<book xmlns="http://docbook.org/ns/docbook" xmlns:xl="http://www.w3.org/1999/xlink" version="5.0">
    <title>CP-2237</title>
    <subtitle>Propagate VOI and Palette Color support from blending to advanced blending presentation state</subtitle>
    <chapter>
        <title></title>
        <informaltable rules="all" frame="box">
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Status</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>Final Text</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Date of Last Update</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>2023/01/19</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Person Assigned</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>David Clunie</para><para><link
                    xl:href="mailto:dclunie@dclunie.com"/></para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submitter Name</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>Markus Daniel Herrmann</para><para><link
                    xl:href="mailto:MDHERRMANN@MGH.HARVARD.EDU"/></para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submission Date</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>2022/06/03</para></td>
            </tr>
        </informaltable>
        <informaltable rules="all" frame="box">
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Correction Number CP-2237</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Log Summary: Propagate VOI and Palette Color support from blending to advanced blending presentation state</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">
                    <para>Name of Standard</para>
                    <para>PS3.3, PS3.4 2022e</para>
                </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Rationale for Correction:</para>
                    <para>The Advanced Blending Presentation State introduced in Sup 189 failed to include key features of the Blending Presentation State,
                    including definition of grayscale transformations and application of palette color LUTs, as defined for the blending pipeline in PS3.4.
                    This greatly limits its utility for sources that are single channel (grayscale structural or parametric).</para>
                    <para>Add the Softcopy VOI LUT Sequence and the Palette Color Lookup Table Sequence within each Item of the Advanced Blending Sequence, and clarify the PS3.4 text regarding grayscale processing and pseudo-coloring.</para>
                    <para>The windowing operation is necessary since its effect is that pixels values below the window and above the window are treated as though they were the same as the low and high values of the window, respectively, i.e., the pixels are still displayed. This is distinct from the threshold operation that excludes pixels entirely (renders them transparently) if they are outside the threshold.</para>
                    <para>The existing thresholding and new windowing operations are not mutually exclusive, since the former may be used for region exclusion (using a pixel padding value) as opposed to contrast adjustment.</para>
                    <para>The additional information is specified in the Advanced Blending Presentation State Module, rather than the Advanced Blending Presentation State Display Module, because it describes the transformation of the input, which is then referenced in the blending operation for display.</para>
                    <para>This pattern is also consistent with that used in the Pseudo-Color Softcopy Presentation State IOD.</para>
                    <para>The new Attributes are added as Type 1C rather than 3, since they have behavioral implications and need to be supported by display SCPs if present; to our knowledge adoption of this recently added SOP class has been limited so this should not break any implementations.</para>
                    <para>This CP depends on CP 2236 "Need to be able to select optical path in advanced blending presentation state", since different VOI and Palette Color LUTs may be applicable to different optical paths in the same image.</para>
                </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">Correction Wording:</td>
            </tr>
        </informaltable>
    </chapter>
    <chapter>
        <title/>

      <informaltable rules="all" frame="box">
        <tr>
          <td>
            <para><emphasis role="italic">Amend DICOM PS3.3 as follows (changes to existing text
              are bold and <emphasis role="bold"><emphasis role="underline"
                  >underlined</emphasis></emphasis> for additions and <emphasis role="bold"
                  ><emphasis role="strikethrough">struckthrough</emphasis></emphasis> for
              removals):</emphasis></para>
          </td>
        </tr>
      </informaltable>
        
<section label="A.33.7" status="3" xml:id="sect_A.33.7">
                <title>Advanced Blending Presentation State IOD</title>
                <section label="A.33.7.1" status="4" xml:id="sect_A.33.7.1">
                    <title>Advanced Blending Presentation State IOD Description</title>
                    <para xml:id="para_efb3c3ca-35a3-4d3f-8d7d-17f8b07049e7">The Advanced Blending Presentation State Information Object Definition (IOD) specifies information that may be used to blend two or more sets of <emphasis role="bold"><emphasis role="underline">grayscale or color </emphasis></emphasis>images that are referenced from within the IOD for the purpose of presentation (display)<emphasis role="bold"><emphasis role="underline"> as a color image</emphasis></emphasis>.</para>
                    <para xml:id="para_e88fdbce-9cda-4223-8281-a0727fce9445">It includes capabilities for specifying:</para>
                    <orderedlist numeration="loweralpha">
                        <listitem>
                            <para xml:id="para_57d7e2d5-944a-43ec-b07c-c170e7766a73">output color space in PCS-Values</para>
                        </listitem>
                        <listitem>
                        <para xml:id="para_951abb38-38c9-4fdb-a1ed-6eb4cb924948">optional thresholds to restrict contributing areas of an input</para>
                    </listitem>
                    <listitem>
                        <para><emphasis role="bold"><emphasis role="underline">optional windowing (VOI LUT) to be applied to grayscale images before pseudo-colorization</emphasis></emphasis></para>
                    </listitem>
                    <listitem>
                        <para><emphasis role="bold"><emphasis role="underline">optional pseudo-color palette to be applied to grayscale images</emphasis></emphasis></para>
                    </listitem>
                        <listitem>
                            <para xml:id="para_44c64125-cc17-43c0-8167-074bbcd0636c">definition of blending control values for the different inputs</para>
                        </listitem>
                        <listitem>
                            <para xml:id="para_15678658-d88d-449a-8a95-a8d86d5e0544">selection of the area of the output images to display and whether to rotate or flip it</para>
                        </listitem>
                        <listitem>
                            <para xml:id="para_6efc7ffc-cd29-4323-901a-217cb3b7cb4b">image and display relative annotations, including graphics, text and overlays</para>
                        </listitem>
                    </orderedlist>
                </section>
            </section>        
        
            <section label="C.11.33" status="3" xml:id="sect_C.11.33">
                <title>Advanced Blending Presentation State Module</title>
                <para xml:id="para_4beebafe-8a92-462c-a72e-147ba7a68e18">
                    <xref linkend="table_C.11.33-1" xrefstyle="select: label"/> specifies the Attributes of the <xref linkend="sect_C.11.33" xrefstyle="select: title"/>, which describe one or more inputs optionally one or more sets of registration objects, and the color<emphasis role="bold"><emphasis role="underline">, grayscale transformations</emphasis></emphasis> and thresholds to be applied to them, for the purpose of blending.</para>
                <table frame="box" label="C.11.33-1" rules="all" xml:id="table_C.11.33-1">
                        <caption>Advanced Blending Presentation State Module Attributes</caption>
                        <thead>
                            <tr valign="top">
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_ab5ff56e-0711-4ded-a949-fd52fad348de">Attribute Name</para>
                                </th>
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_a2bafb6e-47d6-4342-aef0-3bbbb4dead81">Tag</para>
                                </th>
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_3ac0a9ad-3de6-4431-8d51-c40714c4ec24">Type</para>
                                </th>
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_5d930fc2-bf28-4add-b7cb-c4d899d9d7dd">Attribute Description</para>
                                </th>
                            </tr>
                        </thead>
                        <tbody>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_0be2c5cf-d4bc-43cd-81e2-5a416d5df0c6">Advanced Blending Sequence</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_e761556f-079c-40b2-826c-8f440f63f593">(0070,1B01)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_a1b89367-bbbc-46bd-b27e-3a8e1fd34713">1</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_71cd464a-4a8e-48e0-8492-fc85ecc104d8">A Sequence of Items identifying the inputs and describing transformations of them.</para>
                                    <para xml:id="para_c37b26fb-3245-4063-8b9e-e5ccfb1f4743">One or more Items shall be included in this Sequence.</para>
                                    <para xml:id="para_43be7394-845e-4cbb-b9ba-1eae570fc6da">See <xref linkend="sect_C.11.33.1.1" xrefstyle="select: label"/>.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_118a3aad-ca17-4050-94fa-e4d32ccf66a3">&gt;Blending Input Number</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_2298f437-2c6c-450a-836a-c890fcb07ee6">(0070,1B02)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_96663a71-317c-4338-9671-90e5d0f0ddb8">1</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_d5fab946-0f1e-4bee-b549-e565ff0484f3">Identification number of the input. Values shall be ordinal numbers starting from 1 and monotonically increasing by 1 within the Advanced Blending Presentation State Instance.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_1012b696-75ef-4246-96b5-f744455a8c62">&gt;Study Instance UID</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_326c3b02-59ba-4089-9d8e-df11dec4f478">(0020,000D)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_9314c7b3-aa10-41d8-976e-922a9778cdb2">1</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_a98d8f7f-a337-4370-9f79-2a2029d0a3b7">Unique identifier for the Study.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_e4a5e800-3a94-4c81-95b2-1d4f7f5b9d56">&gt;Series Instance UID</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_cccf15d0-0f37-4ab4-9226-34f3b2c724e8">(0020,000E)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_03fc6a8e-0166-4226-9d0a-4833d5bed282">1</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_218c6076-248a-413a-b03c-72acb7d0bb30">Unique identifier of a Series that is part of the Study defined by the Study Instance UID (0020,000D)</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_0e6c2e84-f3d8-4f67-b93a-14c9653776a8">&gt;Referenced Image Sequence</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_6572fd00-da42-413f-b9f8-854d4640c2a0">(0008,1140)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_49b11f0b-98bb-4920-966e-84283ff00a8a">1C</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_65ccfa74-d054-41b8-8925-c9bb44e703fc">The set of images comprising this input. One or more Items shall be included in this Sequence.</para>
                                    <para xml:id="para_8dd972dd-5b82-4c89-a545-b347f3ba2f55">Required if the input is not the entire set of Instances in the Series.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="3" rowspan="1">
                                    <para xml:id="para_695e19e7-42d6-42ef-9f1a-1e556b3cd148">
                                        <emphasis role="italic">&gt;&gt;Include <xref linkend="table_10-3" xrefstyle="select: label quotedtitle"/>
                                        </emphasis>
                                    </para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_edc4a695-07ea-4fcd-86b2-bfb526a1bc5d">
                                        <emphasis role="italic"/>
                                    </para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para><emphasis role="bold"><emphasis role="underline">&gt;&gt;Referenced Optical Path Identifier</emphasis></emphasis></para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para><emphasis role="bold"><emphasis role="underline">(006A,000E)</emphasis></emphasis></para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para><emphasis role="bold"><emphasis role="underline">1C</emphasis></emphasis></para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para><emphasis role="bold"><emphasis role="underline">A single value that identifies the optical path that is the input.</emphasis></emphasis></para>
                                    <para><emphasis role="bold"><emphasis role="underline">Refers to the value in Optical Path Identifier (0048,0106) within the Optical Path Sequence (0048,0105) in the images to which this Presentation State applies.</emphasis></emphasis></para>
                                    <para><emphasis role="bold"><emphasis role="underline">Required if there is more than one Optical Path in the images to which this Presentation State applies.</emphasis></emphasis></para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_b2943a62-3acf-4109-a56b-d801baca80a4">&gt;Referenced Spatial Registration Sequence</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_ea4ed2b5-8f4c-4c28-b00f-40679c790898">(0070,0404)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_6e4f1441-e75f-4ffd-86fd-9627bfb6ec69">1C</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_4e395a4f-bac7-415b-a9d2-75f040a58e61">A reference to a Spatial Registration Instance that is used to register the referenced inputs.</para>
                                    <para xml:id="para_485f6b4a-9443-4f59-a2d7-484b028f1fe3">Only one Item shall be included in this Sequence.</para>
                                    <para xml:id="para_9d1e96e8-d97a-4182-8021-8319f7ef7b20">Required if the Frame of Reference UID (0020,0052) value of the Images referenced by the Referenced Image Sequence (0008,1140) of this Item does not match the Frame of Reference UID (0020,0052) value of this Presentation State Instance. May be present otherwise.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="3" rowspan="1">
                                    <para xml:id="para_53a03122-fe35-4c66-a83d-1be5ca16d82f">
                                        <emphasis role="italic">&gt;&gt;Include <xref linkend="table_C.17-3" xrefstyle="select: label quotedtitle"/>
                                        </emphasis>
                                    </para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_5b4c7c6e-62c6-49ca-ac48-6e2947587b10">
                                        <emphasis role="italic"/>
                                    </para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="3" rowspan="1">
                                    <para xml:id="para_704e3095-1c09-4c07-b082-d64ae4fda75f">
                                        <emphasis role="italic">&gt;Include <xref linkend="table_C.11.33.1-1" xrefstyle="select: label quotedtitle"/>
                                        </emphasis>
                                    </para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_c08e49c2-da0f-4e1d-8410-16143902fd95">
                                        <emphasis role="italic"/>
                                    </para>
                                </td>
                            </tr>
                            
