  <title>Digital Imaging and Communications in Medicine (DICOM)</title>
  <subtitle>Sup 201 - Retirement of Radiation Dose Module from Modality Performed Procedure Step</subtitle>
  <info>
    <author>
      <orgname>DICOM Standards Committee - Working Group 6 - Base Standard</orgname>
      <address>
            <street>1300 N. 17th Street Suite 1752</street>
            <city>Rosslyn</city>
            <state>VA</state>
            <postcode>22209</postcode>
            <country>USA</country>
         </address>
    </author>
    <copyright>
      <year>2017</year>
      <holder>NEMA</holder>
    </copyright>
    <releaseinfo>Status: Final Text</releaseinfo>
    <pubdate>2017/09/15</pubdate>
    <legalnotice>
      <para>This supplement is prepared pursuant to work item: 2016-12-A</para>
    </legalnotice>
  </info>
 
  
  <chapter>
    <title>Scope and Field of Application</title>
    <para>This Supplement retires the Radiation Dose Module from the Modality Performed Procedure Step SOP Class.</para>
    <para>DICOM periodically retires various features that have not been used widely in an interoperable manner, or have been superseded by more robust services. Retirement does not require any implementations to stop using them. It does mean DICOM will no longer maintain them. </para>
    <para>The Radiation Dose Structured Report (RDSR) is the preferred mechanism for exchanging radiation dose exposure information (introduced for XA in Sup 94 in 2005 and CT in Sup 127 in 2007), and the continuing presence of the Radiation Dose Module of the Modality Performed Procedure Step SOP Class in the standard is causing confusion (e.g., as is apparent in the recent draft ICRP report on DRLs, "http://www.icrp.org/page.asp?id=256").</para>
    <para>The optional Radiation Dose Module of the Modality Performed Procedure Step SOP Class was published in 1998, but has not been adopted widely and neither provides a means of persistently storing nor managing the more highly structured radiation dose information that is now commonplace.</para>
    <para>The Radiation Dose Structured Report (RDSR) mechanism has long since replaced MPPS as the mechanism for interchange of radiation dose information for all X-Ray modalities, such as CT, XA/XRF and projection radiography, including mammography.</para>
    <para>RDSR, not MPPS, is used in the IHE Radiation Exposure Monitoring (REM) profile and is the preferred mechanism for submission to registries.</para>
    <para>The RDSR retains the ability to describe an MPPS as the source of information, to allow the conversion from legacy equipment.</para>
  </chapter>
  
  <chapter>
    <title/>
    <informaltable rules="all" frame="box">
      <tr>
        <td>
          <para><emphasis role="italic">Amend DICOM PS3.2 as follows (changes to existing text are bold and <emphasis role="bold"><emphasis role="underline">underlined</emphasis></emphasis> for additions and <emphasis role="bold"><emphasis role="strikethrough">struckthrough</emphasis></emphasis> for removals):</emphasis></para>
        </td>
      </tr>
    </informaltable>
    
