General Parameters
|
Max PDU Receive Size
|
Yes
|
128 KB
|
Max PDU Send Size(If the receiver supports a smaller Max PDU Receive Size then the Max PDU Send Size will be reduced accordingly for the duration of the Association. Max PDU Receive Size information is exchanged during DICOM Association Negotiation in the Maximum Length Sub-Item of the A-ASSOCIATION-RQ and A-ASSOCIATE-AC)
|
Yes
|
128 KB
|
Time-out waiting for a acceptance or rejection response to an Association Request (Application level Timeout).
|
Yes
|
20 s
|
Time-out waiting for a response to an Association release request (Application level Timeout)
|
Yes
|
30 s
|
Time-out waiting for completion of a TCP/IP connect request (Low-level timeout)
|
Yes
|
20 s
|
Time-out awaiting a Response to a DIMSE Request (Low level Timeout)
|
Yes
|
360 s
|
Time-out for waiting for data between TCP/IP-packets (Low level Timeout)
|
Yes
|
30 s
|
Maximum number of simultaneous Associations
|
Yes
|
8
|
Supported Transfer Syntaxes
|
Yes
|
Implicit VR Little Endian
Explicit VR Little Endian
|
Print Server Management
|
Default Print parameters: Max density, Min Density, Contrast, Border Density, Trim, Magnification type, Smoothing factor, Polarity, Number of Copies, Cropping Algorithm, Orientation.
|
Yes
|
Configurable
|
Number of times a failed print-job may be retried
|
Yes
|
3
|
Delay between retrying failed print-jobs
|
Yes
|
60s
|
Printer Bit-depth Configurable: 8 or 12
|
Yes
|
12
|
Custom Format
|
No
|
NA
|
Media Type: Transparent (Film), Reflective (Paper)
|
Yes
|
Transparent
|
Media size Configurable:
8IN X 10IN
11IN X 14IN
14IN X 14IN
14IN X 17IN
|
Yes
|
14IN X 17IN
|
Maximum number of printable pixel matrix per supported Media size; see Note
|
No
|
8x10 - 2286x2836
11x14 - 4096x3195
14x14 - 4096x4108
14x17 - 4096x5120
|
Maximum number of collated films in a film session
|
Yes
|
12
|
Support N-EVENT-REPORT (On/Off for either Printer, Print Job or both).
|
Yes
|
On
|
Handling of print jobs when requested Media Type and/or Film Size are not currently installed. The options are:
-
Queue the print job until the film matching the requested Media Type and/or Film Size is loaded.
-
Print on the film currently loaded in the printer.
|
Yes
|
Print on available media
|
Print SCP time-out waiting for a SCU confirmation to a Print Status N-EVENT-REPORT
|
Yes
|
60 s
|
Print SCP time-out waiting for a SCU confirmation to a Print Job N-EVENT-REPORT
|
Yes
|
60 s
|
Supported Transfer Syntaxes (separately configurable for each remote SCU printer)
|
Yes
|
Implicit VR Little Endian
Explicit VR Little Endian
|