                            
                            
                            
                            
                            
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para><emphasis role="bold"><emphasis role="underline">&gt;Softcopy VOI LUT Sequence</emphasis></emphasis></para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para><emphasis role="bold"><emphasis role="underline">(0028,3110)</emphasis></emphasis></para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para><emphasis role="bold"><emphasis role="underline">1C</emphasis></emphasis></para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para><emphasis role="bold"><emphasis role="underline">Defines a Sequence of VOI LUTs or Window Centers and Widths and to which images and frames they apply.</emphasis></emphasis></para>
                                    <para><emphasis role="bold"><emphasis role="underline">No more than one VOI LUT Sequence containing a single Item or one pair of Window Center/Width values shall be specified for each image or frame.</emphasis></emphasis></para>
                                    <para><emphasis role="bold"><emphasis role="underline">One or more Items shall be included in this Sequence.</emphasis></emphasis></para>
                                <para><emphasis role="bold"><emphasis role="underline">Required if a VOI LUT is to be applied to referenced image(s) and the Threshold Sequence (0070,1B11) is not present. May be present otherwise if the referenced image(s) are grayscale.</emphasis></emphasis></para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para><emphasis role="bold"><emphasis role="underline">&gt;&gt;Referenced Image Sequence</emphasis></emphasis></para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para><emphasis role="bold"><emphasis role="underline">(0008,1140)</emphasis></emphasis></para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para><emphasis role="bold"><emphasis role="underline">1C</emphasis></emphasis></para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para><emphasis role="bold"><emphasis role="underline">Sequence of Items identifying images and optical paths that are defined in the enclosing Item of Advanced Blending Sequence (0070,1B01), to which this VOI LUT or Window Center and Width applies.</emphasis></emphasis></para>
                                    <para><emphasis role="bold"><emphasis role="underline">One or more Items shall be included in this Sequence.</emphasis></emphasis></para>
                                    <para><emphasis role="bold"><emphasis role="underline">Required if the VOI LUT transformation in this Item does not apply to all the images and frames and optical paths listed in the enclosing Item.</emphasis></emphasis></para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="3" rowspan="1">
                                    <para>
                                    <emphasis role="bold"><emphasis role="underline"><emphasis role="italic">&gt;&gt;&gt;Include <xref linkend="table_10-3" xrefstyle="select: label quotedtitle"/>
                                            </emphasis></emphasis></emphasis>
                                </para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para/>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para><emphasis role="bold"><emphasis role="underline">&gt;&gt;&gt;Referenced Optical Path Identifier</emphasis></emphasis></para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para><emphasis role="bold"><emphasis role="underline">(006A,000E)</emphasis></emphasis></para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para><emphasis role="bold"><emphasis role="underline">1C</emphasis></emphasis></para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para><emphasis role="bold"><emphasis role="underline">A single value that identifies the optical path that is the input.</emphasis></emphasis></para>
                                    <para><emphasis role="bold"><emphasis role="underline">Refers to the value in Optical Path Identifier (0048,0106) within the Optical Path Sequence (0048,0105) in the images to which this Presentation State applies.</emphasis></emphasis></para>
                                    <para><emphasis role="bold"><emphasis role="underline">Required if there is more than one Optical Path in the images to which the this reference applies.</emphasis></emphasis></para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="3" rowspan="1">
                                    <para>
                                    <emphasis role="bold"><emphasis role="underline"><emphasis role="italic">&gt;&gt;Include <xref linkend="table_C.11-2b" xrefstyle="select: label quotedtitle"/>
                                            </emphasis></emphasis></emphasis>
                                </para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para/>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para><emphasis role="bold"><emphasis role="underline">&gt;Palette Color Lookup Table Sequence</emphasis></emphasis></para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para><emphasis role="bold"><emphasis role="underline">(0048,0120)</emphasis></emphasis></para>
                                    </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para><emphasis role="bold"><emphasis role="underline">1C</emphasis></emphasis></para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para><emphasis role="bold"><emphasis role="underline">For grayscale input images, provides a transform of the pixel values into an RGB color representation.</emphasis></emphasis></para>
                                    <para><emphasis role="bold"><emphasis role="underline">Only a single Item is permitted in this Sequence.</emphasis></emphasis></para>
                                <para><emphasis role="bold"><emphasis role="underline">Required if the input is a grayscale image and a palette color transformation is to be applied rather than the default transformation to equal RGB values. See <xref linkend="sect_N.2.6" xrefstyle="select: label quotedtitle"/> in PS3.4 <!--<olink targetdoc="PS3.4" targetptr="sect_N.2.6" xrefstyle="template:Section %n “%t” in PS3.4"/>-->.</emphasis></emphasis></para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="3" rowspan="1">
                                    <para>
                                    <emphasis role="bold"><emphasis role="underline"><emphasis role="italic">&gt;&gt;Include <xref linkend="table_C.7-22a" xrefstyle="select: label quotedtitle"/>
                                            </emphasis></emphasis></emphasis>
                                </para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para/>
                                </td>
                            </tr>
                            
