    <title>CP-2356</title>
    <subtitle>Add Ethnic Group Code Sequence</subtitle>
    <chapter>
        <title></title>
        <informaltable rules="all" frame="box">
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Status</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>Final Text</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Date of Last Update</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>2024/06/11</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Person Assigned</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>David Clunie</para><para><link
                    xl:href="mailto:dclunie@dclunie.com"/></para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submitter Name</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>Ulrike Wagner</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submission Date</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>2023/03/14</para></td>
            </tr>
        </informaltable>
        <informaltable rules="all" frame="box">
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Correction Number CP-2356</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Log Summary: Add Ethnic Group Code Sequence</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">
                    <para>Name of Standard</para>
                    <para>PS3.3, PS3.6, PS3.15, PS3.16 2024b</para>
                </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Rationale for Correction:</para>
                    <para>The ethnic group or race of a patient is only described as a text string rather then a coded concept, when there are well-defined coded concepts available, as used in SR. Add a code sequence attribute.</para>
                    <para>Allow a multiplicity of 1-n for mixed races, since some of the other informatics standards like (<link xmlns:xlink="http://www.w3.org/1999/xlink" xlink:href="http://v2plus.hl7.org/2021Jan/segment-definition/PID.html">HL7 V2 PID-10</link>) allow for more than one value (with the proviso that "the second triplet ... is reserved for governmentally assigned codes"), though others do not (like <link xmlns:xlink="http://www.w3.org/1999/xlink" xlink:href="https://build.fhir.org/ig/HL7/US-Core/StructureDefinition-us-core-race.html">HL7 FHIR US Extension</link>). This is consistent with the recently announced revised <link xmlns:xlink="http://www.w3.org/1999/xlink" xlink:href="http://spd15revision.gov/">US OMB Statistical Policy Directive 15</link> that envisages use of multiple values and also removes the distinction between race and ethnic group, and Hispanic ethnic group in particular.</para>
                    <para>Note that in DICOM the concept "ethnic group" is used in a general sense, and is equivalent to the social construct of a group, <link xmlns:xlink="http://www.w3.org/1999/xlink" xlink:href="http://snomed.info/id/415229000">SCT:415229000 Racial Group</link>, and not the historically-used US-specific <link xmlns:xlink="http://www.w3.org/1999/xlink" xlink:href="https://wonder.cdc.gov/wonder/help/populations/bridged-race/Directive15.html">US OMB Statistical Policy Directive 15</link> (Hispanic or not) manner, though it is not forbidden to also convey the latter concept as a pre-coordinated value.</para>
                    <para>Also change context group reference in SR subject context from Defined to Baseline, to allow other codes to be used as necessary.</para>
                    <para>This CP does not attempt to extend the current CID 6099 Racial Group that is used as a value set beyond a small number of additional values, since that is a more complicated exercise. For reference, see also the <link xmlns:xlink="http://www.w3.org/1999/xlink" xlink:href="https://www.hl7.org/fhir/R4/v2/0005/index.html">HL7 V2 Table 0005</link>, and the <link xmlns:xlink="http://www.w3.org/1999/xlink" xlink:href="https://www.cdc.gov/phin/resources/vocabulary/documents/PH_RaceAndEthnicity_CDC_v1.2.xlsx">CDC Race &amp; Ethnicity Code System</link>.</para>
                    <para>The revised SPD 15 also has sub-categories that are nation-specific (e.g., "Japanese") and perhaps be mapped to children of (365450009, SCT, "Finding of place of origin (finding))" though there are not many of these (yet?). No action is taken in this CP on that.</para>
                    <para><emphasis role="italic">[Ed. Note: SNOMED CRS request# 960501 to add 414408004.]</emphasis></para>
                </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">Correction Wording:</td>
            </tr>
        </informaltable>
    </chapter>
    <chapter>
        <title/>

      <informaltable rules="all" frame="box">
        <tr>
          <td>
            <para><emphasis role="italic">Amend DICOM PS3.3 as follows (changes to existing text
              are bold and <emphasis role="bold"><emphasis role="underline"
                  >underlined</emphasis></emphasis> for additions and <emphasis role="bold"
                  ><emphasis role="strikethrough">struckthrough</emphasis></emphasis> for
              removals):</emphasis></para>
          </td>
        </tr>
      </informaltable>
        