                    <table frame="box" label="C.4.2-10" rules="all" xml:id="table_C.4.2-10">
                  <caption>Supported N-SET/N-CREATE Attributes for MPPS</caption>
                  <thead>
                    <tr valign="top">
                      <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_eca3f35f-f25d-4a33-bef0-a4b5de8c7bdd">Attribute Name</para>
                      </th>
                      <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_22d4c6ed-8709-4165-b045-2bfe136e4fff">Tag</para>
                      </th>
                      <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_be2d394b-6ab7-4997-80d2-d0aa23bda96e">N-Create</para>
                      </th>
                      <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_8ff23db7-9bb1-43ad-8c9e-f2937d736972">N-Set</para>
                      </th>
                      <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_73bbbd36-c7c2-41cc-988b-65047e10014b">Database Updates</para>
                      </th>
                    </tr>
                  </thead>
                  <tbody>
                    <tr valign="top">
                      <td align="left" colspan="5" rowspan="1">
                        <para xml:id="para_ace38eb1-2b50-4470-ae3f-ff32505126e1"><emphasis role="bold"><emphasis role="strikethrough">Radiation Dose</emphasis></emphasis></para>
                      </td>
                    </tr>
                    <tr valign="top">
                      <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_9a4df036-5651-4d87-9003-bf32c33d4bda"><emphasis role="bold"><emphasis role="strikethrough">Anatomic Structure, Space or Region Sequence</emphasis></emphasis></para>
                      </td>
                      <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_b4f60f9c-4cd8-4d45-aaca-3df9fb965e22"><emphasis role="bold"><emphasis role="strikethrough">(0008,2229)</emphasis></emphasis></para>
                      </td>
                      <td align="center" colspan="1" rowspan="1"/>
                      <td align="center" colspan="1" rowspan="1"/>
                      <td align="center" colspan="1" rowspan="1"/>
                    </tr>
                    <tr valign="top">
                      <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_879595df-3d07-48e5-92a3-a2aa2b562f74"><emphasis role="bold"><emphasis role="strikethrough">&gt;Code Value</emphasis></emphasis></para>
                      </td>
                      <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_1401c0d0-217e-4d72-9c7a-312dcb202eb1"><emphasis role="bold"><emphasis role="strikethrough">(0008,0100)</emphasis></emphasis></para>
                      </td>
                      <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_8f5ab13a-7b46-43f6-aac0-c46b096299a8"><emphasis role="bold"><emphasis role="strikethrough">Y</emphasis></emphasis></para>
                      </td>
                      <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_7672f586-4f20-46fa-8d1e-c6d5c0a7462d"><emphasis role="bold"><emphasis role="strikethrough">Y</emphasis></emphasis></para>
                      </td>
                      <td align="center" colspan="1" rowspan="1"/>
                    </tr>
                    <tr valign="top">
                      <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_fce16a0d-9f00-4746-b27f-b3fa91c4f376"><emphasis role="bold"><emphasis role="strikethrough">&gt;Coding Scheme Designator</emphasis></emphasis></para>
                      </td>
                      <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_a8cffe00-a965-46ea-b6d9-7e3c9ef58693"><emphasis role="bold"><emphasis role="strikethrough">(0008,0102)</emphasis></emphasis></para>
                      </td>
                      <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_5e462004-3ccc-476e-ba37-5530fed61b89"><emphasis role="bold"><emphasis role="strikethrough">Y</emphasis></emphasis></para>
                      </td>
                      <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_39ae7e63-cc3f-48e8-aaac-0b2163442ea3"><emphasis role="bold"><emphasis role="strikethrough">Y</emphasis></emphasis></para>
                      </td>
                      <td align="center" colspan="1" rowspan="1"/>
                    </tr>
                    <tr valign="top">
                      <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_50237d8b-981b-4a01-ad4c-01135348cecc"><emphasis role="bold"><emphasis role="strikethrough">&gt;Code Meaning</emphasis></emphasis></para>
                      </td>
                      <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_5f769dc8-d25a-48c8-a942-58cffc41f826"><emphasis role="bold"><emphasis role="strikethrough">(0008,0104)</emphasis></emphasis></para>
                      </td>
                      <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_fca248de-c181-44e5-9571-4f8e3ace770d"><emphasis role="bold"><emphasis role="strikethrough">Y</emphasis></emphasis></para>
                      </td>
                      <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_9d7e2402-9469-4d33-bee1-8ba31218ad08"><emphasis role="bold"><emphasis role="strikethrough">Y</emphasis></emphasis></para>
                      </td>
                      <td align="center" colspan="1" rowspan="1"/>
                    </tr>
                    <tr valign="top">
                      <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_60eaf5d9-c857-4416-a5a8-455bb6037fdf"><emphasis role="bold"><emphasis role="strikethrough">Total Time of Fluoroscopy</emphasis></emphasis></para>
                      </td>
                      <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_b022219a-f172-4f8d-8b18-a962b839e8b0"><emphasis role="bold"><emphasis role="strikethrough">(0040,0300)</emphasis></emphasis></para>
                      </td>
                      <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_dcdcfc20-bfe1-4de9-af7d-3d96496d537e"><emphasis role="bold"><emphasis role="strikethrough">Y</emphasis></emphasis></para>
                      </td>
                      <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_b2b2ad27-aafd-41fd-9c5e-89eb27487f51"><emphasis role="bold"><emphasis role="strikethrough">Y</emphasis></emphasis></para>
                      </td>
                      <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_e0d6e946-5b06-45e0-ad3d-f4041dbba9bf"><emphasis role="bold"><emphasis role="strikethrough">Stored in Procedure Techniques table</emphasis></emphasis></para>
                      </td>
                    </tr>
                    <tr valign="top">
                      <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_1cd49317-73d8-4994-aaa6-292d15618a76"><emphasis role="bold"><emphasis role="strikethrough">Total Number of Exposures</emphasis></emphasis></para>
                      </td>
                      <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_40dd5594-c454-48db-a2cf-fbe862165d03"><emphasis role="bold"><emphasis role="strikethrough">(0040,0301)</emphasis></emphasis></para>
                      </td>
                      <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_ed6cb50c-ee4b-4392-b889-ee224d5df4ce"><emphasis role="bold"><emphasis role="strikethrough">Y</emphasis></emphasis></para>
                      </td>
                      <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_0cb94ce1-1cab-473e-b7d3-c128de0fb290"><emphasis role="bold"><emphasis role="strikethrough">Y</emphasis></emphasis></para>
                      </td>
                      <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_aa99dba5-7048-45ba-bf5c-f95609ac2394"><emphasis role="bold"><emphasis role="strikethrough">Stored in Procedure Techniques table</emphasis></emphasis></para>
                      </td>
                    </tr>
                    <tr valign="top">
                      <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_454fdef0-1077-47f7-88da-a1559863c49e"><emphasis role="bold"><emphasis role="strikethrough">Distance Source to Detector</emphasis></emphasis></para>
                      </td>
                      <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_f5580342-5655-405c-9c4f-1ecd9d40b673"><emphasis role="bold"><emphasis role="strikethrough">(0018,1110)</emphasis></emphasis></para>
                      </td>
                      <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_2f8d4e49-67ed-4f1c-8eeb-f5b1c8435375"><emphasis role="bold"><emphasis role="strikethrough">Y</emphasis></emphasis></para>
                      </td>
                      <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_ab7444d4-dcdf-4459-a120-f6ec654dd2b7"><emphasis role="bold"><emphasis role="strikethrough">Y</emphasis></emphasis></para>
                      </td>
                      <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_f5341476-89a2-449f-b689-a3f5d1946f69"><emphasis role="bold"><emphasis role="strikethrough">Stored in Procedure Techniques table</emphasis></emphasis></para>
                      </td>
                    </tr>
                    <tr valign="top">
                      <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_4b7cfe0d-86f1-411c-afd0-3e40a3f410d5"><emphasis role="bold"><emphasis role="strikethrough">Distance Source to Entrance</emphasis></emphasis></para>
                      </td>
                      <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_9e46c86c-2d0c-464f-91c9-9cf63adb529c"><emphasis role="bold"><emphasis role="strikethrough">(0040,0306)</emphasis></emphasis></para>
                      </td>
                      <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_ca6fd521-fe1d-43eb-974b-33b8999e6b7d"><emphasis role="bold"><emphasis role="strikethrough">Y</emphasis></emphasis></para>
                      </td>
                      <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_76e0225e-84e6-4bc6-be7c-52f36e073663"><emphasis role="bold"><emphasis role="strikethrough">Y</emphasis></emphasis></para>
                      </td>
                      <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_a33354da-1b1c-4d92-8653-d420f799e35f"><emphasis role="bold"><emphasis role="strikethrough">Stored in Procedure Techniques table</emphasis></emphasis></para>
                      </td>
                    </tr>
                    <tr valign="top">
                      <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_ecb35241-d040-47ee-a16e-9563a75af7ce"><emphasis role="bold"><emphasis role="strikethrough">Entrance Dose</emphasis></emphasis></para>
                      </td>
                      <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_7017b1df-77cf-448a-bb20-670809ab73a5"><emphasis role="bold"><emphasis role="strikethrough">(0040,0302)</emphasis></emphasis></para>
                      </td>
                      <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_8b826d59-0efa-4e53-9403-7ea4bf38b710"><emphasis role="bold"><emphasis role="strikethrough">Y</emphasis></emphasis></para>
                      </td>
                      <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_0ed021a2-0e63-420a-a5d5-058b339de116"><emphasis role="bold"><emphasis role="strikethrough">Y</emphasis></emphasis></para>
                      </td>
                      <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_bacf39f4-9466-4b47-b455-308bc21ac3eb"><emphasis role="bold"><emphasis role="strikethrough">Stored in Procedure Techniques table</emphasis></emphasis></para>
                      </td>
                    </tr>
                    <tr valign="top">
                      <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_35d4ab54-0a0a-43c4-a6d5-97ed2fb85659"><emphasis role="bold"><emphasis role="strikethrough">Entrance Dose in mGy</emphasis></emphasis></para>
                      </td>
                      <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_231e894d-1a84-4bd4-8e1b-a9b7bfa00e87"><emphasis role="bold"><emphasis role="strikethrough">(0040,8302)</emphasis></emphasis></para>
                      </td>
                      <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_98236713-d734-4103-b238-009f887f8362"><emphasis role="bold"><emphasis role="strikethrough">Y</emphasis></emphasis></para>
                      </td>
                      <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_610b6a41-8918-4975-bcff-809d83866f35"><emphasis role="bold"><emphasis role="strikethrough">Y</emphasis></emphasis></para>
                      </td>
                      <td align="center" colspan="1" rowspan="1"/>
                    </tr>
                    <tr valign="top">
                      <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_5fadd79e-e895-411d-90cd-75eda443f51f"><emphasis role="bold"><emphasis role="strikethrough">Exposed Area</emphasis></emphasis></para>
                      </td>
                      <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_fc6585f5-98a7-4119-b21e-da284ff6ae0b"><emphasis role="bold"><emphasis role="strikethrough">(0040,0303)</emphasis></emphasis></para>
                      </td>
                      <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_907ce19d-e9ce-4e82-96b3-ca86e30c2713"><emphasis role="bold"><emphasis role="strikethrough">Y</emphasis></emphasis></para>
                      </td>
                      <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_866d6971-b16c-402b-aa1f-900d130722ab"><emphasis role="bold"><emphasis role="strikethrough">Y</emphasis></emphasis></para>
                      </td>
                      <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_e5ba05e7-c182-4b34-bd4f-3c708a9ec308"><emphasis role="bold"><emphasis role="strikethrough">Stored in Procedure Techniques table</emphasis></emphasis></para>
                      </td>
                    </tr>
                    <tr valign="top">
                      <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_cc8a4e7f-59d3-4b15-8416-0a380c4b9ae0"><emphasis role="bold"><emphasis role="strikethrough">Image Area Dose Product</emphasis></emphasis></para>
                      </td>
                      <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_49ace80a-b916-4014-978b-ab5eac9b5047"><emphasis role="bold"><emphasis role="strikethrough">(0018,115E)</emphasis></emphasis></para>
                      </td>
                      <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_645bbda5-bfeb-48ae-9d9d-5ecbadc19c73"><emphasis role="bold"><emphasis role="strikethrough">Y</emphasis></emphasis></para>
                      </td>
                      <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_1fd6894e-f2e4-4131-8edd-58824c0891fa"><emphasis role="bold"><emphasis role="strikethrough">Y</emphasis></emphasis></para>
                      </td>
                      <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_49957212-6ed9-4cb3-9bac-34ae4e503720"><emphasis role="bold"><emphasis role="strikethrough">Stored in Procedure Techniques table</emphasis></emphasis></para>
                      </td>
                    </tr>
                    <tr valign="top">
                      <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_32251dcd-9935-4b4c-96d6-f5291162041c"><emphasis role="bold"><emphasis role="strikethrough">Comments on Radiation Dose</emphasis></emphasis></para>
                      </td>
                      <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_d1f0b6d0-c7aa-4317-9907-c1810f02d5ee"><emphasis role="bold"><emphasis role="strikethrough">(0040,0310)</emphasis></emphasis></para>
                      </td>
                      <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_4227a194-8dda-4119-a210-1728a8be86b9"><emphasis role="bold"><emphasis role="strikethrough">Y</emphasis></emphasis></para>
                      </td>
                      <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_8fb006d6-7ebc-47dc-a518-6bec0d69a015"><emphasis role="bold"><emphasis role="strikethrough">Y</emphasis></emphasis></para>
                      </td>
                      <td align="center" colspan="1" rowspan="1"/>
                    </tr>
                  </tbody>
                </table>
    
