DICOM PS3.2 2025e - Conformance

N.7.3.2 DIMSE Services

N.7.3.2.1 Basic Worklist Management Service
N.7.3.2.1.1 SCU of the Modality Worklist Information Model Find SOP Class - C-FIND

Table N.7-10 lists the Status Codes that the SCU of the Modality Worklist Information Model Find SOP Class supports for the C-FIND message and defines the application behavior when encountering the listed Status Codes.

[Describe the behavior of the application when it receives any of the Status Codes listed in the table below, e.g., displaying and logging the error code or retrying the request. For each additional Status Code supported add a row to the table.

In the "Other Status Codes" row document the behavior of the application in case it encounters an unknown Status Code.]

Table N.7-10. Status Codes for C-FIND of the Modality Worklist Information Model SOP Class - SCU

Service Status

Further Meaning

Status Code

Behavior

Success

Matching is complete - No final identifier is supplied

0000

Failure

Refused: Out of Resources

A700

SOP Class Not Supported

0122

Error: Data Set does not match SOP Class

A900

Error: Unable to process

C000-CFFF

Cancel

Matching terminated due to cancel

FE00

Pending

Matches are continuing - Current Match is supplied and any Optional Keys were supported in the same manner as Required Keys.

FF00

Matches are continuing - Warning that one or more Optional Keys were not supported for existence for this Identifier

FF01

-

Other Status Codes

anything else


N.7.3.2.1.2 SCP of the Modality Worklist Information Model Find SOP Class - C-FIND

Table N.7-11 lists the Status Codes that the SCP of the Modality Worklist Information Model Find SOP Class supports for the C-FIND message and defines conditions in which the listed Status Codes are sent.

[List the Attributes that are used to further detail the Status Codes in the "Related Fields Columns". Use N/A if there are no related fields used. Describe the condition which causes the application to send the specific Status Codes. For each other Status Code used add a row to the table.]

Table N.7-11. Status Codes for C-FIND of the Modality Worklist Information Model SOP Class - SCP

Service Status

Further Meaning

Status Code

Related Fields

Condition (and Comments on Related Fields)

Success

Matching is complete - No final identifier is supplied

0000

Failure

Refused: Out of Resources

A700

SOP Class Not Supported

0122

Error: Data Set does not match SOP Class

A900

Error: Unable to process

C000-CFFF

Cancel

Matching terminated due to cancel

FE00

Pending

Matches are continuing - Current Match is supplied and any Optional Keys were supported in the same manner as Required Keys.

FF00

Matches are continuing - Warning that one or more Optional Keys were not supported for existence for this Identifier

FF01


DICOM PS3.2 2025e - Conformance