                
        <section label="C.2.3" status="3" xml:id="sect_C.2.3">
                <title>Patient Demographic Module</title>
                <table frame="box" label="C.2-3" rules="all" xml:id="table_C.2-3">
                    <caption>Patient Demographic Module Attributes</caption>
                    <thead>
                        <tr valign="top">
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_2f485598-511d-492e-a3c7-79bfdd81ce68">Attribute Name</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_523b87fe-c055-4338-9cb9-ff0cf657b37e">Tag</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_e435214b-d609-43f9-be65-c66cd7d255ef">Attribute Description</para>
                            </th>
                        </tr>
                    </thead>
                    <tbody>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_9bc6680c-befc-4fbb-b598-f38cd006d6e2">...</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_19cb32a0-0a70-4d9e-89f2-bb5991d80b03"/>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_1d9c8bed-5a01-4e96-9e19-5f020061b6b9"/>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_a6ea0f92-5321-4fc4-9779-e52e3f5afb37">Ethnic Group</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_d050d530-cd2a-4164-abf4-5edbf90dde9c">(0010,2160)</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_ac3b855e-33c7-4b46-840e-0f0f4268a8a1">Ethnic group or race of Patient.</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1">
                                <para><emphasis role="bold"><emphasis role="underline">Ethnic Group Code Sequence</emphasis></emphasis></para>
                            </td>
                            <td align="center" colspan="1">
                                <para><emphasis role="bold"><emphasis role="underline">(0010,2161)</emphasis></emphasis></para>
                            </td>
                            <td align="left" colspan="1">
                                <para><emphasis role="bold"><emphasis role="underline">Ethnic group or race of Patient.</emphasis></emphasis></para>
                            <para><emphasis role="bold"><emphasis role="underline">One or more Items are permitted in this Sequence.</emphasis></emphasis></para>
                            </td>
                        </tr>
                        <tr valign="top">
                        <td align="left" colspan="2" rowspan="1">
                            <para>
                                <emphasis role="italic"><emphasis role="bold"><emphasis role="underline">&gt;Include Table 8.8-1 "Code Sequence Macro Attributes"<!--<xref linkend="table_8.8-1" xrefstyle="select: label quotedtitle"/>-->
                                </emphasis></emphasis></emphasis>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para>
                                <emphasis role="italic"><emphasis role="bold"><emphasis role="underline">B<olink targetdoc="PS3.16" targetptr="sect_CID_6099" xrefstyle="select: labelnumber quotedtitle"/>.</emphasis></emphasis></emphasis>
                            </para>
                        </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1">
                                <para>...</para>
                            </td>
                            <td align="center" colspan="1">
                             <para/>
                            </td>
                            <td align="left" colspan="1">
                                <para/>
                            </td>
                        </tr>
                    </tbody>
                </table>
            </section>
        
        <section label="C.7.1.1" status="4" xml:id="sect_C.7.1.1">
                    <title>Patient Module</title>
                    <table frame="box" label="C.7-1" rules="all" xml:id="table_C.7-1">
                        <caption>Patient Module Attributes</caption>
                        <thead>
                            <tr valign="top">
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_b3c32db2-8602-4023-89bb-3856b3c51c8e">Attribute Name</para>
                                </th>
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_2f7d731b-1fef-4474-81fb-bd2eb47efb40">Tag</para>
                                </th>
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_8d036e51-8ab1-43d3-8460-77d97c7a5ff9">Type</para>
                                </th>
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_53c874ba-0179-437b-86d0-406e872b2065">Attribute Description</para>
                                </th>
                            </tr>
                        </thead>
                        <tbody>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_5bf1f026-8432-4ae7-b5cd-1935b608d4cd">...</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_353153a4-93d9-455d-9cd3-4f9c0e6e5c3a"/>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_e4638255-baff-4e4c-aa74-ed52d49beb0e"/>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_427a23ce-994c-4eb2-9cab-0c07583ec129"/>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_1f7608c5-0f7c-4762-9200-04a602f2cb06">Ethnic Group</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_a0c22dd7-0ab0-49a0-95e2-6c0c472d780d">(0010,2160)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_79cb0f04-62bc-4b14-aaa0-21280ba10032">3</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_b9146e82-c7a9-44ff-9775-e0d8482195dc">Ethnic group or race of the Patient.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1">
                                    <para><emphasis role="bold"><emphasis role="underline">Ethnic Group Code Sequence</emphasis></emphasis></para>
                                </td>
                                <td align="center" colspan="1">
                                    <para><emphasis role="bold"><emphasis role="underline">(0010,2161)</emphasis></emphasis></para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para><emphasis role="bold"><emphasis role="underline">3</emphasis></emphasis></para>
                                </td>
                                <td align="left" colspan="1">
                                    <para><emphasis role="bold"><emphasis role="underline">Ethnic group or race of Patient.</emphasis></emphasis></para>
                                    <para><emphasis role="bold"><emphasis role="underline">One or more Items are permitted in this Sequence.</emphasis></emphasis></para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="3" rowspan="1">
                                    <para>
                                        <emphasis role="italic"><emphasis role="bold"><emphasis role="underline">&gt;Include Table 8.8-1 "Code Sequence Macro Attributes"<!--<xref linkend="table_8.8-1" xrefstyle="select: label quotedtitle"/>-->
                                        </emphasis></emphasis></emphasis>
                                    </para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para>
                                        <emphasis role="italic"><emphasis role="bold"><emphasis role="underline">B<olink targetdoc="PS3.16" targetptr="sect_CID_6099" xrefstyle="select: labelnumber quotedtitle"/>.</emphasis></emphasis></emphasis>
                            </para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_76c086db-a2d6-4fc4-a972-aa17309cb654">...</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_ecc4b464-145e-41d1-a0f7-1da3f3b132f4"/>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_1ce2285f-6014-4620-a482-683a4f86bc21"/>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_f4631c33-1fec-4136-841a-72846b2899fa"/>
                                </td>
                            </tr>
                        </tbody>
                    </table>
                </section>
        
