DICOM PS3.2 2022b - Conformance |
---|
SOP Class : Basic Grayscale Image Box SOP
Attribute : Bits Stored (0028,0028)
Extensions : 8-16 bits stored are supported.
DICOM only specifies 8 and 12 for number of bits stored. The EXAMPLE-PRINT-SERVER-MANAGEMENT supports the number of bits stored to be from 8 through 16 bits.
SOP Class : Basic Grayscale Image Box SOP
Attribute : High Bit (0028,0028)
Extensions : High Bit positions 7 - 15 are supported.
DICOM specifies that the high bit must be the 7th or 11th bit (for 8 or 12 bits stored, respectively). The EXAMPLE-PRINT-SERVER-MANAGEMENT supports the high bit to be the number of bits stored minus one. For example, if the number of bits stored is 13, the high bit is 12.
DICOM PS3.2 2022b - Conformance |
---|