                            
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_a570d20f-c787-4eda-9d68-ee4dae60be4b">&gt;Time Series Blending</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_ef70950e-5ce0-4859-bbee-894b0cd0d392">(0070,1B07)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_105c857c-a35b-4806-8d62-cbc5b276db67">1C</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_472f86ea-0034-449b-8075-41019a6f7f74">Whether this Series is a time series and every time point in this Series will be blended with the single volumes in the other Series specified in the Advanced Blending Sequence (0070,1B01).</para>
                                    <variablelist spacing="compact">
                                        <title>Enumerated Values:</title>
                                        <varlistentry>
                                            <term>TRUE</term>
                                            <listitem>
                                                <para xml:id="para_a2bca44d-a18a-444c-bd3a-cc6cc6d2a623"/>
                                            </listitem>
                                        </varlistentry>
                                        <varlistentry>
                                            <term>FALSE</term>
                                            <listitem>
                                                <para xml:id="para_acc62819-b9a4-4d88-a587-ab1f7a7fe7f7"/>
                                            </listitem>
                                        </varlistentry>
                                    </variablelist>
                                    <para xml:id="para_4f5e4850-d979-4c52-85a0-feff68c6bfd3">Only a single Item in the Sequence may have the value TRUE</para>
                                    <para xml:id="para_69794620-41be-450b-8792-a2a5f533ceae">Required if the input is a Time Series and all time points need to be blended with the other Series. May be present otherwise</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_73327640-db45-4182-b5d6-b9f561ded9ef">&gt;Geometry for Display</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_d2ed7c94-11a3-42ec-aaee-f6d1ab91c1cc">(0070,1B08)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_95aa3373-6b34-4c63-97b7-e8b29f74ffbe">1C</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_9c8d3b23-ce73-4c4a-9937-a3f18e20b86c">Whether the geometry of this Series is used as the geometry for the blending operation.</para>
                                    <para xml:id="para_24ca481a-74b3-4940-895b-7e3b8f8da983">See <xref linkend="sect_C.11.33.1.1" xrefstyle="select: label"/>.</para>
                                    <variablelist spacing="compact">
                                        <title>Enumerated Values:</title>
                                        <varlistentry>
                                            <term>TRUE</term>
                                            <listitem>
                                                <para xml:id="para_b41771dd-c384-45d7-b9e2-48711cec9e9e"/>
                                            </listitem>
                                        </varlistentry>
                                        <varlistentry>
                                            <term>FALSE</term>
                                            <listitem>
                                                <para xml:id="para_8bf0935e-d2eb-42ab-a653-b5bd2ccf5eff"/>
                                            </listitem>
                                        </varlistentry>
                                    </variablelist>
                                    <para xml:id="para_9ec418f1-d126-4f4f-b6c8-da08d7b55336">Only a single Item in the Sequence may have the value TRUE</para>
                                    <para xml:id="para_4169114f-c4af-4b43-8984-0fc18bb94ce6">Required if the geometry of this input is to be used as the geometry for the blending operation. May be present otherwise.</para>
                                </td>
                            </tr>
                        </tbody>
                    </table>
                <section label="C.11.33.1" status="3" xml:id="sect_C.11.33.1">
                    <title>Advanced Blending Presentation State Module Attributes</title>
                    <section label="C.11.33.1.1" status="3" xml:id="sect_C.11.33.1.1">
                        <title>Advanced Blending Sequence</title>
                        <para xml:id="para_877108a1-cb61-475f-8de5-bda527bda3e1">The images to be blended are specified in the Advanced Blending Sequence (0070,1B01).</para>
                        <para xml:id="para_3fc868c0-b1d1-44ae-9809-2b49ca39dc90">Geometry refers to the following Attributes:</para>
                        <itemizedlist>
                            <listitem>
                                <para xml:id="para_202d2643-8bb9-42bc-95db-87fc4d23475f">Number of Frames (0028,0008)</para>
                            </listitem>
                            <listitem>
                                <para xml:id="para_236aa23e-cb70-4bf0-80d5-267437d55fc9">Rows (0028,0010)</para>
                            </listitem>
                            <listitem>
                                <para xml:id="para_f6f98386-5c6b-4f55-9847-ff0e3ece9e4a">Columns (0028,0011)</para>
                            </listitem>
                            <listitem>
                                <para xml:id="para_b2fda60d-2104-45cc-a0b5-a7afd144630d">Pixel Aspect Ratio (0028,0034)</para>
                            </listitem>
                            <listitem>
                                <para xml:id="para_2abaa79d-b9a5-4b4a-b638-4e38e2817a12">Imager Pixel Spacing (0018,1164) (if applicable)</para>
                            </listitem>
                            <listitem>
                                <para xml:id="para_42288f66-7d8a-4400-a016-e32feadc73eb">Nominal Scanned Pixel Spacing (0018,2010) (if applicable)</para>
                            </listitem>
                            <listitem>
                                <para xml:id="para_cfeaddf8-adc4-4c48-84da-993d8d5d1822">Slice Thickness (0018,0050)</para>
                            </listitem>
                            <listitem>
                                <para xml:id="para_1580d6d7-a9d6-4765-a212-876b7a723b0a">Spacing Between Slices (0018,0088)</para>
                            </listitem>
                        </itemizedlist>
                        <para xml:id="para_5325c33e-11bb-4979-9917-5ade2b473961">The geometry of the output images shall match the geometry of the input specified as having the Geometry for Display (0070,1B08) as TRUE. If no input has Geometry for Display (0070,1B08) equal TRUE then the application shall choose which geometry to use.</para>
                        <para xml:id="para_79edc38a-ae9d-4878-b5f8-2509dc259349">If all inputs do not have the same frames, resolution, etc., the application determines the appropriate pixels to be blended. This implies that the blending result may vary between different applications. If consistent blending results are desired the inputs should be resampled in advance such that they have the same frames, resolution, etc.</para>
                        <para xml:id="para_51fccc03-a33e-4750-b39b-42333a1068e0">This Module specifies no explicit relationship (such as pairing or ordering) between the sets of images and frames defined in the Sequence Attributes. This Module does not define how the images are spatially related, and what re-sampling, if any, needs to be performed before the images are blended for rendering.</para>
                        <para xml:id="para_6fee476d-4829-4bc7-a109-4a70a6eabe18">It is expected that blending takes place between pixels at the same position in space.</para>
                        <para xml:id="para_909f4d81-bd10-4aff-8c37-f8a56be3e96b">The images in the image sets may share the same Frame of Reference, in which case the rendering application can spatially relate the image sets based on their Image Position (Patient) (0020,0032) and Image Orientation (Patient) (0020,0037) Attributes.</para>
                        <para xml:id="para_c143b853-499c-4a4a-9464-2aabbb1b5c0f">If a spatial registration object is included in the Advanced Blending Sequence (0070,1B01) it shall be applied to the referenced input even if the Frame of Reference is the same, as small corrections might have taken place during post-processing.</para>
                        <para xml:id="para_18304dff-76f8-4421-8625-ce2f2a9ed62a">If they are both missing the application will decide how to blend the inputs.</para>
                        <note>
                            <para xml:id="para_94b29b2b-87d6-47bc-9f53-5384c3a9b78b">The underlying image for a superimposed segmentation image need not be the source image for the segmentation.</para>
                        </note>
                    </section>
                    <section label="C.11.33.1.2" status="3" xml:id="sect_C.11.33.1.2">
                        <title>Threshold Sequence Macro</title>
                        <para xml:id="para_404b0ef7-914c-4ff9-abb7-3e1613618ce8">The Threshold Sequence is defining the values of the image that are used or ignored.</para>
                        <table frame="box" label="C.11.33.1-1" rules="all" xml:id="table_C.11.33.1-1">
                            <caption>Threshold Sequence Macro Attributes</caption>
                            <thead>
                                <tr valign="top">
                                    <th align="center" colspan="1" rowspan="1">
                                        <para xml:id="para_f57cb389-7b2c-4175-8647-a1b530e15a63">Attribute Name</para>
                                    </th>
                                    <th align="center" colspan="1" rowspan="1">
                                        <para xml:id="para_07713e1c-f153-45e9-bd99-a989257a5dfa">Tag</para>
                                    </th>
                                    <th align="center" colspan="1" rowspan="1">
                                        <para xml:id="para_4463f51f-0d1a-44ef-95f2-c68fea41f40e">Type</para>
                                    </th>
                                    <th align="center" colspan="1" rowspan="1">