              <table frame="box" label="C.8.1-1" rules="all" xml:id="table_C.8.1-1">
            <caption>Attributes in MPPS IOD Used By DICOMRis Applications</caption>
            <thead>
              <tr valign="top">
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_d1de0470-5f8f-4a16-9f07-a5589af6f474">Attribute Name</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_e9ec0841-3203-4da7-9d8c-ba7e80999d53">Tag</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_871ab15b-9862-49aa-8d7d-fe9011261993">Database Updates</para>
                </th>
              </tr>
            </thead>
            <tbody>
              <tr valign="top">
                <td align="left" colspan="3" rowspan="1">
                  <para xml:id="para_4d82d94d-4cbf-4585-9cf7-a3464d56eed0"><emphasis role="bold"><emphasis role="strikethrough">Radiation Dose</emphasis></emphasis></para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_57b4dd04-4a6d-4364-b664-8706276613cd"><emphasis role="bold"><emphasis role="strikethrough">Total Time of Fluoroscopy</emphasis></emphasis></para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_600b0e41-162f-4527-88ae-da6dc86014c7"><emphasis role="bold"><emphasis role="strikethrough">(0040,0300)</emphasis></emphasis></para>
                </td>
                <td align="left" colspan="1" rowspan="7">
                  <para xml:id="para_fb7f8e79-8768-4b58-a3fb-829a5a77a4bb"><emphasis role="bold"><emphasis role="strikethrough">Values are needed for these attributes so that dose exposure data can be displayed by the RisView application</emphasis></emphasis></para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_0bb7e2a3-8e09-4090-b5f3-c7771390f66a"><emphasis role="bold"><emphasis role="strikethrough">Total Number of Exposures</emphasis></emphasis></para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_8183ce7b-726d-4878-ba05-c3f365a3778b"><emphasis role="bold"><emphasis role="strikethrough">(0040,0301)</emphasis></emphasis></para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_4f156e74-c178-4ad4-8c88-048f34333ae3"><emphasis role="bold"><emphasis role="strikethrough">Distance Source to Detector</emphasis></emphasis></para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_f476b540-4e77-40e7-afbd-f0ecbeb1009d"><emphasis role="bold"><emphasis role="strikethrough">(0018,1110)</emphasis></emphasis></para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_b671c4ed-fa88-46fe-bdbd-4b0fcf58cfec"><emphasis role="bold"><emphasis role="strikethrough">Distance Source to Entrance</emphasis></emphasis></para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_27c4391f-f0bb-4dd1-bcf6-d134fa962b79"><emphasis role="bold"><emphasis role="strikethrough">(0040,0306)</emphasis></emphasis></para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_ae502153-240b-4cf6-b053-95e3fdc43966"><emphasis role="bold"><emphasis role="strikethrough">Entrance Dose</emphasis></emphasis></para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_717a2687-0367-4c53-a2aa-439d65d363f1"><emphasis role="bold"><emphasis role="strikethrough">(0040,0302)</emphasis></emphasis></para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_0798c494-6ef4-43f1-a827-10751dfc5896"><emphasis role="bold"><emphasis role="strikethrough">Exposed Area</emphasis></emphasis></para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_e522c78f-3d16-4f19-91ef-3548a8bf7903"><emphasis role="bold"><emphasis role="strikethrough">(0040,0303)</emphasis></emphasis></para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_938d3e87-5011-4205-a3a8-7144121159bf"><emphasis role="bold"><emphasis role="strikethrough">Image Area Dose Product</emphasis></emphasis></para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_a1e99c5c-cc49-4ebd-ae74-cd290391415f"><emphasis role="bold"><emphasis role="strikethrough">(0018,115E)</emphasis></emphasis></para>
                </td>
              </tr>
            </tbody>
          </table>


    
    
    <informaltable rules="all" frame="box">
      <tr>
        <td>
          <para><emphasis role="italic">Amend DICOM PS3.3 as follows (changes to existing text are bold and <emphasis role="bold"><emphasis role="underline">underlined</emphasis></emphasis> for additions and <emphasis role="bold"><emphasis role="strikethrough">struckthrough</emphasis></emphasis> for removals):</emphasis></para>
        </td>
      </tr>
    </informaltable>
    
    <section label="B.17" status="2" xml:id="sect_B.17">
      <title>Modality Performed Procedure Step Information Object Definition</title>
      <section label="B.17.1" status="3" xml:id="sect_B.17.1">
        <title>IOD Description</title>
        <para xml:id="para_c8fb8e16-ccc6-4ee2-8403-251bc80ccea2">A "Modality Performed Procedure Step Information Object Definition" is an abstraction of the information that describes the activities, conditions and results of an imaging procedure performed on a modality. It contains information about the Modality Performed Procedure Step (MPPS) and its relations to other Information Entities of the DICOM real-world model as introduced in this Part.</para>
        <para xml:id="para_fbec0f96-fc8d-48ec-a79a-b3414db1fe88">A Modality Performed Procedure Step is related to the actual imaging procedure carried out at the modality. Other types of Performed Procedure Steps, e.g., reporting or image processing, are not covered by the Modality Performed Procedure Step IOD. The information gathered includes data about the performance of the procedure itself<emphasis role="bold"><emphasis role="strikethrough">, radiation dose values to which the patient has been exposed,</emphasis></emphasis> and data for billing and material management. The Modality Performed Procedure Step IOD includes general PPS modules and image acquisition specific ones, such as Image Acquisition Results<emphasis role="bold"><emphasis role="strikethrough">, Radiation Dose</emphasis></emphasis> and Billing and Material Management.</para>
      </section>
      <section label="B.17.2" status="3" xml:id="sect_B.17.2">
        <title>IOD Modules</title>
        <para xml:id="para_7ffdb9e9-e535-47b5-ac20-c35191428ffd">
          <xref linkend="table_B.17.2-1" xrefstyle="select: label"/> lists the modules that make up the Modality Performed Procedure Step IOD.</para>
        <table frame="box" label="B.17.2-1" rules="all" xml:id="table_B.17.2-1">
          <caption>Modality Performed Procedure Step IOD Modules</caption>
          <thead>
            <tr valign="top">
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_e8f2e2cb-af0e-4ffe-ac70-8c8a073244b5">Module</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_b1bd312e-bf23-4e6f-9c30-b1f42280e1cc">Reference</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_0221d9df-b338-4e07-8b30-a17f7395c899">Module Description</para>
              </th>
            </tr>
          </thead>
          <tbody>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c37d5f87-a6e0-43f8-9c16-0997a42870e2">SOP Common</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e9c7d941-9bf5-45e1-9db5-3fce0e3ef48d">
                  C.12.1
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_84179146-2a99-4891-b842-886268db55fd">Contains SOP common information</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6004eb25-71c6-4ea1-b677-5a9172196479">Performed Procedure Step Relationship</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_20a1a07a-8d2f-4127-b802-2992185c245b">
                  C.4.13
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_86a6af5b-66a0-4af0-8d77-a0f368c777de">References the related SOPs and IEs.</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_407e5644-4453-444d-9547-bc0be4ce3a09">Performed Procedure Step Information</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1007dff8-3457-4efd-a3c1-26c16ae765c2">
                  C.4.14
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_193810b2-2794-4661-abb7-d9f070bd4700">Includes identifying and status information as well as place and time</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f5b8fea0-802d-4201-ac49-16c8186ed889">Image Acquisition Results</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_26a5dcce-b1cd-4a88-8508-edbe0157e5df">
                  C.4.15
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b4f400ed-0a81-4fe6-a42b-089a2bee6253">Identifies Series and Images related to this PPS and specific image acquisition conditions.</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_74bfb7f8-d98c-4258-9ab7-eca2e21d8986"><emphasis role="bold"><emphasis role="strikethrough">Radiation Dose</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0b78877e-8984-4168-81e1-36dd1f10a5a8">
                  <emphasis role="bold"><emphasis role="strikethrough"><xref linkend="sect_C.4.16" xrefstyle="select: labelnumber"/></emphasis></emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0e67b51b-b12b-4fc5-878b-b04bc490e696"><emphasis role="bold"><emphasis role="strikethrough">Contains radiation dose information related to this Performed Procedure Step.</emphasis></emphasis></para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6ab8bcc3-f9b7-40ae-80d2-dbcd5e8ca008">Billing and Material Management Codes</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_9793c797-e733-4dc4-b75f-cf3d94ae6151">
                  C.4.17
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_271d94b7-643e-4f2e-a772-7ccf5e047107">Contains codes for billing and material management.</para>
              </td>
            </tr>
          </tbody>
        </table>
        <note>
          
