General Parameters
|
Time-out waiting for acceptance or rejection Response to an Association Open Request
|
Yes
|
30 Seconds
|
Time-out waiting for response to TCP/IP connect() request.
|
Yes
|
15 Seconds
|
Time-out for waiting for data between TCP/IP packets. (Low-level timeout)
|
Yes
|
15 Seconds
|
Time-out waiting for a response to a DIMSE Request
|
Yes
|
30 Seconds
|
Time-out waiting for the next DIMSE Request
|
Yes
|
60 Seconds
|
Debugging Capabilities
|
Hex Dump DIMSE Messages
|
Yes
|
Off
|
Hex Dump Association Messages
|
Yes
|
Off
|
TCP/IP Settings
|
TCP/IP Send Buffer
|
Yes
|
65535 Bytes
|
TCP//IP Receive Buffer
|
Yes
|
65535 Bytes
|
PacketFilter
|
Yes
|
On. This option enables running of tcpdump utility from the command line to capture TCP packet headers/contents
|
DICOMSRV Parameters
|
Maximum Number of Simultaneous Associations
|
Yes
|
20
|
Maximum Number of Associations to a given device
|
Yes
|
3
|
Maximum PDU size the AE can receive
|
Yes
|
65536 Bytes
|
Maximum PDU size the AE can send
|
No
|
The lower of the value above and the max PDU size specified by the Remote AE in the Association Request
|
Validation of DICOM Service Messages
|
Yes
|
Validate messages and log validation errors. Do not automatically return error for all validation errors
|
Modality Worklist Parameters
|
Maximum Number of Matches for an MWL Request
|
Yes
|
100
|
Time period after Scheduled Date/Time to leave SPS entries in the SPS Worklist
|
Yes
|
2880 min
|
State of Parent Procedure that causes deletion of child SPS Entries
|
Yes
|
PROCEDURE STARTED
|
Supported Transfer Syntaxes
|
Yes
|
Explicit VR Little Endian
Implicit VR Little Endian
|
Modality Performed Procedure Step Parameters
|
Generate charges based on supplies specified in MPPS transactions
|
Yes
|
Off
|
Purge Period for MPPS transactions in final state
|
Yes
|
30 days
|
State to automatically set procedures to for a given AE on receipt of matching N-CREATE
|
Yes
|
|
State to automatically set procedures to for a given AE on receipt of matching N-SET COMPLETED
|
Yes
|
|
State to automatically set procedures to for a given AE on receipt of matching N-SET DISCONTINUED
|
Yes
|
|
Flag specifying support for IHE Trauma cases for a given AE
|
Yes
|
false
|
Patient ID Range to be used for Patient Registration for IHE Trauma case
|
Yes
|
|
Default Procedure Code to be used for orders for IHE Trauma cases
|
Yes
|
|
Supported Transfer Syntaxes
|
Yes
|
Explicit VR Little Endian
Implicit VR Little Endian
|