General Parameters
|
Max PDU Receive Size
|
Yes
|
65536 Bytes(64 kB)
|
Max PDU Send Size(larger PDUs will never be sent, even if the receiver supports a larger Max PDU Receive Size. If the receiver supports a smaller Max PDU Receive Size then the Max PDU Send Size will be reduced accordingly for the duration of the Association. Max PDU Receive Size information is exchanged during DICOM Association Negotiation in the Maximum Length Sub-Item of the A-ASSOCIATION-RQ and A-ASSOCIATE-AC)
|
No
|
65536 Bytes(64 kB)
|
Time-out waiting for a acceptance or rejection response to an Association Request (Application Level Timeout)
|
Yes
|
15 s
|
Time-out waiting for a response to an Association release request (Application Level Timeout)
|
Yes
|
30 s
|
Time-out waiting for completion of a TCP/IP connect request (Low-level timeout)
|
Yes
|
15 s
|
Time-out awaiting a Response to a DIMSE Request (Low-Level Timeout)
|
Yes
|
360 s
|
Time-out for waiting for data between TCP/IP-packets (Low Level Timeout)
|
Yes
|
30 s
|
Storage Parameters
|
Storage SCU time-out waiting for a response to a C-STORE-RQ
|
Yes
|
120 s
|
Number of times a failed send job may be retried
|
Yes
|
0 (Failed send jobs are not retried)
|
Delay between retrying failed send jobs
|
Yes
|
60 s
|
Maximum number of simultaneously initiated Associations by the Storage AE
|
Yes
|
1
|
Supported Transfer Syntaxes (separately configurable for each remote AE)
|
Yes
|
Implicit VR Little Endian
Explicit VR Little Endian
|
Storage Commitment Parameters
|
Timeout waiting for a Storage Commitment Notification (maximum duration of applicability for a Storage Commitment Transaction UID).
|
Yes
|
24 hours
|
Maximum number of simultaneously accepted Associations by the Storage AE
|
Yes
|
5
|
Delay association release after sending a Storage Commitment Request (wait for a Storage Commitment Notification over the same association).
|
Yes
|
120 s
|
Modality Worklist Parameters
|
Modality Worklist SCU time-out waiting for the final response to a C-FIND-RQ
|
Yes
|
600 s
|
Maximum number of Worklist Items
|
Yes
|
100
|
Supported Transfer Syntaxes for Modality Worklist
|
Yes
|
Implicit VR Little Endian
Explicit VR Little Endian
|
Delay between automatic Worklist Updates
|
Yes
|
10 mins
|
Query Worklist for specific Scheduled Station AE Title
|
Yes
|
EXINTMOD_WFL
|
Query Worklist for specific Modality Value
|
Yes
|
RF
|
MPPS Parameters
|
MPPS SCU time-out waiting for a response to a N-CREATE-RQ
|
Yes
|
60 s
|
MPPS SCU time-out waiting for a response to a N-SET-RQ
|
Yes
|
30 s
|
Supported Transfer Syntaxes for MPPS
|
Yes
|
Implicit VR Little Endian
Explicit VR Little Endian
|
Print Parameters
|
Print SCU time-out waiting for a response to a N-CREATE-RQ
|
Yes
|
60 s
|
Print SCU time-out waiting for a response to a N-SET-RQ
|
Yes
|
30 s
|
Print SCU time-out waiting for a response to a N-ACTION-RQ
|
Yes
|
360s
|
Supported Transfer Syntaxes (separately configurable for each remote printer)
|
Yes
|
Implicit VR Little Endian
Explicit VR Little Endian
|
Number of times a failed print-job may be retried
|
Yes
|
0 (Failed send jobs are not retried)
|
Delay between retrying failed print-jobs
|
Yes
|
60 s
|
Printer correction LUT (separately configurable for each remote printer)
|
Yes
|
Identity LUT
|