          
          <para xml:id="para_68985c23-b55e-47b5-9e77-2b70de3a7aac">The <xref linkend="sect_C.4.16" xrefstyle="select: title"/>
            <emphasis role="bold"><emphasis role="strikethrough">does not have meaning if the modality does not generate ionizing radiation or if the generator does not provide the area dose product</emphasis><emphasis role="underline"> has been retired. See PS3.3 2017c</emphasis></emphasis>.</para>
          
          
        </note>
      </section>
    </section>
    
    <section label="C.4.16" status="3" xml:id="sect_C.4.16">
      <title>Radiation Dose Module<emphasis role="underline"> (Retired)</emphasis></title>
      <para><emphasis role="bold"><emphasis role="underline">This Module has been retired. See PS3.3 2017c.</emphasis></emphasis></para>
      <para xml:id="para_ed68bd24-839f-4335-968b-a13259c9b7a6"><emphasis role="bold"><emphasis role="strikethrough">
            <xref linkend="table_C.4-16" xrefstyle="select: label"/> defines the Attributes that may be used to communicate information related to radiation dose values. The attributes are intended to enable the Information System to store Patient exposure to ionizing radiation for legal purposes. Though these attributes are not intended to be used to accurately calculate volume dose distribution, they may serve for some quality control purposes.</emphasis></emphasis></para>
      <para xml:id="para_df7ef6f4-a53c-4b4c-8873-57fbc50585a0"><emphasis role="bold"><emphasis role="strikethrough">This module provides a means to communicate radiation dose values but DICOM does not define any requirements for the accuracy of these values, which may be defined in other professional, national or international standards.</emphasis></emphasis></para>
      <para xml:id="para_cd5c883f-ae79-4dcd-a118-d97bfe67272e"><emphasis role="bold"><emphasis role="strikethrough">The scope of the attributes contained in this module covers the entire acquisition that comprises the Modality Performed Procedure Step. Attributes that relate to single images, such as mAs or kVp, may be included in the Image IODs. It is beyond the scope of DICOM to define what attributes may be required to calculate or estimate area dose product values.</emphasis></emphasis></para>
      <note>
        <para xml:id="para_61580bad-aa02-49da-810c-4226f6bcc241"><emphasis role="bold"><emphasis role="strikethrough">The X-Ray Radiation Dose SR SOP Class provides a more comprehensive means of reporting radiation dose. Such a dose report may be referenced in Section C.4.15<!--<xref linkend="sect_C.4.15" xrefstyle="select: title"/>-->.</emphasis></emphasis></para>
      </note>
      <table frame="box" label="C.4-16" rules="all" xml:id="table_C.4-16">
        <caption><emphasis role="bold"><emphasis role="strikethrough">Radiation Dose Module Attributes</emphasis></emphasis></caption>
        <thead>
          <tr valign="top">
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_322e11d6-91c6-4c0b-8779-5e41e1d9364d"><emphasis role="bold"><emphasis role="strikethrough">Attribute Name</emphasis></emphasis></para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_960405ab-cbed-4e96-b2bb-a9106d70c2d7"><emphasis role="bold"><emphasis role="strikethrough">Tag</emphasis></emphasis></para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_3cd79036-6d02-4c4f-a463-9a01adff8500"><emphasis role="bold"><emphasis role="strikethrough">Attribute Description</emphasis></emphasis></para>
            </th>
          </tr>
        </thead>
        <tbody>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_886d6286-ed6d-432c-8ffb-f2c990d21860"><emphasis role="bold"><emphasis role="strikethrough">Anatomic Structure, Space or Region Sequence</emphasis></emphasis></para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_2059194e-d0eb-44be-b146-172ec468e746"><emphasis role="bold"><emphasis role="strikethrough">(0008,2229)</emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_7f19f448-adde-464f-9277-7af3ae352114"><emphasis role="bold"><emphasis role="strikethrough">Anatomic structure, space or region that has been exposed to ionizing radiation.</emphasis></emphasis></para>
              <para xml:id="para_92fce6f2-457c-4bb3-8154-ccaa8dda2b82"><emphasis role="bold"><emphasis role="strikethrough">Zero or one Item shall be included in this Sequence.</emphasis></emphasis></para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="2" rowspan="1">
              <para xml:id="para_5c2fe048-88bd-401a-b4bb-0472ad046493">
                <emphasis role="bold"><emphasis role="strikethrough"><emphasis role="italic">&gt;Include Table 8.8-1
                    </emphasis></emphasis></emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2a0b4a61-f6c2-4909-877b-d67c89dcc3d8">
                <emphasis role="bold"><emphasis role="strikethrough"><emphasis role="italic">No Baseline CID is defined.</emphasis></emphasis></emphasis>
              </para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_847285b4-fae4-4a9c-a3c2-1c7c0bc45f16"><emphasis role="bold"><emphasis role="strikethrough">Total Time of Fluoroscopy</emphasis></emphasis></para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_7c1d64e2-4027-49ad-99ff-8ec8dca4457e"><emphasis role="bold"><emphasis role="strikethrough">(0040,0300)</emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_03b0cdce-9f95-4b9d-86f7-9e900d99b038"><emphasis role="bold"><emphasis role="strikethrough">Total duration of X-Ray exposure during fluoroscopy in seconds (pedal time) during this Performed Procedure Step.</emphasis></emphasis></para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f45fc3e4-e0e4-4388-9c2b-cd05f101d0de"><emphasis role="bold"><emphasis role="strikethrough">Total Number of Exposures</emphasis></emphasis></para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_675df48a-f29b-4e89-aaac-d8a8acd7fd13"><emphasis role="bold"><emphasis role="strikethrough">(0040,0301)</emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_bb92f6d1-cc12-485e-ba96-0dbb4924e0ac"><emphasis role="bold"><emphasis role="strikethrough">Total number of exposures made during this Performed Procedure Step. The number includes non-digital and digital expo­sures.</emphasis></emphasis></para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ad9f9d7a-bc0e-4e80-a437-06d6c5fbb30a"><emphasis role="bold"><emphasis role="strikethrough">Distance Source to Detector</emphasis></emphasis></para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_b3b1866f-b449-4f6d-a850-f3ca17c12d5a"><emphasis role="bold"><emphasis role="strikethrough">(0018,1110)</emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4d0cdfb2-35ae-45eb-be94-cd1a95364d96"><emphasis role="bold"><emphasis role="strikethrough">Distance in mm from the source to detector center.</emphasis></emphasis></para>
              <note>
                <para xml:id="para_5579b837-25e1-490a-a24c-05ca3c296a39"><emphasis role="bold"><emphasis role="strikethrough">This value is traditionally referred to as Source Image Receptor Distance (SID).</emphasis></emphasis></para>
              </note>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ac011a3e-3907-45c8-8468-cd87bc7345ef"><emphasis role="bold"><emphasis role="strikethrough">Distance Source to Entrance</emphasis></emphasis></para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_31c0e84b-56ac-4595-ac85-c15273cffe6d"><emphasis role="bold"><emphasis role="strikethrough">(0040,0306)</emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ddbd68ef-f234-4f5d-a1ef-6a0ff02d557b"><emphasis role="bold"><emphasis role="strikethrough">Distance in mm from the source to the surface of the patient closest to the source during this Performed Procedure Step.</emphasis></emphasis></para>
              <note>
                <para xml:id="para_c7c67f6d-c2b4-4c69-bc42-ae6d4db198b6"><emphasis role="bold"><emphasis role="strikethrough">This may be an estimated value based on assumptions about the patient's body size and habitus.</emphasis></emphasis></para>
              </note>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2ca8dd1d-74a2-419c-8f1f-0fb8246b58c0"><emphasis role="bold"><emphasis role="strikethrough">Entrance Dose</emphasis></emphasis></para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_29d2ff91-cbbe-4579-b21b-caac20863c51"><emphasis role="bold"><emphasis role="strikethrough">(0040,0302)</emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d504731b-ecdf-405e-b2da-f217b1e9aa42"><emphasis role="bold"><emphasis role="strikethrough">Average entrance dose value measured in dGy at the surface of the patient­ during this Performed Procedure Step.</emphasis></emphasis></para>
              <note>
                <para xml:id="para_855bc487-3ff4-4a11-869a-0878caea7125"><emphasis role="bold"><emphasis role="strikethrough">This may be an estimated value based on assumptions about the patient's body size and habitus.</emphasis></emphasis></para>
              </note>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_89753a73-70ef-48a9-9dea-d9f3be095606"><emphasis role="bold"><emphasis role="strikethrough">Entrance Dose in mGy</emphasis></emphasis></para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_f57f8001-ca22-46e4-8cb8-a11b6bbb94cb"><emphasis role="bold"><emphasis role="strikethrough">(0040,8302)</emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_223315c7-547d-4cf8-9d0d-cbb38cdaea22"><emphasis role="bold"><emphasis role="strikethrough">Average entrance dose value measured in mGy at the surface of the patient­ during this Performed Procedure Step.</emphasis></emphasis></para>
              <note>
                <para xml:id="para_320c486f-9e74-45a9-bb8a-f00ea591b878"><emphasis role="bold"><emphasis role="strikethrough">This may be an estimated value based on assumptions about the patient's body size and habitus.</emphasis></emphasis></para>
