<book xmlns="http://docbook.org/ns/docbook" xmlns:xl="http://www.w3.org/1999/xlink" version="5.0">
  <title>Digital Imaging and Communications in Medicine (DICOM)</title>
  <subtitle>Sup 172 - Parametric Map Storage</subtitle>
  <info>
    <author>
      <orgname>DICOM Standards Committee - Working Group 18 - Research and Clinical Trials</orgname>
      <address>
        <street>1300 N. 17th Street Suite 1752</street>
        <city>Rosslyn</city>
        <state>VA</state>
        <postcode>22209</postcode>
        <country>USA</country>
      </address>
    </author>
    <copyright>
      <year>2014</year>
      <holder>NEMA</holder>
    </copyright>
    <releaseinfo>Status: Final Text</releaseinfo>
    <pubdate>2014/11/11</pubdate>
    <legalnotice>
      <para>This supplement was prepared pursuant to work item: 2014-04-B</para>
    </legalnotice>
  </info>
    
  <chapter>
    <title>Scope and Field of Application</title>
    <para>This Supplement defines use-cases, image IODs and SOP Classes for storage of information
      related to quantitative parametric maps.</para>
    <para>Parametric maps include quantitative derived images computed from acquired images, whether
      they be used for clinical, clinical trial or basic research purposes. DICOM objects are needed
      so that they can be stored in the PACS and wherever possible visualized in conjunction with
      images for review, and preserve patient (or research subject) identification, management, and
      acquisition and processing technique information.</para>
    <para>Parametric maps are encoded as images just like any other. The distinguishing
      characteristic of a "parametric map" is that each pixel value has a "meaning" in the real
      world, and the name "parametric" alludes to the meaning being the parameter of something, such
      as a pharmacokinetic model. In general, the term "parametric map" is loosely used to refer
      to any derived image whose pixel values have a quantitative meaning, as opposed to acquired
      images, whose pixel values may or may not have meaning (e.g., CT HU versus arbitrary MR
      signal intensity), or derived images whose pixel values are dimensionless.</para>
    <para>Examples of parametric maps include:</para>
    <itemizedlist>
      <listitem>
        <para>parameters of pharmacokinetic models, such as Ktrans of the Tofts model used in
          DCE-MRI</para>
      </listitem>
      <listitem>
        <para>regional cerebral blood volume (RCBV) or flow (RCBF) used in perfusion CT</para>
      </listitem>
      <listitem>
        <para>normalized values based on body size, such as SUVbw used in PET</para>
      </listitem>
      <listitem>
        <para>diffusion parameters such as Apparent Diffusion Coefficient (ADC) used in MR
          DWI</para>
      </listitem>
    </itemizedlist>
    <para>Some of these applications, such as SUV and ADC, have been in widespread use already, and
      use ordinary modality-specific derived image IODs for encoding, with quite satisfactory
      (interoperable) results in most cases, and it is not the purpose of this Supplement to replace
      that common usage.</para>
    <para>Rather, the goals are to define:</para>
    <itemizedlist>
      <listitem>
        <para>additional attributes, modules, acquisition context, code sets and scaling functions
          that may optionally be used to extend existing image IODs</para>
      </listitem>
      <listitem>
        <para>a new IOD to address the need for encoding some parametric maps with floating point
          "pixels"</para>
      </listitem>
      <listitem>
        <para>the basic encoding mechanism changes needed to support floating point "pixels"</para>
      </listitem>
    </itemizedlist>
    <para>Floating point "pixels" are necessary in some cases for subsequent calculations and to
      prevent the accumulation of rounding and quantization errors, as well as to provide a full
      fidelity record of the work performed. In many cases, scientific processing applications have
      been reluctant to sacrifice intermediate precision between processing steps, and have avoided
      the use of the DICOM Standard as a consequence.</para>
    <para>The IOD defined in this supplement uses the enhanced multi-frame encoding.</para>
  </chapter>
    
  <chapter label="DICOM PS3.2" xml:id="chapter_PS3.2">
    <title>Conformance</title>
    <section label="A" xml:id="chapter_PS3.2_A">
      <title>DICOM Conformance Statement Template (Normative)</title>
      <informaltable rules="all" frame="box">
        <tr>
          <td>
            <para><emphasis role="italic">Amend DICOM PS 3.2 - Conformance - Annex A - DICOM Conformance Statement Template (Normative) as follows:</emphasis></para>
          </td>
        </tr>
      </informaltable>
      
      <table frame="box" label="A.1-2" rules="all" xml:id="table_PS3.2_A.1-2">
        <caption>UID Values</caption>
        <thead>
          <tr valign="top">
            <th align="center" colspan="1" rowspan="1">
              <para>UID Value</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para>UID Name</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para>Category</para>
            </th>
          </tr>
        </thead>
        <tbody>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>...</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>...</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>...</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline"
                    >1.2.840.10008.5.1.4.1.1.30</emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">Parametric Map Storage SOP Class</emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">Transfer</emphasis></emphasis></para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>...</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>...</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>...</para>
            </td>
          </tr>
        </tbody>
      </table>
      
    </section>
  </chapter>
  
  <chapter label="DICOM PS3.3" xml:id="chapter_PS3.3">
    <title>Information Object Definitions</title>
    <section label="A" xml:id="chapter_PS3.3_A">
      <title>Composite Information Object Definitions (Normative)</title>
      <informaltable rules="all" frame="box">
        <tr>
          <td>
            <para><emphasis role="italic">Amend DICOM PS 3.3 - Information Object Definitions - Annex A - Composite Information Object Definitions (Normative), to add Parametric Map IOD Modules to A.1.4 Overview of the Composite IOD Module Content.</emphasis></para>
          </td>
        </tr>
      </informaltable>
      
      <informaltable rules="all" frame="box">
        <tr>
          <td>
            <para><emphasis role="italic">Amend DICOM PS 3.3 - Information Object Definitions - Annex A - Composite Information Object Definitions (Normative) as follows, to add a new Parametric Map IOD:</emphasis></para>
          </td>
        </tr>
      </informaltable>
      
      <section label="A.75" status="2" xml:id="sect_PS3.3_A.75">
        <title>Parametric Map IOD</title>
        <section label="A.75.1" status="3" xml:id="sect_PS3.3_A.75.1">
          <title>Parametric Map IOD Description</title>
          <para>The Parametric Map Information Object Definition (IOD) specifies a multi-frame image
            representing pixels with real world values. Parametric Maps are either integer or
            floating point.</para>
          <para>The Parametric Map IOD does not include the full set of acquisition parameters of any
            acquired images from which they were derived, e.g., cardiac phase. An application
            rendering or processing the Parametric Map may need to access the source images for such
            information.</para>
          <para>The Parametric Map IOD requires the presence of VOI LUT (window) information with
            the intent that at a minimum the image be renderable without special processing. The
            output space is defined as P-Values to achieve consistency.</para>
          <note>
            <itemizedlist>
              <listitem>
                <para>The VOI LUT mechanism specifically supports floating point values, and there is no
                  expectation that it be limited to integer input or output ranges.</para>
              </listitem>
              <listitem>
                <para>Even though the output of the VOI LUT is not constrained to be integer values,
                  implicit scaling of the output range to the input range of an integer-based Palette Color LUT
                  can be used to apply pseudo-color for display. No pseudo-color
                  mapping information is encoded with the image. This may be applied at the discretion of the
                  receiving application whether it uses a separate DICOM Color Palette IOD or some other
                  mechanism.</para>
              </listitem>
            </itemizedlist>
          </note>
          <para>The Parametric Map IOD encodes one or more parameters as an image. Other Image IODs
            may be used to encode related information, and instances of them may be referenced from
            the Parametric Map, as the source from which a parameter was derived, or some other
            relationship.</para>
          <note>
            <itemizedlist>
              <listitem>
                <para>The Blending Presentation State IOD may be used to describe how (selected
                  frames of) a Parametric Map instance may be superimposed on, say, frames of
                  acquired images for anatomical reference encoded as instances of other Image IODs,
                  as well as the relative opacity and pseudo-color applicable to the overlying
                  frames.</para>
              </listitem>
              <listitem>
                <para>Commonality of the same Frame of Reference UID also allows an application to
                  relate Parametric Map instances and other Image instances, in the absence of
                  explicit references.</para>
              </listitem>
              <listitem>
                <para>The Parametric Map IOD is not restricted to encoding only a single parameter
                  in one instance. Since it is a multi-frame object, and since the type of parameter
                  is encoded in the Real World Value Mapping functional group macro, which may vary
                  on a per-frame basis, the parameter may also vary from frame-to-frame. In such
                  cases, the Multi-Frame Dimension Module may be used to highlight this, by
                  specifying the Quantity Definition Sequence (0040,9220) as a Dimension
                  Index.</para>
              </listitem>
            </itemizedlist>
          </note>
        </section>
        <section label="A.75.2" status="3" xml:id="sect_PS3.3_A.75.2">
          <title>Parametric Map IOD Entity-Relationship Model</title>
          <para>The E-R Model in Section A.1.2 depicts those components of the DICOM Information Model that directly reference the Parametric Map IOD. The Parametric Map is a kind of Image.</para>
        </section>
        <section label="A.75.3" status="3" xml:id="sect_PS3.3_A.75.3">
          <title>Parametric Map IOD Module Table</title>
          <table frame="box" label="A.75-1" rules="all" xml:id="table_PS3.3_A.75-1">
            <caption>Parametric Map IOD Modules</caption>
            <thead>
              <tr valign="top">
                <th align="center" colspan="1" rowspan="1">
                  <para>IE</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para>Module</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para>Reference</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para>Usage</para>
                </th>
              </tr>
            </thead>
            <tbody>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="2">
                  <para>Patient</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>Patient</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>
                    C.7.1.1
                  </para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>M</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para>Clinical Trial Subject</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>
                    C.7.1.3
                  </para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>U</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="3">
                  <para>Study</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>General Study</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>
                    C.7.2.1
                  </para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>M</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para>Patient Study</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>
                    C.7.2.2
                  </para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>U</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para>Clinical Trial Study</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>
                    C.7.2.3
                  </para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>U</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="3">
                  <para>Series</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>General Series</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>
                    C.7.3.1
                  </para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>M</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para>Parametric Map Series</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>
                    C.8.32.1
                  </para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>M</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para>Clinical Trial Series</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>
                    C.7.3.2
                  </para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>U</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para>Frame of Reference</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>Frame of Reference</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>
                    C.7.4.1
                  </para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>M</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="2">
                  <para>Equipment</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>General Equipment</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>
                    C.7.5.1
                  </para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>M</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para>Enhanced General Equipment</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>
                    C.7.5.2
                  </para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>M</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="16">
                  <para>Image</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>General Image</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>
                    C.7.6.1
                  </para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>M</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para>Image Pixel</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para> C.7.6.3</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>C - Required if integer pixels</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para>Floating Point Image Pixel</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>C.7.6.24</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>C - Required if 32 bit floating point pixels</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para>Double Floating Point Image Pixel</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>C.7.6.25</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>C - Required if 64 bit floating point pixels</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para>Parametric Map Image</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>
                    <xref linkend="sect_PS3.3_C.8.32.2" xrefstyle="select: labelnumber"/>
                  </para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>M</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para>Multi-frame Functional Groups</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>
                    C.7.6.16
                  </para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>M</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para>Multi-frame Dimension</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>
                    C.7.6.17
                  </para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>M</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para>Cardiac Synchronization</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>
                    C.7.6.18.1
                  </para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>U</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para>Respiratory Synchronization</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>
                    C.7.6.18.2
                  </para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>U</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para>Bulk Motion Synchronization</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>
                    C.7.6.18.3
                  </para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>U</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para>Acquisition Context</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>
                    C.7.6.14
                  </para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>M</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para>Device</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>
                    C.7.6.12
                  </para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>U</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para>Specimen</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>
                    C.7.6.22
                  </para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>U</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para>Common Instance Reference</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>
                    C.12.2
                  </para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>C - Required if Referenced Image Functional Group (C.7.6.16.2.5) or
                    Derivation Image Functional Group (C.7.16.2.6) is present.</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para>SOP Common</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>
                    C.12.1
                  </para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>M</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para>Frame Extraction</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>
                    C.12.3
                  </para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>C - Required if the SOP Instance was created in response to a Frame-Level retrieve request</para>
                </td>
              </tr>
            </tbody>
          </table>
        </section>
        <section label="A.75.4" status="3" xml:id="sect_PS3.3_A.75.4">
          <title>Parametric Map IOD Content Constraints</title>
          <para>Either the Pixel Data Module or the Floating Point Pixel Data Module or the Double
            Floating Point Image Pixel Module is required, and only one of these shall be
            present.</para>
          <note><para>I.e., it is not permitted to have more than one of Pixel Data Provider URL (0028,7FE0), Pixel
              Data (7FE0,0010), Float Pixel Data (7FE0,0008) or Double Float Pixel Data (7FE0,0009)
              in a Standard Extended SOP Class in the top level Data Set. See PS3.5</para></note>
          <para>The VOI LUT Module shall not be present.</para>
          <para>The Modality LUT Module shall not be present.</para>
          <para>The Overlay Module shall not be present.</para>
          <para>The Supplemental Palette Color LUT Module shall not be present.</para>
        </section>
        <section label="A.75.5" status="3" xml:id="sect_PS3.3_A.75.5">
          <title>Parametric Map Functional Groups</title>
          <para><xref linkend="table_PS3.3_A.75-2" xrefstyle="select: label"/> specifies the use of the Functional Group Macros used in the Multi-frame Functional Group Module for the Parametric Map IOD.</para>
          <table frame="box" label="A.75-2" rules="all" xml:id="table_PS3.3_A.75-2">
            <caption>Parametric Map Functional Group Macros</caption>
            <thead>
              <tr valign="top">
                <th align="center" colspan="1" rowspan="1">
                  <para>Functional Group Macro</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para>Section</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para>Usage</para>
                </th>
              </tr>
            </thead>
            <tbody>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para>Pixel Measures</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>
                    C.7.6.16.2.1
                  </para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>M</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para>Plane Position (Patient)</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>
                    C.7.6.16.2.3
                  </para>
                </td>
                <td align="left" colspan="1" rowspan="1">M</td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para>Plane Orientation (Patient)</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>
                    C.7.6.16.2.4
                  </para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>M</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para>Referenced Image</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>
                    C.7.6.16.2.5
                  </para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>U</para></td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para>Derivation Image</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>
                    C.7.6.16.2.6
                  </para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>C - Required if the image or frame has been derived from another SOP Instance.</para></td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para>Frame Content</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>
                    C.7.6.16.2.2
                  </para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>M - Shall not be used as a Shared Functional Group</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para>Cardiac Synchronization</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>
                    C.7.6.16.2.7
                  </para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>U</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para>Frame Anatomy</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>
                    C.7.6.16.2.8
                  </para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>M</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para>Identity Pixel Value Transformation</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>
                    C.7.6.16.2.9b
                  </para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>M</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para>Frame VOI LUT With LUT Macro</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>
                    C.7.6.16.2.10b
                  </para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>M</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para>Real World Value Mapping</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>
                    C.7.6.16.2.11
                  </para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>M</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para>Contrast/Bolus Usage</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>
                    C.7.6.16.2.12
                  </para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>U</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para>Respiratory Synchronization</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>
                    C.7.6.16.2.17
                  </para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>U</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para>Parametric Map Frame Type</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>C.8.32.3.1</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>M</para>
                </td>
              </tr>
            </tbody>
          </table>
          <section label="A.75.5.1" status="4" xml:id="sect_PS3.3_A.75.5.1">
            <title>Parametric Map Functional Groups Description</title>
            <para>For the Derivation Image Functional Group Macro, the Baseline CID for:</para>
            <itemizedlist>
              <listitem><para>Purpose of Reference Sequence (0040,A170) is <xref linkend="sect_CID_7222" xrefstyle="select: labelnumber quotedtitle"/>.</para></listitem>
              <listitem><para>Derivation Code Sequence (0008,9215) is CID 7203 "Image Derivation".</para></listitem>
            </itemizedlist>
            <para>For the Real World Value Mapping Functional Group Macro, which defines the type of quantity, the method of generation and the units for the pixel values, the Baseline CID for:</para>
            <itemizedlist>
              <listitem>
                <para>Concept Name Code Sequence of Quantity Definition Sequence (0040,9220) is CID
                  9000 "Physical Quantity Descriptors" (CP 1387).</para>
              </listitem>
              <listitem>
                <para>Concept Code Sequence for Concept Name of (G-C1C6, SRT, "Quantity") of
                  Quantity Definition Sequence (0040,9220) is CID 7180 "Abstract Multi-dimensional
                  Image Model Component Semantics" (CP 1387).</para>
              </listitem>
              <listitem>
                <para>Measurement Units Code Sequence (0040,08EA) is CID 7181 "Abstract Multi-dimensional Image Model Component Units".</para>
              </listitem>
            </itemizedlist>
          </section>
        </section>
      </section>
      