                                        <para xml:id="para_f061c1d1-7352-4335-8f5c-f4fc5ecdcdd4">Attribute Description</para>
                                    </th>
                                </tr>
                            </thead>
                            <tbody>
                                <tr valign="top">
                                    <td align="left" colspan="1" rowspan="1">
                                        <para xml:id="para_0244a679-2ee6-4f49-8fb8-264d9f6cb610">Threshold Sequence</para>
                                    </td>
                                    <td align="center" colspan="1" rowspan="1">
                                        <para xml:id="para_18e839e2-7733-4a28-8513-e3861324c38f">(0070,1B11)</para>
                                    </td>
                                    <td align="center" colspan="1" rowspan="1">
                                        <para xml:id="para_13ee584e-2357-4378-954b-c136c5cce469">1C</para>
                                    </td>
                                    <td align="left" colspan="1" rowspan="1">
                                        <para xml:id="para_dc7c878b-9519-45ae-872b-a105ea96adbd">Threshold specification for the image</para>
                                        <para xml:id="para_0b690d86-ba88-48fe-a778-b7a6d7a55a14">One or more Items shall be included in this Sequence.</para>
                                        <para xml:id="para_d5b9c1bd-db47-4c27-9faf-971074887fe3">Required if Threshold needs to be applied<emphasis role="bold"><emphasis role="underline"> and the Softcopy VOI LUT Sequence (0028,3110) is not present. May be present otherwise if Threshold needs to be applied</emphasis></emphasis>.</para>
                                    </td>
                                </tr>
                                <tr valign="top">
                                    <td align="left" colspan="1" rowspan="1">
                                        <para xml:id="para_a705002e-ac0d-4d60-8eb6-69131b0c11e1">&gt;Threshold Type</para>
                                    </td>
                                    <td align="center" colspan="1" rowspan="1">
                                        <para xml:id="para_c37cbade-367c-48d4-abc0-50149777943f">(0070,1B13)</para>
                                    </td>
                                    <td align="center" colspan="1" rowspan="1">
                                        <para xml:id="para_63809ccd-4469-45d0-9bd6-52ed78290d46">1</para>
                                    </td>
                                    <td align="left" colspan="1" rowspan="1">
                                        <para xml:id="para_20e1f9be-04f4-483d-bdfe-03228e6de205">Type of threshold comparison used to determine the presence of a pixel value in the output.</para>
                                        <para xml:id="para_c7d7e2b9-c60a-46de-8957-00d622b213e3">See <xref linkend="sect_C.11.33.1.2.1" xrefstyle="select: label"/>.</para>
                                    </td>
                                </tr>
                                <tr valign="top">
                                    <td align="left" colspan="1" rowspan="1">
                                        <para xml:id="para_d1fb8d69-9da7-428c-b7b0-1048b9a71dca">&gt;Threshold Value Sequence</para>
                                    </td>
                                    <td align="center" colspan="1" rowspan="1">
                                        <para xml:id="para_6fa13330-ad0c-4e42-ba36-aa5c6c966198">(0070,1B12)</para>
                                    </td>
                                    <td align="center" colspan="1" rowspan="1">
                                        <para xml:id="para_bd16ddce-408b-4c62-8e2a-a5145221c657">1</para>
                                    </td>
                                    <td align="left" colspan="1" rowspan="1">
                                        <para xml:id="para_c7cea7d9-1923-4954-8700-bd874218b2dc">Values for thresholding.</para>
                                        <para xml:id="para_b9249f3d-e6eb-472b-992b-7b86ea9b28da">If the Threshold Type (0070,1B13) is GREATER_OR_EQUAL, LESS_OR_EQUAL, GREATER_THAN or LESS_THAN only a single Item shall be included in this Sequence.</para>
                                        <para xml:id="para_a8de41d5-e35c-4bb8-a583-da815677fa39">If the Threshold Type (0070,1B13) is RANGE_INCL or RANGE_EXCL, exactly two Items shall be included in this Sequence, the first of which is less than or equal to the second.</para>
                                    </td>
                                </tr>
                                <tr valign="top">
                                    <td align="left" colspan="1" rowspan="1">
                                        <para xml:id="para_ae790ae5-ec0c-4abc-baeb-040380612ee9">&gt;&gt;Threshold Value</para>
                                    </td>
                                    <td align="center" colspan="1" rowspan="1">
                                        <para xml:id="para_54ae8d85-4261-4d70-9d63-24a4b87893cc">(0070,1B14)</para>
                                    </td>
                                    <td align="center" colspan="1" rowspan="1">
                                        <para xml:id="para_cda7ca17-7172-40a0-9d76-42469d1ac4ca">1</para>
                                    </td>
                                    <td align="left" colspan="1" rowspan="1">
                                        <para xml:id="para_591fc6de-94a3-45c2-a970-7d1be27959e7">Value for the threshold.</para>
                                        <para xml:id="para_1b548b61-0d6e-45f3-81f4-e05afb7245de">See <xref linkend="sect_C.11.33.1.2.1" xrefstyle="select: label"/>.</para>
                                        <note>
                                            <para xml:id="para_59751ffb-7eff-4e30-83bf-b3d5a6a0b96d">VR of Threshold Value is FD and if this is different than the VR of the pixels then format conversion is needed.</para>
                                        </note>
                                    </td>
                                </tr>
                            </tbody>
                        </table>
                        <section label="C.11.33.1.2.1" status="5" xml:id="sect_C.11.33.1.2.1">
                            <title>Threshold</title>
                            <para xml:id="para_0502cfe7-5349-47ac-bed1-258a48867734">The Threshold Type (0070,1B13) defines the pixel values that will be shown. Values that are not inside any of the specified Threshold Sequence (0070,1B11) Items shall be treated as padding pixels.</para>
                            <para xml:id="para_99aec10b-9edc-4a67-8ba4-26b6bf48f380">To describe a threshold that consists of more than one range, multiple Items are specified in the Threshold Sequence (0070,1B11) Items.</para>
                            <para xml:id="para_78d7b0d3-c98b-49a7-bd91-1c538882b60a">When more than one Item is specified in the Threshold Sequence (0070,1B11) the pixel shall be shown if any Item specifies that the pixel shall be shown.</para>
                            <para xml:id="para_0716c7ff-f0e6-4cee-bbb9-56d906b8d1a1">The number of Items in the Threshold Value Sequence (0070,1B12) and the use of Threshold Value (0070,1B14) depends on the value of the Threshold Type (0070,1B13) as follows:</para>
                            <variablelist spacing="compact">
                                <title>Enumerated Values for Threshold Type (0070,1B13):</title>
                                <varlistentry>
                                    <term>RANGE_INCL</term>
                                    <listitem>
                                        <para xml:id="para_380f7126-063d-441e-91ea-3595a1e18c03">A pixel value shall be shown when the value lies between the specified values or is equal to one of the specified values.</para>
                                    </listitem>
                                </varlistentry>
                                <varlistentry>
                                    <term>RANGE_EXCL</term>
                                    <listitem>
                                        <para xml:id="para_7923083c-a7f4-4d27-9b9f-9c6524e0f67f">A pixel value shall be shown when the value lies outside (i.e. not between) the specified values.</para>
                                    </listitem>
                                </varlistentry>
                                <varlistentry>
                                    <term>GREATER_OR_EQUAL</term>
                                    <listitem>
                                        <para xml:id="para_7e61dee0-350f-4e94-afcb-470214440b82">A pixel value shall be shown when the value is greater than or equal to the specified value.</para>
                                    </listitem>
                                </varlistentry>
                                <varlistentry>
                                    <term>LESS_OR_EQUAL</term>
                                    <listitem>
                                        <para xml:id="para_b5f455b2-2432-44c4-8668-4af160adc212">A pixel value shall be shown when the value is less than or equal to the specified value.</para>
                                    </listitem>
                                </varlistentry>
                                <varlistentry>
                                    <term>GREATER_THAN</term>
                                    <listitem>
                                        <para xml:id="para_a6804bf3-20bc-4e0f-8250-6bd9af87fa75">A pixel value shall be shown when the value is greater than the specified value.</para>
                                    </listitem>
                                </varlistentry>
                                <varlistentry>
                                    <term>LESS_THAN</term>
                                    <listitem>
                                        <para xml:id="para_8c64fa86-585a-4933-bbc3-1b116cbd2ebf">A pixel value shall be shown when the value is less than the specified value.</para>
                                    </listitem>
                                </varlistentry>
                            </variablelist>
                        </section>
                    </section>
                </section>
            </section>