              </note>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f74f0043-2c7f-43d9-bf97-e23e55c9783c"><emphasis role="bold"><emphasis role="strikethrough">Exposed Area</emphasis></emphasis></para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_5aa51c68-c95d-456c-bd60-12110f85e60d"><emphasis role="bold"><emphasis role="strikethrough">(0040,0303)</emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_746b854b-278a-4d7a-a01a-26c6e7239b74"><emphasis role="bold"><emphasis role="strikethrough">Typical dimension of the exposed area at the detector plane. If Rectangular: row dimension followed by column; if Round: diameter. Measured in mm.</emphasis></emphasis></para>
              <note>
                <orderedlist numeration="arabic">
                  <listitem>
                    <para xml:id="para_5391d83b-d335-48f9-acda-732ecd02de7d"><emphasis role="bold"><emphasis role="strikethrough">This may be an estimated value based on assumptions about the patient's body size and habitus.</emphasis></emphasis></para>
                  </listitem>
                  <listitem>
                    <para xml:id="para_6e72d2ea-3c0e-44fe-a380-ccc6da94ae1c"><emphasis role="bold"><emphasis role="strikethrough">This attribute is used in the Section C.8.7.8 with units in cm (see Section C.8.7.8, Table C.8-33).</emphasis></emphasis></para>
                  </listitem>
                </orderedlist>
              </note>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d83adb57-bc35-43bc-ac31-82dae053c07b"><emphasis role="bold"><emphasis role="strikethrough">Image and Fluoroscopy Area Dose Product</emphasis></emphasis></para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_02463679-82db-49ac-aeac-aa9fb2615ee9"><emphasis role="bold"><emphasis role="strikethrough">(0018,115E)</emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a4dce90e-2796-44db-b828-8b78fe4f4648"><emphasis role="bold"><emphasis role="strikethrough">Total area-dose-product to which the patient was exposed, accumulated over the complete Performed Procedure Step and measured in dGy*cm*cm, including fluoroscopy.</emphasis></emphasis></para>
              <note>
                <orderedlist numeration="arabic">
                  <listitem>
                    <para xml:id="para_31c3cae6-7fe0-4030-a44f-c73d41e64314"><emphasis role="bold"><emphasis role="strikethrough">The sum of the area dose product of all images of a Series or a Study may not result in the total area dose product to which the patient was exposed.</emphasis></emphasis></para>
                  </listitem>
                  <listitem>
                    <para xml:id="para_0e2c0a13-c275-4ba3-b5ee-d232696b40eb"><emphasis role="bold"><emphasis role="strikethrough">This may be an estimated value based on assumptions about the patient's body size and habitus.</emphasis></emphasis></para>
                  </listitem>
                </orderedlist>
              </note>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c0448e39-f0ea-40a7-a672-cecbb85b9c00"><emphasis role="bold"><emphasis role="strikethrough">Comments on Radiation Dose</emphasis></emphasis></para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_95c4a00f-72e1-4bec-99cc-8886c6cd985a"><emphasis role="bold"><emphasis role="strikethrough">(0040,0310)</emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_cd81ef72-67f7-4421-8dde-18d2a08cef66"><emphasis role="bold"><emphasis role="strikethrough">User-defined comments on any special conditions related to radiation dose encountered during this Performed Procedure Step.</emphasis></emphasis></para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_cdfb6b0c-fcb2-4051-9e7b-e7df99cadbf8"><emphasis role="bold"><emphasis role="strikethrough">Exposure Dose Sequence</emphasis></emphasis></para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_67a074ad-275c-425d-a93e-42916ebe4c24"><emphasis role="bold"><emphasis role="strikethrough">(0040,030E)</emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5d169884-4a2f-4a22-8448-df494bfbab23"><emphasis role="bold"><emphasis role="strikethrough">Exposure Dose Sequence will contain Total Number of Exposures (0040,0301) items plus an item for each fluoroscopy episode not already counted as an exposure.</emphasis></emphasis></para>
              <para xml:id="para_45b9e069-6a30-4ced-bd7f-efff8adf4c97"><emphasis role="bold"><emphasis role="strikethrough">Zero or more Items shall be included in this Sequence.</emphasis></emphasis></para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a9c86758-3f3b-418f-ac9f-8c803bf7c491"><emphasis role="bold"><emphasis role="strikethrough">&gt;Radiation Mode</emphasis></emphasis></para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_9885f933-d405-42b9-a082-5513ab73fdb2"><emphasis role="bold"><emphasis role="strikethrough">(0018,115A)</emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_966ec00f-a338-4497-864e-3ffc8e115a8a"><emphasis role="bold"><emphasis role="strikethrough">Specifies X-Ray radiation mode.</emphasis></emphasis></para>
              <variablelist spacing="compact">
                <title><emphasis role="bold"><emphasis role="strikethrough">Enumerated Values:</emphasis></emphasis></title>
                <varlistentry>
                  <term><emphasis role="bold"><emphasis role="strikethrough">CONTINUOUS</emphasis></emphasis></term>
                  <listitem>
                    <para xml:id="para_78835a05-ce6f-4cac-add2-3168b8134b27"/>
                  </listitem>
                </varlistentry>
                <varlistentry>
                  <term><emphasis role="bold"><emphasis role="strikethrough">PULSED</emphasis></emphasis></term>
                  <listitem>
                    <para xml:id="para_3cdc623a-2cc6-4137-a5bd-49e9decc5a60"/>
                  </listitem>
                </varlistentry>
              </variablelist>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_963b29e4-2a5a-4e08-a159-54355de797fc"><emphasis role="bold"><emphasis role="strikethrough">&gt;KVP</emphasis></emphasis></para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_01a23d8d-f769-4c69-ac3e-a75760c78937"><emphasis role="bold"><emphasis role="strikethrough">(0018,0060)</emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ab17d2ce-f2c3-4e76-ae95-ab456f1ca618"><emphasis role="bold"><emphasis role="strikethrough">Peak kilo voltage output of the x-ray generator used. An average in the case of fluoroscopy (continuous radiation mode).</emphasis></emphasis></para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9b8f8328-b061-45b7-ab2c-5e7ae31dfecf"><emphasis role="bold"><emphasis role="strikethrough">&gt;X-Ray Tube Current in µA</emphasis></emphasis></para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_a78add94-f2fa-4501-ab3c-14ac458b3789"><emphasis role="bold"><emphasis role="strikethrough">(0018,8151)</emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_685e0608-f7c9-4759-b3f3-1bfd18f656eb"><emphasis role="bold"><emphasis role="strikethrough">X-Ray Tube Current in µA. An average in the case of fluoroscopy (continuous radiation mode).</emphasis></emphasis></para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d5dc1fb9-90d4-4869-a254-c8b422936926"><emphasis role="bold"><emphasis role="strikethrough">&gt;Exposure Time</emphasis></emphasis></para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_a98b0b57-5341-4438-99dc-ca55f21dc5ec"><emphasis role="bold"><emphasis role="strikethrough">(0018,1150)</emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c8dd8a4d-021c-4053-87b2-0d3ad9a64e0b"><emphasis role="bold"><emphasis role="strikethrough">Time of x-ray exposure or fluoroscopy in msec.</emphasis></emphasis></para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3e9808a3-6325-4110-a10e-b9ce0b068095"><emphasis role="bold"><emphasis role="strikethrough">&gt;Filter Type</emphasis></emphasis></para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_6b2a34b0-0f67-47fa-9512-10217276fdc6"><emphasis role="bold"><emphasis role="strikethrough">(0018,1160)</emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_26936504-8d2c-40ab-9a01-298db01eb7c9"><emphasis role="bold"><emphasis role="strikethrough">Type of filter(s) inserted into the X-Ray beam (e.g., wedges). See Section C.8.7.10 and Section C.8.15.3.9 (for enhanced CT) for Defined Terms.</emphasis></emphasis></para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_83234385-ea0b-4439-a49f-7a3ab0a216ad"><emphasis role="bold"><emphasis role="strikethrough">&gt;Filter Material</emphasis></emphasis></para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_287df542-fb6f-4f06-a8f1-cd79428d741b"><emphasis role="bold"><emphasis role="strikethrough">(0018,7050)</emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_127b6fd6-1cba-4433-a2a4-3cc27f729459"><emphasis role="bold"><emphasis role="strikethrough">The X-Ray absorbing material used in the filter. May be multi-valued. See Section C.8.7.10 and Section C.8.15.3.9 (for enhanced CT) for Defined Terms.</emphasis></emphasis></para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d215efa9-5253-476f-96ff-070b20074e84"><emphasis role="bold"><emphasis role="strikethrough">&gt;Comments on Radiation Dose</emphasis></emphasis></para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_19515f92-16a2-4a71-943e-4fc95aec19e4"><emphasis role="bold"><emphasis role="strikethrough">(0040,0310)</emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_bbca8b38-9862-4552-ad07-5e08ca2efbe5"><emphasis role="bold"><emphasis role="strikethrough">User-defined comments on any special conditions related to radiation dose encountered during during the episode described by this Exposure Dose Sequence Item.</emphasis></emphasis></para>
            </td>
          </tr>
        </tbody>
      </table>
      <note>
        <orderedlist numeration="arabic">
          <listitem>
            
