DICOM PS3.2 2026b - Conformance
N.7.3.2.8 Print Management Service
N.7.3.2.8.1 SCU of the Basic Film Session SOP Class
N.7.3.2.8.1.1 SCU of the Basic Film Session SOP Class - N-CREATE

Table N.7-52 lists the Status Codes that the SCU of the Basic Film Session SOP Class supports for the N-CREATE message and defines the application behavior when encountering the listed Status Codes.

[Describe the behavior of the application when it receives any of the Status Codes listed in the table below, e.g., displaying and logging the error code or retrying the request. For each additional Status Code supported add a row to the table.

In the "Other Status Codes" row document the behavior of the application in case it encounters an unknown Status Code.]

Table N.7-52. Status Codes for N-CREATE of the Basic Film Session SOP Class - SCU

Service Status

Further Meaning

Status Code

Behavior

Success

Success

0000

Warning

Attribute List warning

0107

Attribute Value out of range

0116

Memory allocation not supported

B600

Failure

No such Attribute

0105

Invalid Attribute Value

0106

Processing Failure

0110

Duplicate SOP Instance

0111

Invalid SOP Instance

0117

No such SOP Class

0118

Missing Attribute

0120

Missing Attribute Value

0121

Refused: Not authorized

0124

Duplicate invocation

0210

Unrecognized operation

0211

Mistyped argument

0212

Resource limitation

0213

-

Other Status Codes

anything else


N.7.3.2.8.1.2 SCU of the Basic Film Session SOP Class - N-SET

Table N.7-53 lists the Status Codes that the SCU of the Basic Film Session SOP Class supports for the N-SET message and defines the application behavior when encountering the listed Status Codes.

[Describe the behavior of the application when it receives any of the Status Codes listed in the table below, e.g., displaying and logging the error code or retrying the request. For each additional Status Code supported add a row to the table.

In the "Other Status Codes" row document the behavior of the application in case it encounters an unknown Status Code.]

Table N.7-53. Status Codes for N-SET of the Basic Film Session SOP Class - SCU

Service Status

Further Meaning

Status Code

Behavior

Success

Success

0000

Warning

Attribute List warning

0107

Attribute Value out of range

0116

Memory allocation not supported

B600

Failure

No such Attribute

0105

Invalid Attribute Value

0106

Processing Failure

0110

Duplicate SOP Instance

0111

No such SOP Instance

0112

Invalid SOP Instance

0117

No such SOP Class

0118

Class Instance Conflict

0119

Missing Attribute

0120

Missing Attribute Value

0121

Refused: Not authorized

0124

Duplicate invocation

0210

Unrecognized operation

0211

Mistyped argument

0212

Resource limitation

0213

-

Other Status Codes

anything else


N.7.3.2.8.1.3 SCU of the Basic Film Session SOP Class - N-DELETE

Table N.7-54 lists the Status Codes that the SCU of the Basic Film Session SOP Class supports for the N-DELETE message and defines the application behavior when encountering the listed Status Codes.

[Describe the behavior of the application when it receives any of the Status Codes listed in the table below, e.g., displaying and logging the error code or retrying the request. For each additional Status Code supported add a row to the table.

In the "Other Status Codes" row document the behavior of the application in case it encounters an unknown Status Code.]

Table N.7-54. Status Codes for N-DELETE of the Basic Film Session SOP Class - SCU

Service Status

Further Meaning

Status Code

Behavior

Success

Success

0000

Failure

Processing Failure

0110

Invalid SOP Instance

0117

No such SOP Class

0118

Class Instance Conflict

0119

Refused: Not authorized

0124

Duplicate invocation

0210

Unrecognized operation

0211

Mistyped argument

0212

Resource limitation

0213

-

Other Status Codes

anything else


N.7.3.2.8.1.4 SCU of the Basic Film Session SOP Class - N-ACTION

Table N.7-55 lists the Status Codes that the SCU of the Basic Film Session SOP Class supports for the N-ACTION message and defines the application behavior when encountering the listed Status Codes.

[Describe the behavior of the application when it receives any of the Status Codes listed in the table below, e.g., displaying and logging the error code or retrying the request. For each additional Status Code supported add a row to the table.

In the "Other Status Codes" row document the behavior of the application in case it encounters an unknown Status Code.]

Table N.7-55. Status Codes for N-Action of the Basic Film Session SOP Class - SCU

Service Status

Further Meaning

Status Code

Behavior

Success

Film belonging to the film session are accepted for printing; if supported, the Print Job SOP Instance is created

0000

Warning

Film session printing (collation) is not supported

B601

Film Session SOP Instance hierarchy does not contain Image Box SOP Instances (empty page)

B602

Image size is larger than image box size, the image has been demagnified.

B604

Image size is larger than the Image Box size. The Image has been cropped to fit.

B609

Image size or Combined Print Image size is larger than the Image Box size. Image or Combined Print Image has been decimated to fit.