        <informaltable rules="all" frame="box">
            <tr>
                <td>
                    <para><emphasis role="italic">DICOM PS3.3 for reference unchanged:</emphasis></para>
                </td>
            </tr>
        </informaltable>
        
        
        <section label="C.11.34" status="3" xml:id="sect_C.11.34">
                <title>Advanced Blending Presentation State Display Module</title>
                <para xml:id="para_5edaf6bd-a6bc-47af-aa6c-cab50b41febe">The Advanced Blending Presentation State Display Module specifies the input and the method used for each blending step.</para>
                <table frame="box" label="C.11.34.1-1" rules="all" xml:id="table_C.11.34.1-1">
                    <caption>Advanced Blending Presentation State Display Module Attributes</caption>
                    <thead>
                        <tr valign="top">
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_095e3e97-b92e-4154-af06-7012d0b8bef7">Attribute Name</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_1b6c65ee-6b19-4964-8285-c0e57489b3ab">Tag</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_e8c09675-f232-4834-8863-432c3aa27698">Type</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_d5e15001-0b1e-44cb-9fc0-f19ea6d870c4">Attribute Description</para>
                            </th>
                        </tr>
                    </thead>
                    <tbody>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_8910b4ba-c8f8-4506-a9b7-b6a08a03b527">Pixel Presentation</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_e100ee11-3470-4607-bf1d-1de4dcb2e00c">(0008,9205)</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_f66ecca6-f7db-4d69-a0c8-4e9752dcec71">1</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_d5a62674-561f-4c94-8cce-ebdbbb0cdd89">Grayscale or color space of the Presentation State output.</para>
                                <variablelist spacing="compact">
                                    <title>Enumerated Values:</title>
                                    <varlistentry>
                                        <term>TRUE_COLOR</term>
                                        <listitem>
                                            <para xml:id="para_29e1fe17-4bbc-4b15-9aa2-b2764f7a28c3">Output consists of PCS-Values</para>
                                        </listitem>
                                    </varlistentry>
                                </variablelist>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_c40dc970-0aa5-4c94-b5c3-75ce9dca81c2">Blending Display Sequence</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_1fd5eaa0-1b32-4fb9-8b4e-bc694ca4088b">(0070,1B04)</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_5a1c3cc9-7f1f-4189-8e84-ba6f42c60afd">1</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_bf07aa02-185c-415b-943c-416e817bd357">The blending operations and the input series to be used. Each Item results in a single RGB output that may be reused in a following step.</para>
                                <para xml:id="para_be93e906-f1ca-4f5f-966f-5e85b76fd1a6">One or more Items shall be included in this Sequence.</para>
                                <note>
                                    <para xml:id="para_403a3656-5384-4bdf-a484-6c7556882aa9">The order of operations is determined by the specified Blending Input Number (0070,1B02) for each blending step. The displayed output will be the result of the blending step that is not assigned a Blending Input Number (0070,1B02)</para>
                                </note>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_2fb72055-4388-4fbd-bc8e-8fe866877004">&gt;Blending Display Input Sequence</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_cfda1ade-f4a4-4b0e-977e-48415a299aad">(0070,1B03)</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_4060c9cd-5269-4cf3-9328-308e86dd89c6">1</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_5d4e6d6c-b16f-496e-954b-13a184e13cf4">Each Item is an input series that will be used in the blending operation.</para>
                                <para xml:id="para_7ef0726a-005d-45c0-9cfa-6a7ed64488a0">The order of Items is significant</para>
                                <para xml:id="para_09a6283a-b5e1-4409-a7c0-8103ad8287ee">Two Items shall be included if the blending mode is FOREGROUND.</para>
                                <para xml:id="para_4017cecc-dff0-4774-90be-6a7f511a2d14">One or more Items shall be included if the blending mode is EQUAL.</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_a82a4640-bf27-4b75-ba0b-abc62f815b10">&gt;&gt;Blending Input Number</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_84457399-6332-4a2d-9762-e2e14db670b4">(0070,1B02)</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_863b43fb-6b2e-48ef-9efd-d303e65e46b1">1</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_7b8389fd-6890-463d-a49d-9395829a35b0">Identification number of the input series to which the Blending information must be applied.</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_c13bd5fc-d1fc-4644-9d3e-b40a95533491">&gt;Relative Opacity</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_e132015b-3cb1-4cc7-94e7-41315f5696ee">(0070,0403)</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_68f07392-f1a4-40f8-801d-a96a9764f57d">1C</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_4204c91c-e257-45a1-9a04-50292547d2ea">Specifies Relative Opacity for the visible pixels of the set referenced by the first Blending Input Number (0070,1B02)</para>
                                <para xml:id="para_8da890d5-6f7e-4e37-b176-7517b4638810">Required if Blending Mode (0070,1B06) is equal to FOREGROUND</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_2232bc1a-d704-4eec-9844-ee7d4ac6f6e0">&gt;Blending Mode</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_5c6a5173-5b57-41b6-b46d-d35ca8c30e8c">(0070,1B06)</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_e56ea411-3336-4724-9d33-addb10707a0d">1</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_5cc92d2d-f080-4c75-8080-0a82d5ea5a5b">Describes the method for weighting the different input images during the blending operation:</para>
                                <variablelist spacing="compact">
                                    <title>Enumerated Values:</title>
                                    <varlistentry>
                                        <term>EQUAL</term>
                                        <listitem>
                                            <para xml:id="para_02f0ca0f-bbc9-4db0-8d25-805ef016b26d"/>
                                        </listitem>
                                    </varlistentry>
                                    <varlistentry>
                                        <term>FOREGROUND</term>
                                        <listitem>
                                            <para xml:id="para_8154e66d-1bd8-470e-8414-aa445e8fce95"/>
                                        </listitem>
                                    </varlistentry>
                                </variablelist>
                                <para xml:id="para_d16bb421-d539-459d-9450-10024142c949">See <olink targetdoc="PS3.4" targetptr="sect_N.2.6" xrefstyle="template:Section %n in PS3.4"/>.</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_b5ec0744-aafb-4bf3-aca1-6c3bf68a42fe">&gt;Blending Input Number</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_64987c50-1f71-4ea4-aff0-8fb379360113">(0070,1B02)</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_87c981bf-4495-4743-bf53-09499338d9e3">1C</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_1ebc82a7-bb74-42ec-a0b8-1fa040331a53">Identification of the result as input for a subsequent blending operation.</para>
                                <para xml:id="para_c8392688-8606-45b3-8b40-d6c6fbd1b14c">Required if the result is used for further Blending.</para>
                                <note>
                                    <para xml:id="para_36f01d2f-a0ae-45bb-b486-96d9bdd8497d">If this Attribute is not present the Item describes the final output of the blending process.</para>
                                </note>
                            </td>
                        </tr>
                    </tbody>
                </table>
            </section>
        
      <section label="C.7.9" status="3" xml:id="sect_C.7.9">
                <title>Palette Color Lookup Table Module</title>
          <para>...</para>
                <table frame="box" label="C.7-22a" rules="all" xml:id="table_C.7-22a">
                    <caption>Palette Color Lookup Table Macro Attributes</caption>
                    <thead>
                        <tr valign="top">
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_5baedec1-ab08-4504-a897-912bf7a2197b">Attribute Name</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_29dc4762-aca5-4551-bfe2-5481a7470201">Tag</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_cc35e274-60c1-43fa-a192-1972a9de183b">Type</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_75558297-30f6-4002-aed2-05c691ec74d3">Attribute Description</para>
                            </th>
                        </tr>
                    </thead>
                    <tbody>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_e4ab131b-dee8-4b76-bad7-4fd30df8ea79">Red Palette Color Lookup Table Descriptor</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_9a8dde96-a358-4a76-b629-3bc58648571e">(0028,1101)</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_c7338a19-26d8-4af9-9528-395955f952c8">1</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_b2799463-c18b-441a-a242-54d450a3e2c5">Specifies the format of the Red Palette Color Lookup Table Data (0028,1201). See <xref linkend="sect_C.7.6.3.1.5" xrefstyle="select: label"/> for further explanation.</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_8e4b2a05-89a1-4fb1-a2a8-1ada7231be22">Green Palette Color Lookup Table Descriptor</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_6589d7bd-52f5-4f6f-9518-376c3bcf66b4">(0028,1102)</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_1e84806d-f480-4029-bb56-aa46b7b3f886">1</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_9513e3dc-1e89-4b67-b5ec-e1cfa6b5b777">Specifies the format of the Green Palette Color Lookup Table Data (0028,1202). See <xref linkend="sect_C.7.6.3.1.5" xrefstyle="select: label"/> for further explanation.</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_3a3336f3-3a0d-4183-8425-599e7077e879">Blue Palette Color Lookup Table Descriptor</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_d1e1278f-767e-4e02-bb8d-2192c67d884d">(0028,1103)</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_37ccb353-4b27-457b-b5a5-2bfe318e192e">1</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_c39c2f9e-5b9d-4429-a52f-a9a93ca94e40">Specifies the format of the Blue Palette Color Lookup table Data (0028,1203). See <xref linkend="sect_C.7.6.3.1.5" xrefstyle="select: label"/> for further explanation.</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_495a143e-0e9d-48f7-806e-ee1e1d0eaf37">Palette Color Lookup Table UID</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_04e273c0-66a7-47e7-adcc-24f27c3b0230">(0028,1199)</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_82f389a7-a4d0-437c-b4a3-0ef34ebf3393">3</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_58c7e439-c35f-4821-829b-516f02c4f491">Palette Color Lookup Table UID. See <xref linkend="sect_C.7.9.1" xrefstyle="select: label"/> for further explanation.</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_30efbc10-2ba4-4f2b-b184-6a5ee92c72d4">Red Palette Color Lookup Table Data</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_181d145e-3bea-48ec-ab36-3e68424fed53">(0028,1201)</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_9064b619-dba1-4bcf-adca-c388f857c48c">1C</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_5b4ff6e4-f797-4161-ad59-6b30e8c2ccd4">Red Palette Color Lookup Table Data. Required if segmented data is NOT used in an Image IOD or Color Palette IOD, or if the IOD is a Presentation State IOD. See <xref linkend="sect_C.7.6.3.1.6" xrefstyle="select: label"/> for further explanation.</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_6ad29a05-6f88-451d-8614-e7d1d577281b">Green Palette Color Lookup Table Data</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_73d637ce-8566-4806-8696-99d6e150f533">(0028,1202)</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_3936d05f-19cf-4d13-9006-22f302aa5325">1C</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_103d12c2-b21c-4f1f-a0da-fa4aa39f47eb">Green Palette Color Lookup Table Data. Required if segmented data is NOT used in an Image IOD or Color Palette IOD, or if the IOD is a Presentation State IOD. See <xref linkend="sect_C.7.6.3.1.6" xrefstyle="select: label"/> for further explanation.</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_2ee679bb-62c5-4ee7-b42b-2444c7f978ab">Blue Palette Color Lookup Table Data</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_26082704-aac2-4621-a284-24dbbdefac0f">(0028,1203)</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_b06032da-e11f-4ce0-a12b-93de37ab733a">1C</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_7a7b4b13-0be6-45e9-8ab2-76adcd717232">Blue Palette Color Lookup Table Data. Required if segmented data is NOT used in an Image IOD or Color Palette IOD, or if the IOD is a Presentation State IOD. See <xref linkend="sect_C.7.6.3.1.6" xrefstyle="select: label"/> for further explanation.</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_fb67d1ac-7c13-49ed-a2db-e271f4d1820f">Segmented Red Palette Color Lookup Table Data</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_28a23325-e24a-4b58-8b30-a1dcae204797">(0028,1221)</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_5a268220-5c17-4eee-92ec-c6301c2fe257">1C</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_4bb50d53-a9b5-40e2-9b5b-b26d3cb5e333">Segmented Red Palette Color Lookup Table Data. Required if segmented data is used in an Image IOD or Color Palette IOD; shall not be present in a Presentation State IOD. See <xref linkend="sect_C.7.9.2" xrefstyle="select: label"/> for further explanation.</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_b128854b-5337-4561-b8bb-21fb4eac429b">Segmented Green Palette Color Lookup Table Data</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_9b354717-b47e-4b7d-8726-c7772521a4f7">(0028,1222)</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_cd11bc94-f14b-4098-ac75-8de8d0fadcd1">1C</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_7dde8e4c-ef72-4fb9-acda-6f6cdc1d67ca">Segmented Green Palette Color Lookup Table Data. Required if segmented data is used in an Image IOD or Color Palette IOD; shall not be present in a Presentation State IOD See <xref linkend="sect_C.7.9.2" xrefstyle="select: label"/> for further explanation.</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_5177db41-e98c-4f8d-a528-ce05e67ee723">Segmented Blue Palette Color Lookup Table Data</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_435fca5a-41e5-4dde-86c7-3e100c485613">(0028,1223)</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_669d621e-2fb9-4c2c-95ea-4a69070c13a6">1C</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_a6dd4dad-719b-482d-a237-093741ad8200">Segmented Blue Palette Color Lookup Table Data. Required if segmented data is used in an Image IOD or Color Palette IOD; shall not be present in a Presentation State IOD. See <xref linkend="sect_C.7.9.2" xrefstyle="select: label"/> for further explanation.</para>
                            </td>
                        </tr>
                    </tbody>
                </table>
            </section>
      