        <section label="C.13.2" status="3" xml:id="sect_C.13.2">
            <title>Basic Film Session Relationship Module</title>
            <table frame="box" label="C.13-2" rules="all" xml:id="table_C.13-2">
                <caption>Basic Film Session Relationship Module Attributes</caption>
                <thead>
                    <tr valign="top">
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_7972c820-2467-4aa4-bb80-548f30f88d84">Attribute Name</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_82e34827-47ad-43df-bfe8-3b3630e1b60c">Tag</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_f857f4f1-69c0-4157-89c3-5721d88d9662">Attribute Description</para>
                        </th>
                    </tr>
                </thead>
                <tbody>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_8b1ef631-8796-4338-a741-2b236a88da7c">...</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_7dd50da6-b5db-43e8-9355-71d2a847d477"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_28c9b1d6-a2b9-4ae8-b64d-a04dbd5ffe06"/>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_a8f0aad0-ed3d-4304-adf6-d31a906fb442">Proposed Study Sequence</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_7935f665-6ff7-4489-ab21-6dc55dfb6f3d">(2130,00A0)</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_f0f4d56f-1af1-4602-b51e-30e6d2041a27">Attributes that may be used to identify and describe the Patient, Study and Series that are the subject of this Film Session.</para>
                            <para xml:id="para_b42eff0c-d7bd-4a54-8901-33ae5b3f2ebc">Only a single Item is permitted in this Sequence.</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_fae14a11-0c4a-4189-9e94-6ae385e68a90">...</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_ae8f0361-9d88-45ba-b449-a455f257ff45"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_7c1c5009-0c9f-4c16-a1a8-85f1ab0df66d"/>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_776a3d76-61c4-462a-b629-35bcdd194fc5">&gt;Ethnic Group</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_62e268ad-c2aa-46cd-8370-b13e5d568e9c">(0010,2160)</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_46e55cc1-b077-408c-956a-bd4002a0c25e">See <xref linkend="sect_C.2.3" xrefstyle="select: label title"/> for description.</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">&gt;Ethnic Group Code Sequence</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">(0010,2161)</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">See <xref linkend="sect_C.2.3" xrefstyle="select: label title"/> for description.</emphasis></emphasis></para>
                            <para><emphasis role="bold"><emphasis role="underline">One or more Items are permitted in this Sequence.</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="2" rowspan="1">
                            <para>
                                <emphasis role="italic"><emphasis role="bold"><emphasis role="underline">&gt;&gt;Include Table 8.8-1 "Code Sequence Macro Attributes"<!--<xref linkend="table_8.8-1" xrefstyle="select: label quotedtitle"/>-->
                                        </emphasis></emphasis></emphasis>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para>
                                <emphasis role="italic"><emphasis role="bold"><emphasis role="underline">B<olink targetdoc="PS3.16" targetptr="sect_CID_6099" xrefstyle="select: labelnumber quotedtitle"/>.</emphasis></emphasis></emphasis>
                            </para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_ce742bb1-e81a-4784-aa3e-0ce6ecf02147">...</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_9922b9e1-891e-4031-88c3-5e047bfbeaa1"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_c4b79dd2-914d-4bba-80dd-04b351e43395"/>
                        </td>
                    </tr>
                </tbody>
            </table>
        </section>
        
        
        <informaltable rules="all" frame="box">
            <tr>
                <td>
                    <para><emphasis role="italic">Add new data element to DICOM PS3.6 as follows:</emphasis></para>
                </td>
            </tr>
        </informaltable>
        
