DICOM PS3.2 2026c - Conformance
N.7.3.2.5 Storage Service
N.7.3.2.5.1 SCU of the Storage SOP Classes - C-STORE

Table N.7-42 lists the Status Codes that the SCU of the Storage SOP Class supports for the C-STORE message and defines the application behavior when encountering the listed Status Codes.

[Describe the behavior of the application when it receives any of the Status Codes listed in the table below, e.g., displaying and logging the error code or retrying the request. For each additional Status Code supported add a row to the table.

In the "Other Status Codes" row document the behavior of the application in case it encounters an unknown Status Code.]

Table N.7-42. Status Codes C-STORE for the Storage SOP Classes - SCU

Service Status

Further Meaning

Status Code

Behavior

Success

Success

0000

Warning

Coercion of Data Elements

B000

Data Set does not match SOP Class

B007

Elements Discarded

B006

Failure

SOP Class not supported

0112

Invalid SOP Instance

0117

Duplicate invocation

0210

Unrecognized operation

0211

Mistyped argument

0212

Not authorized

0214

Out of Resources

A700-A7FF

Data Set does not match SOP Class

A900-A9FF

Cannot Understand

C000-CFFF

-

Other Status Codes

anything else


N.7.3.2.5.2 SCP of the Storage SOP Classes - C-STORE

Table N.7-43 lists the Status Codes that the SCP of the Storage SOP Classes supports for the C-STORE message and defines conditions in which the listed Status Codes are sent.

[Describe the condition which causes the application to send the specific Status Codes. For each other Status Code used add a row to the table.]

[List the Attributes that are used to further detail the Status Codes in the "Related Fields Columns". Use N/A if there are no related fields used.Further comments regarding the Related Fields can be provided in the "Condition" Column]

Table N.7-43. Status Codes C-STORE of the Storage SOP Classes - SCP

Service Status

Further Meaning

Status Codes

Related Fields

Condition (and Comments on Related fields)

Success

Success

0000

Warning

Coercion of Data Elements

B000

Data Set does not match SOP Class

B007

Elements Discarded

B006

Refused

Refused: Out of Resources

A700

Failure

Error: Data Set does not match SOP Class

A901

Error: Cannot understand

C000


DICOM PS3.2 2026c - Conformance