      
    </section>
    
    <section label="C" xml:id="chapter_PS3.3_C">
      <title>Information Module Definitions (Normative)</title>
      
      <section>
        <title/>
        <informaltable rules="all" frame="box">
          <tr>
            <td>
              <para><emphasis role="italic">Amend DICOM PS 3.3 - Information Object Definitions - to
                add note to Pixel Padding Value in General Equipment Module:</emphasis></para>
            </td>
          </tr>
        </informaltable>
      </section>
      
        <section label="C.7.5.1" status="4" xml:id="sect_C.7.5.1">
          <title>General Equipment Module</title>
          <para>...</para>
          <table frame="box" label="C.7-8" rules="all" xml:id="table_C.7-8">
            <caption>General Equipment Module Attributes</caption>
            <thead>
              <tr valign="top">
                <th align="center" colspan="1" rowspan="1">
                  <para>Attribute Name</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para>Tag</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para>Type</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para>Attribute Description</para>
                </th>
              </tr>
            </thead>
            <tbody>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para>...</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para>...</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para>...</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>...</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para>Pixel Padding Value</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para>(0028,0120)</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para>1C</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>Single pixel value or one limit (inclusive) of a range of pixel values used in an image to pad to rectangular format or to signal background that may be suppressed. See <xref linkend="sect_C.7.5.1.1.2" xrefstyle="select: label"/> for further explanation.</para>
                  <para>Required if Pixel Padding Range Limit (0028,0121) is present and either
                  Pixel Data (7FE0,0010) or Pixel Data Provider URL (0028,7FE0) is present. May be
                  present otherwise if Pixel Data (7FE0,0010) or Pixel Data Provider URL (0028,7FE0)
                  is present.</para>
                  <note>
                    <orderedlist numeration="arabic">
                      <listitem>
                        <para>The Value Representation of this Attribute is determined by the value of Pixel Representation (0028,0103).</para>
                      </listitem>
                      <listitem>
                        <para>This Attribute is not used in Presentation State Instances; there is no means in a Presentation State to "override" any Pixel Padding Value specified in the referenced images.</para>
                      </listitem>
                      <listitem>
                        <para>This Attribute does apply to RT Dose and Segmentation instances, since they include Pixel Data.</para>
                      </listitem>
                      <listitem>
                        <para><emphasis role="bold"><emphasis role="underline">This Attribute does
                            not apply when Float Pixel Data (7FE0,0008) or Double Float Pixel Data
                            ((7FE0,0009) are used instead of Pixel Data (7FE0,0010); Float Pixel
                            Padding Value (0028,0122) or Double Float Pixel Padding Value
                            (0028,0123), respectively, are used instead, and defined at the Image,
                            not the Equipment, level.</emphasis></emphasis></para>
                      </listitem>
                    </orderedlist>
                  </note>
                </td>
              </tr>
            </tbody>
          </table>
          <section label="C.7.5.1.1" status="5" xml:id="sect_C.7.5.1.1">
            <title>General Equipment Attribute Descriptions</title>
          
              <para>...</para>
            
            <section label="C.7.5.1.1.2" status="6" xml:id="sect_C.7.5.1.1.2">
              <title>Pixel Padding Value and Pixel Padding Range Limit</title>
              <para>Pixel Padding Value (0028,0120) is used to pad grayscale images (those with a Photometric Interpretation of MONOCHROME1 or MONOCHROME2) to rectangular format. The native format of some images is not rectangular. It is common for devices with this format to pad the images to the rectangular format required by the DICOM Standard with a specific pixel value that is not contained in the native image. Further, when resampling, such as after spatial registration, padding may need to be used to fill previously non-existent pixels.</para>
              <para>Pixel Padding Value (0028,0120) and Pixel Padding Range Limit (0028,0121) are also used to identify pixels to be excluded from the normal grayscale rendering pipeline for other reasons, such as suppression of background air. Pixel Padding Range Limit (0028,0121) is defined in the Image Pixel Module.</para>
              <note>
                <orderedlist numeration="arabic">
                  <listitem>
                    <para>The "native image" is that which is being padded to the required rectangular format, e.g., the area within the circular reconstruction perimeter of a CT image, or the subset of the rectangular area that contains useful image information.</para>
                  </listitem>
                  <listitem>
                    <para>The pixel padding value is explicitly described in order to prevent display applications from taking it into account when determining the dynamic range of an image, since the Pixel Padding Value will be outside the range between the minimum and maximum values of the pixels in the native image</para>
                  </listitem>
                  <listitem>
                    <para>No pixels in the native image will have a value equal to Pixel Padding Value.</para>
                  </listitem>
                </orderedlist>
              </note>
              <para>Pixel Padding Value (0028,0120) specifies either a single value of this padding value, or when combined with Pixel Padding Range Limit (0028,0121), a range of values (inclusive) that are padding.</para>
              <para>The values of Pixel Padding Value (0028,0120) and Pixel Padding Range Limit (0028,0121) shall be valid values within the constraints defined by Bits Allocated (0028,0100), Bits Stored (0028,0101), and High Bit (0028,0102).</para>
              <para>Pixel Padding Value (0028,0120) and Pixel Padding Range Limit (0028,0121) shall not be present when padding is performed but the pixel value used for padding does occur in the native image.</para>
              <para>If Photometric Interpretation (0028,0004) is MONOCHROME2, Pixel Padding Value (0028,0120) shall be less than (closer to or equal to the minimum possible pixel value) or equal to Pixel Padding Range Limit (0028,0121). If Photometric Interpretation (0028,0004) is MONOCHROME1, Pixel Padding Value (0028,0120) shall be greater than (closer to or equal to the maximum possible pixel value) or equal to Pixel Padding Range Limit (0028,0121).</para>
              <note>
                <orderedlist numeration="arabic">
                  <listitem>
                    <para>When the relationship between pixel value and X-Ray Intensity is unknown, it is recommended that the following values be used to pad with black when the image is unsigned:</para>
                    <itemizedlist>
                      <listitem>
                        <para>0 if Photometric Interpretation (0028,0004) is MONOCHROME2.</para>
                      </listitem>
                      <listitem>
                        <para>2<superscript>Bits Stored</superscript> - 1 if Photometric Interpretation (0028,0004) is MONOCHROME1.</para>
                      </listitem>
                    </itemizedlist>
                    <para>and when the image is signed:</para>
                    <itemizedlist>
                      <listitem>
                        <para>-2<superscript>Bits Stored-1</superscript> if Photometric Interpretation (0028,0004) is MONOCHROME2.</para>
                      </listitem>
                      <listitem>
                        <para>2<superscript>Bits Stored-1</superscript> - 1 if Photometric Interpretation (0028,0004) is MONOCHROME1.</para>
                      </listitem>
                    </itemizedlist>
                  </listitem>
                  <listitem>
                    <para>For projection radiography, when the relationship between pixel value and X-Ray Intensity is known (for example as defined by Pixel Intensity Relationship (0028,1040) and Pixel Intensity relationship Sign (0028,1041)), it is recommended that a pixel value equivalent to, or rendered similarly to, air (least X-Ray absorbance) be used for padding. However, if such a value may occur in the native image, the Pixel Padding Value (0028,0120) Attribute itself should not be sent.</para>
                    <para>E.g., for an XRF image obtained with an image intensifier, if air is black then a padded perimeter, if any, should also appear black. Typically though, if unpadded, this area would be collimated with a circular collimator, in which case the pixels would appear natively as white (greatest X-Ray absorbance) and a circular shutter would be necessary to neutralize them as black. Whether collimated areas are detected and treated as padded, or neutralized with shutters is at the discretion of the application. See also the Display Shutter Module Section C.7.6.11.</para>
                  </listitem>
                  <listitem>
                    <para>The conditional requirement for the Pixel Padding Value Range Limit (0028,0121) in the Image Pixel Module means that it shall not be present unless Pixel Padding Value (0028,0120) is also present.</para>
                  </listitem>
                  <listitem>
                    <para>The range of values to be suppressed between Pixel Padding Value (0028,0120) and Pixel Padding Value Range Limit (0028,0121) is specified as being inclusive, that is the values themselves as well as all values between are suppressed.</para>
                  </listitem>
                  <listitem>
                    <para>When Pixel Padding Value Range Limit (0028,0121) is present, but not supported by a rendering application, the constraint that Pixel Padding Value (0028,0120) is closest to the "blackest" value, which is typically the most frequently occurring background pixel, will most often result in an acceptable display, permitting "backward compatibility" in the majority of cases.</para>
                  </listitem>
                </orderedlist>
              </note>
              <para>When modifying equipment changes the pixel padding value in the image, it shall change the values of Pixel Padding Value (0028,0120) and Pixel Padding Range Limit (0028,0121), if present. If modifying equipment changes the pixel padding values in the image to values present in the native image, the attribute Pixel Padding Value (0028,0120) and Pixel Padding Range Limit (0028,0121) shall be removed.</para>
              <note>
                <orderedlist numeration="arabic">
                  <listitem>
                    <para>For example, if a CT image containing signed values from -1024 to 3191 and a Pixel Padding Value of -2000 and a Rescale Intercept of 0 is converted to an unsigned image with a Rescale Intercept of -1024 by adding 1024 to all pixels and clipping all more negative pixels to 0, then the padding pixels will be indistinguishable from some of the modified native image pixels, and hence Pixel Padding Value (0028,0120) needs to be removed.</para>
                  </listitem>
                  <listitem>
                    <para>If the modification involves lossy compression, which may result in changes to the pixel values, then the application of Pixel Padding Value and Pixel Padding Range Limit may result in a different appearance, and hence these attributes may need different values also.</para>
                  </listitem>
                </orderedlist>
              </note>
            </section>
            
          </section>
        </section>

      
      <section>
        <title/>
        <informaltable rules="all" frame="box">
          <tr>
            <td>
              <para><emphasis role="italic">Amend DICOM PS 3.3 - Information Object Definitions - to
                  add new Sections C.7.6.24  and C.7.6.25, with support for Float and Double Float
                  Pixel Data:</emphasis></para>
            </td>
          </tr>
        </informaltable>
      </section>
      