B60A

Failure

Processing failure

0110

No such SOP Instance

0112

No Such Argument

0114

Invalid argument Value

0115

Invalid SOP Instance

0117

No such SOP Class

0118

Class-Instance conflict

0119

No Such Action

0123

Refused: Not authorized

0124

Duplicate invocation

0210

Unrecognized operation

0211

Mistyped argument

0212

Resource limitation

0213

Failed: Film Session SOP Instance hierarchy does not contain Film Box SOP Instances

C600

Failed: Unable to create Print Job SOP Instance; print queue is full

C601

Failed: Image size is larger than image box size

C603

Failed: Combined Print Image size is larger than the Image Box size

C613

-

Other Status Codes

anything else


N.7.3.2.8.2 SCU of the Basic Film Box Session SOP Class
N.7.3.2.8.2.1 SCU of the Basic Film Box Session SOP Class - N-CREATE

Table N.7-52 lists the Status Codes that the SCU of the Basic Film Box SOP Class supports for the N-CREATE message and defines the application behavior when encountering the listed Status Codes.

[Describe the behavior of the application when it receives any of the Status Codes listed in the table below, e.g., displaying and logging the error code or retrying the request. For each additional Status Code supported add a row to the table.

In the "Other Status Codes" row document the behavior of the application in case it encounters an unknown Status Code.]

Table N.7-56. Status Codes for N-CREATE of the Basic Film Box SOP Class - SCU

Service Status

Further Meaning

Status Code

Behavior

Success

Success

0000

Warning

Attribute List warning

0107

Attribute Value out of range

0116

Requested Min Density or Max Density outside of printer's operating range

B605

Failure

No such Attribute

0105

Invalid Attribute Value

0106

Processing Failure

0110

Duplicate SOP Instance

0111

Invalid SOP Instance

0117

No such SOP Class

0118

Missing Attribute

0120

Missing Attribute Value

0121

Refused: Not authorized

0124

Duplicate invocation

0210

Unrecognized operation

0211

Mistyped argument

0212

Resource limitation

0213

There is an existing Film Box that has not been printed and N-ACTION at the Film Session level is not supported. A new Film Box will not be created when a previous Film Box has not been printed

C616

-

Other Status Codes

anything else


N.7.3.2.8.2.2 SCU of the Basic Film Box Session SOP Class - N-SET

Table N.7-57 lists the Status Codes that the SCU of the Basic Film Box SOP Class supports for the N-SET message and defines the application behavior when encountering the listed Status Codes.

[Describe the behavior of the application when it receives any of the Status Codes listed in the table below, e.g., displaying and logging the error code or retrying the request. For each additional Status Code supported add a row to the table.

In the "Other Status Codes" row document the behavior of the application in case it encounters an unknown Status Code.]

Table N.7-57. Status Codes for N-SET of the Basic Film Box SOP Class - SCU

Service Status

Further Meaning

Status Code

Behavior

Success

Success

0000

Warning

Attribute List warning

0107

Attribute Value out of range

0116

Requested Min Density or Max Density outside of printer's operating range

B605

Failure

No such Attribute

0105

Invalid Attribute Value

0106

Processing Failure

0110

Duplicate SOP Instance

0111

Invalid SOP Instance

0117

No such SOP Class

0118

Class Instance Conflict

0119

Missing Attribute

0120

Missing Attribute Value

0121

Refused: Not authorized

0124

Duplicate invocation

0210

Unrecognized operation

0211

Mistyped argument

0212

Resource limitation

0213

There is an existing Film Box that has not been printed and N-ACTION at the Film Session level is not supported. A new Film Box will not be created when a previous Film Box has not been printed

C616

-

Other Status Codes

anything else


N.7.3.2.8.2.3 SCU of the Basic Film Box Session SOP Class - N-DELETE

Table N.7-58 lists the Status Codes that the SCU of the Basic Film Box SOP Class supports for the N-DELETE message and defines the application behavior when encountering the listed Status Codes.

[Describe the behavior of the application when it receives any of the Status Codes listed in the table below, e.g., displaying and logging the error code or retrying the request. For each additional Status Code supported add a row to the table.

In the "Other Status Codes" row document the behavior of the application in case it encounters an unknown Status Code.]

Table N.7-58. Status Codes for N-DELETE of the Basic Film Box SOP Class - SCU

Service Status

Further Meaning

Status Code

Behavior

Success

Success

0000

Failure

Processing Failure

0110

Invalid SOP Instance

0117

No such SOP Class

0118

Class Instance Conflict

0119

Refused: Not authorized

0124

Duplicate invocation

0210

Unrecognized operation

0211

Mistyped argument

0212

Resource limitation

0213

-

Other Status Codes

anything else


N.7.3.2.8.2.4 SCU of the Basic Film Box Session SOP Class - N-ACTION

Table N.7-59 lists the Status Codes that the SCU of the Basic Film Box SOP Class supports for the N-ACTION message and defines the application behavior when encountering the listed Status Codes.

[Describe the behavior of the application when it receives any of the Status Codes listed in the table below, e.g., displaying and logging the error code or retrying the request. For each additional Status Code supported add a row to the table.

In the "Other Status Codes" row document the behavior of the application in case it encounters an unknown Status Code.]

Table N.7-59. Status Codes for N-ACTION of the Basic Film Box SOP Class - SCU

Service Status

Further Meaning

Status Code

Behavior

Success

Success

0000

Warning

Film Box SOP Instance hierarchy does not contain Image Box SOP Instances (empty page)

B603

Image size is larger than Image Box size. The image has been demagnified.

B604

Image size is larger than Image Box size. The image has been cropped to fit.

B609

Image size or Combined Print Image Size is larger than Image Box size. The image or combined Print Image has been decimated to fit.

