<book xmlns="http://docbook.org/ns/docbook" xmlns:xl="http://www.w3.org/1999/xlink" version="5.0">
  <title>CP-934</title>
  <subtitle>Fix Inconsistency of C-FIND only Number of Patient Related XXX attributes</subtitle>
  <chapter>
    <title/>
    <informaltable rules="all" frame="box">
      <tr>
        <td rowspan="1" colspan="1" valign="top">
          <para>Status</para>
        </td>
        <td rowspan="1" colspan="1" valign="top">
          <para>Final Text</para>
        </td>
      </tr>
      <tr>
        <td rowspan="1" colspan="1" valign="top">
          <para>Date of Last Update</para>
        </td>
        <td rowspan="1" colspan="1" valign="top">
          <para>2015/06/02</para>
        </td>
      </tr>
      <tr>
        <td rowspan="1" colspan="1" valign="top">
          <para>Person Assigned</para>
        </td>
        <td rowspan="1" colspan="1" valign="top">
          <para>David Clunie</para>
          <para><link xmlns:xlink="http://www.w3.org/1999/xlink"
              xlink:href="mailto:dclunie@dclunie.com"/></para>
        </td>
      </tr>
      <tr>
        <td rowspan="1" colspan="1" valign="top">
          <para>Submitter Name</para>
        </td>
        <td rowspan="1" colspan="1" valign="top">
          <para>David Harvey</para>
          <para><link xmlns:xlink="http://www.w3.org/1999/xlink"
            xlink:href="mailto:dave@medicalconnections.co.uk"/></para>
        </td>
      </tr>
      <tr>
        <td rowspan="1" colspan="1" valign="top">
          <para>Submission Date</para>
        </td>
        <td rowspan="1" colspan="1" valign="top">
          <para>2008/08/28</para>
        </td>
      </tr>
    </informaltable>
    <informaltable rules="all" frame="box">
      <tr>
        <td rowspan="1" colspan="1" valign="top">
          <para>Correction Number CP-934</para>
        </td>
      </tr>
      <tr>
        <td rowspan="1" colspan="1" valign="top">
          <para>Log Summary: Fix Inconsistency of C-FIND only Number of Patient Related XXX attributes</para>
        </td>
      </tr>
      <tr>
        <td rowspan="1" colspan="1" valign="top">
          <para>Name of Standard</para>
          <para>PS3.4 2015b</para>
        </td>
      </tr>
      <tr>
        <td rowspan="1" colspan="1" valign="top"><para>Rationale for Correction:</para>
          <para>PS3.4 is inconsistent about whether or not the following attributes are allowed in a study root, study level request.</para>
          <itemizedlist>
            <listitem><para>Number of Patient Related Studies (0020,1200)</para></listitem>
            <listitem><para>Number of Patient Related Series (0020,1202)</para></listitem>
            <listitem><para>Number of Patient Related Instances (0020,1204)</para></listitem>
          </itemizedlist>
          <para>Table C.3-1 states that they represent "The number of (studies/series/composite object instances) that match the Patient level Query/Retrieve search criteria",
            i.e. explicitly tying them to the PATIENT level, but Table C.6-5 lists them as available (optional) for Study Root, Study Level queries.</para>
          <para>Remove the attributes from the Study Root Study Level table.</para>
        </td>
      </tr>
      <tr>
        <td rowspan="1" colspan="1" valign="top">Correction Wording:</td>
      </tr>
    </informaltable>
  </chapter>
  <chapter>
    <title/>
    <section>
      <title/>
      <informaltable rules="all" frame="box">
        <tr>
          <td>
            <para><emphasis role="italic">Amend DICOM PS3.4 as follows:</emphasis></para>
          </td>
        </tr>
      </informaltable>
    </section>
    
    <section label="C.3.4" status="3" xml:id="sect_C.3.4">
      <title>Additional Query/Retrieve Attributes</title>
      <para>Some optional Attributes that may be used in Query/Retrieve Information Models that are not Attributes of an Information Object Definition and, therefore, are not defined in PS3.3. These Attributes are defined in <xref linkend="table_C.3-1" xrefstyle="select: label"/>.</para>
      <table frame="box" label="C.3-1" rules="all" xml:id="table_C.3-1">
        <caption>Additional Query/Retrieve Attributes</caption>
        <thead>
          <tr valign="top">
            <th align="center" colspan="1" rowspan="1">
              <para>
                <emphasis role="bold">Attribute Name</emphasis>
              </para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para>
                <emphasis role="bold">Tag</emphasis>
              </para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para>
                <emphasis role="bold">Attribute Description</emphasis>
              </para>
            </th>
          </tr>
        </thead>
        <tbody>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>Number of Patient Related Studies</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>(0020,1200)</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>The number of studies that match the Patient level Query/Retrieve search criteria</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>Number of Patient Related Series</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>(0020,1202)</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>The number of series that match the Patient level Query/Retrieve search criteria</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>Number of Patient Related Instances</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>(0020,1204)</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>The number of Composite Object Instances that match the Patient level Query/Retrieve search criteria</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>Number of Study Related Series</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>(0020,1206)</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>The number of series that match the Study level Query/Retrieve search criteria</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>Number of Series Related Instances</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>(0020,1209)</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>The number of Composite Object Instances in a Series that match the Series level Query/Retrieve search criteria</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>Number of Study Related Instances</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>(0020,1208)</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>The number of Composite Object Instances that match the Study level Query/Retrieve search criteria</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>Modalities in Study</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>(0008,0061)</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>All of the distinct values used for Modality (0008,0060) in the Series of the Study.</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>SOP Classes in Study</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>(0008,0062)</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>The SOP Classes contained in the Study.</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>Alternate Representation Sequence</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>(0008,3001)</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>A Sequence of Items, each identifying an alternate encoding of an image that matches the Instance level Query/Retrieve search criteria (see Section C.6.1.1.5.1)</para>
            </td>
          </tr>
        </tbody>
      </table>
      <para>If the SCP manages images in multiple alternate encodings, only one of the alternate encodings of an image is included in the number of object instances.</para>
      <para>...</para>
      <para/>
    </section>
    
              <section label="C.6.1.1.2" status="5" xml:id="sect_C.6.1.1.2">
            <title>Patient Level</title>