      <section label="C.11.2" status="3" xml:id="sect_C.11.2">
                <title>VOI LUT Module</title>
                <para xml:id="para_427f5eb6-cfd8-4e9e-8047-e9b3afde44c0">...</para>
                
                <table frame="box" label="C.11-2b" rules="all" xml:id="table_C.11-2b">
                    <caption>VOI LUT Macro Attributes</caption>
                    <thead>
                        <tr valign="top">
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_d61a80d9-c22c-40b6-9a70-da1abd7d8159">Attribute Name</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_5ff85ee1-618c-427d-b75b-cf0014082389">Tag</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_b167d9fa-04ba-40e4-9465-9e8ba7c29a29">Type</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_51fb6b43-54d3-4c9b-85b0-962e1efc5984">Attribute Description</para>
                            </th>
                        </tr>
                    </thead>
                    <tbody>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_6cf459c0-7173-4ade-a142-9a28ee76ced9">VOI LUT Sequence</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_0d25a715-e25c-45e2-ba38-744d0db5a094">(0028,3010)</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_8594621f-b321-4047-8276-28d77fb9208e">1C</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_e9d7f59c-dae8-43aa-9257-a5fa93ce87d2">Defines a Sequence of VOI LUTs.</para>
                                <para xml:id="para_7dfd3858-7bea-4c48-b69e-e04cf2be8f40">One or more Items shall be included in this Sequence.</para>
                                <para xml:id="para_bec9ea36-dd19-4de8-84ef-55f5e23aac94">Required if Window Center (0028,1050) is not present. May be present otherwise.</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_8a2bfbb9-3274-4501-a2d5-94c62259463c">&gt;LUT Descriptor</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_9d4d0522-cb7b-4603-b55e-a54156b60283">(0028,3002)</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_c1877125-2538-436b-a1c6-c735d1f1f397">1</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_e3a75469-b733-42c9-930c-908f99ebb5de">Specifies the format of the LUT Data in this Sequence.</para>
                                <para xml:id="para_9ec79c72-d1e5-4ed4-a77c-6d75a2a1a938">See <xref linkend="sect_C.11.2.1.1" xrefstyle="select: label"/> for further explanation.</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_19017ebb-706f-4867-95a3-8e80326fa0c9">&gt;LUT Explanation</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_a6dfd43d-7154-4d72-a106-104d9e0e8eb4">(0028,3003)</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_313b5383-6bc8-45a4-9696-29df190b152e">3</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_e5bf817d-9db5-45d4-a05e-55cca8df9728">Free form text explanation of the meaning of the LUT.</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_b8bd4f5c-b956-4898-af4b-5a15e1686c7a">&gt;LUT Data</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_94ddbc95-5145-40a0-98e0-3c5e8855eee1">(0028,3006)</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_81405fa2-da86-4fdf-a6f4-b6b9ec4175a9">1</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_23b298c5-7b67-4821-9f44-4f67827307b7">LUT Data in this Sequence.</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_ebfe0812-3506-4441-90ca-9d771244fd6d">Window Center</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_a479cde9-702a-4918-9127-ee84675abdd1">(0028,1050)</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_79dea68f-6f1a-4a66-927f-bb8724bfc2b8">1C</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_87f0bc34-2bec-4b70-8e14-61fe3ea63316">Window Center for display.</para>
                                <para xml:id="para_69a7673b-8827-444e-b7e9-ed827d88e883">See <xref linkend="sect_C.11.2.1.2" xrefstyle="select: label"/> for further explanation.</para>
                                <para xml:id="para_1c4f5fdf-8a57-4099-815b-a0e948a4847d">Required if VOI LUT Sequence (0028,3010) is not present. May be present otherwise.</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_1c20a9b7-de6c-4cf0-b6cb-fa8c1dd52cee">Window Width</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_c53fd4df-e371-42a8-b534-a89e39a7b015">(0028,1051)</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_74968fe9-0cac-4e60-b676-f8e7fe5ff85f">1C</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_ae20e976-7415-4c12-ae38-f45e88c5cc95">Window Width for display. See <xref linkend="sect_C.11.2.1.2" xrefstyle="select: label"/> for further explanation.</para>
                                <para xml:id="para_a010e7fe-73aa-4c7d-8699-9af7e685379a">Required if Window Center (0028,1050) is present.</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_35fcdca6-f277-426e-b323-aba0ce360fe9">Window Center &amp; Width Explanation</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_de6be26d-aa06-4925-b2bd-bcaa76908b10">(0028,1055)</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_97b53b00-bab7-4c08-b96c-d28955546a27">3</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_8e00e5fa-f6fe-4eb6-8042-e3aa5969271a">Free form explanation of the meaning of the Window Center and Width. Multiple values correspond to multiple Window Center and Width values.</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_adc6cae2-7c25-4e26-8da3-828c456a3163">VOI LUT Function</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_f7f3bcf1-f019-4f87-8e54-149b0d4e9137">(0028,1056)</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_de3379f7-24c9-4774-8ce1-6a15538db0a3">3</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_4936889c-430b-4ec8-8009-cb3da34249c4">Describes a VOI LUT function to apply to the values of Window Center (0028,1050) and Window Width (0028,1051).</para>
                                <para xml:id="para_b58b7a4e-cf27-45f2-ade8-b9318f231053">See <xref linkend="sect_C.11.2.1.3" xrefstyle="select: label"/> for further explanation.</para>
                                <variablelist spacing="compact">
                                    <title>Defined Terms:</title>
                                    <varlistentry>
                                        <term>LINEAR</term>
                                        <listitem>
                                            <para xml:id="para_e6775251-42ef-4fc3-9c9c-84e41f62892b"/>
                                        </listitem>
                                    </varlistentry>
                                    <varlistentry>
                                        <term>LINEAR_EXACT</term>
                                        <listitem>
                                            <para xml:id="para_50f27c93-4e1b-4062-8247-26baaec58f92"/>
                                        </listitem>
                                    </varlistentry>
                                    <varlistentry>
                                        <term>SIGMOID</term>
                                        <listitem>
                                            <para xml:id="para_20a01234-c3b3-4083-abeb-556b03e1f3db"/>
                                        </listitem>
                                    </varlistentry>
                                </variablelist>
                                <para xml:id="para_75b1cfa0-ea7a-4290-a324-10947a739587">When this Attribute is not present, the interpretation of the values of Window Center (0028,1050) and Window Width (0028,1051) is linear as in <xref linkend="sect_C.11.2.1.2" xrefstyle="select: label"/>.</para>
                            </td>
                        </tr>
                    </tbody>
                </table>
                
            </section>

        <section label="10.3" status="2" xml:id="sect_10.3">
            <title>Image SOP Instance Reference Macro</title>
            <table frame="box" label="10-3" rules="all" xml:id="table_10-3">
                <caption>Image SOP Instance Reference Macro Attributes</caption>
                <thead>
                    <tr valign="top">
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_c004ece0-f1f7-4597-b689-6669d88c6579">Attribute Name</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_bd9d5b65-657d-40b0-b455-18d7b573e06f">Tag</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_89897585-548c-4f02-8dc6-0f4380e6e794">Type</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_7018df71-786b-45ee-8068-23e2bd049403">Attribute Description</para>
                        </th>
                    </tr>
                </thead>
                <tbody>
                    <tr valign="top">
                        <td align="left" colspan="3" rowspan="1">
                            <para xml:id="para_ae25377d-28bd-43a9-8653-b600c99dd819">
                                <emphasis role="italic">Include <xref linkend="table_10-11" xrefstyle="select: label quotedtitle"/>
                                </emphasis>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_88a4c8be-9e35-40b3-86f7-19d6dd498365"/>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_2789c5e8-a253-4237-8d95-72cdc43d1edf">Referenced Frame Number</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_58e622b7-4a47-4938-96dd-d3073aeea4b0">(0008,1160)</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_6acdc911-a087-43b4-a068-a8e756eaa132">1C</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_26b17b47-2064-4754-a98d-288d36b104d8">Identifies the frame numbers within the Referenced SOP Instance to which the reference applies. The first frame shall be denoted as frame number 1.</para>
                            <note>
                                <para xml:id="para_68450e47-9448-4ac5-82e8-4015a854968a">This Attribute may be multi-valued.</para>
                            </note>
                            <para xml:id="para_01a320b0-09fc-4c2e-a478-1eaa82de3416">Required if the Referenced SOP Instance is a multi-frame image and the reference does not apply to all frames, and Referenced Segment Number (0062,000B) is not present.</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_bbb881e3-b0b5-4a6d-bcb5-f4ffd5fbe837">Referenced Segment Number</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_d69702bd-c6a6-4cbd-86e0-7d71a791dc6c">(0062,000B)</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_3a684360-c4f6-4da5-bd83-8eb296e386d2">1C</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_d27bd98a-7385-426d-baa4-2fb5d7572d0c">Identifies the Segment Number to which the reference applies.</para>
                            <para xml:id="para_435cfa21-4755-4a64-847f-643b0ac2f953">Required if the Referenced SOP Instance is a Segmentation or Surface Segmentation and the reference does not apply to all segments and Referenced Frame Number (0008,1160) is not present.</para>
                        </td>
                    </tr>
                </tbody>
            </table>
        </section>
        