B60A

Failure

Processing failure

0110

No such SOP Instance

0112

No Such Argument

0114

Invalid argument Value

0115

Invalid SOP Instance

0117

No such SOP Class

0118

Class-Instance conflict

0119

No Such Action

0123

Refused: Not authorized

0124

Duplicate invocation

0210

Unrecognized operation

0211

Mistyped argument

0212

Resource limitation

0213

Unable to create Print Job SOP Instance; print queue is full.

C602

Image size is larger than Image Box size.

C603

Combined Print Image Size is larger than Image Box size.

C613

-

Other Status Codes

anything else


N.7.3.2.8.3 SCU of the Basic Grayscale Image Box SOP Class - N-SET

Table N.7-60 lists the Status Codes that the SCU of the Basic Grayscale Image Box SOP Class supports for the N-SET message and defines the application behavior when encountering the listed Status Codes.

[Describe the behavior of the application when it receives any of the Status Codes listed in the table below, e.g., displaying and logging the error code or retrying the request. For each additional Status Code supported add a row to the table.

In the "Other Status Codes" row document the behavior of the application in case it encounters an unknown Status Code.]

Table N.7-60. Status Codes for N-SET of the Grayscale Image Box SOP Class - SCU

Service Status

Further Meaning

Status Code

Behavior

Success

Success

0000

Warning

Image size is larger than Image Box size. The image has been demagnified.

B604

Requested Min Density or Max Density outside of printer's operating range.

B605

Image size is larger than Image Box size. The image has been cropped to fit.

B609

Image size or Combined Print Image Size is larger than Image Box size. The image or combined Print Image has been decimated to fit.

B60A

Failure

No such Attribute

0105

Invalid Attribute Value

0106

Processing Failure

0110

Duplicate SOP Instance

0111

Invalid SOP Instance

0117

No such SOP Class

0118

Class Instance Conflict

0119

Missing Attribute

0120

Missing Attribute Value

0121

Refused: Not authorized

0124

Duplicate invocation

0210

Unrecognized operation

0211

Mistyped argument

0212

Resource limitation

0213

Image size is larger than Image Box size.

C603

Insufficient memory in printer to store the image.

C605

Combined Print Image Size is larger than Image Box size.

C613

-

Other Status Codes

anything else


N.7.3.2.8.4 SCU of the Basic Color Image Box SOP Class - N-SET

Table N.7-61 lists the Status Codes that the SCU of the Basic Color Image Box SOP Class supports for the N-SET message and defines the application behavior when encountering the listed Status Codes.

[Describe the behavior of the application when it receives any of the Status Codes listed in the table below, e.g., displaying and logging the error code or retrying the request. For each additional Status Code supported add a row to the table.

In the "Other Status Codes" row document the behavior of the application in case it encounters an unknown Status Code.]

Table N.7-61. Status Codes for N-SET of the Color Image Box SOP Class - SCU

Service Status

Further Meaning

Status Code

Behavior

Success

Success

0000

Warning

Image size is larger than Image Box size. The image has been demagnified.

B604

Image size is larger than Image Box size. The image has been cropped to fit.

B609

Image size or Combined Print Image Size is larger than Image Box size. The image or combined Print Image has been decimated to fit.

B60A

Failure

No such Attribute

0105

Invalid Attribute Value

0106

Processing Failure

0110

Duplicate SOP Instance

0111

Invalid SOP Instance

0117

No such SOP Class

0118

Class Instance Conflict

0119

Missing Attribute

0120

Missing Attribute Value

0121

Refused: Not authorized

0124

Duplicate invocation

0210

Unrecognized operation

0211

Mistyped argument

0212

Resource limitation

0213

Image size is larger than Image Box size.

C603

Insufficient memory in printer to store the image.

C605

Combined Print Image Size is larger than Image Box size.

C613

-

Other Status Codes

anything else


N.7.3.2.8.5 SCU of the Printer SOP Class
N.7.3.2.8.5.1 SCU of the Printer SOP Class - N-EVENT-REPORT

Table N.7-62 lists the Status Codes that the SCU of Printer SOP Class supports for the N-EVENT-REPORT message and defines the application behavior when encountering the listed Status Codes.

[Describe the behavior of the application when it receives any of the Status Codes listed in the table below, e.g., displaying and logging the error code or retrying the request. For each additional Status Code supported add a row to the table.

In the "Other Status Codes" row document the behavior of the application in case it encounters an unknown Status Code.]

Table N.7-62. Status Codes for N-EVENT-REPORT of the Printer SOP Class - SCU

Service Status

Further Meaning

Status Code

Behavior

Success

Success

0000

Failure

Processing Failure

0110

No such SOP Instance

0112

No Such Event Type

0113

No Such Argument

0114

Invalid Argument Value

0115

Invalid SOP Instance

0117

No such SOP Class

0118

Class-Instance conflict

0119

Duplicate invocation

0210

Unrecognized operation

0211

Mistyped argument

0212

Resource limitation

0213

-

Other Status Codes

anything else


N.7.3.2.8.5.2 SCU of the Printer SOP Class - N-GET

Table N.7-63 lists the Status Codes that the SCU of the Printer SOP Class supports for the N-GET message and defines the application behavior when encountering the listed Status Codes.

[Describe the behavior of the application when it receives any of the Status Codes listed in the table below, e.g., displaying and logging the error code or retrying the request. For each additional Status Code supported add a row to the table.

In the "Other Status Codes" row document the behavior of the application in case it encounters an unknown Status Code.]

