DICOM PS3.2 2022b - Conformance

C.8 Annexes

C.8.1 IOD Contents

C.8.1.1 Created SOP Instances

DICOMRis does not create SOP instances

C.8.1.2 Usage of Attributes From Received IODs

Fields from MPPS such as technique and supplies and how they are used.

Table C.8.1-1. Attributes in MPPS IOD Used By DICOMRis Applications

Attribute Name

Tag

Database Updates

SOP Common Module

Performed Procedure Step Relationship Module

Scheduled Step Attribute Sequence

(0040,0270)

>Accession Number

(0008,0050)

These attributes need to match values in the DICOMRis database so other data contained in MPPS messages e.g., Dose and Materials data, can update the database and be displayed by the RisView application as described below

Patient ID

(0010,0020)

Performed Procedure Step Information

Performed Station AE Title

(0040,0241)

This attribute is used by the MppsSrv application as a key into the DICOM Configuration database to determine if the procedure referenced by the MPPS message should automatically have its state changed

Performed Procedure Step Description

(0040,0254)

Values for these attributes are accessible using the RisView application if they have been stored

Procedure Code Sequence

(0008,1032)

>Code Value

(0008,0100)

>Coding Scheme Designator

(0008,0102)

>Code Meaning

(0008,0104)

Image Acquisition Results

Performed Protocol Code Sequence

(0040,0260)

Values for these attributes are required if the RisView application is to display the protocol used to perform the procedure

>Code Value

(0008,0100)

>Coding Scheme Designator

(0008,0102)

>Code Meaning

(0008,0104)

Performed Series Sequence

(0040,0340)

>Protocol Name

(0018,1030)

>Operator's Name

(0008,1070)

Can be displayed by RisView application

>Retrieve AE Title

(0008,0054)

Billing and Material Management Code

Billing Procedure Step Sequence

Values are needed for these attributes so that the RisView application can display actual supplies and film used to perform a given procedure rather than default values associated with the given procedure in the DICOMRis database

The values of these attributes may also be used to generate charges if the site is configured for charging based on MPPS

>Code Value

(0008,0100)

>Coding Scheme Designator

(0008,0102)

>Code Meaning

(0008,0104)

Film Consumption Sequence

(0040,0321)

> Number of Films

(2100,0170)

> Medium Type

(2000,0030)

> Film Size ID

(2010,0050)

Billing Supplies and Devices Sequence

(0040,0384)

>Billing Item Sequence

(0040,0296)

>>Code Value

(0008,0100)

>>Coding Scheme Designator

(0008,0102)

>>Code Meaning

(0008,0104)

>Quantity Sequence

(0040,0293)

>>Quantity

(0040,0294)

>>Measuring Units Sequence

(0040,0295)

>>>Code Value

(0008,0100)

>>>Coding Scheme Designator

(0008,0102)

>>>Code Meaning

(0008,0104)


C.8.1.3 Attribute Mapping

The mapping between attributes received via HL7 from the HIS and those supplied in Modality Worklist is configurable. The default mapping is contained in the table below. Empty cells indicate that there is no mapping for the specific attribute

Table C.8.1-2. HL7/Modality Worklist Attribute Mapping

DICOM Attribute

DICOM Tag

HL7 Attribute Name

HL7 Segment

Notes

Scheduled Procedure Step

Scheduled Procedure Step Sequence

(0040,0100)

>Scheduled Station AE Title

(0040,0001)

DICOMRis generated

>Scheduled Procedure Step Start Date

(0040,0002)

Quantity/Timing

ORM OBR:27

DICOMRis generated

>Scheduled Procedure Step Start Time

(0040,0003)

Quantity/Timing

ORM OBR:27

DICOMRis generated

>Modality

(0008,0060)

DICOMRis generated

>Scheduled Performing Physician's Name

(0040,0006)

Technician

ORM OBR:34

>Scheduled Procedure Step Description

(0040,0007)

DICOMRis generated

>Scheduled Station Name

(0040,0010)

DICOMRis generated

>Scheduled Procedure Step Location

(0040,0011)

DICOMRis generated

>Scheduled Protocol Code Sequence

(0040,0008)

>>Code Value

(0008,0100)

DICOMRis generated

>>Coding Scheme Designator

(0008,0102)

DICOMRis generated

>>Code Meaning

(0008,0104)

DICOMRis generated

>Pre-Medication

(0040,0012)

DICOMRis generated

>Scheduled Procedure Step ID

(0040,0009)

DICOMRis generated

>Requested Contrast Agent

(0032,1070)

DICOMRis generated

>Scheduled Procedure Step Status

(0040,0020)

DICOMRis generated

>Comments on the Scheduled Procedure Step

(0040,0400)

DICOMRis generated

Requested Procedure

Requested Procedure ID

(0040,1001)

DICOMRis generated

Requested Procedure Description

(0032,1060)

