<book xmlns="http://docbook.org/ns/docbook" xmlns:xl="http://www.w3.org/1999/xlink" version="5.0">
    <title>CP-2459</title>
    <subtitle>More data elements to de-identify</subtitle>
    <chapter>
        <title></title>
        <informaltable rules="all" frame="box">
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Status</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>Final Text</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Date of Last Update</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>2025/07/09</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Person Assigned</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>David Clunie</para><para><link
                    xl:href="mailto:dclunie@dclunie.com"/></para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submitter Name</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>David Clunie</para><para><link
                    xl:href="mailto:dclunie@dclunie.com"/></para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submission Date</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>2024/10/23</para></td>
            </tr>
        </informaltable>
        <informaltable rules="all" frame="box">
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Correction Number CP-2459</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Log Summary: More data elements to de-identify</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">
                    <para>Name of Standard</para>
                    <para>PS3.15 2025b</para>
                </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">
                    <para>Rationale for Correction:</para>
                    <para>Some device related identifiers are not included (and should be retained only with retain device identifiers option:</para>
                    <para>(0018,1010) Secondary Capture Device ID</para>
                    <para>(0018,1011) Hardcopy Creation Device ID</para>
                </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">Correction Wording:</td>
            </tr>
        </informaltable>
    </chapter>
    <chapter>
        <title/>

      <informaltable rules="all" frame="box">
        <tr>
          <td>
            <para><emphasis role="italic">Amend DICOM PS3.15 as follows (changes to existing text
              are bold and <emphasis role="bold"><emphasis role="underline"
                  >underlined</emphasis></emphasis> for additions and <emphasis role="bold"
                  ><emphasis role="strikethrough">struckthrough</emphasis></emphasis> for
              removals):</emphasis></para>
          </td>
        </tr>
      </informaltable>
        
       <section label="E.1.1" status="3" xml:id="sect_E.1.1">
        <title>De-identifier</title>
        