            <para xml:id="para_df92c98e-e314-4450-ae6f-d5e82cdb84e5"><emphasis role="bold"><emphasis role="strikethrough">The Anatomic Region may be deduced from attribute values available within the Modality Worklist Manage­ment SOP Class, such as Reason for Service Request, Reasons for Requested Procedure, Scheduled Proce­dure Step Description and Scheduled Protocol Code Sequence.</emphasis></emphasis></para>
          </listitem>
          
              <listitem>
                <para xml:id="para_5b235e7b-fdaf-46c9-a4c5-940b9efae1b5"><emphasis role="bold"><emphasis role="strikethrough">The Image Area Dose Product should take into account collimator position and filters, and the value for the Exposed Area should also take into account collimator position. If the equipment does not provide the Entrance Dose, it may be calculated using Area Dose Product, Exposed Area, SID and an assumed body thickness.</emphasis></emphasis></para>
              </listitem>
          <listitem>
            <para xml:id="para_fa7b25be-1402-43fd-ba53-fe45873af6fe"><emphasis role="bold"><emphasis role="strikethrough">The Distance Source to Detector (0018,1110) and Exposed Area (0040,0303) are only meaningful if they remain constant for all acquisitions during this Performed Procedure Step.</emphasis></emphasis></para>
          </listitem>
        </orderedlist>
      </note>
    </section>
    
    <section label="C.8.7.8" status="4" xml:id="sect_C.8.7.8">
          <title>X-Ray Acquisition Dose Module</title>
          <table frame="box" label="C.8-33" rules="all" xml:id="table_C.8-33">
            <caption>X-Ray Acquisition Dose Module Attributes</caption>
            <thead>
              <tr valign="top">
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_6c60ae2f-e3d6-466e-b4a5-2b6d3f65d828">Attribute Name</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_f412f7a2-9ec0-42ce-8f31-a2129ed16fcb">Tag</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_73df5bd2-9821-459c-bebc-a952db81281d">Type</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_b7e8defa-2074-4313-b41f-ceee6b261136">Attribute Description</para>
                </th>
              </tr>
            </thead>
            <tbody>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_8ef6dcd3-f4c9-4d6d-b9bc-b6ae3d72f40c">...</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_0b0d1220-d58c-4eed-b152-7eb69e859768">...</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_cb83c984-8cf9-4969-aac1-6a3326db43c8">...</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_c1aa3bc3-c61d-45eb-b064-f6d9043ed7d9">...</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_3aa37ec2-8703-4d1a-9fe8-0940684823b1">Exposed Area</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_335bbe7b-dd48-45b3-bdae-d83f918822d1">(0040,0303)</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_b994a6e3-ad9d-472a-805d-d82a19318f2f">3</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_c8062c9b-efa5-4482-8075-5467c69260d1">Typical dimension of the exposed area at the detector plane. If Rectangular: row dimension followed by column; if Round: diameter. Measured in cm.</para>
                  <note>
                    <orderedlist numeration="arabic">
                      <listitem>
                        <para xml:id="para_54eb9a46-b0ed-4608-b584-8bf4c538d37a">The exposed area should be consistent with values specified in the Section C.8.7.3, if present.</para>
                      </listitem>
                      <listitem>
                        <para xml:id="para_d94e1314-1af0-4e69-81b3-7b8cf2c900cc">This may be an estimated value based on assumptions about the patient's body size and habitus.</para>
                      </listitem>
                      <listitem>
                        <para xml:id="para_fd47b876-4803-4e68-860f-2d47bd4f66d3">This attribute <emphasis role="bold"><emphasis role="strikethrough">is</emphasis><emphasis role="underline">was previously</emphasis></emphasis> used in the <xref linkend="sect_C.4.16" xrefstyle="select: title"/> with units in mm <emphasis role="bold"><emphasis role="strikethrough">(see <xref linkend="sect_C.4.16" xrefstyle="select: label"/>, <xref linkend="table_C.4-16" xrefstyle="select: labelquotedtitle"/>)</emphasis></emphasis>.<emphasis role="bold"><emphasis role="underline"> This use has been retired. See PS3.3 2017c.</emphasis></emphasis></para>
                      </listitem>
                    </orderedlist>
                  </note>
                </td>
              </tr>
            </tbody>
          </table>
        </section>
    