            <para><xref linkend="table_C.6-1" xrefstyle="select: label"/> defines the Attributes at the Patient Query/Retrieve level of the Patient Root Query/Retrieve Information Model.</para>
            <note>
              <orderedlist numeration="arabic">
                <listitem>
                  <para>A description of the Attributes of this Information Model is contained in Section C.3 of this part.</para>
                </listitem>
                <listitem>
                  <para>Although the Patient ID may not be globally unique, the Study Instance UID is globally unique ensuring that no two studies may be misidentified.</para>
                </listitem>
              </orderedlist>
            </note>
            <table frame="box" label="C.6-1" rules="all" xml:id="table_C.6-1">
              <caption>Patient Level Attributes for the Patient Root Query/Retrieve Information Model</caption>
              <thead>
                <tr valign="top">
                  <th align="center" colspan="1" rowspan="1">
                    <para>
                      <emphasis role="bold">Attribute Name</emphasis>
                    </para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para>
                      <emphasis role="bold">Tag</emphasis>
                    </para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para>
                      <emphasis role="bold">Type</emphasis>
                    </para>
                  </th>
                </tr>
              </thead>
              <tbody>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>Patient's Name</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0010,0010)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>R</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>Patient ID</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0010,0020)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>U</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>Issuer of Patient ID</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0010,0021)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>O</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>Referenced Patient Sequence</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,1120)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>O</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>&gt;Referenced SOP Class UID</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,1150)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>O</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>&gt;Referenced SOP Instance UID</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,1155)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>O</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>Patient's Birth Date</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0010,0030)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>O</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>Patient's Birth Time</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0010,0032)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>O</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>Patient's Sex</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0010,0040)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>O</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>Other Patient Ids</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0010,1000)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>O</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>Other Patient Names</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0010,1001)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>O</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>Ethnic Group</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0010,2160)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>O</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>Patient Comments</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0010,4000)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>O</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>Number of Patient Related Studies</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0020,1200)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>O</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>Number of Patient Related Series</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0020,1202)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>O</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>Number of Patient Related Instances</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0020,1204)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>O</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para><emphasis role="italic">All other Attributes at Patient Level</emphasis></para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="center" colspan="1" rowspan="1">
                    <para>O</para>
                  </td>
                </tr>
              </tbody>
            </table>
          </section>
          <section label="C.6.1.1.3" status="5" xml:id="sect_C.6.1.1.3">
            <title>Study Level</title>
            <para><xref linkend="table_C.6-2" xrefstyle="select: label"/> defines the keys at the Study Information level of the Patient Root Query/Retrieve Information Model.</para>
            <note>
              <orderedlist numeration="arabic">
                <listitem>
                  <para>A description of the Attributes of this Information Model is contained in Section C.3 of this Part.</para>
                </listitem>
                <listitem>
                  <para>Although the Patient ID may not be globally unique, the Study Instance UID is globally unique ensuring that no two studies may be misidentified.</para>
                </listitem>
              </orderedlist>
            </note>
            <table frame="box" label="C.6-2" rules="all" xml:id="table_C.6-2">
              <caption>Study Level Keys for the Patient Root Query/Retrieve Information Model</caption>
              <thead>
                <tr valign="top">
                  <th align="center" colspan="1" rowspan="1">
                    <para>
                      <emphasis role="bold">Attribute Name</emphasis>
                    </para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para>
                      <emphasis role="bold">Tag</emphasis>
                    </para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para>
                      <emphasis role="bold">Type</emphasis>
                    </para>
                  </th>
                </tr>
              </thead>
              <tbody>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>Study Date</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,0020)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>R</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>Study Time</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,0030)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>R</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>Accession Number</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,0050)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>R</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>Study ID</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0020,0010)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>R</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>Study Instance UID</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0020,000D)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>U</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>Modalities in Study</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,0061)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>O</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>SOP Classes in Study</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,0062)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>O</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>Referring