DICOM PS3.2 2022c - Conformance
I.4.2.2.2 WADO-URI Retrieve Rendered Imaging Document Set

Table I.4.2-6. WADO-URI Retrieve Rendered Imaging Documents Specification

Parameter

Restrictions

Transfer Syntaxes Supported

Restricted to transfer syntaxes supported by the hosting EXAMPLE-PACS-ARCHIVE

SOP Class restrictions

Restricted to SOP classes supported by the hosting EXAMPLE-PACS-ARCHIVE

Size restriction

Restricted to sizes supported by the hosting EXAMPLE-PACS-ARCHIVE

Rendered formats available

Supports JPEG and PDF for IMAGE IODS, and PDF for non-IMAGE IODS.

Rows restrictions

Must be in range 16 - 32767

Columns restrictions

Must be in range 16 - 32767

Region restrictions

None

Window Center restrictions

Whole window must be in the range of image pixel values.

Window Width restrictions

Must be greater than 4 and whole window must be in the range of image pixel values.

Image Quality restrictions

None

Anonymization

Supports the DICOM Basic Application Level Confidentiality Profile plus the Retain Patient Characteristics option.

Annotation Restrictions

None

Compression available

JPEG

Other restrictions

None


DICOM PS3.2 2022c - Conformance