DICOM PS3.2 2022b - Conformance |
---|
STORAGE-SCU attempts to initiate a new association for each instance it attempts to transfer.
For each instance selected from the user interface to be transferred, a single attempt will be made to transmit it to the selected remote AE. If the send fails, for whatever reason, no retry will be performed, and an attempt will be made to send the next instance.
Table D.4.2-16. Proposed Presentation Contexts for STORAGE-SCU and Receive Storage Request
See Table D.4.2-12 |
See Table D.4.2-12 |
||||
STORAGE-SCU will propose Presentation Contexts only for the SOP Class of the instance that is to be transferred.
For that SOP Class, STORAGE-SCU will propose multiple Presentation Contexts, one for each of the supported Transfer Syntaxes, and an additional Presentation Context with all of the supported Transfer Syntaxes, in order to determine which Transfer Syntaxes the remote SCP supports, and which it prefers.
STORAGE-SCU prefers explicit Transfer Syntaxes. If offered a choice of Transfer Syntaxes in the accepted Presentation Contexts, it will apply the following priority to the choice of Presentation Context to use for the C-STORE operation:
DICOM PS3.2 2022b - Conformance |
---|