DICOMRis generated

Requested Procedure Code Sequence

(0032,1064)

>Code Value

(0008,0100)

Universal Service Id

ORM OBR:44

The value in the HL7 attribute is mapped to one or more procedure codes in the DICOMRis database. The mapping is configurable

>Coding Scheme Designator

(0008,0102)

Universal Service Id

ORM OBR:44

Maps to a site-defined Coding Scheme, the LOINC Coding Scheme or the DICOMRis internal Coding Scheme

>Code Meaning

(0008,0008)

DICOMRis generated

Study Instance UID

(0020,000D)

DICOMRis generated

Referenced Study Sequence

(0008,1110)

>Referenced SOP Class UID

(0008,1150)

DICOMRis generated

>Referenced SOP Instance UID

(0008,1155)

DICOMRis generated

Requested Procedure Priority

(0040,1003)

ORM OBR:27

Patient Transport Arrangements

(0040,1004)

ORM OBR:30

Reason for the Requested Procedure

(0040,0040)

DICOMRis generated

Imaging Service Request

Accession Number

(0008,0050)

DICOMRis generated

Requesting Physician

(0032,1032)

ORM OBR:16

Referring Physician's Name

(0008,0090)

ORM PV1:8

Reason for the Imaging Service

Request

(0040,2001)

Reason for Study

ORM OBR:31

Order Entered By

(0040,2008)

Entered By

ORM ORC:10

Order Enterer's Location

(0040,2009)

Entering Organization

ORM ORC:17

Visit Identification

Admission ID

(0038,0010)

ADT PID:3

Admitting Diagnosis Description

(0008,1080)

ADT DG1:4

Admitting Diagnoses Code Sequence

(0008,1084)

>Code Value

(0008,0008)

ADT DG1:3

>Coding Scheme Designator

(0008,0008)

ADT DG1:2

>Code Meaning

(0008,0008)

Patient Identification

Patient's Name

(0010,0010)

ADT PID:5

Patient ID

(0010,0020)

ADT PID:3

Patient Demographics

Patients Birth Date

(0010,0030)

ADT PID:7

Patient's Sex

(0010,0040)

ADT PID:8

Patient's Weight

(0010,1030)

ADT OBX:5

Ethnic Group

(0010,2160)

ADT PID:10

Patient Comment

(0010,4000)

ORM NTE:3

Patient Medical

Patient State

(0038,0500)

Danger Code

ORM OBR:12

Pregnancy Status

(0010,21C0)

Filler Field 1

ORM OBR:20

Medical Alerts

(0010,2000)

Relevant Clinical Information

ORM OBR:13

Allergies

(0010,2110)

ADT AL1:3

Last Menstrual Date

(0010,21D0)

Filler Field 1

ORM OBR:20


C.8.1.4 Coerced/Modified Fields

Table C.8.1-3. Coerced Fields for Modality Performed Procedure Step

Attribute Name

Tag

Coercion Conditions

Performed Procedure Step Relationship Module

Scheduled Step Attribute Sequence

(0040,0270)

>Accession Number

(0008,0050)

Procedure Step has been placed in the Exception queue due to failure to match DICOMRis database. User enters a corrected value for Accession number through the RisView application

>Placer Order Number/Imaging Service Request

(0040,2006)

Value for this attribute is coerced when the value does not match the corresponding value in the DICOMRis database. This may occur when the step is initially processed or during exception resolution

>Filler Order Number/Imaging Service Request

(0040,2007)

As for attribute Placer Order Number/Imaging Service Request

>Requested Procedure ID

(0040,1001)

As for attribute Placer Order Number/Imaging Service Request

>Requested Procedure Description

(0032,1060)

As for attribute Placer Order Number/Imaging Service Request

>Scheduled Procedure Step ID

(0040,0009)

As for attribute Placer Order Number/Imaging Service Request

>Scheduled Procedure Step Description

(0040,0007)

As for attribute Placer Order Number/Imaging Service Request

>Scheduled Protocol Code Sequence

(0040,0008)

As for attribute Placer Order Number/Imaging Service Request

>>Code Value

(0008,0100)

As for attribute Placer Order Number/Imaging Service Request

>>Coding Scheme designator

(0008,0102)

As for attribute Placer Order Number/Imaging Service Request

>>Code Meaning

(0008,0104)

As for attribute Placer Order Number/Imaging Service Request

Patient Name

(0010,0010)

As for attribute Placer Order Number/Imaging Service Request

Patient ID

(0010,0020)

Procedure Step has been placed in the Exception queue due to failure to match DICOMRis database. User enters a corrected value for Patient ID through the RisView application

Patient's Birth Date

(0010,0030)

As for attribute Placer Order Number/Imaging Service Request

Patient's Sex

(0010,0040)

As for attribute Placer Order Number/Imaging Service Request


DICOM PS3.2 2022b - Conformance