Table N.7-63. Status Codes for N-GET of the Printer SOP Class - SCU

Service Status

Further Meaning

Status Code

Behavior

Success

Success

0000

Warning

Attribute List warning

0107

Failure

Processing Failure

0110

No such SOP Instance

0112

Invalid SOP Instance

0117

No such SOP Class

0118

Class-Instance conflict

0119

Refused: Not authorized

0124

Duplicate invocation

0210

Unrecognized operation

0211

Mistyped argument

0212

Resource limitation

0213

-

Other Status Codes

anything else


N.7.3.2.8.6 SCU of the Basic Annotation Box SOP Class - N-SET

Table N.7-64 lists the Status Codes that the SCU of the Basic Annotation Box SOP Class supports for the N-SET message and defines the application behavior when encountering the listed Status Codes.

[Describe the behavior of the application when it receives any of the Status Codes listed in the table below, e.g., displaying and logging the error code or retrying the request. For each additional Status Code supported add a row to the table.

In the "Other Status Codes" row document the behavior of the application in case it encounters an unknown Status Code.]

Table N.7-64. Status Codes for N-SET of the Basic Annotation Box SOP Class - SCU

Service Status

Further Meaning

Status Code

Behavior

Success

Success

0000

Failure

No such Attribute

0105

Invalid Attribute Value

0106

Processing Failure

0110

Duplicate SOP Instance

0111

Invalid SOP Instance

0117

No such SOP Class

0118

Class Instance Conflict

0119

Missing Attribute

0120

Missing Attribute Value

0121

Refused: Not authorized

0124

Duplicate invocation

0210

Unrecognized operation

0211

Mistyped argument

0212

Resource limitation

0213

-

Other Status Codes

anything else


N.7.3.2.8.7 SCU of the Print Job SOP Class
N.7.3.2.8.7.1 SCU of the Print Job SOP Class - N-EVENT-REPORT

Table N.7-65 lists the Status Codes that the SCU of the Print Job SOP Class supports for the N-EVENT-REPORT message and defines the application behavior when encountering the listed Status Codes.

[Describe the behavior of the application when it receives any of the Status Codes listed in the table below, e.g., displaying and logging the error code or retrying the request. If any other Status Code is supported add it to the table.] In the "other Status Codes" row document the behavior of the application in case it encounters and unknown Status Code.]

Table N.7-65. Status Codes N-EVENT-REPORT of the Print Job SOP Class - SCU

Service Status

Further Meaning

Status Code

Behavior

Success

Success

0000

Failure

Processing Failure

0110

No such SOP Instance

0112

No Such Event Type

0113

No Such Argument

0114

Invalid Argument Value

0115

Invalid SOP Instance

0117

No such SOP Class

0118

Class-Instance conflict

0119

Duplicate invocation

0210

Unrecognized operation

0211

Mistyped argument

0212

Resource limitation

0213

-

Other Status Codes

anything else


N.7.3.2.8.7.2 SCU of the Print Job SOP Class - N-GET

Table N.7-66 lists the Status Codes that the SCU of Print Job SOP Class supports for the N-GET message and defines the application behavior when encountering the listed Status Codes.

[Describe the behavior of the application when it receives any of the Status Codes listed in the table below, e.g., displaying and logging the error code or retrying the request. For each additional Status Code supported add a row to the table.

In the "Other Status Codes" row document the behavior of the application in case it encounters an unknown Status Code.]

Table N.7-66. Status Codes for N-GET of the Print Job SOP Class - SCU

Service Status

Further Meaning

Status Code

Behavior

Success

Success

0000

Warning

Attribute List warning

0107

Failure

Processing Failure

0110

No such SOP Instance

0112

Invalid SOP Instance

0117

No such SOP Class

0118

Class-Instance conflict

0119

Refused: Not authorized

0124

Duplicate invocation

0210

Unrecognized operation

0211

Mistyped argument

0212

Resource limitation

0213

-

Other Status Codes

anything else


N.7.3.2.8.8 SCU of the Presentation LUT SOP Class
N.7.3.2.8.8.1 SCU of the Presentation LUT SOP Class - N-CREATE

Table N.7-67 lists the Status Codes that the SCU of the Presentation LUT SOP Class supports for the N-CREATE message and defines the application behavior when encountering the listed Status Codes.

[Describe the behavior of the application when it receives any of the Status Codes listed in the table below, e.g., displaying and logging the error code or retrying the request. For each additional Status Code supported add a row to the table.

In the "Other Status Codes" row document the behavior of the application in case it encounters an unknown Status Code.]

Table N.7-67. Status Codes N-CREATE of the Presentation LUT SOP Class - SCU

Service Status

Further Meaning

Status Code

Behavior

Success

Success

0000

Warning

Attribute List warning

0107

Attribute Value out of range

0116

Requested Min Density or Max Density outside of printer's operating range

B605

Failure

No such Attribute

0105

Invalid Attribute Value

0106

Processing Failure

0110

Duplicate SOP Instance

0111

Invalid SOP Instance

0117

No such SOP Class

0118

Missing Attribute

0120

Missing Attribute Value

0121

Refused: Not authorized

0124

Duplicate invocation

0210

Unrecognized operation

0211

Mistyped argument

0212

Resource limitation

0213

-

Other Status Codes

anything else


N.7.3.2.8.8.2 SCU of the Presentation LUT SOP Class - N-DELETE

Table N.7-68 lists the Status Codes that the SCU of the Presentation LUT SOP Class supports for the N-DELETE message and defines the application behavior when encountering the listed Status Codes.