    
    
    <informaltable rules="all" frame="box">
      <tr>
        <td>
          <para><emphasis role="italic">Amend DICOM PS3.4 as follows (changes to existing text are bold and <emphasis role="bold"><emphasis role="underline">underlined</emphasis></emphasis> for additions and <emphasis role="bold"><emphasis role="strikethrough">struckthrough</emphasis></emphasis> for removals):</emphasis></para>
        </td>
      </tr>
    </informaltable>
    
    <section label="F.7.2.1.1" status="5" xml:id="sect_F.7.2.1.1">
            <title>Modality Performed Procedure Step Subset Specification</title>
            <table frame="box" label="F.7.2-1" rules="all" xml:id="table_F.7.2-1">
              <caption>Modality Performed Procedure Step SOP Class N-CREATE, N-SET and Final State Attributes</caption>
              <thead>
                <tr valign="top">
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_e2bd9ea4-5712-4b7d-9f69-85ab8426b822">
                      <emphasis role="bold">Attribute Name</emphasis>
                    </para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_0c7b14f0-1cee-4b72-975c-d70b9b5afeec">
                      <emphasis role="bold">Tag</emphasis>
                    </para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_26479eae-896d-45bd-989f-859f8ce4df4b">
                      <emphasis role="bold">Req. Type N-CREATE (SCU/SCP)</emphasis>
                    </para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_c70f35bd-6a56-4861-b1a8-89fca7a3af64">
                      <emphasis role="bold">Req. Type N-SET (SCU/SCP)</emphasis>
                    </para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_6e2040b9-4191-4654-a70d-fdbd495a0efe">
                      <emphasis role="bold">Requirement Type Final State (see Note 1)</emphasis>
                    </para>
                  </th>
                </tr>
              </thead>
              <tbody>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_31260377-9a55-4915-8c33-013fbf2ca1b8">...</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">...</td>
                  <td align="center" colspan="1" rowspan="1">...</td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_f60ed9c4-edcd-48d5-8e2c-0321ae2690c2">...</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="5" rowspan="1">
                    <para xml:id="para_ceefb217-4953-4faa-af37-c04451861a3f">
                      <emphasis role="bold">Image Acquisition Results</emphasis>
                    </para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_0cae8ec4-4c1b-4a10-88cd-3d74e955801e">...</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_fb789c5c-87ae-410d-b85f-9b62125ea8ca">...</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_442f9575-158c-4f9a-934f-089deeb59c1b">...</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_4c821033-2f46-4e99-bfa6-e4fbf40b8430">...</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_0f522284-153b-41b3-99ca-0b0918cc8f52">
                <emphasis role="italic">All other Attributes of the <emphasis role="bold"><emphasis role="strikethrough"><olink targetdoc="PS3.3" targetptr="sect_C.4.16" xrefstyle="select: title"/> and </emphasis></emphasis><olink targetdoc="PS3.3" targetptr="sect_C.4.17" xrefstyle="select: title"/>
                </emphasis>
              </para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_44decf17-697c-42be-9132-73dc47d7a2bb">3/3</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_c1c53a40-dbcd-4955-9791-b98f7ad22975">3/3</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>
              </tbody>
            </table>
            <note>
              <orderedlist numeration="arabic">
                <listitem>
                  <para xml:id="para_4879c763-d0e7-421a-a8a4-e5d501d3fb64">...</para>
                </listitem>
                <listitem>
                  <para xml:id="para_0b0bec76-21aa-4b9d-8e58-ac893f763192">...</para>
                </listitem>
                <listitem>
                  <para xml:id="para_b37799e8-c862-474a-a35b-b51afa339459">Attributes (0040,1006) Placer Order Number/Procedure and (0040,1007) Filler Order Number/Procedure were previously defined in DICOM. They are now retired (see PS3.3-1998).</para>
                </listitem>
                <listitem>
                  <para xml:id="para_a8a94640-ac78-41cd-8350-4105884b812b">Attributes (0040,2006) and (0040,2007) were previously defined in DICOM. They are now retired (see PS3.3-1998).</para>
                </listitem>
                <listitem>
            <para xml:id="para_900335a8-3b3a-40f7-b1f6-829ff9774694">...</para>
          </listitem>
          <listitem>
            <para><emphasis role="bold"><emphasis role="underline">The Radiation Dose Module was previously defined in DICOM. This is now retired (see PS3.3-2017c).</emphasis></emphasis></para>
          </listitem>
              </orderedlist>
            </note>
          </section>
    
    <section label="F.8.2.1.1" status="5" xml:id="sect_F.8.2.1.1">
            <title>Modality Performed Procedure Step Retrieve IOD Subset Specifications</title>
            <table frame="box" label="F.8.2-1" rules="all" xml:id="table_F.8.2-1">
              <caption>Modality Performed Procedure Step Retrieve SOP Class N-GET Attributes</caption>
              <thead>
                <tr valign="top">
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_ff4f326c-4b5c-426f-bc86-17fb609296d1">
                      <emphasis role="bold">Attribute Name</emphasis>
                    </para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_9668ed3d-6442-4bf5-95b1-a693aa439bc9">
                      <emphasis role="bold">Tag</emphasis>
                    </para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_8a47739c-1a44-4a87-a0c4-f8828a4ba636">
                      <emphasis role="bold">Requirement Type (SCU/SCP)</emphasis>
                    </para>
                  </th>
                </tr>
              </thead>
              <tbody>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_08afcf1c-2bfd-4b59-bbd0-777caeba97c6">...</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_e046b25a-0192-4c2b-91f4-45460323f1d4">...</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">...</td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="3" rowspan="1">
                    <para xml:id="para_c7a7776b-9315-4efd-a576-54a91c41c2ee">
                      <emphasis role="bold">Image Acquisition Results</emphasis>
                    </para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_22fe7b39-1fed-4a43-9f6e-29ab7ff92def">...</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_f65b4a79-d9f6-4649-b8b1-8c848cccb26d">...</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_8b07bfdc-8c50-4bc8-9fa1-631dab3b800e">...</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_dff8bfd9-d4e3-419d-8046-c501999bbbd1">
                <emphasis role="italic">All other Attributes of the <emphasis role="bold"><emphasis role="strikethrough"><olink targetdoc="PS3.3" targetptr="sect_C.4.16" xrefstyle="select: title"/> and </emphasis></emphasis><olink targetdoc="PS3.3" targetptr="sect_C.4.17" xrefstyle="select: title"/>
                </emphasis>
              </para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_3062c894-f238-4ed3-93b0-0c4f65a32c13">3/3</para>
                  </td>
                </tr>
              </tbody>
            </table>
            <note>
              <orderedlist numeration="arabic">
                <listitem>
                  <para xml:id="para_25d44b0a-1db5-499b-8265-ca0d66934b19">Attributes (0040,1006) Placer Order Number/Procedure and (0040,1007) Filler Order Number/Procedure were previously defined in DICOM. They are now retired (see PS3.3-1998).</para>
                </listitem>
                <listitem>
            <para xml:id="para_1c1b4abd-5a51-400a-9920-f07e15155c4e">Attributes (0040,2006) and (0040,2007) were previously defined in DICOM. They are now retired (see PS3.3-1998).</para>
          </listitem>
          <listitem>
            <para><emphasis role="bold"><emphasis role="underline">The Radiation Dose Module was previously defined in DICOM. This is now retired (see PS3.3-2017c).</emphasis></emphasis></para>
          </listitem>
              </orderedlist>
            </note>
          </section>
    