        <section label="6" status="1" xml:id="chapter_6">
            <title>Registry of DICOM Data Elements</title>
            <table frame="box" label="6-1" rules="all" xml:id="table_6-1">
                <caption>Registry of DICOM Data Elements</caption>
                <thead>
                    <tr valign="top">
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_3e8c52d5-3582-4ec4-a7e6-acd7604caf32">
                                <emphasis role="bold">Tag</emphasis>
                            </para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_ac298c3b-bcdd-462b-9240-c38e738bfb7c">
                                <emphasis role="bold">Name</emphasis>
                            </para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_99f6b0ff-e1b0-484e-83b5-c3003d61704c">
                                <emphasis role="bold">Keyword</emphasis>
                            </para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_42b3621d-8b30-46eb-8992-10bb21310ac7">
                                <emphasis role="bold">VR</emphasis>
                            </para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_551ea829-a949-4bfd-a201-bb102024e64c">
                                <emphasis role="bold">VM</emphasis>
                            </para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_3a18713f-2e81-4664-b9a1-e4e3d41957ab"/>
                        </th>
                    </tr>
                </thead>
                <tbody>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_2c36579f-3dad-4b41-9f0a-b488d9a5be90">(0010,2160)</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_96279fc7-e8b4-416c-af6f-ced7cdd4a031">Ethnic Group</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_1693d091-2a4b-47c9-85f6-61fb6bbdcb6f">Ethnic​Group</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_390361b5-9b5c-45d4-9004-a99dac2e2749">SH</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_82f0863f-ea57-4b67-9cd3-d285d7ce67fa">1</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_f5e29409-1c72-4a74-b8e2-82a6a7bfa68a"/>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">(0010,2161)</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Ethnic Group Code Sequence</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Ethnic​Group​Code​Sequence</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">SQ</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">1</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline"/></emphasis></para>
                        </td>
                    </tr>
                    
                </tbody>
            </table>
        </section>
        
        
        
        <informaltable rules="all" frame="box">
            <tr>
                <td>
                    <para><emphasis role="italic">Amend DICOM PS3.15 as follows (changes to existing text are bold and <emphasis role="bold"><emphasis role="underline">underlined</emphasis></emphasis> for additions and <emphasis role="bold"><emphasis role="strikethrough">struckthrough</emphasis></emphasis> for removals):</emphasis></para>
                </td>
            </tr>
        </informaltable>
        
        <section label="E.1.1" status="3" xml:id="sect_E.1.1">
            <title>De-identifier</title>
        <table frame="box" label="E.1-1" rules="all" xml:id="table_E.1-1">
            <caption>Application Level Confidentiality Profile Attributes</caption>
            <thead>
                <tr valign="top">
                    <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_c5e6e8ba-34e0-4e31-90b5-bd9ed7086ddd">
                            <emphasis role="bold">Attribute Name</emphasis>
                        </para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_d02d5580-8438-4a00-8c30-ba93530bcd31">
                            <emphasis role="bold">Tag</emphasis>
                        </para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_705936dc-2c42-4198-b406-ba0e61b35e99">
                            <emphasis role="bold">Retd. (from <olink targetdoc="PS3.6" targetptr="PS3.6" xrefstyle="select: labelnumber"/>)</emphasis>
                        </para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_a8387cac-386f-4ad9-96b3-b3dcb1b1010c">
                            <emphasis role="bold">In Std. Comp. IOD (from <olink targetdoc="PS3.3" targetptr="PS3.3" xrefstyle="select: labelnumber"/>)</emphasis>
                        </para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_fc0ff9a1-5780-4d06-861a-24ae0ba38b47">
                            <emphasis role="bold">Basic Prof.</emphasis>
                        </para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_19ddb67a-a5d6-474d-a13f-8df27d28b707">
                            <emphasis role="bold">Rtn. Safe Priv. Opt.</emphasis>
                        </para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_5b335431-5276-4cc0-bddf-93b6d107226e">
                            <emphasis role="bold">Rtn. UIDs Opt.</emphasis>
                        </para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_8317e0cf-0080-48f9-8040-1f49a4478c0f">
                            <emphasis role="bold">Rtn. Dev. Id. Opt.</emphasis>
                        </para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_1ba661a2-a572-420a-95c1-bc1bf694a70e">
                            <emphasis role="bold">Rtn. Inst. Id. Opt.</emphasis>
                        </para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_796841da-6b27-4558-99c9-99503edccc6c">
                            <emphasis role="bold">Rtn. Pat. Chars. Opt.</emphasis>
                        </para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_edac8181-290b-4f9c-945d-5dfc7b1a43fc">
                            <emphasis role="bold">Rtn. Long. Full Dates Opt.</emphasis>
                        </para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_6305752d-56e7-4f4d-b9d8-e5052fcd6eea">
                            <emphasis role="bold">Rtn. Long. Modif. Dates Opt.</emphasis>
                        </para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_84f3b128-8fe5-45bc-b04e-abc483377b06">
                            <emphasis role="bold">Clean Desc. Opt.</emphasis>
                        </para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_0adc5d55-1196-4265-8ee1-b205103c96f1">
                            <emphasis role="bold">Clean Struct. Cont. Opt.</emphasis>
                        </para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_5e74e547-b6fd-410e-9c09-b271915c3eeb">
                            <emphasis role="bold">Clean Graph. Opt.</emphasis>
                        </para>
                    </th>
                </tr>
            </thead>
            <tbody>
        