        <table frame="box" label="E.1-1" rules="all" xml:id="table_E.1-1">
            <caption>Application Level Confidentiality Profile Attributes</caption>
            <thead>
                <tr valign="top">
                    <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_c5e6e8ba-34e0-4e31-90b5-bd9ed7086ddd">
                            <emphasis role="bold">Attribute Name</emphasis>
                        </para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_d02d5580-8438-4a00-8c30-ba93530bcd31">
                            <emphasis role="bold">Tag</emphasis>
                        </para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_705936dc-2c42-4198-b406-ba0e61b35e99">
                            <emphasis role="bold">Retd. (from <olink targetdoc="PS3.6" targetptr="PS3.6" xrefstyle="select: labelnumber"/>)</emphasis>
                        </para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_a8387cac-386f-4ad9-96b3-b3dcb1b1010c">
                            <emphasis role="bold">In Std. Comp. IOD (from <olink targetdoc="PS3.3" targetptr="PS3.3" xrefstyle="select: labelnumber"/>)</emphasis>
                        </para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_fc0ff9a1-5780-4d06-861a-24ae0ba38b47">
                            <emphasis role="bold">Basic Prof.</emphasis>
                        </para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_19ddb67a-a5d6-474d-a13f-8df27d28b707">
                            <emphasis role="bold">Rtn. Safe Priv. Opt.</emphasis>
                        </para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_5b335431-5276-4cc0-bddf-93b6d107226e">
                            <emphasis role="bold">Rtn. UIDs Opt.</emphasis>
                        </para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_8317e0cf-0080-48f9-8040-1f49a4478c0f">
                            <emphasis role="bold">Rtn. Dev. Id. Opt.</emphasis>
                        </para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_1ba661a2-a572-420a-95c1-bc1bf694a70e">
                            <emphasis role="bold">Rtn. Inst. Id. Opt.</emphasis>
                        </para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_796841da-6b27-4558-99c9-99503edccc6c">
                            <emphasis role="bold">Rtn. Pat. Chars. Opt.</emphasis>
                        </para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_edac8181-290b-4f9c-945d-5dfc7b1a43fc">
                            <emphasis role="bold">Rtn. Long. Full Dates Opt.</emphasis>
                        </para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_6305752d-56e7-4f4d-b9d8-e5052fcd6eea">
                            <emphasis role="bold">Rtn. Long. Modif. Dates Opt.</emphasis>
                        </para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_84f3b128-8fe5-45bc-b04e-abc483377b06">
                            <emphasis role="bold">Clean Desc. Opt.</emphasis>
                        </para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_0adc5d55-1196-4265-8ee1-b205103c96f1">
                            <emphasis role="bold">Clean Struct. Cont. Opt.</emphasis>
                        </para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_5e74e547-b6fd-410e-9c09-b271915c3eeb">
                            <emphasis role="bold">Clean Graph. Opt.</emphasis>
                        </para>
                    </th>
                </tr>
            </thead>
            <tbody>
                <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_b5881e2a-62a6-4d6c-8b75-e0de9fb4f891">Device Alternate Identifier</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_5d3bc77a-8f1b-4072-a194-2a3aaa39bf47">(3010,001B)</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_19b26edb-7187-4461-8f7a-0e2dc3ac1be0">N</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_ba1d36f1-d411-42ea-983c-812795c034b3">Y</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_a6f152a2-03c3-4b0b-b6ba-a323c8207e75">Z</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_59cbbd3d-eadc-45d2-96e1-84f6940d55f2">Device Description</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_1bcc1d64-ecb7-42d4-9d0e-4574e9ceed36">(0050,0020)</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_f44b4a12-5d76-410b-b2fe-59e89104c785">N</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_2b5c32cf-6ee2-4b45-ad9d-2e477314fa09">Y</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_6bffb001-a05f-4a01-a0f3-d325819eb788">X</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_a290ca30-c265-47c1-8f51-7f0a48447b68">K</para>
                    </td>
                    <td/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_6ad8d7eb-0bde-4f01-8369-3a7cd7bf48e2">Device Label</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_2aced51b-017a-46a4-8ded-bf41b2fd2a40">(3010,002D)</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_53e1a3b6-3ccb-4c36-b586-493d8c621e9f">N</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_cf35a9a6-a2bd-491f-8696-655302a42d1d">Y</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_9c01acbb-ea50-4bbc-9fd7-ae89d86169d0">D</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_8f8118f9-7573-4785-9c24-8157a81462b5">K</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_ff4e909e-b2ea-4cc2-adc5-6fd23268d1d9">Device Serial Number</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_59c171d5-7e7a-47f9-90d5-e136701d867d">(0018,1000)</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_df0910f7-f2b8-4cfc-b488-6c674ebf511f">N</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_a17138ca-ff70-44d6-8ce5-a18094511f03">Y</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_1acb8726-68ae-4b7f-8787-0ef53bac9102">X/Z/D</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_498fa51e-aeea-4b47-86f9-5246a30d6013">K</para>
                    </td>
                    <td/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_44013946-d8ae-4a7e-aa41-08e5578086ca">Device Setting Description</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_9996ea1c-0e1d-4b4e-922e-c3fa2431f4fb">(0016,004B)</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_21dd9cb5-f835-48a6-b474-066b5399d2f9">N</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_c8d80c89-9593-4a15-bec6-098490d94db6">Y</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_6b041717-975e-4e94-a31f-fb54244f20bf">X</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_34a1fb2c-de39-4fd0-ba98-1d6665a0bea4">C</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_476dea1c-549a-4487-a181-5640902b285f">Device UID</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_48adb2ab-ec21-4333-a7d3-4f85eb94194a">(0018,1002)</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_bbc6a8fb-1f6f-4be8-9adf-2b3f8b09ca5d">N</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_0f90c0bc-2506-441f-98a5-d443f84aa413">Y</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_42727097-70f2-4517-b513-5687c45155dc">U</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_20626606-5c7f-4790-aee7-d86d1a658021">K</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_7640238b-d372-42f3-9eb4-0a43aa3fd792">K</para>
                    </td>
                    <td/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                        <para><emphasis role="bold"><emphasis role="underline">Hardcopy Creation Device ID</emphasis></emphasis></para>
                    </td>
                    <td align="center" colspan="1" rowspan="1"><emphasis role="bold"><emphasis role="underline">
                        <para>(0018,1011)</para>
                    </emphasis></emphasis></td>
                    <td align="center" colspan="1" rowspan="1">
                        <para><emphasis role="bold"><emphasis role="underline">Y</emphasis></emphasis></para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                        <para><emphasis role="bold"><emphasis role="underline">N</emphasis></emphasis></para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                        <para><emphasis role="bold"><emphasis role="underline">X</emphasis></emphasis></para>
                    </td>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1">
                        <para><emphasis role="bold"><emphasis role="underline">K</emphasis></emphasis></para>
                    </td>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_0b8d6659-fef0-4d8f-8d61-81225bc393a2">Manufacturer's Device Class UID</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_5f257738-9e5a-4e83-8bda-9419d5775b4e">(0018,100B)</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_eaddc3a2-6d24-40a9-bcae-6e0f973fa9b8">N</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_2722469d-38aa-4f4d-b689-76c09bb6d479">Y</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_3d8f49a3-7787-4bf5-b847-df2acd0d2cb2">U</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_f3f101e0-f091-4bbd-a5c6-0f6787cdad81">K</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_ce4296c5-6085-45a1-a12c-eab3361de6c9">K</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_e77fd488-4825-406b-971a-51a2b0e36c91">Manufacturer's Device Identifier</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_628cc8eb-e633-49bb-ad99-77238736c59e">(3010,0043)</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_965a5b8e-3ac3-4348-8517-232a862a723a">N</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_0f6e6437-b49e-495a-899d-fca5dc9205f4">Y</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_73ec787d-2e79-42aa-916d-694d61124ede">Z</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_b417dd34-5d08-4355-a448-76bd20ee9d8a">K</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                        <para><emphasis role="bold"><emphasis role="underline">Secondary Capture Device ID</emphasis></emphasis></para>
                    </td>
                    <td align="center" colspan="1" rowspan="1"><emphasis role="bold"><emphasis role="underline">
                                    <para>(0018,1010)</para>
                                </emphasis></emphasis></td>
                    <td align="center" colspan="1" rowspan="1">
                        <para><emphasis role="bold"><emphasis role="underline">N</emphasis></emphasis></para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                        <para><emphasis role="bold"><emphasis role="underline">Y</emphasis></emphasis></para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                        <para><emphasis role="bold"><emphasis role="underline">X</emphasis></emphasis></para>
                    </td>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1">
                        <para><emphasis role="bold"><emphasis role="underline">K</emphasis></emphasis></para>
                    </td>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                </tr>
            </tbody>
        </table>
       </section>
        
