<book xmlns="http://docbook.org/ns/docbook" xmlns:xl="http://www.w3.org/1999/xlink" version="5.0">
    <title>CP-2215</title>
    <subtitle>Measurements and Categorical Statements on Patients, Studies and Series</subtitle>
    <chapter>
        <title></title>
        <informaltable rules="all" frame="box">
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Status</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>Final Text</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Date of Last Update</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>2022/12/01</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Person Assigned</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>David Clunie</para><para><link
                    xl:href="mailto:dclunie@dclunie.com"/></para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submitter Name</para></td>
                <td rowspan="1" colspan="1" valign="top">Imaging Data Commons (IDC)</td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submission Date</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>2022/02/11</para></td>
            </tr>
        </informaltable>
        <informaltable rules="all" frame="box">
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Correction Number CP-2215</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Log Summary: Measurements and Categorical Statements on Patients, Studies and Series</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">
                    <para>Name of Standard</para>
                    <para>PS3.6, PS3.16 2022d</para>
                </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">
                    <para>Rationale for Correction:</para>
                    <para>Some AI/ML and other image processing applications as well as human observers make categorical and numerical statements about patients, studies and series rather than images, frames or ROIs. It is useful to be able to include these in TID 1500 reports.</para>
                    <para>Update TID 1501 to allow references to patients, studies and series.</para>
                    <para><emphasis role="italic">[Ed.Note: CRS Request #773892 for 260893003.]</emphasis></para>
                </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">Correction Wording:</td>
            </tr>
        </informaltable>
    </chapter>
    <chapter>
        <title/>

        <informaltable rules="all" frame="box">
            <tr>
                <td>
                    <para><emphasis role="italic">Amend PS3.16 as follows (changes to existing text are bold and <emphasis role="bold"><emphasis role="underline">underlined</emphasis></emphasis> for additions and <emphasis role="bold"><emphasis role="strikethrough">struckthrough</emphasis></emphasis> for removals):</emphasis></para>
                </td>
            </tr>
        </informaltable>

        <section label="TID 1501" status="3" xml:id="sect_TID_1501">
            <title>Measurement and Qualitative Evaluation Group</title>
            <para xml:id="para_114cfb65-4682-44de-a13f-c1d06e3dbb6e">This Template groups measurements and/or qualitative evaluations into a Measurement Group.</para>
            <para xml:id="para_603f1d16-c3e0-4d64-8ce4-e00dc881174b">Each Measurement Group is identified by Tracking ID and UIDs, and may be described as having being made at a particular time point.</para>
            <para xml:id="para_0707a5ac-1bc8-458c-b2ee-72289f6e123a">Measurement groups may contain various common measurement modifiers that are shared by all measurements in the group, such as method and finding site.</para>
            <variablelist spacing="compact">
                <varlistentry>
                    <term>
                        <emphasis role="bold">Type:</emphasis>
                    </term>
                    <listitem>
                        <para xml:id="para_cc679c37-9059-4b23-8836-3fdf0046a0d4">
                            <emphasis role="bold">Extensible</emphasis>
                        </para>
                    </listitem>
                </varlistentry>
                <varlistentry>
                    <term>
                        <emphasis role="bold">Order:</emphasis>
                    </term>
                    <listitem>
                        <para xml:id="para_ea8614a5-4694-4cee-be77-09a96c05720c">
                            <emphasis role="bold">Non-Significant</emphasis>
                        </para>
                    </listitem>
                </varlistentry>
                <varlistentry>
                    <term>
                        <emphasis role="bold">Root:</emphasis>
                    </term>
                    <listitem>
                        <para xml:id="para_9ad71d29-1bd6-44df-b282-0b50a7388d3b">
                            <emphasis role="bold">No</emphasis>
                        </para>
                    </listitem>
                </varlistentry>
            </variablelist>
            <table frame="box" label="TID 1501" rules="all" xml:id="table_TID_1501" width="1191">
                <caption>Measurement and Qualitative Evaluation Group</caption>
                <thead>