        <section label="10.8" status="2" xml:id="sect_10.8">
            <title>SOP Instance Reference Macro</title>
            <table frame="box" label="10-11" rules="all" xml:id="table_10-11">
                <caption>SOP Instance Reference Macro Attributes</caption>
                <thead>
                    <tr valign="top">
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_5a42bead-86fd-4f7b-8207-2e881965ab73">Attribute Name</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_70ae57b5-e439-4067-bc95-e80d14d1fa04">Tag</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_d34eb0c1-3332-4ca0-a083-ab76f3cb6533">Type</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_1de58578-d160-4168-9c8c-d2466626ce85">Attribute Description</para>
                        </th>
                    </tr>
                </thead>
                <tbody>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_01f3f1cd-dd4d-471c-b080-9961faf6d1fb">Referenced SOP Class UID</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_e2fab2a1-95b0-4b70-a566-6d84cf79331f">(0008,1150)</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_dc719629-f544-459b-93f4-ea556d10f880">1</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_63809a86-e616-4736-811b-d8de9a2a076c">Uniquely identifies the referenced SOP Class.</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_ee639991-ef92-4300-9a16-339d6e7d18b0">Referenced SOP Instance UID</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_cbe8fcac-5c05-4c18-ae71-d9bb8875962b">(0008,1155)</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_917913b4-a3fa-45eb-ab44-e0d4142b5626">1</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_e163fe69-a141-4d4a-8f47-dfaa443fdddb">Uniquely identifies the referenced SOP Instance.</para>
                        </td>
                    </tr>
                </tbody>
            </table>
        </section>
        
        <section label="C.17.2.1" status="4" xml:id="sect_C.17.2.1">
                    <title>Hierarchical SOP Instance Reference Macro</title>
                    <table frame="box" label="C.17-3" rules="all" xml:id="table_C.17-3">
                        <caption>Hierarchical SOP Instance Reference Macro Attributes</caption>
                        <thead>
                            <tr valign="top">
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_ab31d31f-96a1-4843-a3a1-c1fc96ec2e36">Attribute Name</para>
                                </th>
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_90a9e3e5-d0b8-4bec-bfbc-d39a189e3de3">Tag</para>
                                </th>
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_92b7deb4-fcb1-4d14-a831-3348425175bd">Type</para>
                                </th>
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_852eb5da-ce8d-462c-bfc0-2f88a8303a8a">Attribute Description</para>
                                </th>
                            </tr>
                        </thead>
                        <tbody>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_37d20de1-7c12-4af4-9d26-e81050acab54">Study Instance UID</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_63e61323-8580-4647-8b33-a185645dbb0f">(0020,000D)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_3af87ee9-6da4-449a-86a3-e34fc50c2b84">1</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_2846a110-ed5e-4c0f-8fa2-073c43d2d932">Unique identifier for the Study</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_a999e6f9-96a7-4aaa-9129-5e21b5280182">Referenced Series Sequence</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_2ee322d5-3db8-41fc-b032-fed8a21355ba">(0008,1115)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_a404ba42-072a-4d97-bd15-72c2db31db86">1</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_db0350e9-b0c3-46d8-b36c-083dfa0402fe">Sequence of Items where each Item includes the Attributes of a Series containing referenced Composite Object(s).</para>
                                    <para xml:id="para_94cde6b0-d53b-4c3a-a3eb-947f3cf969a1">One or more Items shall be included in this Sequence</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="3" rowspan="1">
                                    <para xml:id="para_5c3fd5f8-cab1-4912-8349-5ce7372a3a23">
                                        <emphasis role="italic">&gt;Include <xref linkend="table_C.17-3a" xrefstyle="select: label quotedtitle"/>
                                        </emphasis>
                                    </para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_251fe7c7-be85-4637-8058-448f6339365c"/>
                                </td>
                            </tr>
                        </tbody>
                    </table>
                    <table frame="box" label="C.17-3a" rules="all" xml:id="table_C.17-3a">
                        <caption>Hierarchical Series Reference Macro Attributes</caption>
                        <thead>
                            <tr valign="top">
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_6a21e836-e4bb-472f-9ae9-b632b9369882">Attribute Name</para>
                                </th>
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_6e27a4d7-a525-45b0-b4f0-084dc87b0ebc">Tag</para>
                                </th>
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_bde49c44-5ccd-4549-93c8-a8db2b2d3b73">Type</para>
                                </th>
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_bc6cbc22-6903-4a52-a3da-b92244086057">Attribute Description</para>
                                </th>
                            </tr>
                        </thead>
                        <tbody>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_2782fb56-eda6-476a-becb-1e0d7c0b98c4">Series Instance UID</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_c7694a90-3a89-4e9a-93bc-f8251734fd15">(0020,000E)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_724b2474-b0ba-4f06-9848-0e6fffdc25b3">1</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_2b365808-d4ca-41d3-bc2c-74fce7b0701c">Unique identifier of a Series that is part of this Study and contains the referenced Composite Object(s).</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_654303f6-344c-4ef1-b11e-71c6e5ab6eec">...</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_232d4920-991c-4591-ba15-a17dbce55e99"/>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_2d40f134-4499-439f-b067-8779cd253b5b"/>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_65fb5683-5c64-476c-8649-14e7eabead7f"/>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_0e49ff09-aa93-4ee0-96cd-6ea0bc1613bb">Referenced SOP Sequence</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_9b0f2527-ccda-4d77-b8b1-cdf47695f2cb">(0008,1199)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_d6240d1a-96d1-40d6-bbe3-e9e530340baf">1</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_12a432ca-5708-4fb4-8aa8-55f09f10349d">References to Composite Object SOP Class/SOP Instance pairs that are part of the Study defined by Study Instance UID and the Series defined by Series Instance UID (0020,000E).</para>
                                    <para xml:id="para_036aa0bb-9e3a-4ce9-ae37-5812e7f927e3">One or more Items shall be included in this Sequence.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="3" rowspan="1">
                                    <para xml:id="para_2ff94818-ad49-4bde-bf9a-607e66e896c4">
                                        <emphasis role="italic">&gt;Include <xref linkend="table_10-11" xrefstyle="select: label quotedtitle"/>
                                        </emphasis>
                                    </para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_f1d961b4-261f-41c3-abe0-1c02b37a0054"/>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_f3ef95be-8934-4c89-bd95-3fb3ee7d07b2">...</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_6ec55c5d-7115-44e6-82a8-3a91c432d52b"/>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_97757199-bd1d-44d0-b977-776411e05d27"/>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_75865631-ffc4-4bdd-aa78-2e90f6d64390"/>
                                </td>
                            </tr>
                        </tbody>
                    </table>
                </section>
        
        <informaltable rules="all" frame="box">
            <tr>
                <td>
                    <para><emphasis role="italic">Amend DICOM PS3.4 as follows (changes to existing text
                        are bold and <emphasis role="bold"><emphasis role="underline"
                            >underlined</emphasis></emphasis> for additions and <emphasis role="bold"
                                ><emphasis role="strikethrough">struckthrough</emphasis></emphasis> for
                        removals):</emphasis></para>
                </td>
            </tr>
        </informaltable>
        
