EXAMPLE-RTV-SERVICE provides Standard Conformance to the following SOP Classes:
Table M.4.2-1. SOP Classes for DICOM-RTV AE
SOP Class Name
|
SOP Class UID
|
SCU
|
SCP
|
Video Photographic Image Real-Time Communication
|
1.2.840.10008.10.2
|
Yes
|
No
|
Audio Waveform Real-Time Communication
|
1.2.840.10008.10.3
|
Yes
|
No
|
Rendition Selection Document Real-Time Communication
|
1.2.840.10008.10.4
|
Yes
|
No
|
Some restrictions applies on the Real-Time Communications:
Table M.4.2-2. DICOM-RTV Instances Specification
Category
|
Restrictions
|
Transfer Syntaxes Supported
|
SMPTE ST 2110-20 Uncompressed Progressive Active Video, SMPTE ST 2110-30 PCM Digital Audio
|
Photometric interpretation
|
RGB
|
Bit depth (video)
|
10
|
Number of Waveform Channels
|
2
|
Bit depth (audio)
|
16 (signed 16-bits linear)
|
Sampling Frequency
|
48 kHz
|
Table M.4.2-3. DICOM-RTV Screen Resolutions
Rows
|
Columns
|
Frame rate
|
Video Type
|
Progressive or Interlaced
|
1080
|
1920
|
25
|
25 Hz HD
|
P
|
1080
|
1920
|
29.97, 30
|
30 Hz HD
|
P
|
1080
|
1920
|
25
|
25 Hz HD
|
I
|
1080
|
1920
|
29.97, 30
|
30 Hz HD
|
I
|
720
|
1280
|
25
|
25 Hz HD
|
P
|
720
|
1280
|
29.97, 30
|
30 Hz HD
|
P
|
720
|
1280
|
50
|
50 Hz HD
|
P
|
720
|
1280
|
59.94, 60
|
60 Hz HD
|
P
|
The resolution is automatically determined based on the one provided by the sent video.