                    <tr valign="top">
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_6cf3eeee-bde3-41f2-9cd2-d2a7c2fd15de"/>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_a753304a-333e-44a2-b1de-6a357d827b35">NL</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_9b361704-3515-44f8-bc8d-8b4c7bbd3bd6">Rel with Parent</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_3f88f263-0608-4fda-9eb2-629f8f00ab90">VT</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_472206d9-066c-4c71-9bbd-2114d729f75b">Concept Name</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_8720acfb-54df-451c-b098-dc55f146f6bd">VM</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_789f5c1b-14da-47e9-8b2d-91afa2c9f6ce">Req Type</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_e618c06c-2ed8-4cc8-86d8-570049b0290b">Condition</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_50c97dbc-38b3-4780-8666-e80d0e960f39">Value Set Constraint</para>
                        </th>
                    </tr>
                </thead>
                <tbody>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_1e553294-3b91-4308-9dd5-cab5993fc271">1</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_c32e330b-8290-4360-a024-2bf2bd867a5a"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_7999f1a1-e780-4573-be47-540a1718b75b">CONTAINS</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_1b3f4a25-408c-4c4d-8be6-1d600c149895">CONTAINER</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_1cc549fc-754a-4eaa-a2d3-6291e8750353">EV <link linkend="DCM_125007" xrefstyle="select: nopage">(125007, DCM, "Measurement Group")</link>
                            </para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_fcc4ffad-bb0d-49ad-b7fa-967199743ff2">1</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_b73d3ca3-05ba-48ac-9eab-b2f0e7f0df78">M</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_d35f0bc7-9294-4136-b82c-6d7c23f702b5"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_03be1fcd-9618-44fa-a630-706e645c6bf3"/>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">1a</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">&gt;</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">CONTAINS</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">CODE</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">EV <link linkend="DCM_130780" xrefstyle="select: nopage">(130780, DCM, "Specific observation subject class")</link></emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">1</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">U</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">B<xref linkend="sect_CID_272" xrefstyle="select: labelnumber quotedtitle nopage"/></emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_ff2f07db-7c32-467f-bb12-fe971b409b38">1b</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_6a520a91-4690-4471-b566-ea4eb14543a2">&gt;</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_8115766b-050b-4d5d-852f-e4e581690f79">HAS OBS CONTEXT</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_f646f123-7cea-4bca-a381-029fde6fc59a">TEXT</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_e3542061-a8ad-4a26-a63f-ca9784b061bd">EV <link xl:href="http://ncit.nci.nih.gov/ncitbrowser/ConceptReport.jsp?dictionary=NCI_Thesaurus&amp;code=C67447" xl:show="new">(C67447, NCIt, "Activity Session")</link>
                            </para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_ddaa1d32-2d8c-4385-ba25-45d27dfc1f25">1</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_3ae01147-d156-41b1-b8cf-80bae1934d65">U</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_023d7269-c7d0-45a6-a134-2c3f9b57a170"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_32313350-d035-44c7-a8fc-f965aee4cbf2"/>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_8e488327-ccf3-46ab-a700-fb2b68c49f6f">...</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_fe676738-d128-41c5-8c32-be3078edcce7"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_ad0f94dd-d2f6-499b-bcb5-49ee077ea2f0"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_20588b02-b2b9-45f8-93e8-103c4bcec4dd"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_9c14218d-ff9a-455a-92c9-fd8ec166d83a"/>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_03043cfc-538a-4d2e-8216-4fee57fd0024"/>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_80484d05-5972-495c-9601-26aa6843bbfe"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_d097d9ee-c976-4361-9f72-240a2c5405bc"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_3edc3b48-a8b6-46fb-a053-470c6feadc10"/>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_9d5a36cf-23f5-4c7d-a5d7-7ee4322f1b03">10</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_e58ce1bd-d0ac-448d-9b93-01f903241340">&gt;</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_6177297b-681a-4e6b-b8c7-2881cd7e0ba7">CONTAINS</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_262cd8dd-5c62-46dc-9228-61b78178bd26">INCLUDE</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_4f382802-3d6d-4d33-a37d-5208d882708c">D<xref linkend="sect_TID_300" xrefstyle="select: labelnumber quotedtitle nopage"/>