Physician's Name</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,0090)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>O</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>Study Description</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,1030)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>O</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>Procedure Code Sequence</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,1032)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>O</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>&gt;Code Value</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,0100)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>O</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>&gt;Coding Scheme Designator</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,0102)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>O</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>&gt;Coding Scheme Version</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,0103)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>O</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>&gt;Code Meaning</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,0104)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>O</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>Name of Physician(s) Reading Study</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,1060)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>O</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>Admitting Diagnoses Description</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,1080)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>O</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>Referenced Study Sequence</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,1110)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>O</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>&gt;Referenced SOP Class UID</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,1150)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>O</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>&gt;Referenced SOP Instance UID</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,1155)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>O</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>Patient's Age</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0010,1010)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>O</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>Patient's Size</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0010,1020)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>O</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>Patient's Weight</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0010,1030)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>O</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>Occupation</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0010,2180)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>O</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>Additional Patient History</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0010,21B0)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>O</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>Other Study Numbers</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0020,1070)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>O</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>Number of Study Related Series</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0020,1206)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>O</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>Number of Study Related Instances</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0020,1208)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>O</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para><emphasis role="italic">All other Attributes at Study Level</emphasis></para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="center" colspan="1" rowspan="1">
                    <para>O</para>
                  </td>
                </tr>
              </tbody>
            </table>
          </section>

    
    <section label="C.6.2.1.2" status="5" xml:id="sect_C.6.2.1.2">
            <title>Study Level</title>
            <para><xref linkend="table_C.6-5" xrefstyle="select: label"/> defines the keys at the Study Information level of the Study Root Query/Retrieve Information Model.</para>
            <note>
              <orderedlist numeration="arabic">
                <listitem>
                  <para>A description of the Attributes of this Information Model is contained in Section C.3.</para>
                </listitem>
                <listitem>
                  <para>Although the Patient ID may not be globally unique, the Study Instance UID is globally unique ensuring that no two studies may be misidentified.</para>
                </listitem>
              </orderedlist>
            </note>
            <table frame="box" label="C.6-5" rules="all" xml:id="table_C.6-5">
              <caption>Study Level Keys for the Study Root Query/Retrieve Information Model</caption>
              <thead>
                <tr valign="top">
                  <th align="center" colspan="1" rowspan="1">
                    <para>
                      <emphasis role="bold">Attribute Name</emphasis>
                    </para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para>
                      <emphasis role="bold">Tag</emphasis>
                    </para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para>
                      <emphasis role="bold">Type</emphasis>
                    </para>
                  </th>
                </tr>
              </thead>
              <tbody>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>Study Date</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,0020)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>R</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>Study Time</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,0030)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>R</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>Accession Number</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,0050)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>R</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>Patient's Name</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0010,0010)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>R</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>Patient ID</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0010,0020)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>R</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>Study ID</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0020,0010)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>R</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>Study Instance UID</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0020,000D)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>U</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>Modalities in Study</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,0061)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>O</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>SOP Classes in Study</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,0062)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>O</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>Referring Physician's Name</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,0090)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>O</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>Study