        <informaltable rules="all" frame="box">
            <tr>
                <td>
                    <para><emphasis role="italic">For reference unchanged DICOM PS3.3:</emphasis></para>
                </td>
            </tr>
        </informaltable>
        
        <section label="C.8.6.1" status="4" xml:id="sect_C.8.6.1">
                    <title>SC Equipment Module</title>
                    <table frame="box" label="C.8-24" rules="all" xml:id="table_C.8-24">
                        <caption>SC Equipment Module Attributes</caption>
                        <thead>
                            <tr valign="top">
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_a2430496-9bd6-4540-96d4-612d5ca217a6">Attribute Name</para>
                                </th>
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_6d583be2-8cac-4551-97a9-2aa6860df05c">Tag</para>
                                </th>
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_59bd6861-4b84-466c-b613-7f57e14975fe">Type</para>
                                </th>
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_7154b5b1-5af1-4fa0-b93d-c0d499d1ad4f">Attribute Description</para>
                                </th>
                            </tr>
                        </thead>
                        <tbody>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_2c74bf2a-c1d1-4e7a-89a3-1bab697e6b9e">...</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_c315a999-c8c3-4427-b9f0-23ced3f2f08d"/>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_cbbe1658-6c70-4c17-a833-735eaf2748b7"/>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_70f43095-d31b-41c6-95a1-211b8c084936"/>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_3ee3565b-403e-4493-b6f5-14eea6aeac69">Secondary Capture Device ID</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_c0e75f86-002b-4938-a8a9-89ab11e2525e">(0018,1010)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_5576634e-7d3d-4fc9-8aeb-f3431df0b9a4">3</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_0cd60521-4f23-47db-ba07-b9c7d94e96f3">User defined identification of the device that converted the image.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_8dfde123-6705-4871-a887-218ae4205c8f">Secondary Capture Device Manufacturer</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_64cce583-3e74-4abd-950b-d37bf829d360">(0018,1016)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_86e7788b-13bf-4787-9748-63a1de20ef6d">3</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_af2e32da-6707-49a8-bd37-fef7c875858f">Manufacturer of the Secondary Capture Device.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_3edd8353-d631-4ee8-94fa-4f2a593b7ece">Secondary Capture Device Manufacturer's Model Name</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_f9c05367-9b56-4a8a-be80-440a581c89dd">(0018,1018)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_0698b99f-28dd-46c7-9bd2-b9c8f4c63b1b">3</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_6b39f8f3-cb86-4b9f-98fb-2c24c05c84f1">Manufacturer's model number of the Secondary Capture Device.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_59cfe419-b317-4165-a650-df6f5a6754f0">Secondary Capture Device Software Versions</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_f2af799c-20a9-4ace-ae4c-088e43385d6f">(0018,1019)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_c04c2bc1-23ec-4296-970a-44a70d45e9f8">3</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_ac518335-eb19-4a7b-b156-42be986ab6da">Manufacturer's designation of software version of the Secondary Capture Device.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_9a4e6312-5680-4b1f-a4c5-90dac3e99cfb">...</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_cf5fcb60-e122-42a9-8da8-32332774722b"/>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_90f3919f-b8bb-42fb-a702-4ffb225bf86c"/>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_9d9bdd9c-b791-4918-ba54-7c2828e97933"/>
                                </td>
                            </tr>
                        </tbody>
                    </table>
                </section>
        