                            </para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_8e58ba2c-c810-4780-b9f5-3f7391d17037">1-n</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_680211e9-9f16-46b4-830e-f947d05bb711">U</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_f5faad4f-a65b-4bee-8c06-4b271b8f4559"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_f39a1c62-ce64-4943-a3b6-165b144f07e7">$Measurement = $Measurement</para>
                            <para xml:id="para_793d2241-8a62-4744-babe-23d6f78f652a">$Units = $Units</para>
                            <para xml:id="para_92e6f6ab-8253-4199-b880-698e0b3764e8">$ModType = $ModType</para>
                            <para xml:id="para_51b7a3f0-cef1-41be-ac3c-7328c2a23cf5">$ModValue = $ModValue</para>
                            <para xml:id="para_0195d2d8-4704-46aa-847f-888bf183efef">$Method = $Method</para>
                            <para xml:id="para_a598a4cf-16b8-41e9-8da5-e45fee215ba8">$Derivation = $Derivation</para>
                            <para xml:id="para_a520ab94-54af-48bc-94ad-1c0c012b35a8">$TargetSite = $TargetSite</para>
                            <para xml:id="para_9136f165-0728-41a9-846b-3c467174949e">$TargetSiteMod = $TargetSiteMod</para>
                            <para xml:id="para_618a925c-b9aa-4e68-9be9-69035614020c">$Equation = $Equation</para>
                            <para xml:id="para_d2c54ff0-c4f5-4375-8838-fde39cb9232d">$ImagePurpose = $ImagePurpose</para>
                            <para xml:id="para_5db5f2d5-f0c9-48b9-8259-d2c2b7f6d44c">$WavePurpose = $WavePurpose</para>
                            <para xml:id="para_d16f9c46-d989-48e8-8655-1fc0d42caced">$RefAuthority = $RefAuthority</para>
                            <para xml:id="para_874803a4-4c0f-45a0-b22b-6ee636108a9d">$RangeAuthority = $RangeAuthority</para>
                            <para xml:id="para_5b19ea79-97de-4f5e-87c4-8a04ddcc3740">$DerivationParameter = $DerivationParameter</para>
                            <para xml:id="para_9b9c9301-94db-4c83-af7d-b1da9193fccf">$DerivationParameterUnits = $DerivationParameterUnits</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_fe31cfd0-0ce4-4780-aaa7-eaec116a88fd">10b</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_7013753d-940e-4092-bc23-55abc268ff8d">&gt;</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_80f4174a-a81e-4fac-9333-06a8a1bddd77">CONTAINS</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_b3f8d549-5b2e-4bb3-87b6-7ef1760c3de4">IMAGE</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_e144bb01-a068-450e-9a9f-68975304aa86">$ImagePurpose </para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_790eb5d4-5524-4563-8523-19519e646139">1-n</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_7510fc8c-fb42-4790-8675-ad9dd9748131">U</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_a62bde1f-bd3e-4b39-94bc-af9912cea925"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1"/>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_3ce81373-33d3-4a81-a8ce-f1ffc89e05c0">10c</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_00a43dba-8083-45f3-9460-92fd75b504ff">&gt;</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_389d6072-ed07-460e-89ac-bf24cf4553e1">CONTAINS</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_54da7c3f-713f-4528-81d6-71d5e551cffc">SCOORD</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_79bd3703-3a6d-414b-a222-21c1d01f4685">$ImagePurpose </para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_b7145b36-d956-429f-bdd2-3041646bad46">1-n</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_7488472a-a88f-456c-95dd-67dd7db899da">U</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_51c7419d-7051-4f43-aa9a-d24eda290233"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1"/>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_abbb5fcd-365c-4f29-babc-515564fc2414">10d</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_6cd7dd69-be04-459c-8f30-241156fd980f">&gt;&gt;</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_f08b1a45-d33b-4fda-ba1a-daba9b2c4f92">SELECTED FROM</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_1c03d871-2eea-4422-91fc-ed4df2e27b4b">IMAGE</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_d9062efc-23e3-4bf3-bdd2-8459c68fd729"> </para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_4decf582-9fdc-4952-83ef-b13ab94177c0">1</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_95267483-ea91-4fdd-a0cd-c5a33f7fd39e">M</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_c9bf5d0f-286e-4fb1-b065-480e4b565e09"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1"/>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_90ef13d9-8db0-429c-be95-592ff171a73c">10e</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_559cee77-c204-40e7-84d8-0fa2418bca5e">&gt;</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_f1665735-ade9-4057-ad41-b9fa584db044">CONTAINS</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_5ccc6647-d9c5-4359-83fe-f8ca5a41d029">SCOORD3D</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_8d599de8-6e97-4445-9363-df257c3c105e">$ImagePurpose </para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_fdd0af50-df09-40f8-9dfa-911ef2354a71">1-n</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_2b6b8235-c1c4-4392-ac77-25fb4e340345">U</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_c60b7756-2020-4463-bb06-ecfd12989daa"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1"/>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_9645b24e-0d44-4ba1-9919-a443cf3242fb">10f</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_e4e52a76-bb92-45ee-955d-08eb4ce72f12">&gt;</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_30c55c25-c712-4a33-8450-6377eb2b3076">CONTAINS</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_b42ca300-85e0-4a1a-bbc5-1adcfd576749">WAVEFORM</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_c885e0f4-a70d-4ac2-a18a-bdb017b878f4">$WavePurpose </para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_f0d9847b-dc1a-43d9-87fa-540aa80defdc">1-n</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_213e7795-faa8-42e0-999d-5f6cd9f2f701">U</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_60bca958-e18b-430f-bc22-5a309c0b7d99"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1"/>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_a44180b0-8356-441e-9ea2-3a04c0bf4ac0">10g</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_9e6adf9e-4bad-40bc-b9c8-b1683770114b">&gt;</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_8709a883-35c3-4990-b18f-af7cc3b013d9">CONTAINS</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_faeaa1a6-9819-44ba-9fa3-a91718d67eec">TCOORD</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_866398b7-b872-4153-aae4-306fe44bd89d">$WavePurpose </para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_d93cd573-6391-46ea-800e-762c7bdcf3e7">1-n</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_5f3fe687-2f6d-4ed7-8bdc-01daa57a9248">U</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_7ed9604a-254d-4ac2-aad0-f3d54cecb402"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1"/>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_539521ec-2dc4-4ed2-bffd-5c6b7c1e322a">10h</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_8cc4e3f8-1073-4266-8977-3d5653a13c9b">&gt;&gt;</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_c9b49bf3-db00-447c-a566-03cb44945dc2">SELECTED FROM</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_ed536d4c-b256-4d04-a099-cecbf8116f75">WAVEFORM</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_157e2928-ca01-4541-8b05-f721d2aab1ab"> </para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_c0863e1a-7954-4619-9652-22ae2fe9c769">1</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_dd0e3a6e-6fb5-40f9-9869-07a14f41d048">M</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_2a441ec0-3528-4334-84a5-1f8815d87012"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1"/>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">10i</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">&gt;</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">CONTAINS</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">UIDREF</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">EV <link linkend="DCM_110180" xrefstyle="select: nopage">(110180, DCM, "Study Instance UID")</link></emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">1-n</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">C</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">IFF Row 1a = <link linkend="DCM_113014" xrefstyle="select: nopage">(113014, DCM, "Study")</link></emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para/>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">10j</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">&gt;</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">CONTAINS</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">UIDREF</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">EV <link linkend="DCM_112002" xrefstyle="select: nopage">(112002, DCM, "Series Instance UID")</link></emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">1-n</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">C</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">IFF Row 1a = <link linkend="DCM_113015" xrefstyle="select: nopage">(113015, DCM, "Series")</link></emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para/>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_561fe520-579d-41cd-be3a-81f9167425a3">11</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_67f4a30d-4f88-4719-ae9a-c8d5625d7f62">&gt;</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_65325706-0d92-4dba-8b41-c608cc87e86e">CONTAINS</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_2acb5a0d-c545-4a1e-abb4-0913efdf34ca">CODE</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_d360eea1-f8d6-4f6d-9f53-fc0aa814839c">$QualType</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_1bd12891-64bf-4198-bc36-caeadd08e7eb">1-n</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_4d77d339-02e3-4ae5-b412-059bb6434e3c">U</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_4680e520-4539-4d07-91f9-446e6133607c"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_da4cbd66-66b8-4dbf-884d-50a12e848513">$QualValue</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_72ef9da7-e5d8-4c75-a09f-a1a908ff10f2">11b</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_02f69756-3dd1-4043-bdf6-6a79eafe80fe">&gt;&gt;</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_06e88546-589d-4f5c-8634-3d0aefe2e8e6">HAS