[Describe the behavior of the application when it receives any of the Status Codes listed in the table below, e.g., displaying and logging the error code or retrying the request. For each additional Status Code supported add a row to the table.

In the "Other Status Codes" row document the behavior of the application in case it encounters an unknown Status Code.]

Table N.7-68. Status Codes for N-DELETE of the Presentation LUT SOP Class - SCU

Service Status

Further Meaning

Status Code

Behavior

Success

Success

0000

Failure

Processing Failure

0110

Invalid SOP Instance

0117

No such SOP Class

0118

Class Instance Conflict

0119

Refused: Not authorized

0124

Duplicate invocation

0210

Unrecognized operation

0211

Mistyped argument

0212

Resource limitation

0213

-

Other Status Codes

anything else


N.7.3.2.8.9 SCU of the Printer Configuration Retrieval SOP Class - N-GET

Table N.7-69 lists the Status Codes that the SCU of the Printer Configuration SOP Class supports for the N-GET message and defines the application behavior when encountering the listed Status Codes.

[Describe the behavior of the application when it receives any of the Status Codes listed in the table below, e.g., displaying and logging the error code or retrying the request. For each additional Status Code supported add a row to the table.

In the "Other Status Codes" row document the behavior of the application in case it encounters an unknown Status Code.]

Table N.7-69. Status Codes N-GET of the Printer Configuration Retrieval SOP Class - SCU

Service Status

Further Meaning

Status Code

Behavior

Success

Success

0000

Warning

Attribute List warning

0107

Failure

Processing Failure

0110

No such SOP Instance

0112

Invalid SOP Instance

0117

No such SOP Class

0118

Class-Instance conflict

0119

Refused: Not authorized

0124

Duplicate invocation

0210

Unrecognized operation

0211

Mistyped argument

0212

Resource limitation

0213

-

Other Status Codes

anything else


N.7.3.2.8.10 SCP of the Basic Film Session SOP Class
N.7.3.2.8.10.1 SCP of the Basic Film Session SOP Class - N-CREATE

Table N.7-70 lists the Status Codes that the SCP of the Basic Film Session SOP Class supports for the N-CREATE message and defines conditions in which the listed Status Codes are sent.

[Describe the condition which causes the application to send the specific Status Codes. For each other Status Code used add a row to the table.]

Table N.7-70. Status Codes for N-CREATE of the Basic Film Session SOP Class - SCP

Service Status

Further Meaning

Status Code

Condition

Success

Success

0000

Warning

Memory allocation not supported

B600

Attribute Value out of range

0116

Attribute List warning

0107

Failure

No such Attribute

0105

Invalid Attribute Value

0106

Processing Failure

0110

Duplicate SOP Instance

0111

Invalid SOP Instance

0117

No such SOP Class

0118

Missing Attribute

0120

Missing Attribute Value

0121

Refused: Not authorized

0124

Duplicate invocation

0210

Unrecognized operation

0211

Mistyped argument

0212

Resource limitation

0213


N.7.3.2.8.10.2 SCP of the Basic Film Session SOP Class - N-SET

Table N.7-71 lists the Status Codes that the SCP of the Basic Film Session SOP Class supports for the N-SET message and defines conditions in which the listed Status Codes are sent.

[Describe the condition which causes the application to send the specific Status Codes. For each other Status Code used add a row to the table.]

Table N.7-71. Status Codes for N-SET of the Basic Film Session SOP Class - SCP

Service Status

Further Meaning

Status Code

Condition

Success

Success

0000

Warning

Attribute Value out of range

0116

Attribute List warning

0107

Failure

No such Attribute

0105

Invalid Attribute Value

0106

Processing Failure

0110

Duplicate SOP Instance

0111

Invalid SOP Instance

0117

No such SOP Class

0118

Missing Attribute

0120

Missing Attribute Value

0121

Refused: Not authorized

0124

Duplicate invocation

0210

Unrecognized operation

0211

Mistyped argument

0212

Resource limitation

0213


N.7.3.2.8.10.3 SCP of the Basic Film Session SOP Class - N-DELETE

Table N.7-72 lists the Status Codes that the SCP of the Basic Film Session SOP Class supports for the N-DELETE message and defines conditions in which the listed Status Codes are sent.

[Describe the condition which causes the application to send the specific Status Codes. For each other Status Code used add a row to the table.]

Table N.7-72. Status Codes for N-DELETE of the Basic Film Session SOP Class - SCP

Service Status

Further Meaning

Status Code

Condition

Success

Success

0000

Failure

Processing Failure

0110

Invalid SOP Instance

0117

No such SOP Class

0118

Class Instance Conflict

0119

Refused: Not authorized

0124

Duplicate invocation

0210

Unrecognized operation

0211

Mistyped argument

0212

Resource limitation

0213


N.7.3.2.8.10.4 SCP of the Basic Film Session SOP Class - N-ACTION

Table N.7-73 lists the Status Codes that the SCP of the Basic Film Session SOP Class supports for the N-ACTION message and defines conditions in which the listed Status Codes are sent.

[Describe the condition which causes the application to send the specific Status Codes. For each other Status Code used add a row to the table.]