        
        <section label="N.2.4" status="3" xml:id="sect_N.2.4">
            <title>Blending Transformations</title>
            <para xml:id="para_22f6af77-7bd4-4256-98a5-78950fca4957">The grayscale to color blending transformation model applies only to a pair of grayscale values, one of which is first mapped to color and then superimposed upon the other. The resulting values are device independent color PCS-Values. This process is illustrated in <xref linkend="figure_N.2-3" xrefstyle="select: label"/>.</para>
            <para xml:id="para_44d443a3-515b-496c-a6f5-1bb6ac8d0206">For the purpose of this section, pixels are referred to as stored pixel values and transformations are defined as point operations on these values. However, it is likely that pixels from either or both the superimposed and underlying image sets will have been spatially resampled and hence interpolated or replicated. Such operations do not affect the conceptual pipeline.</para>
            <para xml:id="para_b751caaa-c572-483d-acd7-a87de933f407">
                <figure label="N.2-3" pgwide="1" xml:id="figure_N.2-3">
                    <title>Grayscale to Color Blending Transformation Model</title>
                    <mediaobject>
                        <imageobject>
                            <imagedata fileref="cp2237/PS3.4_N.2-3.svg"/>
                            <!--<imagedata fileref="figures/PS3.4_N.2-3.svg"/>-->
                            <!--<imagedata fileref="part04_fromword_files/image030.png"/>-->
                        </imageobject>
                    </mediaobject>
                </figure>
            </para>
            <section label="N.2.4.1" status="4" xml:id="sect_N.2.4.1">
                <title>Underlying Image Pixels</title>
                <para xml:id="para_8aa0f5fc-b862-4369-ae61-25372f5e2e34">The Modality LUT and VOI LUT transformations are applied to the stored pixel values of the underlying image.</para>
                <para xml:id="para_29f6d31f-c9c6-4163-a43b-feb3bcae6b82">The output range of the VOI LUT transformation depends either on the width of the linear window or the range of output values of the LUT defined by the LUT Descriptor. Conceptually, for the purpose of describing the succeeding blending operation, the smallest pixel value from the range is mapped to 0.0 and the largest pixel value is mapped to 1.0 and all intermediate values are linearly mapped to the [0.0..1.0] interval.</para>
            </section>
            <section label="N.2.4.2" status="4" xml:id="sect_N.2.4.2">
                <title>Superimposed Image Pixels</title>
                <para xml:id="para_36ca4d75-07d8-40c6-a8cb-8a6e7d48a8d0">The Modality LUT and VOI LUT transformations are applied to the stored pixel values of the superimposed image.</para>
                <para xml:id="para_de1f8fed-41f7-4daf-bac5-2d224b15cd94">The full output range of the preceding VOI LUT transformation is implicitly scaled to the entire input range of the Palette Color LUT Transformation.</para>
                <para xml:id="para_e7f65a5f-4643-43e7-9cb3-76b4a8d9177f">The output range of the RGB values in the Palette Color LUT Transformation depends on the range of output values of the LUT defined by the LUT Descriptors. Conceptually, for the purpose of describing the succeeding blending operation, a LUT entry of 0 is mapped to 0.0 and the largest LUT entry possible is mapped to 1.0 and all intermediate values are linearly mapped to the [0.0..1.0] interval.</para>
                <note>
                    <para xml:id="para_d76567cf-1cad-4a20-8a45-7e70e35d7e5f">In practice, the Palette Color LUT output for the superimposed images is encoded in 8 or 16 bits and hence will have a range of 0 to 0xFF or 0xFFFF.</para>
                </note>
                <para xml:id="para_348cabfa-0c9f-4422-a837-170c32f4b92b">The Palette Color LUT used is that encoded in the Blending Presentation State; any Palette Color LUTs or Supplemental Palette Color LUTs in the image instances are ignored.</para>
            </section>
            <section label="N.2.4.3" status="4" xml:id="sect_N.2.4.3">
                <title>Blending Operation</title>
                <para xml:id="para_8eea2685-899f-49f3-af53-952158e5d2c1">The inputs to the blending operation are grayscale values from 0.0 to 1.0 from the underlying image (Y<subscript>u</subscript>) and RGB values from 0.0 to 1.0 from the superimposed image (RGB<subscript>s</subscript>), and an opacity value from 0.0 to 1.0 (A).</para>
                <para xml:id="para_919eb852-d27a-4162-bda8-ee171028b226">The output is a single image containing RGB values (RGB<subscript>o</subscript>) blended as:</para>
                <para xml:id="para_7fcc6b97-8b81-45cc-88b5-77c7c3b35814">R<subscript>o</subscript> = R<subscript>s</subscript> * A + Y<subscript>u</subscript> * (1-A)</para>
                <para xml:id="para_a9b0e0d5-eeeb-4e61-b8f4-2d5ad10427d3">G<subscript>o</subscript> = G<subscript>s</subscript> * A + Y<subscript>u</subscript> * (1-A)</para>
                <para xml:id="para_d0b7a924-e892-4589-a4a0-a0390507069a">B<subscript>o</subscript> = B<subscript>s</subscript> * A + Y<subscript>u</subscript> * (1-A)</para>
            </section>
            <section label="N.2.4.4" status="4" xml:id="sect_N.2.4.4">
                <title>Conversion to Profile Connection Space</title>
                <para xml:id="para_1ddbf3d8-9913-41db-9fec-38652a859da1">The output of the blending operation is implicitly scaled to the gamut of the hypothetical device described by the ICC Input Profile, resulting in PCS-Values.</para>
            </section>
        </section>
        
        
        <section label="N.2.6" status="3" xml:id="sect_N.2.6">
            <title>Advanced Blending Transformations</title>
            <para xml:id="para_0b8b5ee6-83f5-407b-836d-2612a490aa54">The advanced blending transformation model applies to multiple <emphasis role="bold"><emphasis role="underline">RGB </emphasis></emphasis>color inputs<emphasis role="bold"><emphasis role="underline">,</emphasis></emphasis> and uses foreground blending or equal blending.</para>
            <para xml:id="para_afc18214-5674-49fa-a09b-19bf99a033b6">Several transformations in this <emphasis role="bold"><emphasis role="strikethrough">IOD</emphasis><emphasis role="underline">model</emphasis></emphasis> affect the input prior to its use in blending as depicted in <xref linkend="figure_N.2.6-1" xrefstyle="select: label"/>.</para>
            <para><emphasis role="bold"><emphasis role="underline">Grayscale inputs that have grayscale transformations (Modality LUT, VOI LUT) and palette color transformations (Palette Color LUT) specified in the Presentation State shall have those transformations applied, as specified in <xref linkend="sect_N.2.4.2" xrefstyle="select: label"/>, to produce RGB values.</emphasis></emphasis></para>
            <para xml:id="para_26acfd62-9e52-4831-80b8-16b524d046e4">Grayscale inputs that have no associated <emphasis role="bold"><emphasis role="underline">Palette </emphasis></emphasis>Color LUT information <emphasis role="bold"><emphasis role="underline">specified in the Presentation State or in the reference Image Instance</emphasis></emphasis>shall <emphasis role="bold"><emphasis role="strikethrough">have the normal grayscale processing and then</emphasis></emphasis> be converted to a full color image by setting R equals G equals B.</para>
            <para xml:id="para_5560bdc1-c53e-4a34-b558-672e185ff280">
                <figure label="N.2.6-1" pgwide="1" xml:id="figure_N.2.6-1">
                    <title>Color and Threshold Application</title>
                    <mediaobject>
                        <imageobject>
                            <imagedata fileref="cp2237/PS3.4_N.2.6-1.svg"/>
                            <!--<imagedata fileref="figures/PS3.4_N.2.6-1.svg"/>-->
                            <!--<imagedata fileref="part04_fromword_files/sup189_image002.png"/>-->
                        </imageobject>
                    </mediaobject>
                </figure>
            </para>
            <para xml:id="para_06ba5515-747e-4522-9bf3-8f025548804f">Padding pixels in an input are given an opacity value zero and shall be set to 0 for Red, Green, and Blue.</para>
            <para xml:id="para_7a9f8aa0-367b-4560-9e76-f50865029531">The foreground method blends two inputs. The first input uses an opacity of Relative Opacity (0070,0403) and the second input uses an opacity of (1 - Relative Opacity (0070,0403) ).</para>
            <para xml:id="para_3d0ce242-34e9-436a-b34f-d935e2aa859d">If both the inputs are padding values then the result is padding value.</para>
            <para xml:id="para_085385d7-d6ad-42b9-a959-827f9f1836f3">If one of the values is padding value then the result is the non-padding value.</para>
            <para xml:id="para_2407be95-28b7-4983-aad9-3bb6932d88b9">If both pixels have values then result is Relative Opacity * first value + (1 - Relative Opacity) * second value.</para>
            <para xml:id="para_6e9562ff-b3bb-4a33-b102-c04eff529d61">
                <figure label="N.2.6-2" pgwide="1" xml:id="figure_N.2.6-2">
                    <title>Foreground Blending</title>
                    <mediaobject>
                        <imageobject>
                            <imagedata fileref="cp2237/PS3.4_N.2.6-2.svg"/>
                            <!--<imagedata fileref="figures/PS3.4_N.2.6-2.svg"/>-->
                            <!--<imagedata fileref="part04_fromword_files/sup189_image004.png"/>-->
                        </imageobject>
                    </mediaobject>
                </figure>
            </para>
            <para xml:id="para_6fede366-0355-478e-800e-f64353476cba">The Equal blending mode blends two or more inputs where for each pixel location the opacity is calculated as 1.0 divided by the number of non-padding pixels. The result pixel blends all non-padding pixels using the calculated opacity.</para>
            <para xml:id="para_dd913709-9c4d-4d50-b774-e6845e39ee73">If an input pixel value is the padding-value then the Relative Opacity for that input pixel is zero.</para>
            <para xml:id="para_87a2aba1-2726-466f-a3a3-0a494c8a5b05">If an input pixel value is not the padding value then the Relative Opacity for that pixel is 1 / (number of input pixels that are non-padding pixels).</para>
            <para xml:id="para_bd04c033-36fe-46a4-8579-a74c30dae53e">The result value is the sum for all input pixels of the input pixel value * Relative Opacity.</para>
            <para xml:id="para_e335de31-d4e6-46d5-98ea-2f2af913cb43">If all the inputs pixels are padding values then the result is padding value.</para>
            <para xml:id="para_87030dcd-2616-47f6-92e7-9b98f91fcaf0">
                <figure label="N.2.6-3" pgwide="1" xml:id="figure_N.2.6-3">
                    <title>Equal Blending</title>
                    <mediaobject>
                        <imageobject>
                            <imagedata fileref="cp2237/PS3.4_N.2.6-3.svg"/>
                            <!--<imagedata fileref="figures/PS3.4_N.2.6-3.svg"/>-->
                            <!--<imagedata fileref="part04_fromword_files/sup189_image006.png"/>-->
                        </imageobject>
                    </mediaobject>
                </figure>
            </para>
        </section>
        
        
        

        
        
        
        
        
        
        
        
    </chapter>
</book>