CONCEPT MOD</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_9cb382d0-78df-4814-a819-3912a4b35b42">CODE</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_df69c26b-6bd2-4d3b-907d-15020b23c0cf">$QualModType</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_a84e0089-23f4-43b6-9099-88ea1ee38b57">1-n</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_937056df-20fd-49e4-9d88-bd8930066f41">U</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_1f60fda5-07ca-4a05-84ba-b4e2fd2867af"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_e7e61255-2b59-46e0-abd7-58a2c1e3a7f8">$QualModValue</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_1288f892-e9dd-4b87-a348-511374eebcb4">12</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_b19197ad-457d-4166-9ed9-3107b274508e">&gt;</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_4dac1e8d-28a9-4e64-aa9c-0f8c6df27fc1">CONTAINS</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_4e4a327a-8947-4705-8be1-0c37e8359150">TEXT</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_5514b42a-e6e1-4b09-a978-1509da4697bb">$QualType</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_5a32cb44-a90a-410b-b406-9eaa87163b62">1-n</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_38c92490-44a2-4b42-9362-28f0d3f90594">U</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_46ad3fe8-a433-4291-b8ca-0e657cf2fb25"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_7c0daa2f-4d73-494a-a6b5-a2e2f2bc0481"/>
                        </td>
                    </tr>
                </tbody>
            </table>
            <para xml:id="para_c6830046-1dab-4cc0-aaf8-fd119bd48c1e">
                <emphasis role="bold">Content Item Descriptions</emphasis>
            </para>
            <informaltable frame="box" rules="all">
                <tbody>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Row 1a</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Specifies the type of entity that is the subject of this measurement group.</emphasis></emphasis></para>
                            <para><emphasis role="bold"><emphasis role="underline">If this is the entire Patient, then the current <xref linkend="sect_TID_1006" xrefstyle="select: label title"/> specifies that Patient; it defaults to the Attributes of the Patient Module.</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_96280a7d-623e-4e89-a8a0-fd292a80eecc">...</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_e69cade8-84bc-4519-8b50-52ce0104173d"/>
                        </td>
                    </tr>
                    <tr>
                        <td align="left" colspan="1">
                            <para xml:id="para_5608bfbe-583d-4205-9620-8fca26ce6448">Rows 10b, 10c, 10d, 10e, 10f, 10g, 10h</para>
                        </td>
                        <td align="left" colspan="1">
                            <para xml:id="para_bb5afb79-0999-457d-af45-5f22053440b8">Coordinates, image or waveform references that are the subject of this measurement group instead of, or in addition to, coordinate, image or waveform references associated with measurements in Row 10. Note that if an ROI is defined, other templates may be used, such as TID 1410 Planar ROI Measurements and Qualitative Evaluations.</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Row 10i</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">The Study that is the subject of this measurement group.</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Row 10j</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">The Series that is the subject of this measurement group.</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_d76683bc-f523-4ce9-972c-1a7c96c91435">Rows 11, 11b, 12</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_fa3c719e-6cfd-45f9-ba69-33127346b4b6">Allows encoding a flat list of name-value pairs that are coded questions with coded or text answers, for example, to record categorical observations related to the subject of the measurement group. A single level of coded modifiers may be present.</para>
                        </td>
                    </tr>
                </tbody>
            </informaltable>
        </section>

        <section label="CID 272" status="3" xml:id="sect_CID_272">
            <title>Specific Observation Subject Class</title>
            <variablelist spacing="compact">
                <varlistentry>
                    <term>
                        <emphasis role="bold">Resources:</emphasis>
                    </term>
                    <listitem>
                        <para>
                            <emphasis role="bold">