Table N.7-73. Status Codes for N-ACTION of the Basic Film Session SOP Class - SCP

Service Status

Further Meaning

Status Code

Condition

Success

Film belonging to the film session are accepted for printing; if supported, the Print Job SOP Instance is created

0000

Warning

Film session printing (collation) is not supported

B601

Film Session SOP Instance hierarchy does not contain Image Box SOP Instances (empty page)

B602

Image size is larger than image box size, the image has been demagnified.

B604

Image size is larger than the Image Box size. The Image has been cropped to fit.

B609

Image size or Combined Print Image size is larger than the Image Box size. Image or Combined Print Image has been decimated to fit.

B60A

Failure

Failed: Film Session SOP Instance hierarchy does not contain Film Box SOP Instances

C600

Failed: Unable to create Print Job SOP Instance; print queue is full

C601

Failed: Image size is larger than image box size

C603

Failed: Combined Print Image size is larger than the Image Box size

C613


N.7.3.2.8.11 SCP of the Basic Film Box SOP Class
N.7.3.2.8.11.1 SCP of the Basic Film Box SOP Class - N-CREATE

Table N.7-74 lists the Status Codes that the SCP of the Basic Film Box SOP Class supports for the N-CREATE message and defines conditions in which the listed Status Codes are sent.

[Describe the condition which causes the application to send the specific Status Codes. For each other Status Code used add a row to the table.]

Table N.7-74. Status Codes for N-CREATE of the Basic Film Box SOP Class - SCP

Service Status

Further Meaning

Status Code

Condition

Success

Success

0000

Warning

Attribute List warning

0107

Attribute Value out of range

0116

Requested Min Density or Max Density outside of printer's operating range

B605

Failure

No such Attribute

0105

Invalid Attribute Value

0106

Processing Failure

0110

Duplicate SOP Instance

0111

Invalid SOP Instance

0117

No such SOP Class

0118

Missing Attribute

0120

Missing Attribute Value

0121

Refused: Not authorized

0124

Duplicate invocation

0210

Unrecognized operation

0211

Mistyped argument

0212

Resource limitation

0213

There is an existing Film Box that has not been printed and N-ACTION at the Film Session level is not supported. A new Film Box will not be created when a previous Film Box has not been printed

C616


N.7.3.2.8.11.2 SCP of the Basic Film Box SOP Class - N-SET

Table N.7-75 lists the Status Codes that the SCP of the Basic Film Box SOP Class supports for the N-SET message and defines conditions in which the listed Status Codes are sent.

[Describe the condition which causes the application to send the specific Status Codes. For each other Status Code used add a row to the table.]

Table N.7-75. Status Codes for N-SET of the Basic Film Box SOP Class - SCP

Service Status

Further Meaning

Status Code

Condition

Success

Success

0000

Warning

Attribute List warning

0107

Attribute Value out of range

0116

Requested Min Density or Max Density outside of printer's operating range

B605

Failure

No such Attribute

0105

Invalid Attribute Value

0106

Processing Failure

0110

Duplicate SOP Instance

0111

Invalid SOP Instance

0117

No such SOP Class

0118

Class Instance Conflict

0119

Missing Attribute

0120

Missing Attribute Value

0121

Refused: Not authorized

0124

Duplicate invocation

0210

Unrecognized operation

0211

Mistyped argument

0212

Resource limitation

0213

There is an existing Film Box that has not been printed and N-ACTION at the Film Session level is not supported. A new Film Box will not be created when a previous Film Box has not been printed

C616


N.7.3.2.8.11.3 SCP of the Basic Film Box SOP Class - N-DELETE

Table N.7-76 lists the Status Codes that the SCP of the Basic Film Box SOP Class supports for the N-DELETE message and defines conditions in which the listed Status Codes are sent.

[Describe the condition which causes the application to send the specific Status Codes. For each other Status Code used add a row to the table.]

Table N.7-76. Status Codes for N-DELETE of the Basic Film Box SOP Class - SCP

Service Status

Further Meaning

Status Code

Condition

Success

Success

0000

Failure

Processing Failure

0110

Invalid SOP Instance

0117

No such SOP Class

0118

Class Instance Conflict

0119

Refused: Not authorized

0124

Duplicate invocation

0210

Unrecognized operation

0211

Mistyped argument

0212

Resource limitation

0213


N.7.3.2.8.11.4 SCP of the Basic Film Box SOP Class - N-ACTION

Table N.7-77 lists the Status Codes that the SCP of the Basic Film Box SOP Class supports for the N-ACTION message and defines conditions in which the listed Status Codes are sent.

[Describe the condition which causes the application to send the specific Status Codes. For each other Status Code used add a row to the table.]

Table N.7-77. Status Codes for N-ACTION of the Basic Film Box SOP Class - SCP

Service Status

Further Meaning

Status Code

Condition

Success

Success

0000

Warning

Film Box SOP Instance hierarchy does not contain Image Box SOP Instances (empty page)

B603

Image size is larger than Image Box size. The image has been demagnified.

B604

Image size is larger than Image Box size. The image has been cropped to fit.

B609

Image size or Combined Print Image Size is larger than Image Box size. The image or combined Print Image has been decimated to fit.