    <informaltable rules="all" frame="box">
      <tr>
        <td>
          <para><emphasis role="italic">Amend DICOM PS3.6 as follows (changes to existing text are bold and <emphasis role="bold"><emphasis role="underline">underlined</emphasis></emphasis> for additions and <emphasis role="bold"><emphasis role="strikethrough">struckthrough</emphasis></emphasis> for removals):</emphasis></para>
        </td>
      </tr>
    </informaltable>
    
    <table frame="box" label="6-1" rules="all" xml:id="table_6-1">
      <caption>Registry of DICOM Data Elements</caption>
      <thead>
        <tr valign="top">
          <th align="center" colspan="1" rowspan="1">
            <para xml:id="para_3e8c52d5-3582-4ec4-a7e6-acd7604caf32">
              <emphasis role="bold">Tag</emphasis>
            </para>
          </th>
          <th align="center" colspan="1" rowspan="1">
            <para xml:id="para_ac298c3b-bcdd-462b-9240-c38e738bfb7c">
              <emphasis role="bold">Name</emphasis>
            </para>
          </th>
          <th align="center" colspan="1" rowspan="1">
            <para xml:id="para_99f6b0ff-e1b0-484e-83b5-c3003d61704c">
              <emphasis role="bold">Keyword</emphasis>
            </para>
          </th>
          <th align="center" colspan="1" rowspan="1">
            <para xml:id="para_42b3621d-8b30-46eb-8992-10bb21310ac7">
              <emphasis role="bold">VR</emphasis>
            </para>
          </th>
          <th align="center" colspan="1" rowspan="1">
            <para xml:id="para_551ea829-a949-4bfd-a201-bb102024e64c">
              <emphasis role="bold">VM</emphasis>
            </para>
          </th>
          <th align="center" colspan="1" rowspan="1">
            <para xml:id="para_3a18713f-2e81-4664-b9a1-e4e3d41957ab"/>
          </th>
        </tr>
      </thead>
      <tbody>
        
    
    <tr valign="top">
      <td align="center" colspan="1" rowspan="1">
        <para xml:id="para_a7783681-10af-4b13-aa8e-9af80e9e1c07"><emphasis role="italic">(0008,2229)</emphasis></para>
      </td>
      <td align="left" colspan="1" rowspan="1">
        <para xml:id="para_a4ab66e6-a61f-458a-a157-be99bc962d97"><emphasis role="italic">Anatomic Structure, Space or Region Sequence</emphasis></para>
      </td>
      <td align="left" colspan="1" rowspan="1">
        <para xml:id="para_a25a9d79-f768-45ea-ad04-cc4f8ac1af1c"><emphasis role="italic">Anatomic​Structure​Space​Or​Region​Sequence</emphasis></para>
      </td>
      <td align="center" colspan="1" rowspan="1">
        <para xml:id="para_87d25b03-de2d-4a90-a910-8dbdb1f11095"><emphasis role="italic">SQ</emphasis></para>
      </td>
      <td align="center" colspan="1" rowspan="1">
        <para xml:id="para_cd8cbbb0-427d-422c-bcd8-db316cee82e9"><emphasis role="italic">1</emphasis></para>
      </td>
      <td align="center" colspan="1" rowspan="1">
        <para xml:id="para_e82e5941-47ae-4ac8-ac5a-8859971c538e"><emphasis role="italic"><emphasis role="bold"><emphasis role="underline">RET</emphasis></emphasis></emphasis></para>
      </td>
    </tr>
        
        <tr valign="top">
          <td align="center" colspan="1" rowspan="1"><emphasis role="italic">
              <para xml:id="para_a5be9198-1b63-4749-a5fb-6ac8d475c537">(0040,0300)</para>
            </emphasis></td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_beeb938f-d7d3-4c79-8472-7a92f4e62c1d"><emphasis role="italic">Total Time of Fluoroscopy</emphasis></para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b231ec60-b1da-4376-8a82-05d573956071"><emphasis role="italic">Total​Time​Of​Fluoroscopy</emphasis></para>
          </td>
          <td align="center" colspan="1" rowspan="1">
            <para xml:id="para_de15ee30-a35f-4ba4-9e85-e5206a0b9c7f"><emphasis role="italic">US</emphasis></para>
          </td>
          <td align="center" colspan="1" rowspan="1">
            <para xml:id="para_ef7a867f-44fa-4f0d-a1a1-2d072ffc492d"><emphasis role="italic">1</emphasis></para>
          </td>
          <td align="center" colspan="1" rowspan="1">
            <para xml:id="para_d4e211b9-0171-4790-ac93-745ca5bb58c8"><emphasis role="italic"><emphasis role="bold"><emphasis role="underline">RET</emphasis></emphasis></emphasis></para>
          </td>
        </tr>
        <tr valign="top">
          <td align="center" colspan="1" rowspan="1"><emphasis role="italic">
              <para xml:id="para_85d61086-fd9f-4f1a-9429-944044246ae6">(0040,0301)</para>
            </emphasis></td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3b5d0c5e-f5be-44c8-93dd-00411009c34e"><emphasis role="italic">Total Number of Exposures</emphasis></para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e4fc99ca-e737-484b-aed7-9ac36048560c"><emphasis role="italic">Total​Number​Of​Exposures</emphasis></para>
          </td>
          <td align="center" colspan="1" rowspan="1">
            <para xml:id="para_bf391bf4-91b1-4253-b1bf-b0e0dbffd408"><emphasis role="italic">US</emphasis></para>
          </td>
          <td align="center" colspan="1" rowspan="1">
            <para xml:id="para_5fc1c19b-5f15-4d34-8a48-7d864c6c6924"><emphasis role="italic">1</emphasis></para>
          </td>
          <td align="center" colspan="1" rowspan="1">
            <para xml:id="para_58446c5a-2a1d-47a1-95bf-84fd8f46195a"><emphasis role="italic"><emphasis role="bold"><emphasis role="underline">RET</emphasis></emphasis></emphasis></para>
          </td>
        </tr>
        <tr valign="top">
          <td align="center" colspan="1" rowspan="1"><emphasis role="italic">
              <para xml:id="para_b0a497aa-2a9f-4a0c-9cf5-40d68e8cf9ce">(0040,030E)</para>
            </emphasis></td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_1e64018a-8266-4a2c-a15f-f154683f6cd6"><emphasis role="italic">Exposure Dose Sequence</emphasis></para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_253aea41-8a87-479f-8d56-6a06f565e778"><emphasis role="italic">Exposure​Dose​Sequence</emphasis></para>
          </td>
          <td align="center" colspan="1" rowspan="1">
            <para xml:id="para_cbda2dac-026f-440f-86f0-94c800466458"><emphasis role="italic">SQ</emphasis></para>
          </td>
          <td align="center" colspan="1" rowspan="1">
            <para xml:id="para_1958364a-b405-42ea-aab2-accaa8f84e2b"><emphasis role="italic">1</emphasis></para>
          </td>
          <td align="center" colspan="1" rowspan="1">
            <para xml:id="para_01ea19dd-815d-4070-89dc-7558511f0885"><emphasis role="italic"><emphasis role="bold"><emphasis role="underline">RET</emphasis></emphasis></emphasis></para>
          </td>
        </tr>
        
    
      </tbody>
    </table>
    
  </chapter>
  