                                <link xl:href="http://dicom.nema.org/medical/dicom/current/output/chtml/part16/sect_CID_272.html" xl:show="new">HTML</link> | <link xl:href="ftp://medical.nema.org/medical/dicom/resources/valuesets/fhir/json/ValueSet-dicom-cid-272-ScopeofReference.json" xl:show="new">FHIR JSON</link> | <link xl:href="ftp://medical.nema.org/medical/dicom/resources/valuesets/fhir/xml/ValueSet-dicom-cid-272-ScopeofReference.xml" xl:show="new">FHIR XML</link> | <link xl:href="ftp://medical.nema.org/medical/dicom/resources/valuesets/ihe/svs/1.2.840.10008.6.1.1457.xml" xl:show="new">IHE SVS XML</link>
                            </emphasis>
                        </para>
                    </listitem>
                </varlistentry>
                <varlistentry>
                    <term>
                        <emphasis role="bold">Type:</emphasis>
                    </term>
                    <listitem>
                        <para>
                            <emphasis role="bold">Non-Extensible</emphasis>
                        </para>
                    </listitem>
                </varlistentry>
                <varlistentry>
                    <term>
                        <emphasis role="bold">Version:</emphasis>
                    </term>
                    <listitem>
                        <para>
                            <emphasis role="bold">20221201</emphasis>
                        </para>
                    </listitem>
                </varlistentry>
                <varlistentry>
                    <term>
                        <emphasis role="bold">UID:</emphasis>
                    </term>
                    <listitem>
                        <para>
                            <emphasis role="bold">1.2.840.10008.6.1.1457</emphasis>
                        </para>
                    </listitem>
                </varlistentry>
            </variablelist>
            <table frame="box" label="CID 272" rules="all" xml:id="table_CID_272">
                <caption>Specific Observation Subject Class</caption>
                <thead>
                    <tr valign="top">
                        <th align="center" colspan="1" rowspan="1">
                            <para>Coding Scheme Designator</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para>Code Value</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para>Code Meaning</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para>SNOMED-RT ID</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para>UMLS Concept Unique ID</para>
                        </th>
                    </tr>
                </thead>
                <tbody>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para>DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para>
                                <link linkend="DCM_121025" xrefstyle="select: nopage">121025</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para>Patient</para>
                        </td>
                        <td/>
                        <td/>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para>DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para>
                                <link linkend="DCM_113014" xrefstyle="select: nopage">113014</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para>Study</para>
                        </td>
                        <td/>
                        <td/>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para>DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para>
                                <link linkend="DCM_113015" xrefstyle="select: nopage">113015</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para>Series</para>
                        </td>
                        <td/>
                        <td/>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para>NCIt</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para><link xl:href="http://ncit.nci.nih.gov/ncitbrowser/ConceptReport.jsp?dictionary=NCI_Thesaurus&amp;code=C48179" xl:show="new">C48179</link></para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para>Image</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para/>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para>
                                <link xl:href="http://uts.nlm.nih.gov/uts/umls/concept/C1704922" xl:show="new">C1704922</link>
                            </para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para>DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para>
                                <link linkend="DCM_130781" xrefstyle="select: nopage">130781</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para>Image frame</para>
                        </td>
                        <td/>
                        <td/>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para>NCIt</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para><link xl:href="http://ncit.nci.nih.gov/ncitbrowser/ConceptReport.jsp?dictionary=NCI_Thesaurus&amp;code=C85402" xl:show="new">C85402</link></para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para>Imaging Region of Interest</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para/>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para>
                                <link xl:href="http://uts.nlm.nih.gov/uts/umls/concept/C0807727" xl:show="new">C0807727</link>