      <section label="C.7.6.24" status="4" xml:id="sect_PS3.3_C.7.6.24">
        <title>Floating Point Image Pixel Module</title>
        <para><xref linkend="table_PS3.3_C.7.6.24-1" xrefstyle="select: label"/> describes the Floating
          Point Image Pixel Module. This module differs from the Section C.7.6.3 Image Pixel Module in that:</para>
        <itemizedlist>
          <listitem>
            <para>instead of integer stored pixel values, float stored pixel values are used</para>
          </listitem>
          <listitem>
            <para>Bits Stored (0028,0101) and High Bit (0029,0102) are not used because the stored
              pixel values always occupy the entire word</para>
          </listitem>
          <listitem>
            <para>Pixel Representation (0028,0103) is not used because the stored pixel values are
              always signed</para>
          </listitem>
          <listitem>
            <para>Photometric Interpretation is constrained</para>
          </listitem>
          <listitem>
            <para>color palette tables are not used</para>
          </listitem>
          <listitem>
            <para>Pixel Data Provider URL (0028,7FE0) is not used</para>
          </listitem>
        </itemizedlist>
        <table frame="box" label="C.7.6.24-1" rules="all" xml:id="table_PS3.3_C.7.6.24-1">
          <caption>Floating Point Image Pixel Module Attributes</caption>
          <thead>
            <tr valign="top">
              <th align="center" colspan="1" rowspan="1">
                <para>Attribute Name</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>Tag</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>Type</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>Attribute Description</para>
              </th>
            </tr>
          </thead>
          <tbody>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>Samples per Pixel</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>(0028,0002)</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>Number of samples (planes) in this image. See Section C.7.6.3.1.1 for further
                  explanation.</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>Photometric Interpretation</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>(0028,0004)</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>Specifies the intended interpretation of the pixel data.</para>
                <variablelist spacing="compact">
                  <title>Enumerated Values:</title>
                  <varlistentry>
                    <term>MONOCHROME2</term>
                    <listitem>
                      <para/>
                    </listitem>
                  </varlistentry>
                </variablelist>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>Rows</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>(0028,0010)</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>Number of rows in the image.</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">Columns</td>
              <td align="center" colspan="1" rowspan="1">
                <para>(0028,0011)</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>Number of columns in the image.</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>Bits Allocated</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>(0028,0100)</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>Number of bits allocated for each pixel sample. Each sample shall have the
                  same number of bits allocated. See PS3.5 for further explanation.</para>
                <variablelist spacing="compact">
                  <title>Enumerated Values:</title>
                  <varlistentry><term>32</term><listitem><para/></listitem></varlistentry>
                </variablelist>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>Float Pixel Data</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>(7FE0,0008)</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>A data stream of the pixel samples that comprise the Image. The order of
                  pixels sent for each image plane is left to right, top to bottom, i.e., the upper
                  left pixel (labeled 1,1) is sent first followed by the remainder of row 1,
                  followed by the first pixel of row 2 (labeled 2,1) then the remainder of row 2 and
                  so on.</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>Pixel Aspect Ratio</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>(0028,0034)</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1C</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>Ratio of the vertical size and horizontal size of the pixels in the image
                  specified by a pair of integer values where the first value is the vertical pixel
                  size, and the second value is the horizontal pixel size. Required if the aspect
                  ratio values do not have a ratio of 1:1 and the physical pixel spacing is not
                  specified by Pixel Spacing (0028,0030), or Imager Pixel Spacing (0018,1164) or
                  Nominal Scanned Pixel Spacing (0018,2010), either for the entire Image or
                  per-frame in a Functional Group Macro. See Section C.7.6.3.1.7.</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>Float Pixel Padding Value</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>(0028,0122)</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>3</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>One limit (inclusive) of a range of pixel values used in an image to pad to
                  rectangular format or to signal background that may be suppressed.</para>
               </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>Float Pixel Padding Range Limit</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>(0028,0124)</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1C</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>Pixel value that represents one limit (inclusive) of a range of padding values
                  used together with Float Pixel Padding Value (0028,0122).</para>
                <para>Required if Float Pixel Padding Value (0028,0122) is present.</para>
                <note>
                  <orderedlist numeration="arabic">
                    <listitem>
                      <para>If only a single padding value rather than a range is required, then
                        both Float Pixel Padding Value (0028,0122) and Float Pixel Padding Range
                        Limit (0028,0124) will contain the same value.</para>
                    </listitem>
                    <listitem>
                      <para>The general considerations described in <xref linkend="sect_C.7.5.1.1.2"
                          xrefstyle="select: label"/> may be helpful in understanding the
                        corresponding floating point attributes, but are not normative.</para>
                    </listitem>
                  </orderedlist>
                </note>
              </td>
            </tr>
          </tbody>
        </table>
      </section>

      <section label="C.7.6.25" status="4" xml:id="sect_PS3.3_C.7.6.25">
        <title>Double Floating Point Image Pixel Module</title>
        <para><xref linkend="table_PS3.3_C.7.6.25-1" xrefstyle="select: label"/> describes the Floating
          Point Image Pixel Module. This module differs from the Section C.7.6.3 Image Pixel Module in that:</para>
        <itemizedlist>
          <listitem>
            <para>instead of integer stored pixel values, double float stored pixel values are
              used</para>
          </listitem>
          <listitem>
            <para>Bits Stored (0028,0101) and High Bit (0029,0102) are not used because the stored
              pixel values always occupy the entire word</para>
          </listitem>
          <listitem>
            <para>Pixel Representation (0028,0103) is not used because the stored pixel values are
              always signed</para>
          </listitem>
          <listitem>
            <para>Photometric Interpretation is constrained</para>
          </listitem>
          <listitem>
            <para>color palette tables are not used</para>
          </listitem>
          <listitem>
            <para>Pixel Data Provider URL (0028,7FE0) is not used</para>
          </listitem>
        </itemizedlist>
        <table frame="box" label="C.7.6.25-1" rules="all" xml:id="table_PS3.3_C.7.6.25-1">
          <caption>Double Floating Point Image Pixel Module Attributes</caption>
          <thead>
            <tr valign="top">
              <th align="center" colspan="1" rowspan="1">
                <para>Attribute Name</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>Tag</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>Type</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>Attribute Description</para>
              </th>
            </tr>
          </thead>
          <tbody>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>Samples per Pixel</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>(0028,0002)</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>Number of samples (planes) in this image. See Section C.7.6.3.1.1 for further
                  explanation.</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>Photometric Interpretation</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>(0028,0004)</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>Specifies the intended interpretation of the pixel data.</para>
                <variablelist spacing="compact">
                  <title>Enumerated Values:</title>
                  <varlistentry>
                    <term>MONOCHROME2</term>
                    <listitem>
                      <para/>
                    </listitem>
                  </varlistentry>
                </variablelist>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>Rows</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>(0028,0010)</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>Number of rows in the image.</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">Columns</td>
              <td align="center" colspan="1" rowspan="1">
                <para>(0028,0011)</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>Number of columns in the image.</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>Bits Allocated</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>(0028,0100)</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>Number of bits allocated for each pixel sample. Each sample shall have the
                  same number of bits allocated. See PS3.5 for further explanation.</para>
                <variablelist spacing="compact">
                  <title>Enumerated Values:</title>
                  <varlistentry><term>64</term><listitem><para/></listitem></varlistentry>
                </variablelist>
                
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>Double Float Pixel Data</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>(7FE0,0009)</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>A data stream of the pixel samples that comprise the Image. The order of
                  pixels sent for each image plane is left to right, top to bottom, i.e., the upper
                  left pixel (labeled 1,1) is sent first followed by the remainder of row 1,
                  followed by the first pixel of row 2 (labeled 2,1) then the remainder of row 2 and
                  so on.</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>Pixel Aspect Ratio</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>(0028,0034)</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1C</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>Ratio of the vertical size and horizontal size of the pixels in the image
                  specified by a pair of integer values where the first value is the vertical pixel
                  size, and the second value is the horizontal pixel size. Required if the aspect
                  ratio values do not have a ratio of 1:1 and the physical pixel spacing is not
                  specified by Pixel Spacing (0028,0030), or Imager Pixel Spacing (0018,1164) or
                  Nominal Scanned Pixel Spacing (0018,2010), either for the entire Image or
                  per-frame in a Functional Group Macro. See Section C.7.6.3.1.7.</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>Double Float Pixel Padding Value</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>(0028,0123)</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>3</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>One limit (inclusive) of a range of pixel values used in an image to pad to
                  rectangular format or to signal background that may be suppressed.</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>Double Float Pixel Padding Range Limit</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>(0028,0125)</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1C</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>Pixel value that represents one limit (inclusive) of a range of padding values
                  used together with Float Pixel Padding Value (0028,0122).</para>
                <para>Required if Double Float Pixel Padding Value (0028,0123) is present.</para>
                <note>
                  <orderedlist numeration="arabic">
                    <listitem>
                      <para>If only a single padding value rather than a range is required, then
                        both Double Float Pixel Padding Value (0028,0123) and Double Float Pixel
                        Padding Range Limit (0028,0125) will contain the same value.</para>
                    </listitem>
                    <listitem>
                      <para>The general considerations described in <xref linkend="sect_C.7.5.1.1.2"
                        xrefstyle="select: label"/> may be helpful in understanding the
                        corresponding floating point attributes, but are not normative.</para>
                    </listitem>
                  </orderedlist>
                </note>
              </td>
            </tr>
          </tbody>
        </table>
      </section>
            
      <section>
        <title/>
        <informaltable rules="all" frame="box">
          <tr>
            <td>
              <para><emphasis role="italic">Amend DICOM PS 3.3 - Information Object Definitions as follows, to add new Modules and Macros for Parametric Maps:</emphasis></para>
            </td>
          </tr>
        </informaltable>
      </section>
      
      
      