B60A

Failure

Processing failure

0110

No such SOP Instance

0112

No Such Argument

0114

Invalid argument Value

0115

Invalid SOP Instance

0117

No such SOP Class

0118

Class-Instance conflict

0119

No Such Action

0123

Refused: Not authorized

0124

Duplicate invocation

0210

Unrecognized operation

0211

Mistyped argument

0212

Resource limitation

0213

Unable to create Print Job SOP Instance; print queue is full.

C602

Image size is larger than Image Box size.

C603

Combined Print Image Size is larger than Image Box size.

C613


N.7.3.2.8.12 SCP of the Basic Grayscale Image Box SOP Class - N-SET

Table N.7-78 lists the Status Codes that the SCP of the Basic Grayscale Image Box SOP Class supports for the N-SET message and defines conditions in which the listed Status Codes are sent.

[Describe the condition which causes the application to send the specific Status Codes. For each other Status Code used add a row to the table.]

Table N.7-78. Status Codes for N-SET of the Basic Grayscale Image Box SOP Class - SCP

Service Status

Further Meaning

Status Code

Condition

Success

Success

0000

Warning

Image size is larger than Image Box size. The image has been demagnified.

B604

Requested Min Density or Max Density outside of printer's operating range.

B605

Image size is larger than Image Box size. The image has been cropped to fit.

B609

Image size or Combined Print Image Size is larger than Image Box size. The image or combined Print Image has been decimated to fit.

B60A

Failure

No such Attribute

0105

Invalid Attribute Value

0106

Processing Failure

0110

Duplicate SOP Instance

0111

Invalid SOP Instance

0117

No such SOP Class

0118

Class Instance Conflict

0119

Missing Attribute

0120

Missing Attribute Value

0121

Refused: Not authorized

0124

Duplicate invocation

0210

Unrecognized operation

0211

Mistyped argument

0212

Resource limitation

0213

Image size is larger than Image Box size.

C603

Insufficient memory in printer to store the image.

C605

Combined Print Image Size is larger than Image Box size.

C613


N.7.3.2.8.13 SCP of the Basic Color Image Box SOP Class - N-SET

Table N.7-79 lists the Status Codes that the SCP of the Basic Color Image Box SOP Class supports for the N-SET message and defines conditions in which the listed Status Codes are sent.

[Describe the condition which causes the application to send the specific Status Codes. For each other Status Code used add a row to the table.]

Table N.7-79. Status Codes for N-SET of the Basic Color Image Box SOP Class - SCP

Service Status

Further Meaning

Status Code

Condition

Success

Success

0000

Warning

Image size is larger than Image Box size. The image has been demagnified.

B604

Requested Min Density or Max Density outside of printer's operating range.

B605

Image size is larger than Image Box size. The image has been cropped to fit.

B609

Image size or Combined Print Image Size is larger than Image Box size. The image or combined Print Image has been decimated to fit.

B60A

Failure

No such Attribute

0105

Invalid Attribute Value

0106

Processing Failure

0110

Duplicate SOP Instance

0111

Invalid SOP Instance

0117

No such SOP Class

0118

Class Instance Conflict

0119

Missing Attribute

0120

Missing Attribute Value

0121

Refused: Not authorized

0124

Duplicate invocation

0210

Unrecognized operation

0211

Mistyped argument

0212

Resource limitation

0213

Image size is larger than Image Box size.

C603

Insufficient memory in printer to store the image.

C605

Combined Print Image Size is larger than Image Box size.

C613


N.7.3.2.8.14 SCP of the Printer SOP Class
N.7.3.2.8.14.1 SCP of the Printer SOP Class - N-EVENT-REPORT

Table N.7-80 lists the Status Codes that the SCP of the Printer SOP Class supports for the N-EVENT-REPORT message and defines conditions in which the listed Status Codes are sent.

[Describe the condition which causes the application to send the specific Status Codes. For each other Status Code used add a row to the table.]

Table N.7-80. Status Codes for N-EVENT-REPORT of the Printer SOP Class - SCP

Service Status

Further Meaning

Status Code

Condition

Success

Success

0000

Failure

Processing Failure

0110

No such SOP Instance

0112

No Such Event Type

0113

No Such Argument

0114

Invalid Argument Value

0115

Invalid SOP Instance

0117

No such SOP Class

0118

Class-Instance conflict

0119

Duplicate invocation

0210

Unrecognized operation

0211

Mistyped argument

0212

Resource limitation

0213


N.7.3.2.8.14.2 SCP of the Printer SOP Class - N-GET

Table N.7-81 lists the Status Codes that the SCP of the Printer SOP Class supports for the N-GET message and defines conditions in which the listed Status Codes are sent.

[Describe the condition which causes the application to send the specific Status Codes. For each other Status Code used add a row to the table.]

Table N.7-81. Status Codes for N-GET of the Printer SOP Class - SCP

Service Status

Further Meaning

Status Code

Condition

Success

Success

0000

Warning

Attribute List warning

0107

Failure

Processing Failure

0110

No such SOP Instance

0112

Invalid SOP Instance

0117

No such SOP Class

0118

Class-Instance conflict

0119

Refused: Not authorized

0124

Duplicate invocation

0210

Unrecognized operation

0211

Mistyped argument

0212

Resource limitation

0213


N.7.3.2.8.15 SCP the Basic Annotation Box SOP Class - N-SET

Table N.7-82 lists the Status Codes that the SCP of the Basic Annotation Box SOP Class supports for the N-SET message and defines conditions in which the listed Status Codes are sent.

[Describe the condition which causes the application to send the specific Status Codes. For each other Status Code used add a row to the table.]