        <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_cbfc33ea-86f8-4ef1-b9c1-82426ad1f6d0">Ethnic Group</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
                <para xml:id="para_251edca2-386c-4cd9-a041-93e2329ce96a">(0010,2160)</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
                <para xml:id="para_51bac9de-c669-429d-98e0-b62818512c29">N</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
                <para xml:id="para_bf82f485-ce36-4aa5-b908-a67688f6f68f">Y</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
                <para xml:id="para_5ceaea19-e403-45ad-897f-7e212326e56c">X</para>
            </td>
            <td align="center" colspan="1" rowspan="1"/>
            <td align="center" colspan="1" rowspan="1"/>
            <td align="center" colspan="1" rowspan="1"/>
            <td/>
            <td align="center" colspan="1" rowspan="1">
                <para xml:id="para_30416bf2-fd9c-443d-9aea-d1a02df20f60">K</para>
            </td>
            <td align="center" colspan="1" rowspan="1"/>
            <td align="center" colspan="1" rowspan="1"/>
            <td align="center" colspan="1" rowspan="1"/>
            <td align="center" colspan="1" rowspan="1"/>
            <td align="center" colspan="1" rowspan="1"/>
        </tr>
                <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                        <para><emphasis role="bold"><emphasis role="underline">Ethnic Group Code Sequence</emphasis></emphasis></para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                        <para><emphasis role="bold"><emphasis role="underline">(0010,2161)</emphasis></emphasis></para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                        <para><emphasis role="bold"><emphasis role="underline">N</emphasis></emphasis></para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                        <para><emphasis role="bold"><emphasis role="underline">Y</emphasis></emphasis></para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                        <para><emphasis role="bold"><emphasis role="underline">X</emphasis></emphasis></para>
                    </td>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td/>
                    <td align="center" colspan="1" rowspan="1">
                        <para><emphasis role="bold"><emphasis role="underline">K</emphasis></emphasis></para>
                    </td>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                </tr>
        
            </tbody>
        </table>
        </section>
        
        <informaltable rules="all" frame="box">
            <tr>
                <td>
                    <para><emphasis role="italic">Amend DICOM PS3.16 as follows (changes to existing text are bold and <emphasis role="bold"><emphasis role="underline">underlined</emphasis></emphasis> for additions and <emphasis role="bold"><emphasis role="strikethrough">struckthrough</emphasis></emphasis> for removals):</emphasis></para>
                </td>
            </tr>
        </informaltable>
        