      <section label="C.8.32" status="3" xml:id="sect_PS3.3_C.8.32">
        <title>Parametric Map</title>
        <para>This section describes the specific modules for the Parametric Map IOD.</para>
        <section label="C.8.32.1" status="4" xml:id="sect_PS3.3_C.8.32.1">
          <title>Parametric Map Series Module</title>
          <para><xref linkend="table_PS3.3_C.8.32-1" xrefstyle="select: label"/> defines the general Attributes of the Parametric Map Series Module.</para>
          <table frame="box" label="C.8.32-1" rules="all" xml:id="table_PS3.3_C.8.32-1">
            <caption>Parametric Map Series Module Attributes</caption>
            <thead>
              <tr valign="top">
                <th align="center" colspan="1" rowspan="1">
                  <para>Attribute Name</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para>Tag</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para>Type</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para>Attribute Description</para>
                </th>
              </tr>
            </thead>
            <tbody>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para>Modality</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para>(0008,0060)</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para>1</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>Modality Type</para>
                  <note>
                    <orderedlist numeration="arabic">
                      <listitem><para>It is expected that the majority of Parametric Maps will use the appropriate value for the acquisition modality, e.g. "MR", and so no specific Defined Terms or Enumerated Values are specified here.</para></listitem>
                      <listitem><para>If the image is derived from multiple modalities, then a value of "OT" is appropriate.</para></listitem>
                    </orderedlist>
                  </note>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para>Series Number</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para>(0020,0011)</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para>1</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>A number that identifies this Series</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para>Referenced Performed Procedure Step Sequence</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para>(0008,1111)</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para>1C</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>Uniquely identifies the Performed Procedure Step SOP Instance to which the Series is related</para>
                  <para>Only a single Item shall be included in this sequence.</para>
                  <para>Required if a Performed Procedure Step SOP Class was involved in the creation of this Series.</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="3" rowspan="1">
                  <para>
                    <emphasis role="italic">&gt;Include Table 10-11 "SOP Instance Reference Macro"</emphasis>
                  </para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para/>
                </td>
              </tr>
            </tbody>
          </table>
        </section>
        <section label="C.8.32.2" status="4" xml:id="sect_PS3.3_C.8.32.2">
          <title>Parametric Map Image Module</title>
          <para><xref linkend="table_PS3.3_C.8.32-2" xrefstyle="select: label"/> defines the general Attributes of the Parametric Map Image Module.</para>
          <table frame="box" label="C.8.32-2" rules="all" xml:id="table_PS3.3_C.8.32-2">
            <caption>Parametric Map Image Module Attributes</caption>
            <thead>
              <tr valign="top">
                <th align="center" colspan="1" rowspan="1">
                  <para>Attribute Name</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para>Tag</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para>Type</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para>Attribute Description</para>
                </th>
              </tr>
            </thead>
            <tbody>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para>Image Type</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para>(0008,0008)</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para>1</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>Image identification characteristics.</para>
                  <variablelist spacing="compact">
                    <title>Enumerated Values for Value 1:</title>
                    <varlistentry><term>DERIVED</term><listitem><para/></listitem></varlistentry>
                  </variablelist>
                  <variablelist spacing="compact">
                    <title>Enumerated Values for Value 2:</title>
                    <varlistentry><term>PRIMARY</term><listitem><para/></listitem></varlistentry>
                  </variablelist>
                  <para>Value 3 shall be Image Flavor, Defined Terms for which are specified in
                    Section C.8.16.1.3.</para>
                  <para>Value 4 shall be Derived Pixel Contrast, common Defined Terms for which are
                    specified in Section C.8.16.1.4 and MR-specific Defined Terms for which are
                    specified in Section C.8.13.1.1.1.4.</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="3" rowspan="1">
                  <para>
                    <emphasis role="italic">Include Table 10-12 "Content Identification Macro"</emphasis>
                  </para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para/>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para>Samples per Pixel</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para>(0028,0002)</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para>1</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>Number of samples (planes) in this image.</para>
                  <variablelist spacing="compact">
                    <title>Enumerated Values:</title>
                    <varlistentry><term>1</term><listitem><para/></listitem></varlistentry>
                  </variablelist>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para>Photometric Interpretation</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para>(0028,0004)</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para>1</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>Specifies the intended interpretation of the pixel data.</para>
                  <variablelist spacing="compact">
                    <title>Enumerated Values:</title>
                    <varlistentry><term>MONOCHROME2</term><listitem><para/></listitem></varlistentry>
                  </variablelist>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para>Bits Allocated</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para>(0028,0100)</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para>1</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>Number of bits allocated for each pixel sample.</para>
                  <variablelist spacing="compact">
                    <title>Enumerated Values if Pixel Data (7FE0,0010) or Pixel Data Provider URL (0028,7FE0) is present:</title>
                    <varlistentry><term>16</term><listitem><para/></listitem></varlistentry>
                  </variablelist>
                  <variablelist spacing="compact">
                    <title>Enumerated Values if Float Pixel Data (7FE0,0008) is present:</title>
                    <varlistentry><term>32</term><listitem><para/></listitem></varlistentry>
                  </variablelist>
                  <variablelist spacing="compact">
                    <title>Enumerated Values if Double Float Pixel Data (7FE0,0009) is
                      present:</title>
                    <varlistentry><term>64</term><listitem><para/></listitem></varlistentry>
                  </variablelist>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para>Bits Stored</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para>(0028,0101)</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para>1C</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>Number of bits stored for each pixel sample.</para>
                  <variablelist spacing="compact">
                    <title>Enumerated Values:</title>
                    <varlistentry><term>16</term><listitem><para/></listitem></varlistentry>
                  </variablelist>
                  <para>Required if Pixel Data (7FE0,0010) or Pixel Data Provider URL (0028,7FE0) is present.</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para>High Bit</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para>(0028,0102)</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para>1C</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>Most significant bit for pixel sample data.</para>
                  <variablelist spacing="compact">
                    <title>Enumerated Values:</title>
                    <varlistentry><term>15</term><listitem><para/></listitem></varlistentry>
                  </variablelist>
                  <para>Required if Pixel Data (7FE0,0010) or Pixel Data Provider URL (0028,7FE0) is present.</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para>Presentation LUT Shape</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para>(2050,0020)</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para>1</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>Specifies an identity transformation for the Presentation LUT such that the output of all grayscale transformations are defined to be in P-Values.</para>
                  <variablelist spacing="compact">
                    <title>Enumerated Values:</title>
                    <varlistentry>
                      <term>IDENTITY</term>
                      <listitem>
                        <para>output is in P-Values.</para>
                      </listitem>
                    </varlistentry>
                  </variablelist>
                  <note>
                    <para>The intent of specifying this Attribute for Parametric Maps is only to
                      achieve consistency of rendering when displayed as grayscale, not to imply
                      that grayscale contrast of the displayed image is meaningful. It is not
                      applicable when the image is rendered with pseudo-color, for example.</para>
                  </note>
                 </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para>Lossy Image Compression</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para>(0028,2110)</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para>1</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>Specifies whether an Image has undergone lossy compression (at a point in its lifetime), or is derived from lossy compressed images.</para>
                  <variablelist spacing="compact">
                    <title>Enumerated Values:</title>
                    <varlistentry>
                      <term>00</term>
                      <listitem>
                        <para>Image has NOT been subjected to lossy compression.</para>
                      </listitem>
                    </varlistentry>
                    <varlistentry>
                      <term>01</term>
                      <listitem>
                        <para>Image has been subjected to lossy compression.</para>
                      </listitem>
                    </varlistentry>
                  </variablelist>
                  <para>Once this value has been set to 01 it shall not be reset.</para>
                  <para>See <xref linkend="sect_PS3.3_C.8.32.2.1" xrefstyle="select: label"/> and Section C.7.6.1.1.5.</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para>Lossy Image Compression Ratio</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para>(0028,2112)</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para>1C</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>Describes the approximate lossy compression ratio(s) that have been applied to this image.</para>
                  <para>See Section C.7.6.1.1.5 for further explanation. May be multivalued if successive lossy compression steps have been applied.</para>
                  <note>
                    <orderedlist numeration="arabic">
                      <listitem>
                        <para>For example, a compression ratio of 30:1 would be described in this Attribute with a single value of 30.</para>
                      </listitem>
                      <listitem>
                        <para>For historical reasons, the lossy compression ratio may also be described in Derivation Description (0008,2111).</para>
                      </listitem>
                    </orderedlist>
                  </note>
                  <para>Required if present in the source images or this IOD instance has been compressed.</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para>Lossy Image Compression Method</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para>(0028,2114)</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para>1C</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>A label for the lossy compression method(s) that have been applied to this image.</para>
                  <para>See Section C.7.6.1.1.5 for further explanation. May be multivalued if successive lossy compression steps have been applied; the value order shall correspond to the values of Lossy Image Compression Ratio (0028,2112).</para>
                  <note>
                    <para>For historical reasons, the lossy compression method may also be described in Derivation Description (0008,2111).</para>
                  </note>
                  <para>Required if present in the source images or this IOD instance has been compressed. See <xref linkend="sect_PS3.3_C.8.32.2.1" xrefstyle="select: label"/>.</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para>Burned In Annotation</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para>(0028,0301)</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para>1</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>Indicates whether or not image contains sufficient burned in annotation to identify the patient and date the image was acquired.</para>
                  <variablelist spacing="compact">
                    <title>Enumerated Values:</title>
                    <varlistentry>
                      <term>NO</term>
                      <listitem>
                        <para/>
                      </listitem>
                    </varlistentry>
                  </variablelist>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para>Recognizable Visual Features</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para>(0028,0302)</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para>1</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>Indicates whether or not the image contains sufficiently recognizable visual features to allow the image or a reconstruction from a set of images to identify the patient.</para>
                  <variablelist spacing="compact">
                    <title>Enumerated Values:</title>
                    <varlistentry>
                      <term>YES</term>
                      <listitem>
                        <para/>
                      </listitem>
                    </varlistentry>
                    <varlistentry>
                      <term>NO</term>
                      <listitem>
                        <para/>
                      </listitem>
                    </varlistentry>
                  </variablelist>
                 </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para>Content Qualification</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para>(0018,9004)</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para>1</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>Content Qualification Indicator</para>
                  <variablelist spacing="compact">
                    <title>Enumerated Values:</title>
                    <varlistentry>
                      <term>PRODUCT</term>
                      <listitem>
                        <para/>
                      </listitem>
                    </varlistentry>
                    <varlistentry>
                      <term>RESEARCH</term>
                      <listitem>
                        <para/>
                      </listitem>
                    </varlistentry>
                    <varlistentry>
                      <term>SERVICE</term>
                      <listitem>
                        <para/>
                      </listitem>
                    </varlistentry>
                  </variablelist>
                  <para>See Section C.8.13.2.1.1 for further explanation.</para>
                 </td>
              </tr>
            </tbody>
          </table>
          <section label="C.8.32.2.1" status="5" xml:id="sect_PS3.3_C.8.32.2.1">
            <title>Lossy Image Compression and Lossy Image Compression Method</title>
            <para>If Lossy Image Compression (0028,2110) in any of the source images is "01", the value shall be "01" for the Parametric Map instance.</para>
            <para>The process of derivation of a Parametric Map itself is defined not to be lossy compression, even though it involves loss. If the Parametric Map instance is encoded using a lossy compression transfer syntax, then the value shall be set to "01".</para>
            <note>
              <para>To state the obvious, it is not advisable to lossy compress a Parametric Map SOP Instance, since that will alter the quantitative values that are the intent of the object.</para>
            </note>
          </section>
        </section>
        <section label="C.8.32.3" status="4" xml:id="sect_PS3.3_C.8.32.3">
          <title>Parametric Map Functional Group Macros</title>
          <para>The following sections contain Functional Group Macros specific to the Parametric Map IOD.</para>
          <note>
            <para>The attribute descriptions in the Functional Group Macros are written as if they were applicable to a single frame (i.e., the macro is part of the Per-frame Functional Groups Sequence). If an attribute is applicable to all frames (i.e., the macro is part of the Shared Functional Groups Sequence) the phrase "this frame" in the attribute description shall be interpreted to mean "for all frames".</para>
          </note>
          <section label="C.8.32.3.1" status="5" xml:id="sect_PS3.3_C.8.32.3.1">
            <title>Parametric Map Frame Type Macro</title>
            <para><xref linkend="table_PS3.3_C.8.32-3" xrefstyle="select: label"/> specifies the attributes of the Parametric Map Frame Type Functional Group Macro.</para>
            <table frame="box" label="C.8.32-3" rules="all" xml:id="table_PS3.3_C.8.32-3">
              <caption>Parametric MapFrame Type  Macro Attributes</caption>
              <thead>
                <tr valign="top">
                  <th align="center" colspan="1" rowspan="1">
                    <para>Attribute Name</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para>Tag</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para>Type</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para>Attribute Description</para>
                  </th>
                </tr>
              </thead>
              <tbody>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>Parametric Map Frame Type Sequence</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0040,9092)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>1</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Identifies the characteristics of this Parametric Map frame.</para>
                    <para>Only a single Item shall be included in this sequence.</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>&gt;Frame Type</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,9007)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>1</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Type of Frame. A multi-valued attribute analogous to Image Type (0008,0008).</para>
                    <para>Enumerated Values and Defined Terms are the same as those for the four
                      values of Image Type (0008,0008), except that the value MIXED is not allowed.
                      See <xref linkend="sect_PS3.3_C.8.32.2" xrefstyle="select: label"/>.</para>
                  </td>
                </tr>
              </tbody>
            </table>
          </section>
        </section>
      </section>
    </section>
  </chapter>
  
  <chapter label="DICOM PS3.4" xml:id="chapter_PS3.4">
    <title>Service Class Specifications</title>
    <section label="B" xml:id="chapter_PS3.4_B">
      <title>Storage Service Class (Normative)</title>
      <informaltable rules="all" frame="box">
        <tr>
          <td>
            <para><emphasis role="italic">Amend DICOM PS 3.4 - Service Class Specifications - Annex B - Storage Service Class (Normative) as follows:</emphasis></para>
          </td>
        </tr>
      </informaltable>
      
      <section label="B.5" status="2" xml:id="sect_B.5">
        <title>Standard SOP Classes</title>
        <para>...</para>
        <table frame="box" label="B.5-1" rules="all" xml:id="table_B.5-1">
          <caption>Standard SOP Classes</caption>
          <thead>
            <tr valign="top">
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold">SOP Class Name</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold">SOP Class UID</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold">IOD Specification (defined in PS3.3)</emphasis>
                </para>
              </th>
            </tr>
          </thead>
          <tbody>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>...</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>...</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>...</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">Parametric Map Storage</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline"
                      >1.2.840.10008.5.1.4.1.1.30</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">Parametric Map IOD</emphasis></emphasis></para>   <!-- hyperlink this to PS 3.3 section when folded in -->
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>...</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>...</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>...</para>
              </td>
            </tr>
          </tbody>
        </table>
      </section>
      
    </section>
    
    <section label="I" xml:id="chapter_PS3.4_I">
      <title>Media Storage Service Class (Normative)</title>
      <informaltable rules="all" frame="box">
        <tr>
          <td>
            <para><emphasis role="italic">Amend DICOM PS 3.4 - Service Class Specifications - Annex I - Media Storage Service Class (Normative) as follows:</emphasis></para>
          </td>
        </tr>
      </informaltable>
      
      <section label="I.4" status="2" xml:id="sect_I.4">
        <title>Media Storage Standard SOP Classes</title>
        <para>...</para>
        <table frame="box" label="I.4-1" rules="all" xml:id="table_I.4-1">
          <caption>Media Storage Standard SOP Classes</caption>
          <thead>
            <tr valign="top">
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold">SOP Class Name</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold">SOP Class UID</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold">IOD Specification (defined in PS3.3)</emphasis>
                </para>
              </th>
            </tr>
          </thead>
          <tbody>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>...</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>...</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>...</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">Parametric Map Storage</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline"
                      >1.2.840.10008.5.1.4.1.1.30</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">Parametric Map IOD</emphasis></emphasis></para>   <!-- hyperlink this to PS 3.3 section when folded in -->
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>...</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>...</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>...</para>
              </td>
            </tr>
          </tbody>
        </table>
      </section>
      
      
    </section>
    
  </chapter>
  
  <chapter label="DICOM PS3.5" xml:id="chapter_PS3.5">
    <title>Data Structures and Encoding</title>
    
    <section label="3" xml:id="chapter_PS3.5_3">
      <title>Definitions</title>
      
      <informaltable rules="all" frame="box">
        <tr>
          <td>
            <para><emphasis role="italic">Amend DICOM PS 3.5 - Data Structures and Encoding - Section 3 - Definitions as follows:</emphasis></para>
          </td>
        </tr>
      </informaltable>
      
      
      <section xml:id="sect_3.10" label="3.10" status="2">
        <title>DICOM Data Structures and Encoding Definitions</title>
        <para>The following definitions are commonly used in this Standard:</para>
        <para>...</para>
        <para><emphasis role="bold">Pixel Cell</emphasis>: The container for a single Pixel Sample Value that may include unused bits<emphasis role="bold"><emphasis role="strikethrough"> or bits for data other than the Pixel Sample Value (e.g., overlay planes)</emphasis></emphasis>. The size of a Pixel Cell shall be specified by the Bits Allocated (0028, 0100) Data Element.</para>
        <para><emphasis role="bold">Pixel Data</emphasis>: Graphical data (e.g., images<emphasis role="bold"><emphasis role="strikethrough"> or overlays</emphasis></emphasis>) of variable pixel-depth encoded in the Pixel Data<emphasis role="bold"><emphasis role="underline">, Float Pixel Data or Double Float Pixel Data</emphasis></emphasis> Element<emphasis role="bold"><emphasis role="strikethrough">, with Value Representation OW or OB</emphasis></emphasis>.<emphasis role="bold"><emphasis role="strikethrough"> Additional descriptor Data Elements are often used to describe the contents of the Pixel Data element.</emphasis></emphasis></para>
        <para><emphasis role="bold">Pixel Sample Value</emphasis>: A value associated with an individual pixel. An individual pixel consists of one or more Pixel Sample Values (e.g., color images).</para>
        <para>...</para>
      </section>
    </section>
    
    <section xml:id="sect_PS3.5_6" label="PS3.5_6">
      <title></title>
      
      <informaltable rules="all" frame="box">
        <tr>
          <td>
            <para><emphasis role="italic">For reference DICOM PS 3.5 - table defining OF and OD VR:</emphasis></para>
          </td>
        </tr>
      </informaltable>
      
      <section xml:id="sect_PS3.5_6.2" label="PS3.5_6.2" status="2">
      <title>Value Representation (VR)</title>
     <para>...</para>
      <table frame="box" rules="all" xml:id="table_6.2-1" label="6.2-1">
        <caption>DICOM Value Representations</caption>
        <thead>
          <tr valign="top">
            <th align="center" colspan="1" rowspan="1">
              <para>
                <emphasis role="bold">VR Name</emphasis>
              </para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para>
                <emphasis role="bold">Definition</emphasis>
              </para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para>
                <emphasis role="bold">Character Repertoire</emphasis>
              </para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para>
                <emphasis role="bold">Length of Value</emphasis>
              </para>
            </th>
          </tr>
        </thead>
        <tbody>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>...</para>
              <para>...</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>...</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>...</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>...</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>OD</para>
              <para>Other Double String</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>A string of 64-bit IEEE 754:1985 floating point words. OD is a VR that requires byte swapping within each 64-bit word when changing between Little Endian and Big Endian byte ordering (see Section 7.3).</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>not applicable</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>2<superscript>32</superscript>-8 bytes maximum</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>OF</para>
              <para>Other Float String</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>A string of 32-bit IEEE 754:1985 floating point words. OF is a VR that requires byte swapping within each 32-bit word when changing between Little Endian and Big Endian byte ordering (see Section 7.3).</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>not applicable</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>2<superscript>32</superscript>-4 bytes maximum</para>
            </td>
          </tr>
        </tbody>
      </table>
    </section>
      