Table N.7-82. Status Codes for N-SET of the Basic Annotation Box SOP Class - SCP

Service Status

Further Meaning

Status Code

Condition

Success

Success

0000

Failure

No such Attribute

0105

Invalid Attribute Value

0106

Processing Failure

0110

Duplicate SOP Instance

0111

Invalid SOP Instance

0117

No such SOP Class

0118

Class Instance Conflict

0119

Missing Attribute

0120

Missing Attribute Value

0121

Refused: Not authorized

0124

Duplicate invocation

0210

Unrecognized operation

0211

Mistyped argument

0212

Resource limitation

0213


N.7.3.2.8.16 SCP of the Print Job SOP Class
N.7.3.2.8.16.1 SCP of the Print Job SOP Class - N-EVENT-REPORT

Table N.7-83 lists the Status Codes that the SCP of the Print Job SOP Class supports for the N-EVENT-REPORT message and defines conditions in which the listed Status Codes are sent.

[Describe the condition which causes the application to send the specific Status Codes. For each other Status Code used add a row to the table.]

Table N.7-83. Status Codes for N-EVENT-REPORT of the Print Job SOP Class - SCP

Service Status

Further Meaning

Status Code

Condition

Success

Success

0000

Failure

Processing Failure

0110

No such SOP Instance

0112

No Such Event Type

0113

No Such Argument

0114

Invalid Argument Value

0115

Invalid SOP Instance

0117

No such SOP Class

0118

Class-Instance conflict

0119

Duplicate invocation

0210

Unrecognized operation

0211

Mistyped argument

0212

Resource limitation

0213


N.7.3.2.8.16.2 SCP of the Print Job SOP Class - N-GET

Table N.7-84 lists the Status Codes that the SCP of the Print Job SOP Class supports for the N-GET message and defines conditions in which the listed Status Codes are sent.

[Describe the condition which causes the application to send the specific Status Codes. For each other Status Code used add a row to the table.]

Table N.7-84. Status Codes for N-GET of the Print Job SOP Class - SCP

Service Status

Further Meaning

Status Code

Condition

Success

Success

0000

Warning

Attribute List warning

0107

Failure

Processing Failure

0110

No such SOP Instance

0112

Invalid SOP Instance

0117

No such SOP Class

0118

Class-Instance conflict

0119

Refused: Not authorized

0124

Duplicate invocation

0210

Unrecognized operation

0211

Mistyped argument

0212

Resource limitation

0213


N.7.3.2.8.17 SCP of the Presentation LUT SOP Class
N.7.3.2.8.17.1 SCP of the Presentation LUT SOP Class - N-CREATE

Table N.7-85 lists the Status Codes that the SCP of the Presentation LUT SOP Class supports for the N-CREATE message and defines conditions in which the listed Status Codes are sent.

[Describe the condition which causes the application to send the specific Status Codes. For each other Status Code used add a row to the table.]

Table N.7-85. Status Codes for N-CREATE of the Presentation LUT SOP Class - SCP

Service Status

Further Meaning

Status Code

Condition

Success

Success

0000

Warning

Attribute List warning

0107

Attribute Value out of range

0116

Requested Min Density or Max Density outside of printer's operating range

B605

Failure

No such Attribute

0105

Invalid Attribute Value

0106

Processing Failure

0110

Duplicate SOP Instance

0111

Invalid SOP Instance

0117

No such SOP Class

0118

Missing Attribute

0120

Missing Attribute Value

0121

Refused: Not authorized

0124

Duplicate invocation

0210

Unrecognized operation

0211

Mistyped argument

0212

Resource limitation

0213


N.7.3.2.8.17.2 SCP of the Presentation LUT SOP Class - N-DELETE

Table N.7-86 lists the Status Codes that the SCP of the Presentation LUT SOP Class supports for the N-DELETE message and defines conditions in which the listed Status Codes are sent.

[Describe the condition which causes the application to send the specific Status Codes. For each other Status Code used add a row to the table.]

Table N.7-86. Status Codes for N-DELETE of the Presentation LUT SOP Class - SCP

Service Status

Further Meaning

Status Code

Condition

Success

Success

0000

Failure

Processing Failure

0110

Invalid SOP Instance

0117

No such SOP Class

0118

Class Instance Conflict

0119

Refused: Not authorized

0124

Duplicate invocation

0210

Unrecognized operation

0211

Mistyped argument

0212

Resource limitation

0213


N.7.3.2.8.18 SCP of the Printer Configuration Retrieval SOP Class - N-GET

Table N.7-87 lists the Status Codes that the SCP of the Printer Configuration SOP Class supports for the N-GET message and defines conditions in which the listed Status Codes are sent.

[Describe the condition which causes the application to send the specific Status Codes. For each other Status Code used add a row to the table.]

Table N.7-87. Status Codes for N-GET of the Printer Configuration Retrieval SOP Class - SCP

Service Status

Further Meaning

Status Code

Condition

Success

Success

0000

Warning

Attribute List warning

0107

Failure

Processing Failure

0110

No such SOP Instance

0112

Invalid SOP Instance

0117

No such SOP Class

0118

Class-Instance conflict

0119

Refused: Not authorized

0124

Duplicate invocation

0210

Unrecognized operation

0211

Mistyped argument

0212

Resource limitation

0213


DICOM PS3.2 2026b - Conformance