        <section label="TID 1007" status="3" xml:id="sect_TID_1007">
            <title>Subject Context, Patient</title>
            <table frame="box" label="TID 1007" rules="all" xml:id="table_TID_1007">
                <caption>Subject Context, Patient</caption>
                <thead>
                    <tr valign="top">
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_cbfaf626-9734-4d07-a4ac-0aa6958258c5"/>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_3a036833-b7d7-41eb-917b-2ecdcf0d995d">NL</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_901af82f-5f12-455e-96dd-d7660580d658">Rel with Parent</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_5b7de8b7-7530-475a-87e6-8787224443c7">VT</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_aea81ac1-0ded-411c-94bb-f9a48e915cd0">Concept Name</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_9b38dad8-5e65-4e44-9911-505877fa176d">VM</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_80064aea-6c47-48ad-8ebf-a4926020a288">Req Type</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_9af8b701-1cde-485e-9275-6909874ce303">Condition</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_6162a781-0873-4597-824f-bdf777429be7">Value Set Constraint</para>
                        </th>
                    </tr>
                </thead>
                <tbody>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_8019b492-1fd1-44e6-8732-c7f765778080">...</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_3192be22-006e-47b0-9272-80f0d842378d"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_fb168607-e8f0-4e62-8c96-ea881d809fdd"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_e45ac586-76fc-443f-87ec-7ca703334271"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_5ab0ba85-5b97-4079-8945-aac23bdf520e"/>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_44eec935-d3b5-43ee-95ce-62dec7662df9"/>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_8d64e418-4e74-42cb-8db6-00481b9fd2bf"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_bcb64d08-0f26-46fc-b46e-5b47975537db"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_cd49e539-941c-4ec0-9ca6-b4094b8e8ea5"/>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_244187c3-3e03-4f1c-bced-a23bf7b02956">9</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_199ae9a5-2b85-425f-a4a5-b3e5170e32da"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_c8c2b899-f437-4160-a4ed-6d2b1d0fdea8"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_ae38d44b-07c4-4ca7-a8d6-87dedec09304">CODE</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_001845b6-e368-4f4f-bfa6-92a8cb4ed4c8">EV <link xl:href="http://snomed.info/id/415229000" xl:show="new">(415229000, SCT, "Racial group")</link>
                            </para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_d68fd13c-389e-49f3-8a6b-0ae19e59b592">1</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_2c2aa631-9941-426b-9cca-2f3473987b98">U</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_bd544c37-f3b1-4262-a145-833c83ea61f0"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_2c234fea-fdfd-40c9-93a3-8b07ead0a6fe">Defaults to <emphasis role="bold"><emphasis role="underline">Value of Ethnic Group Code Sequence (0010,2161) of the <xref linkend="sect_C.7.1.1" xrefstyle="select: title"/><!--<olink targetdoc="PS3.3" targetptr="sect_C.7.1.1" xrefstyle="select: title"/>-->, or if absent, </emphasis></emphasis>
                                the coded equivalent of the text Value of Ethnic Group (0010,2160) of the <xref linkend="sect_C.7.1.1" xrefstyle="select: title"/><!--<olink targetdoc="PS3.3" targetptr="sect_C.7.1.1" xrefstyle="select: title"/>-->, which is defined as "race or ethnic group"</para>
                            <para xml:id="para_296466ba-c231-41b5-8108-fba2584de6d7"><emphasis role="bold"><emphasis role="strikethrough">D</emphasis><emphasis role="underline">B</emphasis></emphasis><xref linkend="sect_CID_6099" xrefstyle="select: labelnumber quotedtitle nopage"/>
                            </para>
                        </td>
                    </tr>
                </tbody>
            </table>
        </section>
        
        <section label="TID 8300" status="3" xml:id="sect_TID_8300">
              <title>Skin Imaging Acquisition Context</title>
              <table frame="box" label="TID 8300" rules="all" xml:id="table_TID_8300">
                  <caption>Skin Imaging Acquisition Context</caption>
                    <thead>
                        <tr valign="top">
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_ccdf222c-bf7c-4a0d-9d1d-475ee1cd3a6b"/>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_0d0d3d34-da4c-4f44-b548-2d4cf522e4ea">VT</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_32cb7809-d2db-4875-8b9e-3c7366eb41e7">Concept Name</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_b8639655-6406-4b70-8052-acd7867929b8">VM</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_08514e4c-d0e3-4c32-9dd9-980137ae5d88">Req Type</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_6c4fa80b-af80-4bf7-a8de-02f4049966b5">Condition</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_72119e80-00d0-4365-982a-35dc20184bb4">Value Set Constraint</para>
                            </th>
                        </tr>
                    </thead>
                    <tbody>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_fec4c0f4-3905-4dd4-b132-19e9bfeedd1d">2</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_3523d8aa-ce18-45c6-9ee4-b2d1685b2957">CODE</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_efeeaac9-455d-4806-9504-d8b398c4a8ed">EV <link xl:href="http://snomed.info/id/415229000" xl:show="new">(415229000, SCT, "Racial group")</link>
</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_eb202258-2a4d-466b-8c98-4121e7cea779">1</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_6424f7a4-ebf8-407b-bfc6-07f7da5b8787">U</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_59d0261c-8b12-4301-897c-3fb8c6de1fea"/>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_636d23ef-6baf-472f-b10a-0884968269b2"><emphasis role="bold"><emphasis role="strikethrough">D</emphasis><emphasis role="underline">B</emphasis></emphasis><xref linkend="sect_CID_6099" xrefstyle="select: labelnumber quotedtitle nopage"/>
                            </para>
                            </td>
                        </tr>
                        
                        
                        
                        
                        
                        
                        
                        
                        
                        
                        
                        
                        
                        
                        
                        
                        