    </section>
    
    <section xml:id="sect_PS3.5_7" label="PS3.5_7">
      <title>The Data Set</title>
      
      <informaltable rules="all" frame="box">
        <tr>
          <td>
            <para><emphasis role="italic">Amend DICOM PS 3.5 - Data Structures and Encoding - Section 7 - The Data Set as follows:</emphasis></para>
          </td>
        </tr>
      </informaltable>
      
      <section xml:id="sect_PS3.5_7.8.2" label="7.8.2" status="3">
        <title>Encoding of Private Elements</title>
        <para>...</para>
        <para>For a Standard Extended SOP Class the Attributes Pixel Data (07FE,0010)<emphasis
            role="bold"><emphasis role="underline">, Float Pixel Data (7FE0,0008), Double Float
              Pixel Data (7FE0,0009)</emphasis></emphasis>, Waveform Data (5400,1010), and Overlay
          Data (60xx,3000) shall not be included within a Private Sequence Item, nor within a
          standard Sequence Item nested directly or indirectly within a Private Sequence
          Item.</para>
      </section>
    </section>
    
    
    <section label="8" xml:id="chapter_PS3.5_8">
      <title>Encoding of Pixel, Overlay and Waveform Data</title>
      <informaltable rules="all" frame="box">
        <tr>
          <td>
            <para><emphasis role="italic">Amend DICOM PS 3.5 - Data Structures and Encoding - Section 8 - Encoding of Pixel, Overlay and Waveform Data as follows:</emphasis></para>
          </td>
        </tr>
      </informaltable>
      
      <section xml:id="sect_PS3.5_8.1" label="8.1" status="2">
        <title>Pixel and Overlay Data, and Related Data Elements</title>
        <para><emphasis role="bold"><emphasis role="strikethrough">The </emphasis></emphasis>Pixel
          Data <emphasis role="bold"><emphasis role="strikethrough">Element
            </emphasis></emphasis>(7FE0,0010)<emphasis role="bold"><emphasis role="underline">,
              Float Pixel Data (7FE0,0008), Double Float Pixel Data
            (7FE0,0009)</emphasis></emphasis> and Overlay Data <emphasis role="bold"><emphasis
              role="strikethrough">Element </emphasis></emphasis>(60xx,3000) shall be used for the
          exchange of encoded graphical image data. These elements along with additional Data
          Elements, specified as Attributes of the Image Information Entities defined in PS3.3,
          shall be used to describe the way in which the Pixel Data and Overlay Data are encoded and
          shall be interpreted. Finally, depending on the negotiated Transfer Syntax (see Section 10
          and <xref linkend="chapter_PS3.5_A" xrefstyle="template: Annex %n"/>), Pixel Data may be
          compressed.</para>
        <para><emphasis role="bold"><emphasis role="strikethrough">The </emphasis></emphasis>Pixel Data <emphasis role="bold"><emphasis role="strikethrough">Element </emphasis></emphasis>(7FE0,0010) and Overlay Data <emphasis role="bold"><emphasis role="strikethrough">Element </emphasis></emphasis>(60xx,3000) have a VR of OW or OB, depending on the negotiated Transfer Syntax (see <xref linkend="chapter_PS3.5_A" xrefstyle="template: Annex %n"/>). The only difference between OW and OB being that OB, a string of bytes, shall be unaffected by Byte Ordering (see Section 7.3).</para>
        <para><emphasis role="bold"><emphasis role="underline">Float Pixel Data (7FE0,0008) has a
              Value Representation of OF.</emphasis></emphasis></para>
        <para><emphasis role="bold"><emphasis role="underline">Double Float Pixel Data (7FE0,0009)
              has a Value Representation of OD.</emphasis></emphasis></para>
        <para><emphasis role="bold"><emphasis role="underline">For Pixel Data values encoded in OF
              and OD, any value that is permitted by the IEEE 754:1985 may be used, including NaN,
              +ve Infinity and -ve Infinity. See <xref linkend="table_6.2-1"
                xrefstyle="select: label"/></emphasis></emphasis></para>
        <note><para><emphasis role="bold"><emphasis role="underline">Float and double float pixel data values are
                not arbitrarily constrained to finite numbers, since it may be important for the
                application to signal that the result of a calculation that produced a pixel is an
                infinite value or not a number.</emphasis></emphasis></para></note>
        <section xml:id="sect_8.1.1" label="8.1.1" status="3">
          <title>Pixel Data Encoding of Related Data Elements</title>
          <para>Encoded Pixel Data of various bit depths shall be accommodated. The following three Data Elements shall define the Pixel structure:</para>
          <itemizedlist>
            <listitem>
              <para>Bits Allocated (0028,0100)</para>
            </listitem>
            <listitem>
              <para>Bits Stored (0028,0101)</para>
            </listitem>
            <listitem>
              <para>High Bit (0028,0102)</para>
            </listitem>
          </itemizedlist>
          <para>Each Pixel Cell shall contain a single Pixel Sample Value. The size of the Pixel Cell shall be specified by Bits Allocated (0028,0100). Bits Stored (0028,0101) defines the total number of these allocated bits that will be used to represent a Pixel Sample Value. Bits Stored (0028,0101) shall never be larger than Bits Allocated (0028,0100). High Bit (0028,0102) specifies where the high order bit of the Bits Stored (0028,0101) is to be placed with respect to the Bits Allocated (0028,0100) specification. <emphasis role="bold"><emphasis role="strikethrough">Bits not used for Pixel Sample Values can be used for overlay planes described further in PS3.3</emphasis></emphasis>.</para>
          <note>
            <orderedlist>
              <listitem><para>For example, in Pixel Data with 16 bits (2 bytes) allocated, 12 bits stored, and bit 15 specified as the high bit, one pixel sample is encoded in each 16-bit word, with the 4 least significant bits of each word not containing Pixel Data. See <xref linkend="chapter_PS3.5_D" xrefstyle="template: Annex %n"/> for other examples of the basic encoding schemes.</para></listitem>
              <listitem><para><emphasis role="bold"><emphasis role="underline">Formerly, bits not used for Pixel Sample
                Values were described as being usable for overlay planes, but this usage has
                been retired. See PS3.5-2004.</emphasis></emphasis></para></listitem>
            </orderedlist>
            
          </note>
          <para>Restrictions are placed on acceptable Values for Bits Allocated (0028,0100), Bits
            Stored (0028,0101), and High Bit (0028,0102) <emphasis role="bold"><emphasis
              role="underline">for Pixel Data (7FE0,0010) </emphasis></emphasis>and are specified
            in the Information Object Definitions in PS3.3.</para>
          <para><emphasis role="bold"><emphasis role="underline">Restrictions are placed on
                acceptable Values for Bits Allocated (0028,0100) for Float Pixel Data (7FE0,0008) and
                Double Float Pixel Data (7FE0,0009), such that only a single Pixel Cell entirely
                occupies the allocated bits specified by Bits Allocated (0028,0100), hence Bits
                Stored (0028,0101) and High Bit (0028,0102) are not
            sent.</emphasis></emphasis></para>
          <para> Also, the Value Field containing Pixel Data, like all other Value Fields in DICOM,
            shall be an even number of bytes in length. This means that the Value Field may need to
            be padded with data that is not part of the image and shall not be considered
            significant. If needed, the padding bits shall be appended to the end of the Value
            Field, and shall be used only to extend the data to the next even byte increment of
            length.</para>
          <para>In a multi-frame object that is transmitted in Native Format, the individual frames are not padded. The individual frames shall be concatenated and padding bits (if necessary) apply to the complete Value Field.</para>
          <note>
            <para>Receiving applications should be aware that some older applications may send Pixel Data with excess padding, which was not explicitly prohibited in earlier versions of the Standard. Applications should be prepared to accept such Pixel Data elements, but may delete the excess padding. In no case should a sending application place private data in the padding data.</para>
          </note>
          <para>The field of bits representing the value of a Pixel Sample shall be a binary 2's complement integer or an unsigned integer, as specified by the Data Element Pixel Representation (0028,0103). The sign bit shall be the High Bit in a Pixel Sample Value that is a 2's complement integer. The minimum actual Pixel Sample Value encountered in the Pixel Data is specified by Smallest Image Pixel Value (0028,0106) while the maximum value is specified by Largest Image Pixel Value (0028,0107).</para>
        </section>
        <section xml:id="sect_8.1.2" label="8.1.2" status="3">
          <title>Overlay Data Encoding of Related Data Elements</title>
          <para>Encoded Overlay Planes always have a bit depth of 1, and are encoded separately from the Pixel Data in Overlay Data (60xx,3000). The following two Data Elements shall define the Overlay Plane structure:</para>
          <itemizedlist>
            <listitem>
              <para>Overlay Bits Allocated (60xx,0100)</para>
            </listitem>
            <listitem>
              <para>Overlay Bit Position (60xx,0102)</para>
            </listitem>
          </itemizedlist>
          <note>
            <orderedlist numeration="arabic">
              <listitem>
                <para>There is no Data Element analogous to Bits Stored (0028,0101) since Overlay Planes always have a bit depth of 1.</para>
              </listitem>
              <listitem>
                <para>Restrictions on the allowed values for these Data Elements are defined in PS3.3. Formerly overlay data stored in unused bits of Pixel Data (7FE0,0010) was described, and these attributes had meaningful values but this usage has been retired. See PS3.5-2004. For overlays encoded in Overlay Data Element (60xx,3000), Overlay Bits Allocated (60xx,0100) is always 1 and Overlay Bit Position (60xx,0102) is always 0.</para>
              </listitem>
            </orderedlist>
          </note>
          <para>For Overlay Data Element (60xx,3000)...</para>
          <para>...</para>
        </section>
      </section>
      <section xml:id="sect_PS3.5_8.2" label="8.2" status="2">
        <title>Native or Encapsulated Format Encoding</title>
        <para>Pixel data conveyed in the Pixel Data Element (7FE0,0010) may be sent either in a Native (uncompressed) Format or in an Encapsulated Format (e.g., compressed) defined outside the DICOM standard.</para>
        <para><emphasis role="bold"><emphasis role="underline">Pixel Data conveyed in the Float
              Pixel Data (7FE0,0008) or Double Float Pixel Data (7FE0,0009) shall be in a Native
              (uncompressed) Format if encoded in a Standard Transfer
          Syntax.</emphasis></emphasis></para>
        <note>
          <orderedlist>
            <listitem>
              <para><emphasis role="bold"><emphasis role="underline">In future, if Standard Transfer
                    Syntaxes are defined for compression of Float Pixel Data (7FE0,0008) or Double
                    Float Pixel Data (7FE0,0009), this constraint may be relaxed and Encapsulated
                    Format permitted.</emphasis></emphasis></para>
            </listitem>
            <listitem>
              <para><emphasis role="bold"><emphasis role="underline">This constraint does not apply to Private Transfer Syntaxes.</emphasis></emphasis></para>
            </listitem>
          </orderedlist>
          
        </note>
        <para>If Pixel Data <emphasis role="bold"><emphasis role="underline">(7FE0,0010) </emphasis></emphasis>is sent in a Native Format, the Value Representation OW is
          most often required. The Value Representation OB may also be used for Pixel Data <emphasis role="bold"><emphasis role="underline">(7FE0,0010) </emphasis></emphasis>in cases
          where Bits Allocated has a value less than or equal to 8, but only with Transfer Syntaxes
          where the Value Representation is explicitly conveyed (see <xref linkend="chapter_PS3.5_A" xrefstyle="template: Annex %n"/>).</para>
        <note>
          <para>The DICOM default Transfer Syntax (Implicit VR Little Endian) does not explicitly convey Value Representation and therefore the VR of OB may not be used for Pixel Data <emphasis role="bold"><emphasis role="underline">(7FE0,0010) </emphasis></emphasis>when using the default Transfer Syntax.</para>
        </note>
        <para><emphasis role="bold"><emphasis role="underline">Float Pixel Data (7FE0,0008) is sent
              in Native Format; the Value Representation shall be OF, Bits Allocated (0028,0100)
              shall be 32, Bits Stored (0028,0101), High Bit (0028,0102) and Pixel Representation
              (0028,0103) shall not be present.</emphasis></emphasis></para>
        <para><emphasis role="bold"><emphasis role="underline">Double Float Pixel Data (7FE0,0009)
              is sent in Native Format; the Value Representation shall be OD, Bits Allocated
              (0028,0100) shall be 64, Bits Stored (0028,0101) and High Bit (0028,0102) and Pixel
              Representation (0028,0103) shall not be present.</emphasis></emphasis></para>
        <para><emphasis role="bold"><emphasis role="underline">It is not permitted to have more than
              one of Pixel Data Provider URL (0028,7FE0), Pixel Data (7FE0,0010), Float Pixel Data
              (7FE0,0008) or Double Float Pixel Data (7FE0,0009) in the top level Data
              Set.</emphasis></emphasis></para>
        <note>
          <para><emphasis role="bold"><emphasis role="underline">Pixel Data encoded in Float Pixel
                Data (7FE0,0008) or Double Float Pixel Data (7FE0,0009) can be considered as
                consisting of Pixel Cells that entirely occupy the allocated bits, and therefore do
                not cross word boundaries.</emphasis></emphasis></para>
        </note>
        <para>Native format Pixel Cells are encoded as the direct concatenation of the bits of each Pixel Cell, the least significant bit of each Pixel Cell is encoded in the least significant bit of the encoded word or byte, immediately followed by the next most significant bit of each Pixel Cell in the next most significant bit of the encoded word or byte, successively until all bits of the Pixel Cell have been encoded, then immediately followed by the least significant bit of the next Pixel Cell in the next most significant bit of the encoded word or byte. The number of bits of each Pixel Cell is defined by the Bits Allocated (0028,0100) Data Element Value. When a Pixel Cell crosses a word boundary in the OW case, or a byte boundary in the OB case, it shall continue to be encoded, least significant bit to most significant bit, in the next word, or byte, respectively (see <xref linkend="chapter_PS3.5_D" xrefstyle="template: Annex %n"/>). For Pixel Data <emphasis role="bold"><emphasis role="underline">(7FE0,0010) </emphasis></emphasis>encoded with the Value Representation OW, the byte ordering of the resulting 2-byte words is
          defined by the Little Endian or Big Endian Transfer Syntaxes negotiated at the Association Establishment (see <xref linkend="chapter_PS3.5_A" xrefstyle="template: Annex %n"/>).</para>
        <note>
          <orderedlist numeration="arabic">
            <listitem>
              <para>For Pixel Data <emphasis role="bold"><emphasis role="underline">(7FE0,0010) </emphasis></emphasis>encoded with the Value Representation OB, the Pixel Data <emphasis role="bold"><emphasis role="underline">(7FE0,0010) </emphasis></emphasis>encoding is unaffected by Little Endian or Big Endian byte ordering.</para>
            </listitem>
            <listitem>
              <para>If encoding Pixel Data <emphasis role="bold"><emphasis role="underline">(7FE0,0010) </emphasis></emphasis>with a Value for Bits Allocated (0028,0100) not equal to 16 be sure to read and understand <xref linkend="chapter_PS3.5_D" xrefstyle="template: Annex %n"/>.</para>
            </listitem>
          </orderedlist>
        </note>
        <para>If sent in an Encapsulated Format ...</para>
        <para>...</para>
      </section>
      <section xml:id="sect_PS3.5_8.4" label="8.4" status="2">
        <title>Pixel Data Provider Service</title>
        <para>Specific Transfer Syntaxes allow for the pixel data of the message to be replaced with a reference to a pixel data provider service. The pixel data provider service that is referenced supplies the pixel data using a network protocol that is defined outside DICOM.</para>
        <note>
          <para><emphasis role="bold"><emphasis role="underline">The Pixel Data Provider Service is
                not applicable to Pixel Data encoded as Float Pixel Data (7FE0,0008) or Double Float
                Pixel Data (7FE0,0009).</emphasis></emphasis></para>
        </note>
        <para>...</para>
      </section>
      