        
        <informaltable rules="all" frame="box">
            <tr>
                <td>
                    <para><emphasis role="italic">For reference unchanged DICOM PS3.6:</emphasis></para>
                </td>
            </tr>
        </informaltable>
        
        
        <section label="6" status="1" xml:id="chapter_6">
            <title>Registry of DICOM Data Elements</title>
            
            <table frame="box" label="6-1" rules="all" xml:id="table_6-1">
                <caption>Registry of DICOM Data Elements</caption>
                <thead>
                    <tr valign="top">
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_3e8c52d5-3582-4ec4-a7e6-acd7604caf32">
                                <emphasis role="bold">Tag</emphasis>
                            </para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_ac298c3b-bcdd-462b-9240-c38e738bfb7c">
                                <emphasis role="bold">Name</emphasis>
                            </para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_99f6b0ff-e1b0-484e-83b5-c3003d61704c">
                                <emphasis role="bold">Keyword</emphasis>
                            </para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_42b3621d-8b30-46eb-8992-10bb21310ac7">
                                <emphasis role="bold">VR</emphasis>
                            </para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_551ea829-a949-4bfd-a201-bb102024e64c">
                                <emphasis role="bold">VM</emphasis>
                            </para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_3a18713f-2e81-4664-b9a1-e4e3d41957ab"/>
                        </th>
                    </tr>
                </thead>
                <tbody>
        <tr valign="top">
            <td align="center" colspan="1" rowspan="1">
                <para xml:id="para_f1a57acf-0a3e-4f3e-b4b8-4e397e87ff40">
                    <emphasis role="italic">(0018,1011)</emphasis>
                </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c5676893-5f90-4850-9052-a29a0709ebe9">
                    <emphasis role="italic">Hardcopy Creation Device ID</emphasis>
                </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_817f5fee-da4d-4934-ab74-8c4bdc5a2aca">
                    <emphasis role="italic">Hardcopy​Creation​Device​ID</emphasis>
                </para>
            </td>
            <td align="center" colspan="1" rowspan="1">
                <para xml:id="para_736f2604-d90c-4625-aaca-4e1f53bd66a0">
                    <emphasis role="italic">LO</emphasis>
                </para>
            </td>
            <td align="center" colspan="1" rowspan="1">
                <para xml:id="para_7f20f0db-2f53-4151-b580-7a26cadc0e40">
                    <emphasis role="italic">1</emphasis>
                </para>
            </td>
            <td align="center" colspan="1" rowspan="1">
                <para xml:id="para_23a9332d-f3bf-4ac1-a984-54c0d47ac232">
                    <emphasis role="italic">RET (2007)</emphasis>
                </para>
            </td>
        </tr>
                </tbody>
            </table>
        </section>
        
</chapter>
</book>