                    </tbody>
                </table>
                
                
          </section>        
        
        <section label="CID 6099" status="3" xml:id="sect_CID_6099">
            <title>Racial Group</title>
            <note>
                <para xml:id="para_27a9db3c-3467-4d70-a48d-946d65d581c9">This Context Group is intended to be a set of values that is the union of concepts used in various jurisdictions. It does not distinguish race from ethnic group.</para>
            </note>
            <variablelist spacing="compact">
                <varlistentry>
                    <term>
                        <emphasis role="bold">Version:</emphasis>
                    </term>
                    <listitem>
                        <para xml:id="para_d113e7ce-bd7a-4a30-aa23-9bea963c7e8c">
                            <emphasis role="bold"><emphasis role="strikethrough">20201115</emphasis><emphasis role="underline">202240611</emphasis></emphasis>
                        </para>
                    </listitem>
                </varlistentry>
            </variablelist>
            <table frame="box" label="CID 6099" rules="all" xml:id="table_CID_6099">
                <caption>Racial Group</caption>
                <thead>
                    <tr valign="top">
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_1aa4a60c-9d65-41a4-ac82-ba82d1035833">Coding Scheme Designator</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_147cb3d4-e22e-401c-8c9d-772024012b9a">Code Value</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_cf7e1edb-3356-4b77-858a-275e81d3e45a">Code Meaning</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_fef09a07-9652-4b16-925e-9e13016cf548">SNOMED-RT ID</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_65915413-e913-438a-9319-3dafcf365832">UMLS Concept Unique ID</para>
                        </th>
                    </tr>
                </thead>
                <tbody>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_1b7df7ec-5bef-40db-b6dd-feb9e528756d">SCT</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_171a36c6-b841-454d-a0c3-45cfdfd6a151">
                                <link xl:href="http://snomed.info/id/413464008" xl:show="new">413464008</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_b97ccd0a-8a39-4410-aefc-b25c2b1052b0">African race</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_42de4af1-5319-4e28-a4d6-0f0c382c7bed">
                                <link xl:href="http://snomed.info/id/413464008" xl:show="new">S-0004E</link>
                            </para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_baff6e1c-a62f-4040-80e3-1a004e730494">
                                <link xl:href="http://uts.nlm.nih.gov/uts/umls/concept/C0027567" xl:show="new">C0027567</link>
                            </para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_66ce4569-1deb-45fe-b1fb-381b1f270f94">SCT</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_6f2b5379-a3f6-4d92-8d60-e701ecf2f035">
                                <link xl:href="http://snomed.info/id/413582008" xl:show="new">413582008</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_9f440c3d-8d11-4711-94f3-35fd64826314">Asian race</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_03480f6a-386b-46f8-ad9b-b571b2cace86">
                                <link xl:href="http://snomed.info/id/413582008" xl:show="new">S-00051</link>
                            </para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_e87bd2f6-6cd7-4554-a459-783655bef9d8">
                                <link xl:href="http://uts.nlm.nih.gov/uts/umls/concept/C0078988" xl:show="new">C0078988</link>
                            </para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_20c3c10c-5db0-4046-a1d5-6b9661c08f59">SCT</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_eeeda8af-eb47-473c-a618-060632098ef9">
                                <link xl:href="http://snomed.info/id/413773004" xl:show="new">413773004</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_683bd777-687e-40ff-927b-648dbe1f4788">Caucasian race</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_3cdba906-2bf8-4797-a33a-352dbac3f5f8">
                                <link xl:href="http://snomed.info/id/413773004" xl:show="new">S-0003D</link>
                            </para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_d982a7ab-dd97-4f61-b3e1-617ddbbc77d5">
                                <link xl:href="http://uts.nlm.nih.gov/uts/umls/concept/C0007457" xl:show="new">C0007457</link>
                            </para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_13e145ea-e546-40a1-9819-3fbc189107ff">SCT</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_8d0b58a5-206c-49c8-b54d-efe0e76f730c">
                                <link xl:href="http://snomed.info/id/413490006" xl:show="new">413490006</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_1670d8a4-4cb6-438d-8060-8353332f6901">American Indian or Alaska native</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_8ff0f7ab-58e5-446c-8c5d-11f58f5c20f6">
                                <link xl:href="http://snomed.info/id/413490006" xl:show="new">S-0004B</link>
                            </para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_cc492226-4bc7-40d1-bd43-7ba23bacfbb5">
                                <link xl:href="http://uts.nlm.nih.gov/uts/umls/concept/C1515945" xl:show="new">C1515945</link>
                            </para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_b8a8a629-f55a-49b3-a552-867b01e55ede">NCIt</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_94eed133-0241-4b88-9594-faec9d245991">
                                <link xl:href="http://ncit.nci.nih.gov/ncitbrowser/ConceptReport.jsp?dictionary=NCI_Thesaurus&amp;ns=NCI_Thesaurus&amp;code=C41219" xl:show="new">C41219</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_762dab77-53b7-4b99-98e2-041bb88124ad">Native Hawaiian or other Pacific Islander</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_861ace95-1ad9-4ba6-bcf2-a4b813aa7b26"/>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_4dacf13c-8f1d-4070-b58d-eb352993221e">
