DICOM PS3.2 2022d - Conformance

F.4.3.2 Additional Protocols

EXAMPLE-QUERY-RETRIEVE-SERVER conforms to the System Management Profiles listed in Table F.4.3-2. All requested transactions for the listed profiles and actors are supported. It does not support any optional transactions.

Table F.4.3-2. Supported System Management Profiles

Profile Name

Actor

Protocols Used

Optional Transactions

Security Support

Network Address Management

DHCP Client

DHCP

N/A

DNS Client

DNS

N/A


F.4.3.2.1 DHCP

DHCP can be used to obtain TCP/IP network configuration information. The network parameters obtainable via DHCP are shown in Table F.4.3-3. The Default Value column of the table shows the default used if the DHCP server does not provide a value. Values for network parameters set in the Service/Installation tool take precedence over values obtained from the DHCP server. Support for DHCP can be configured via the Service/Installation Tool. The Service/Installation tool can be used to configure the machine name. If DHCP is not in use, TCP/IP network configuration information can be manually configured via the Service/Installation Tool.

Table F.4.3-3. Supported DHCP Parameters

DHCP Parameter

Default Value

IP Address

None

Hostname

Requested machine name

List of NTP servers

Empty list

List of DNS servers

Empty list

Routers

Empty list

Static routes

None

Domain name

None

Subnet mask

Derived from IP Address (see service manual)

Broadcast address

Derived from IP Address (see service manual)

Default router

None

Time offset

Site configurable (from Time zone)

MTU

Network Hardware Dependent

Auto-IP permission

No permission


If the DHCP server refuses to renew a lease on the assigned IP address all active DICOM Associations will be aborted.

DICOM PS3.2 2022d - Conformance