Description</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,1030)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>O</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>Procedure Code Sequence</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,1032)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>O</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>&gt;Code Value</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,0100)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>O</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>&gt;Coding Scheme Designator</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,0102)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>O</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>&gt;Coding Scheme Version</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,0103)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>O</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>&gt;Code Meaning</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,0104)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>O</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>Name of Physician(s) Reading Study</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,1060)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>O</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>Admitting Diagnoses Description</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,1080)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>O</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>Referenced Study Sequence</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,1110)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>O</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>&gt;Referenced SOP Class UID</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,1150)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>O</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>&gt;Referenced SOP Instance UID</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,1155)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>O</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>Referenced Patient Sequence</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,1120)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>O</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>&gt;Referenced SOP Class UID</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,1150)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>O</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>&gt;Referenced SOP Instance UID</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0008,1155)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>O</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>Issuer of Patient ID</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0010,0021)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>O</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>Patient's Birth Date</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0010,0030)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>O</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>Patient's Birth Time</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0010,0032)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>O</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>Patient's Sex</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0010,0040)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>O</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>Other Patient Ids</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0010,1000)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>O</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>Other Patient Names</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0010,1001)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>O</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>Patient's Age</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0010,1010)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>O</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>Patient's Size</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0010,1020)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>O</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>Patient's Weight</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0010,1030)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>O</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>Ethnic Group</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0010,2160)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>O</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>Occupation</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0010,2180)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>O</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>Additional Patient History</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0010,21B0)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>O</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>Patient Comments</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0010,4000)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>O</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>Other Study Numbers</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0020,1070)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>O</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para><emphasis role="bold"><emphasis role="strikethrough">Number of Patient Related
                    Studies</emphasis></emphasis></para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para><emphasis role="bold"><emphasis role="strikethrough"
                  >(0020,1200)</emphasis></emphasis></para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para><emphasis role="bold"><emphasis role="strikethrough"
                >O</emphasis></emphasis></para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para><emphasis role="bold"><emphasis role="strikethrough">Number of Patient Related
                    Series</emphasis></emphasis></para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para><emphasis role="bold"><emphasis role="strikethrough"
                  >(0020,1202)</emphasis></emphasis></para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para><emphasis role="bold"><emphasis role="strikethrough"
                >O</emphasis></emphasis></para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para><emphasis role="bold"><emphasis role="strikethrough">Number of Patient Related
                    Instances</emphasis></emphasis></para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para><emphasis role="bold"><emphasis role="strikethrough"
                  >(0020,1204)</emphasis></emphasis></para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para><emphasis role="bold"><emphasis role="strikethrough"
                >O</emphasis></emphasis></para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>Number of Study Related Series</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0020,1206)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>O</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>Number of Study Related Instances</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>(0020,1208)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>O</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para><emphasis role="italic">All other Attributes at Study Level</emphasis></para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="center" colspan="1" rowspan="1">
                    <para>O</para>
                  </td>
                </tr>
              </tbody>
            </table>
            <note>
              <para>The use of the word "Images" rather than "Composite Object Instances" is historical, and should not be taken to mean that Composite Object Instances of other than image type are not included in the number.</para>
            </note>
          </section>

    
  </chapter>
</book>