section N.5.2.8.1.1 Basic Film Session SOP Class for Basic Grayscale Print Management Meta SOP Class.
Otherwise, copy the Film Session table here and fill in the proper Values.]
N.5.2.8.2.2 Basic Film Box SOP Class
[If your system also supports the Basic Grayscale Print Management Meta SOP Class and the Film Box parameters are identical for color,
see Section N.5.2.8.1.2 Basic Film Box SOP Class for Basic Grayscale Print Management Meta SOP Class.
Otherwise copy the Film Box table here and fill in the proper Values.]
N.5.2.8.2.3 Basic Color Image Box SOP Class
Table N.5-30 lists the supported DIMSE Service for the Basic Color Image Box SOP Class:
Table N.5-30. Services for the Basic Color Image Box SOP Class - SCU
|
DIMSE Service Element
|
Purpose
|
|
N-SET
|
Set each Image Attributes for a previously created Film Box
|
Table N.5-31 lists the supported N-SET Attributes for Basic Color Image Box:
[List the supported Attributes and their possible Values. Provide the default Value when relevant. See example below.]
Table N.5-31. Supported N-SET Attributes for the Basic Color Image Box SOP Class - SCU
|
Attribute Name
|
Tag
|
Values
|
Default
|
|
Magnification Type
|
(2010,0060)
|
<<REPLICATE
BILINEAR
CUBIC
NONE
…>>
|
CUBIC
|
|
Smoothing Type
|
(2010,0080)
|
<possible Values or range>
|
143
|
|
Image Box Position
|
(2020,0010)
|
<possible Position Values or range>
|
|
|
Polarity
|
(2020,0020)
|
<<NORMAL
REVERSE>>
|
NORMAL
|
|
Requested Image Size
|
(2020,0030)
|
width, x-dimension, in mm
|
|
|
Requested Decimate/Crop Behavior
|
(2020,0040)
|
<<DECIMATE
CROP
FAIL>>
|
|
|
Basic Color Image Sequence
|
(2020,0111)
|
|
|
|
>Samples per Pixel
|
(0028,0002)
|
3
|
|
|
>Photometric Interpretation
|
(0028,0004)
|
RGB
|
|
|
>Planar Configuration
|
(0028,0006)
|
1 (frame interleave)
|
|
|
>Rows
|
(0028,0010)
|
|
|
|
>Columns
|
(0028,0011)
|
|
|
|
>Pixel Aspect Ratio
|
(0028,0034)
|
|
1\1
|
|
>Bits Allocated
|
(0028,0100)
|
8
|
|
|
>Bits Stored
|
(0028,0101)
|
8
|
|
|
>High Bit
|
(0028,0102)
|
7
|
|
|
>Pixel Representation
|
(0028,0103)
|
0
|
|
|
>Pixel Data
|
(7FE0,0010)
|
|
|
N.5.2.8.2.4 Printer SOP Class
[If your system also supports the Basic Grayscale Print Management Meta SOP Class, see 'Printer SOP Class' for 'Basic Grayscale Print Management Meta SOP Class' in Section N.5.2.8.1.4. Otherwise copy the Printer SOP Class table here and fill in the proper Values.]