      
    </section>
    
    <section label="A" xml:id="chapter_PS3.5_A">
      <title>Transfer Syntax Specifications (Normative)</title>
      
      <informaltable rules="all" frame="box">
        <tr>
          <td>
            <para><emphasis role="italic">Amend DICOM PS 3.5 - Data Structures and Encoding - Annex
                A - Transfer Syntax Specifications (note that few changes are necessary since the
                new Data Elements have OF and OD VR and are not affected by OB/OW
                concerns):</emphasis></para>
          </td>
        </tr>
      </informaltable>
      
      <section xml:id="sect_PS3.5_A.1" label="A.1" status="2">
        <title>DICOM Implicit VR Little Endian Transfer Syntax</title>
        <para>This Transfer Syntax applies to the encoding of the entire DICOM Data Set. This implies that when a DICOM Data Set is being encoded with the DICOM Implicit VR Little Endian Transfer Syntax the following requirements shall be met:</para>
        <orderedlist numeration="loweralpha">
          <listitem>
            <para>The Data Elements contained in the Data Set structure shall be encoded with Implicit VR (without a VR Field) as specified in Section 7.1.3.</para>
          </listitem>
          <listitem>
            <para>The encoding of the overall Data Set structure (Data Element Tags, Value Length, and Value) shall be in Little Endian as specified in Section 7.3.</para>
          </listitem>
          <listitem>
            <para>The encoding of the Data Elements of the Data Set shall be as follows according to their Value Representations:</para>
            <itemizedlist>
              <listitem>
                <para>For all Value Representations defined in this part, except for the Value
                  Representations OB and OW, the encoding shall be in Little Endian as specified in
                  Section 7.3.</para>
              </listitem>
              <listitem>
                <para>For the Value Representations OB and OW, the encoding shall meet the following
                  specification depending on the Data Element Tag:</para>
              </listitem>
              <listitem>
                <para>Data Element (7FE0,0010) Pixel Data has the Value Representation OW and shall be encoded in Little Endian.</para>
              </listitem>
              <listitem>
                <para>...</para>
              </listitem>
            </itemizedlist>
          </listitem>
        </orderedlist>
        <para>...</para>
      </section>
      <section xml:id="sect_PS3.5_A.2" label="A.2" status="2">
        <title>DICOM Little Endian Transfer Syntax (Explicit VR)</title>
        <para>This Transfer Syntax applies to the encoding of the entire DICOM Data Set. This implies that when a DICOM Data Set is being encoded with the DICOM Little Endian Transfer Syntax the following requirements shall be met:</para>
        <orderedlist numeration="loweralpha">
          <listitem>
            <para>The Data Elements contained in the Data Set structure shall be encoded with Explicit VR (with a VR Field) as specified in Section 7.1.2.</para>
          </listitem>
          <listitem>
            <para>The encoding of the overall Data Set structure (Data Element Tags, Value Length, and Value) shall be in Little Endian as specified in Section 7.3.</para>
          </listitem>
          <listitem>
            <para>The encoding of the Data Elements of the Data Set shall be as follows according to their Value Representations:</para>
            <itemizedlist>
              <listitem>
                <para>For all Value Representations defined in this part, except for the Value Representations OB and OW, the encoding shall be in Little Endian as specified in Section 7.3.</para>
              </listitem>
              <listitem>
                <para>For the Value Representations OB and OW, the encoding shall meet the following specification depending on the Data Element Tag:</para>
                <itemizedlist>
                  <listitem>
                    <para>Data Element (7FE0,0010) Pixel Data</para>
                    <itemizedlist>
                      <listitem>
                        <para>where Bits Allocated (0028,0100) has a value greater than 8 shall have Value Representation OW and shall be encoded in Little Endian;</para>
                      </listitem>
                      <listitem>
                        <para>where Bits Allocated (0028,0100) has a value less than or equal to 8 shall have the Value Representation OB or OW and shall be encoded in Little Endian.</para>
                      </listitem>
                    </itemizedlist>
                  </listitem>
                  <listitem>
                    <para>...</para>
                  </listitem>
                </itemizedlist>
              </listitem>
            </itemizedlist>
          </listitem>
        </orderedlist>
        <para>...</para>
      </section>
      <section xml:id="sect_PS3.5_A.3" label="A.3" status="2">
        <title>DICOM Big Endian Transfer Syntax (Explicit VR)</title>
        <para>This Transfer Syntax applies to the encoding of the entire DICOM Data Set. This implies that when a DICOM Data Set is being encoded with the DICOM Big Endian Transfer Syntax the following requirements shall be met:</para>
        <orderedlist numeration="loweralpha">
          <listitem>
            <para>The Data Elements contained in the Data Set structure shall be encoded with Explicit VR (with a VR Field) as specified in Section 7.1.2.</para>
          </listitem>
          <listitem>
            <para>The encoding of the overall Data Set structure (Data Element Tags, Value Length, and Value) shall be in Big Endian as specified in Section 7.3.</para>
          </listitem>
          <listitem>
            <para>The encoding of the Data Elements of the Data Set shall be as follows according to their Value Representations:</para>
            <itemizedlist>
              <listitem>
                <para>For all Value Representations defined in this part, except for the Value Representations OB and OW, the encoding shall be in Big Endian as specified in Section 7.3.</para>
              </listitem>
              <listitem>
                <para>For the Value Representations OB and OW, the encoding shall meet the following specification depending on the Data Element Tag:</para>
                <itemizedlist>
                  <listitem>
                    <para>Data Element (7FE0,0010) Pixel Data</para>
                    <itemizedlist>
                      <listitem>
                        <para>where Bits Allocated (0028,0100) has a value greater than 8 shall have Value Representation OW and shall be encoded in Big Endian;</para>
                      </listitem>
                      <listitem>
                        <para>where Bits Allocated (0028,0100) has a value less than or equal to 8 shall have the Value Representation OB or OW and shall be encoded in Big Endian.</para>
                      </listitem>
                    </itemizedlist>
                  </listitem>
                  <listitem>
                    <para>...</para>
                  </listitem>
                </itemizedlist>
              </listitem>
            </itemizedlist>
          </listitem>
        </orderedlist>
        <para>...</para>
      </section>
      <section xml:id="sect_PS3.5_A.4" label="A.4" status="2">
        <title>Transfer Syntaxes For Encapsulation of Encoded Pixel Data</title>
        <para>These Transfer Syntaxes apply to the encoding of the entire DICOM Data Set, even
          though the image Pixel Data (7FE0,0010) portion of the DICOM Data Set is the only portion
          that is encoded by an encapsulated format. <emphasis role="underline"><emphasis
              role="bold">These Transfer Syntaxes shall only be used when Pixel Data (7FE0,0010) is
              present in the top level Data Set, and hence shall not be used when Float Pixel Data
              (7FE0,0008) or Double Float Pixel Data (7FE0,0009) are present.
          </emphasis></emphasis>This implies that when a DICOM Message is being encoded according to
          an encapsulation Transfer Syntax the following requirements shall be met:</para>
        <para>...</para>
      </section>
      <section xml:id="sect_PS3.5_A.5" label="A.5" status="2">
        <title>DICOM Deflated Little Endian Transfer Syntax (Explicit VR)</title>
        <para>This Transfer Syntax applies to the encoding of the entire DICOM Data Set.</para>
        <para>The entire Data Set is first encoded according to the rules specified in <xref linkend="sect_PS3.5_A.2" xrefstyle="select: label"/>.</para>
        <para>The entire byte stream is then compressed using the "Deflate" algorithm defined in Internet RFC 1951.</para>
        <para>If the deflate algorithm produces an odd number of bytes then a single trailing NULL byte shall be added after the last byte of the deflated bit stream.</para>
        <note>
          <orderedlist numeration="arabic">
            <listitem>
              <para>The Pixel Data <emphasis role="bold"><emphasis role="underline">in Pixel Data
                  </emphasis></emphasis>(7FE0,0010)<emphasis role="bold"><emphasis role="underline"
                    >, Float Pixel Data (7FE0,0008) or Double Float Pixel Data
                    (7FE0,0009)</emphasis></emphasis> is not handled in any special manner. The
                pixel data is first encoded as sequential uncompressed frames without encapsulation,
                and then is handled as part of the byte stream fed to the "deflate" compressor in
                the same manner as the value of any other attribute.</para>
            </listitem>
            <listitem>
              <para>This transfer syntax is particularly useful for compression of objects without pixel data, such as structured reports. It is not particularly effective at image compression, since any benefit obtained from compressing the non-pixel data is offset by less effective compression of the much larger pixel data.</para>
            </listitem>
            <listitem>
              <para>...</para>
            </listitem>
          </orderedlist>
        </note>
        <para>...</para>
      </section>
      <section xml:id="sect_PS3.5_A.6" label="A.6" status="2">
        <title>DICOM JPIP Referenced Transfer Syntax (Explicit VR)</title>
        <para>This Transfer Syntax applies to the encoding of the entire DICOM Data Set. <emphasis
            role="underline"><emphasis role="bold">This Transfer Syntax shall only be used when
              Pixel Data (7FE0,0010) is present in the top level Data Set, and hence shall not be
              used when Float Pixel Data (7FE0,0008) or Double Float Pixel Data (7FE0,0009) are
              present. </emphasis></emphasis>This implies that when a DICOM Data Set is being
          encoded with the DICOM Little Endian Transfer Syntax the following requirements shall be
          met:</para>
        <orderedlist numeration="loweralpha">
          <listitem>
            <para>The Data Elements contained in the Data Set structure shall be encoded with Explicit VR (with a VR Field) as specified in Section 7.1.2.</para>
          </listitem>
          <listitem>
            <para>The encoding of the overall Data Set structure (Data Element Tags, Value Length, and Value) shall be in Little Endian as specified in Section 7.3.</para>
          </listitem>
          <listitem>
            <para>The encoding of the Data Elements of the Data Set shall be as follows according to their Value Representations:</para>
            <itemizedlist>
              <listitem>
                <para>For all Value Representations defined in this part, except for the Value Representations OB and OW, the encoding shall be in Little Endian as specified in Section 7.3.</para>
              </listitem>
              <listitem>
                <para>For the Value Representations OB and OW, the encoding shall meet the following specification depending on the Data Element Tag:</para>
                <itemizedlist>
                  <listitem>
                    <para>Data Element (7FE0,0010) Pixel Data shall not be present, but rather pixel data shall be referenced via Data Element (0028,7FE0) Pixel Data Provider URL</para>
                  </listitem>
                  <listitem>
                    <para>Overlay data, if present, shall only be encoded in the Overlay Data attribute (60xx,3000), which shall have the Value Representation OB or OW and shall be encoded in Little Endian.</para>
                  </listitem>
                  <listitem>
                    <para>Data Element (0028,0004) Photometric Interpretation shall be limited to the values: MONOCHROME1, MONOCHROME2, YBR_ICT and YBR_RCT.</para>
                  </listitem>
                </itemizedlist>
              </listitem>
            </itemizedlist>
          </listitem>
        </orderedlist>
        <para>This DICOM JPIP Referenced Transfer Syntax shall be identified by a UID of Value "1.2.840.10008.1.2.4.94 ".</para>
      </section>
      <section xml:id="sect_PS3.5_A.7" label="A.7" status="2">
        <title>DICOM JPIP Referenced Deflate Transfer Syntax (Explicit VR)</title>
        <para>This Transfer Syntax applies to the encoding of the entire DICOM Data Set.</para>
        <para>The entire Data Set is first encoded according to the rules specified in <xref linkend="sect_PS3.5_A.6" xrefstyle="select: label"/>.</para>
        <para>The entire byte stream is then compressed using the "Deflate" algorithm defined in Internet RFC 1951.</para>
        <para>This DICOM JPIP Referenced Deflate Transfer Syntax shall be identified by a UID of Value "1.2.840.10008.1.2.4.95".</para>
      </section>
      