                                <link xl:href="http://uts.nlm.nih.gov/uts/umls/concept/C1513907" xl:show="new">C1513907</link>
                            </para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_9c1d235d-7e50-49b2-ad02-e3a9da2b6e96">SCT</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_4b36765b-78fb-4e32-a767-40c96d02f217">
                                <link xl:href="http://snomed.info/id/413581001" xl:show="new">413581001</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_70084131-80f6-4b5a-8626-1252a769e3ef">Asian or Pacific Islander race</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_1909f60c-4e17-434d-aa8d-a821c89dfb04">
                                <link xl:href="http://snomed.info/id/413581001" xl:show="new">S-0004C</link>
                            </para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_fc752c74-3efd-4b19-ab18-f3a9e0d8acbe">
                                <link xl:href="http://uts.nlm.nih.gov/uts/umls/concept/C1531604" xl:show="new">C1531604</link>
                            </para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_110d5b8e-0733-4096-b7df-3d097cf93503">SCT</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_0808d906-de10-4253-94d6-530d6fd9a8d6">
                                <link xl:href="http://snomed.info/id/413600007" xl:show="new">413600007</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_349f0c15-d549-4063-8f15-8a261773b9f3">Australian aborigine race</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_56bf2c38-154e-48ae-9ecb-cb71e4cc8461">
                                <link xl:href="http://snomed.info/id/413600007" xl:show="new">S-00052</link>
                            </para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_29b49ce9-e504-4dc7-9d8b-8010ff2f9884">
                                <link xl:href="http://uts.nlm.nih.gov/uts/umls/concept/C0337948" xl:show="new">C0337948</link>
                            </para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_9c1388c5-ff11-4f56-a3f2-f2ec3f259d05">SCT</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_bf9d913d-daa1-4ea0-9cd4-0f13db1500e6">
                                <link xl:href="http://snomed.info/id/414481008" xl:show="new">414481008</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_8f1e7781-27ea-471d-9104-fbb42f36ed92">Indian race</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_a1cf3459-1170-4f99-90da-c71cc5bb28b6">
                                <link xl:href="http://snomed.info/id/414481008" xl:show="new">S-0003E</link>
                            </para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_d06d222d-3223-4b64-9ab7-9b4476ce9f86">
                                <link xl:href="http://uts.nlm.nih.gov/uts/umls/concept/C1524069" xl:show="new">C1524069</link>
                            </para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">SCT</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1">
                            <para>
                                <link xl:href="http://snomed.info/id/414408004" xl:show="new"><emphasis role="bold"><emphasis role="underline">414408004</emphasis></emphasis></link>
                            </para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Hispanic</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1">
                            <para>
                                <link xl:href="http://snomed.info/id/414408004" xl:show="new"><emphasis role="bold"><emphasis role="underline">S-00045</emphasis></emphasis></link>
                            </para>
                        </td>
                        <td align="center" colspan="1">
                            <para>
                                <link xl:href="http://uts.nlm.nih.gov/uts/umls/concept/C0086409" xl:show="new"><emphasis role="bold"><emphasis role="underline">C0086409</emphasis></emphasis></link>
                            </para>
                        </td>
                    </tr>
                    <tr>
                        <td align="center" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">UMLS</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1">
                            <para>
                                <link xl:href="http://uts.nlm.nih.gov/uts/umls/concept/C5690844" xl:show="new"><emphasis role="bold"><emphasis role="underline">C5690844</emphasis></emphasis></link>
                            </para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Middle Eastern or North African</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1">
                            <para/>
                        </td>
                        <td align="center" colspan="1">
                            <para>
                                <link xl:href="http://uts.nlm.nih.gov/uts/umls/concept/C5690844" xl:show="new"><emphasis role="bold"><emphasis role="underline">C5690844</emphasis></emphasis></link>
                            </para>
                            </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_b49de16e-7d12-4d0b-9eb3-3dcf01dd7453">SCT</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_0dba2368-8d47-45af-b590-2b86b7320504">
                                <link xl:href="http://snomed.info/id/414752008" xl:show="new">414752008</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_85722c9a-eafd-4727-a02c-ec235bc48bee">Mixed racial group</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_66097b2b-9e18-4506-81c6-b1375e99631e">
                                <link xl:href="http://snomed.info/id/414752008" xl:show="new">S-00043</link>
                            </para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_9d6c4089-de88-4c28-9d81-38314ca93d08">
                                <link xl:href="http://uts.nlm.nih.gov/uts/umls/concept/C0682081" xl:show="new">C0682081</link>
                            </para>
                        </td>
                    </tr>
                </tbody>
            </table>
        </section>
         
</chapter>
</book>