DICOM PS3.2 2026c - Conformance
N.7.3.2.7 Query/Retrieve Service
N.7.3.2.7.1 SCU of the Query/Retrieve FIND SOP Classes - C-FIND

Table N.7-48 lists the Status Codes that the SCU of any of the Query/Retrieve FIND SOP Class supports for the C-FIND message and defines the application behavior when encountering the listed Status Codes.

[Describe the behavior of the application when it receives any of the Status Codes listed in the table below, e.g., displaying and logging the error code or retrying the request. For each additional Status Code supported add a row to the table.

In the "Other Status Codes" row document the behavior of the application in case it encounters an unknown Status Code.]

Table N.7-48. Status Codes C-FIND for Query/Retrieve FIND SOP Classes - SCU

Service Status

Further Meaning

Status Code

Behavior

Success

Matching is complete - No final identifier is supplied

0000

Failure

Refused: Out of Resources

A700

Error: Data Set does not match SOP Class

A900

Error: Unable to process

C000-CFFF

SOP Class Not Supported

0122

Cancel

Matching terminated due to cancel

FE00

Pending

Matches are continuing - Current Match is supplied and any Optional Keys were supported in the same manner as Required Keys.

FF00

Matches are continuing - Warning that one or more Optional Keys were not supported for existence for this Identifier

FF01

-

Other Status Codes

anything else


N.7.3.2.7.2 SCU of the Query/Retrieve MOVE SOP Classes - C-MOVE

Table N.7-49 lists the Status Codes that the SCU of any of the Query/Retrieve MOVE SOP Class supports for the C-MOVE message and defines the application behavior when encountering the listed Status Codes.

[Describe the behavior of the application when it receives any of the Status Codes listed in the table below, e.g., displaying and logging the error code or retrying the request. For each additional Status Code supported add a row to the table.

In the "Other Status Codes" row document the behavior of the application in case it encounters an unknown Status Code.]

Table N.7-49. Status Codes C-MOVE for Query/Retrieve MOVE SOP Classes - SCU

Service Status

Further Meaning

Status Codes

Related Fields

Behavior

Success

Sub-operations Complete - No Failures

0000

(0000,1020)

(0000,1021)

(0000,1022)

(0000,1023)

Warning

Sub-operations Complete - One or more Failures

B000

(0000,1020)

(0000,1022)

(0000,1023)

Failed

Out of Resources - Unable to calculate number of matches

A701

(0000,0902)

Out of Resources - Unable to perform sub-operations

A702

(0000,1020)

(0000,1021)

(0000,1022)

(0000,1023)

Move Destination unknown

A801

(0000,0902)

Data Set does not match SOP Class

A900

(0000,0901)

(0000,0902)

Unable to process

Cxxx

(0000,0901)

(0000,0902)

Cancel

Sub-operations terminated due to Cancel Indication

FE00

(0000,1020)

(0000,1021)

(0000,1022)

(0000,1023)

Pending

Sub-operations are continuing

FF00

(0000,1020)

(0000,1021)

(0000,1022)

(0000,1023)

-

Other Status Codes

anything else

-


N.7.3.2.7.3 SCP of the Query/Retrieve FIND SOP Classes - C-FIND

Table N.7-50 lists the Status Codes that the SCP of any of the Query/Retrieve FIND SOP Classes supports for the C-FIND message and defines conditions in which the listed Status Codes are sent.

[List the Attributes that are used to further detail the Status Codes in the "Related Fields Columns". Use N/A if there are no related fields used. Describe the condition which causes the application to send the specific Status Codes. For each other Status Code used add a row to the table.]

Table N.7-50. Status Codes C-FIND for Query/Retrieve FIND SOP Classes - SCP

Service Status

Further Meaning

Status Code

Related Fields

Condition (and Comments on Related Fields)

Success

Matching is complete - No final identifier is supplied

0000

Failure

Refused: Out of Resources

A700

Error: Data Set does not match SOP Class

A900

Error: Unable to process

C000

SOP Class Not Supported

0122

Cancel

Matching terminated due to cancel

FE00

Pending

Matches are continuing - Current Match is supplied and any Optional Keys were supported in the same manner as Required Keys.

FF00

Matches are continuing - Warning that one or more Optional Keys were not supported for existence for this Identifier

FF01


N.7.3.2.7.4 SCP of the Query/Retrieve MOVE SOP Classes - C-MOVE

Table N.7-51 lists the Status Codes that the SCP of any of the Query/Retrieve MOVE SOP Classes supports for the C-MOVE message and defines conditions in which the listed Status Codes are sent.

[Describe the condition which causes the application to send the specific Status Codes. For each other Status Code used add a row to the table.]

[Describe the action on the C-STORE sub-operation due to above mentioned conditions. Mention what happens to the C-STORE sub-operation when the specific condition occurs.]

Table N.7-51. Status Codes C-MOVE for Query/Retrieve MOVE SOP Classes - SCP

Service Status

Further Meaning

Status Codes

Related Fields sent in the response

Condition

Action on the Store due the condition.

Success

Sub-operations Complete - No Failures

0000

(0000,1020)

(0000,1021)

(0000,1022)

(0000,1023)

Warning

Sub-operations Complete - One or more Failures

B000

(0000,1020)

(0000,1022)

(0000,1023)

Failed

Out of Resources - Unable to calculate number of matches

A701

(0000,0902)

Out of Resources - Unable to perform sub-operations

A702

(0000,1020)

(0000,1021)

(0000,1022)

(0000,1023)

Move Destination unknown

A801

(0000,0902)

Data Set does not match SOP Class

A900

(0000,0901)

(0000,0902)

Unable to process

Cxxx

(0000,0901)

(0000,0902)

Cancel

Sub-operations terminated due to Cancel Indication

FE00

(0000,1020)

(0000,1021)

(0000,1022)

(0000,1023)

Pending

Sub-operations are continuing

FF00

(0000,1020)

(0000,1021)

(0000,1022)

(0000,1023)


DICOM PS3.2 2026c - Conformance