      
    </section>
    
    <section label="D" xml:id="chapter_PS3.5_D">
      <title>Examples of Various Pixel Data and Overlay Encoding Schemes (Informative)</title>
      
      <informaltable rules="all" frame="box">
        <tr>
          <td>
            <para><emphasis role="italic">Amend DICOM PS 3.5 - Data Structures and Encoding - Annex
              D - Examples of Various Pixel Data and Overlay Encoding Schemes as
              follows:</emphasis></para>
          </td>
        </tr>
      </informaltable>

      <section xml:id="sect_PS3.5_D.3" label="D.3" status="2">
        <title>Examples of Float and Double Float Pixel Data</title>
        
        <para>Float Pixel Data having the Value Representation OF always has 32 bits allocated; the resulting byte streams pictured in <xref linkend="figure_PS3.5_D.3-1" xrefstyle="select: label"/> are as they would be transmitted across a network and/or stored on media.</para>
        <para>
          <figure xml:id="figure_PS3.5_D.3-1" label="D.3-1" pgwide="1">
            <title>Sample Float Pixel Data Byte Streams for VR = OF</title>
            <mediaobject>
              <imageobject>
                <imagedata fileref="sup172figures/PS3.5_D.3-1.svg"/>
              </imageobject>
            </mediaobject>
          </figure>
        </para>
        <para>Double Float Pixel Data having the Value Representation OD always has 64 bits allocated; the resulting byte streams pictured in <xref linkend="figure_PS3.5_D.3-2" xrefstyle="select: label"/> are as they would be transmitted across a network and/or stored on media.</para>
        <para>
          <figure xml:id="figure_PS3.5_D.3-2" label="D.3-2" pgwide="1">
            <title>Sample Float Pixel Data Byte Streams for VR = OD</title>
            <mediaobject>
              <imageobject>
                <imagedata fileref="sup172figures/PS3.5_D.3-2.svg"/>
              </imageobject>
            </mediaobject>
          </figure>
        </para>
      </section>
      
    </section>
    
  </chapter>
  
  
  <chapter label="DICOM PS3.6" xml:id="chapter_PS3.6">
    <title>Data Dictionary</title>
    <section label="6" xml:id="chapter_PS3.6_6">
      <title>Registry of DICOM Data Elements</title>
      <informaltable rules="all" frame="box">
        <tr>
          <td>
            <para><emphasis role="italic">Amend DICOM PS 3.6 - Data Dictionary - Section 6 - Registry of DICOM Data Elements as follows:</emphasis></para>
          </td>
        </tr>
      </informaltable>
      
      <table rules="all" frame="box" xml:id="table_6-1" label="6-1">
        <caption>Registry of DICOM Data Elements</caption>
        <thead>
          <tr valign="top">
            <th align="center" colspan="1" rowspan="1">
              <para>
                <emphasis role="bold">Tag</emphasis>
              </para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para>
                <emphasis role="bold">Name</emphasis>
              </para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para>
                <emphasis role="bold">Keyword</emphasis>
              </para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para>
                <emphasis role="bold">VR</emphasis>
              </para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para>
                <emphasis role="bold">VM</emphasis>
              </para>
            </th>
            <td align="center" colspan="1" rowspan="1">
              <para/>
            </td>
          </tr>
        </thead>
        <tbody>
          <tr valign="top">
            <td align="center" colspan="1" rowspan="1">
              <para>...</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>...</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>...</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>...</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>...</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para/>
            </td>
          </tr>
          <tr valign="top">
            <td align="center" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline"
                  >(0040,9220)</emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">Quantity Definition Sequence</emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">Quantity​Definition​Sequence</emphasis></emphasis></para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">SQ</emphasis></emphasis></para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">1</emphasis></emphasis></para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline"></emphasis></emphasis></para>
            </td>
          </tr>
          <tr valign="top">
            <td align="center" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline"
                  >(0040,9092)</emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">Parametric Map Frame Type Sequence</emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">Parametric&#x200b;​Map​&#x200b;Frame&#x200b;Type&#x200b;Sequence</emphasis></emphasis></para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">SQ</emphasis></emphasis></para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">1</emphasis></emphasis></para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline"></emphasis></emphasis></para>
            </td>
          </tr>
          <tr valign="top">
            <td align="center" colspan="1" rowspan="1">
              <para>...</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>...</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>...</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>...</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>...</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para/>
            </td>
          </tr>
          <tr valign="top">
            <td align="center" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline"
                  >(0028,0122)</emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">Float Pixel Padding Value</emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">Float&#x200b;Pixel&#x200b;Padding&#x200b;Value</emphasis></emphasis></para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">FL</emphasis></emphasis></para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">1</emphasis></emphasis></para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline"></emphasis></emphasis></para>
            </td>
          </tr>
          <tr valign="top">
            <td align="center" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline"
                  >(0028,0123)</emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">Double Float Pixel Padding Value</emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">Double&#x200b;Float&#x200b;Pixel&#x200b;Padding&#x200b;Value</emphasis></emphasis></para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">FD</emphasis></emphasis></para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">1</emphasis></emphasis></para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline"></emphasis></emphasis></para>
            </td>
          </tr>
          <tr valign="top">
            <td align="center" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline"
                  >(0028,0124)</emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">Float Pixel Padding Range Limit</emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">Float&#x200b;Pixel&#x200b;Padding&#x200b;Range&#x200b;Limit</emphasis></emphasis></para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">FL</emphasis></emphasis></para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">1</emphasis></emphasis></para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline"></emphasis></emphasis></para>
            </td>
          </tr>
          <tr valign="top">
            <td align="center" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline"
                  >(0028,0125)</emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">Double Float Pixel Padding Range Limit</emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">Double&#x200b;Float&#x200b;Pixel&#x200b;Padding&#x200b;Range&#x200b;Limit</emphasis></emphasis></para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">FD</emphasis></emphasis></para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">1</emphasis></emphasis></para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline"></emphasis></emphasis></para>
            </td>
          </tr>
          <tr valign="top">
            <td align="center" colspan="1" rowspan="1">
              <para>(7FE0,0010)</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Pixel Data</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Pixel​&#x200b;Data</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>OB or OW</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>1</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para/>
            </td>
          </tr>
          <tr valign="top">
            <td align="center" colspan="1" rowspan="1">
              <para>
                <emphasis role="bold"><emphasis role="underline">(7FE0,0008)</emphasis></emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>
                <emphasis role="bold"><emphasis role="underline">Float Pixel Data</emphasis></emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>
                <emphasis role="bold"><emphasis role="underline">Float​&#x200b;Pixel&#x200b;​Data</emphasis></emphasis>
              </para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>
                <emphasis role="bold"><emphasis role="underline">OF</emphasis></emphasis>
              </para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>
                <emphasis role="bold"><emphasis role="underline">1</emphasis></emphasis>
              </para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>
                <emphasis role="bold"><emphasis role="underline"></emphasis></emphasis>
              </para>
            </td>
          </tr>
          <tr valign="top">
            <td align="center" colspan="1" rowspan="1">
              <para>
                <emphasis role="bold"><emphasis role="underline">(7FE0,0009)</emphasis></emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>
                <emphasis role="bold"><emphasis role="underline">Double Float Pixel Data</emphasis></emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>
                <emphasis role="bold"><emphasis role="underline">Double&#x200b;Float&#x200b;Pixel​&#x200b;Data</emphasis></emphasis>
              </para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>
                <emphasis role="bold"><emphasis role="underline">OD</emphasis></emphasis>
              </para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>
                <emphasis role="bold"><emphasis role="underline">1</emphasis></emphasis>
              </para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>
                <emphasis role="bold"><emphasis role="underline"></emphasis></emphasis>
              </para>
            </td>
          </tr>
          <tr valign="top">
            <td align="center" colspan="1" rowspan="1">
              <para>
                <emphasis role="italic">(7FE0,0020)</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>
                <emphasis role="italic">Coefficients SDVN</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>
                <emphasis role="italic">Coefficients​SDVN</emphasis>
              </para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>
                <emphasis role="italic">OW</emphasis>
              </para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>
                <emphasis role="italic">1</emphasis>
              </para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>
                <emphasis role="italic">RET</emphasis>
              </para>
            </td>
          </tr>
        </tbody>
      </table>
    </section>
    
    <section label="A" xml:id="chapter_PS3.6_A">
      <title>Registry of DICOM Unique Identifiers (UIDs) (Normative)</title>
      <informaltable rules="all" frame="box">
        <tr>
          <td>
            <para><emphasis role="italic">Amend DICOM PS 3.6 - Data Dictionary - Annex A - Registry of DICOM Unique Identifiers (UIDs) as follows:</emphasis></para>
          </td>
        </tr>
      </informaltable>
      
      <table rules="all" frame="box" xml:id="table_A-1" label="A-1">
        <caption>UID Values</caption>
        <thead>
          <tr valign="top">
            <th align="center" colspan="1" rowspan="1">
              <para>
                <emphasis role="bold">UID Value</emphasis>
              </para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para>
                <emphasis role="bold">UID NAME</emphasis>
              </para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para>
                <emphasis role="bold">UID TYPE</emphasis>
              </para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para>
                <emphasis role="bold">Part</emphasis>
              </para>
            </th>
          </tr>
        </thead>
        <tbody>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>...</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>...</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>...</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>...</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">1.2.840.10008.5.1.4.1.1.30</emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">Parametric Map Storage</emphasis></emphasis></para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">SOP Class</emphasis></emphasis></para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">PS 3.4</emphasis></emphasis></para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>...</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>...</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>...</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>...</para>
            </td>
          </tr>
        </tbody>
      </table>
    </section>
    
    <section>
      <title/>
      <table rules="all" frame="box" xml:id="table_A-3" label="A-3">
        <caption>Context Group UID Values</caption>
        <thead>
          <tr valign="top">
            <th align="center" colspan="1" rowspan="1">
              <para>
                <emphasis role="bold">Context UID</emphasis>
              </para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para>
                <emphasis role="bold">Context Identifier</emphasis>
              </para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para>
                <emphasis role="bold">Context Group Name</emphasis>
              </para>
            </th>
          </tr>
        </thead>
        <tbody>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">1.2.840.10008.6.1.1009</emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">7222</emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">Parametric Map Derivation Image Purpose of Reference</emphasis></emphasis></para>
            </td>
          </tr>
        </tbody>
      </table>
    </section>
    
  </chapter>
  
  <chapter label="DICOM PS3.16" xml:id="chapter_PS3.16">
    <title>Content Mapping Resource</title>
    
    <section label="B" xml:id="chapter_PS3.16_B">
      <title>DCMR Context Groups (Normative)</title>
      
      <section>
        <title/>
        <informaltable rules="all" frame="box">
          <tr>
            <td>
              <para><emphasis role="italic">Amend DICOM PS 3.16 - Content Mapping Resource - Annex B - DCMR Context Groups (Normative) as follows, to add new Context Groups for Parametric Maps:</emphasis></para>
            </td>
          </tr>
        </informaltable>
      </section>
      
      <section label="CID 7222" status="3" xml:id="sect_CID_7222">
        <title>Parametric Map Derivation Image Purpose of Reference</title>
        <variablelist>
          <varlistentry>
            <term>
              <emphasis role="bold">Type:</emphasis>
            </term>
            <listitem>
              <para>
                <emphasis role="bold">Extensible</emphasis>
              </para>
            </listitem>
          </varlistentry>
          <varlistentry>
            <term>
              <emphasis role="bold">Version:</emphasis>
            </term>
            <listitem>
              <para>
                <emphasis role="bold">20141110</emphasis>
              </para>
            </listitem>
          </varlistentry>
        </variablelist>
        <table rules="all" frame="box" label="CID 7222" xml:id="table_CID_7222">
          <caption>Parametric Map Derivation Image Purpose of Reference</caption>
          <thead>
            <tr valign="top">
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold">Coding Scheme Designator</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold">Code Value</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold">Code Meaning</emphasis>
                </para>
              </th>
            </tr>
          </thead>
          <tbody>
            <tr valign="top">
              <td align="center" colspan="1" rowspan="1">
                <para>DCM</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>121322</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>Source Image for Image Processing Operation</para>
              </td>
            </tr>
          </tbody>
        </table>
      </section>
      
      
    </section>
    
  </chapter>
  
  <chapter label="DICOM PS3.17" xml:id="chapter_PS3.17">
    <title>Explanatory Information</title>
    <section>
      <title/>
      <informaltable rules="all" frame="box">
        <tr>
          <td>
            <para><emphasis role="italic">Amend DICOM PS 3.17 - Explanatory Information to add new Annex QQQ - Parametric Maps as follows:</emphasis></para>
          </td>
        </tr>
      </informaltable>
      
      <section label="QQQ" status="1" xml:id="PS3.17_QQQ">
        <title>Parametric Maps (Informative)</title>
        <section>
          <title/>
          <informaltable rules="all" frame="box">
            <tr>
              <td>
                <para><emphasis role="italic">Amend DICOM PS 3.17 - Explanatory Information to add new Annex XXX:</emphasis></para>
              </td>
            </tr>
          </informaltable>
        </section>
        <section label="QQQ.1" status="2" xml:id="sect_PS3.17_QQQ.1">
          <title></title>
          <para>This Annex contains examples of the use of the Parametric Map IOD.</para>
          <section label="QQQ.1.1" status="3" xml:id="sect_PS3.17_QQQ.1.1">
            <title></title>
            <para>This Section contains an example of the use of the Parametric Map IOD to encode
              Ktrans for a Dynamic Contrast Enhanced (DCE) MR.</para>
            <para>The frames comprise a single traversal of a regularly sampled 3D volume, described
              as a single stack and a single quantity, with dimensions of Stack ID, In-Stack
              Position Number and Quantity. A reference is also provided to the (single entire
              multi-frame) MR image from which the parametric map was derived. Only the Frame
              Content Sequence and Plane Position Sequence vary per-frame; all other functional
              groups are shared in this example.</para>
            <informaltable frame="box" rules="all">
              <thead>
                <tr valign="top">
                  <th align="center" colspan="1" rowspan="1">
                    <para>
                      <emphasis role="bold">Nesting</emphasis>
                    </para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para>
                      <emphasis role="bold">Attribute</emphasis>
                    </para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para>
                      <emphasis role="bold">Tag</emphasis>
                    </para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para>
                      <emphasis role="bold">VR</emphasis>
                    </para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para>
                      <emphasis role="bold">VL (hex)</emphasis>
                    </para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para>
                      <emphasis role="bold">Value</emphasis>
                    </para>
                  </th>
                </tr>
              </thead>
              <tbody>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Specific Character Set</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,0005)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>CS</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>000a</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>ISO_IR 100</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Image Type</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,0008)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>CS</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0022</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>DERIVED\SECONDARY\PERFUSION\KTRANS</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Instance Creation Date</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,0012)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>DA</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0008</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>20140312</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Instance Creation Time</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,0013)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>TM</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>000a</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>141900.944</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Instance Creator UID</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,0014)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>UI</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0016</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>1.3.6.1.4.1.5962.99.3</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>SOP Class UID</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,0016)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>UI</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0016</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>1.3.6.1.4.1.5962.301.9</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>SOP Instance UID</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,0018)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>UI</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>003e</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>1.3.6.1.4.1.5962.99.1.3078904268.1788845519.1394648340940.2.0</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Study Date</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,0020)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>DA</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0008</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>20140312</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Series Date</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,0021)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>DA</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0008</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>20140312</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Content Date</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,0023)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>DA</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0008</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>20140312</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Study Time</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,0030)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>TM</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>000a</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>141900.944</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Series Time</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,0031)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>TM</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>000a</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>141900.944</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Content Time</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,0033)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>TM</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>000a</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>141900.944</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Accession Number</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,0050)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>SH</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0000</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para></para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Modality</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,0060)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>CS</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0002</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>MR</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Manufacturer</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,0070)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>LO</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0000</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para></para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Referring Physician's Name</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,0090)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>PN</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0008</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Doe^John</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Study Description</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,1030)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>LO</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>002C</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Dynamic magnetic resonance imaging of pelvis</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Procedure Code Sequence</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,1032)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>SQ</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>ffffffff</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para></para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>%item</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Code Value</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,0100)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>SH</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0008</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>P5-70694</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Coding Scheme Designator</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,0102)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>SH</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0004</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>SRT</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Code Meaning</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,0104)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>LO</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>002C</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Dynamic magnetic resonance imaging of pelvis</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>%enditem</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>%endseq</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Series Description</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,103E)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>LO</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0010</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>PK Model Results</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Patient's Name</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0010,0010)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>PN</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0008</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Doe^Jane</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Patient ID</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0010,0020)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>LO</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0004</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>1234</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Patient's Birth Date</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0010,0030)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>DA</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0000</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para></para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Patient's Sex</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0010,0040)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>CS</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0000</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para></para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Study Instance UID</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0020,000d)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>UI</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>003e</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>1.3.6.1.4.1.5962.99.1.3078904268.1788845519.1394648340940.4.0</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Series Instance UID</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0020,000e)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>UI</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>003e</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>1.3.6.1.4.1.5962.99.1.3078904268.1788845519.1394648340940.3.0</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Study ID</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0020,0010)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>SH</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0004</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>5678</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Series Number</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0020,0011)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>IS</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0004</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>100</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Instance Number</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0020,0013)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>IS</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0002</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>1</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Patient Orientation</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0020,0020)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>CS</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0000</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para></para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Frame of Reference UID</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0020,0052)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>UI</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>003e</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>1.3.6.1.4.1.5962.99.1.3078904268.1788845519.1394648340940.5.0</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Position Reference Indicator</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0020,1040)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>LO</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0000</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para></para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Dimension Organization Sequence</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0020,9221)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>SQ</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>ffffffff</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para></para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>%item</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Dimension Organization UID</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0020,9164)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>UI</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>003e</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>1.3.6.1.4.1.5962.99.1.3078904268.1788845519.1394648340940.6.0</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>%enditem</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>%endseq</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Dimension Index Sequence</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0020,9222)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>SQ</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>ffffffff</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para></para>
                  </td>
                </tr>

                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>%item</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Dimension Organization UID</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0020,9164)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>UI</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>003e</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>1.3.6.1.4.1.5962.99.1.3078904268.1788845519.1394648340940.6.0</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Dimension Index Pointer</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0020,9165)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>AT</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0004</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>(0020,9056)</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Functional Group Pointer</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0020,9167)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>AT</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0004</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>(0040,9096)</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Dimension Description Label</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0020,9421)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>LO</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0008</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Stack ID</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>%enditem</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>%item</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Dimension Organization UID</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0020,9164)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>UI</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>003e</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>1.3.6.1.4.1.5962.99.1.3078904268.1788845519.1394648340940.6.0</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Dimension Index Pointer</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0020,9165)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>AT</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0004</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>(0020,9057)</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Functional Group Pointer</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0020,9167)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>AT</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0004</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>(0020,9111)</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Dimension Description Label</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0020,9421)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>LO</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0018</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>In-Stack Position Number</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>%enditem</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>%item</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Dimension Organization UID</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0020,9164)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>UI</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>003e</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>1.3.6.1.4.1.5962.99.1.3078904268.1788845519.1394648340940.6.0</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Dimension Index Pointer</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0020,9165)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>AT</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0004</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>(0040,9220) (the Quantity Definition Code Sequence (CP 1387))</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Functional Group Pointer</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0020,9167)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>AT</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0004</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>(0020,9111)</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Dimension Description Label</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0020,9421)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>LO</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0008</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Quantity</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>%enditem</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>%endseq</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>

                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Samples per Pixel</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0028,0002)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>US</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0002</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0001</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Photometric Interpretation</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0028,0004)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>CS</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>000c</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>MONOCHROME2</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Number of Frames</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0028,0008)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>IS</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0002</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>14</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Rows</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0028,0010)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>US</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0002</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>256 dec</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Columns</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0028,0011)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>US</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0002</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>256 dec</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Bits Allocated</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0028,0100)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>US</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0002</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0020 hex</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Burned In Annotation</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0028,0301)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>CS</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0002</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>NO</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Recognizable Visual Features</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0028,0302)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>CS</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0002</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>NO</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Lossy Image Compression</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0028,2110)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>CS</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0002</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>00</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Presentation LUT Shape</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(2050,0020)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>CS</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0008</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>IDENTITY</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Shared Functional Groups Sequence</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(5200,9229)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>SQ</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>ffffffff</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para></para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>%item</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>
                
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Derivation Image Sequence</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,9124)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>SQ</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>ffffffff</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para></para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>%item</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Source Image Sequence</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,2112)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>SQ</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>ffffffff</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para></para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>%item</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>
                
                
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Referenced SOP Class UID</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,1150)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>UI</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>001C</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>1.2.840.10008.5.1.4.1.1.4.1</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Referenced SOP Instance UID</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,1155)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>UI</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>002E</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>1.3.6.1.4.1.5962.1.1.0.0.0.1410021852.13877.0</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Spatial Locations Preserved</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0028,135A)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>CS</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0004</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>YES</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Purpose of Reference Code Sequence</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0040,A170)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>SQ</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>ffffffff</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para></para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>%item</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Code Value</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,0100)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>SH</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0006</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>121322</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Coding Scheme Designator</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,0102)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>SH</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0004</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>DCM</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Code Meaning</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,0104)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>LO</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>002C</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Source image for image processing operation</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>%enditem</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>%endseq</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>%enditem</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>%endseq</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>
                
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Derivation Code Sequence</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,9215)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>SQ</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>ffffffff</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para></para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>%item</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Code Value</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,0100)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>SH</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0006</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>113066</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Coding Scheme Designator</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,0102)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>SH</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0004</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>DCM</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Code Meaning</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,0104)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>LO</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>000C</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Time Course of Signal</para>
                  </td>
                </tr>
                         
                        
                        
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>%enditem</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>%endseq</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>%enditem</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>%endseq</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>
                

                                
                                
                                
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Frame Anatomy Sequence</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0020,9071)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>SQ</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>ffffffff</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para></para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>%item</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Anatomic Region Sequence</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,2218)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>SQ</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>ffffffff</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para></para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>%item</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Code Value</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,0100)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>SH</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0008</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>T-9200B </para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Coding Scheme Designator</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,0102)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>SH</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0004</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>SRT</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Code Meaning</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,0104)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>LO</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0008</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Prostate</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>%enditem</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>%endseq</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Frame Laterality</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0020,9072)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>CS</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0002</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>U </para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>%enditem</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>%endseq</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>
                
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Plane Orientation Sequence</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0020,9116)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>SQ</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>ffffffff</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para></para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>%item</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Image Orientation (Patient)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0020,0037)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>DS</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0064</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0.99979773312597\​-.0160528955995\​.012115996823878\​.012116000683426\​0.96149705857037\​.274548008348208</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>%enditem</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>%endseq</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>                
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Pixel Measures Sequence</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0028,9110)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>SQ</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>ffffffff</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para></para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>%item</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Slice Thickness</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0018,0050)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>DS</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0010</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>5.9999942779541</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Pixel Spacing</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0028,0030)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>DS</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0022</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>1.01559996604919\1.01560020446777</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>%enditem</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>%endseq</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Frame VOI LUT Sequence</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0028,9132)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>SQ</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>ffffffff</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para></para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>%item</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Window Center</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0028,1050)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>DS</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0004</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>2.5</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Window Width</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0028,1051)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>DS</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0002</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>5</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>VOI LUT Function</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0028,1056)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>CS</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>000c</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>LINEAR_EXACT</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>%enditem</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>%endseq</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Pixel Value Transformation Sequence</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0028,9145)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>SQ</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>ffffffff</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para></para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>%item</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Rescale Intercept</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0028,1052)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>DS</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0002</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Rescale Slope</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0028,1053)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>DS</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0002</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>1</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Rescale Type</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0028,1054)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>LO</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0002</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>US</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>%enditem</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>%endseq</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Real World Value Mapping Sequence</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0040,9096)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>SQ</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>ffffffff</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para></para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>%item</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>LUT Explanation</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0028,3003)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>LO</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0006</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Ktrans</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Measurement Units Code Sequence</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0040,08ea)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>SQ</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>ffffffff</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para></para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>%item</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Code Value</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,0100)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>SH</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0004</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>/min</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Coding Scheme Designator</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,0102)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>SH</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0004</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>UCUM</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Code Meaning</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,0104)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>LO</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0004</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>/min</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>%enditem</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>%endseq</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>LUT Label</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0040,9210)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>SH</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0006</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Ktrans</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Real World Value Last Value Mapped</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0040,9211)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>XS</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0002</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0005</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Real World Value First Value Mapped</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0040,9216)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>XS</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0002</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0000</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Real World Value Intercept</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0040,9224)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>FD</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0008</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Real World Value Slope</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0040,9225)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>FD</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0008</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>1</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Quantity Definition Code Sequence (CP 1387)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0040,9220)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>SQ</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>ffffffff</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para></para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>%item</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Value Type</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0040,a040)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>CS</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0004</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>CODE</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Concept Name Code Sequence</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0040,a043)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>SQ</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>ffffffff</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para></para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>%item</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Code Value</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,0100)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>SH</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0006</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>G-C1C6</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Coding Scheme Designator</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,0102)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>SH</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0004</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>SRT</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Code Meaning</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,0104)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>LO</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0008</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Quantity</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>%enditem</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>%endseq</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Concept Code Sequence</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0040,a168)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>SQ</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>ffffffff</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para></para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>%item</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Code Value</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,0100)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>SH</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0006</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>dd2d60 (CP1391)</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Coding Scheme Designator</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,0102)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>SH</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0004</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>DCM </para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Code Meaning</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,0104)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>LO</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0006</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Ktrans</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>%enditem</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>%endseq</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>%enditem</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>%item</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Value Type</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0040,a040)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>CS</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0004</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>CODE</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Concept Name Code Sequence</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0040,a043)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>SQ</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>ffffffff</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para></para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>%item</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Code Value</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,0100)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>SH</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0006</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>G-C036</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Coding Scheme Designator</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,0102)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>SH</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0004</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>SRT</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Code Meaning</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,0104)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>LO</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0012</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Measurement Method</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>%enditem</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>%endseq</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Concept Code Sequence</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0040,a168)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>SQ</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>ffffffff</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para></para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>%item</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Code Value</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,0100)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>SH</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0006</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>dd2d71 (CP1391)</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Coding Scheme Designator</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,0102)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>SH</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0004</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>DCM </para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Code Meaning</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,0104)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>LO</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0014</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Standard Tofts Model</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>%enditem</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>%endseq</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>%enditem</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>%endseq</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>%enditem</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>%endseq</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>


                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Parametric Map Frame Type Sequence</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0040,9092)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>SQ</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>ffffffff</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para></para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>%item</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Frame Type</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,9007)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>CS</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0022</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>DERIVED\SECONDARY\PERFUSION\KTRANS</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>%enditem</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>%endseq</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>
                



                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>%enditem</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>%endseq</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Per-frame Functional Groups Sequence</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(5200,9230)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>SQ</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>ffffffff</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para></para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>%item</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Frame Content Sequence</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0020,9111)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>SQ</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>ffffffff</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para></para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>%item</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Stack ID</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0020,9056)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>SH</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0002</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>1</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>In-Stack Position Number</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0020,9057)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>UL</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0004</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0001</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Dimension Index Values</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0020,9157)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>UL</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>000C</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>00000001,00000001,00000001</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>%enditem</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>%endseq</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Plane Position Sequence</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0020,9113)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>SQ</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>ffffffff</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para></para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>%item</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Image Position (Patient)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0020,0032)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>DS</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>0032</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>-153.28300476074\-111.93399810791\-54.366100311279</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>%enditem</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>%endseq</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>%enditem</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>%</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>...</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>%endseq</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Float Pixel Data</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(7FE0,0008)</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>OF</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>380000</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>[]</para>
                  </td>
                </tr>
              </tbody>
            </informaltable>
          </section>
        </section>
        
        
      </section>
    </section>
  </chapter>
  
  
  
</book>