                            </para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para>SCT</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para>
                                <link xl:href="http://snomed.info/id/260893003" xl:show="new">260893003</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para>Waveform</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para>
                                <link xl:href="http://snomed.info/id/260893003" xl:show="new">G-C0AE</link>
                            </para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para>
                                <link xl:href="http://uts.nlm.nih.gov/uts/umls/concept/C0450448" xl:show="new">C0450448</link>
                            </para>
                        </td>
                    </tr>
                </tbody>
            </table>
        </section>

        <section label="D" status="1" xml:id="chapter_D">
            <title>DICOM Controlled Terminology Definitions (Normative)</title>
            <table frame="box" label="D-1" rules="all" xml:id="table_D-1">
                <caption>DICOM Controlled Terminology Definitions (Coding Scheme Designator "DCM" Coding Scheme Version "01")</caption>
                <thead>
                    <tr valign="top">
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_f05607a1-90fe-4408-ad56-d194c3203ae5">Code Value</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_022556ee-7c69-4138-a152-b3fa16bb86b2">Code Meaning</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_e0270183-1962-48d5-877f-b63624671b37">Definition</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_5aa10e80-ebb2-45b9-a2a9-fe4b3eb4497c">Notes</para>
                        </th>
                    </tr>
                </thead>
                <tbody>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_112002">112002</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_77c202a0-f270-4474-b04f-ec590997bb89">Series Instance UID</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_17a049ee-79b8-41cc-97e1-bb3ca1fb91a7">A unique identifier for a series of DICOM SOP instances.</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_9057cced-ca72-4574-8854-d654050b301a"/>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_110180">110180</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_eb954a32-1cf2-4eb0-a5c9-0f22b4f3a1f8">Study Instance UID</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_a259e12f-b1e0-4ea7-92f9-04ab3a2ec453">ParticipantObjectID type: Study Instance UID.</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_ea0e7f33-a9ee-4288-8b12-df9baab0ac8e"/>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_121025">121025</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_91910766-4e9e-4e0e-b033-054c811ed643">Patient</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_df262078-1dcf-432c-95c1-0985be72754d">A patient is the subject of observations.</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_17c2a41f-4bf2-45f9-8e55-7aa059bad605"/>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_113014">113014</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_69b27730-b8b4-4cb9-acce-854c1601bc21">Study</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_fd7404bc-2f64-4e3d-b280-3ed3bbc74004">A study is a collection of one or more series of medical images, presentation states, and/or SR documents that are logically related for the purpose of diagnosing a patient. A study may include composite instances that are created by a single modality, multiple modalities or by multiple devices of the same modality. [From <olink targetdoc="PS3.3" targetptr="sect_A.1.2.2" xrefstyle="select: docname label quotedtitle"/>]</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_f48d83b4-3045-4d9a-bbaf-30e6e7e20bd9"/>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_113015">113015</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_5e43391e-ab27-489e-9de1-403defd641d3">Series</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_88b1c660-326e-49bd-9dc7-50eb0783f51c">A distinct logical set used to group composite instances. All instances within a Series are of the same modality, in the same Frame of Reference (if any), and created by the same equipment. [See <olink targetdoc="PS3.3" targetptr="sect_A.1.2.3" xrefstyle="select: docname label quotedtitle"/>]</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_845fb170-0f54-4bbf-950a-5fbd731ee936"/>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_121024">121024</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_9fb60554-30fd-48cf-9b8f-4cba2beeb261">Subject Class</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_2883e0ef-1021-4467-a6a5-3791bde7bf06">Type of observation subject.</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_b4321a86-499c-41f6-95b8-b213259e1111"/>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_130780"><emphasis role="bold"><emphasis role="underline">130780</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Specific observation subject class</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">The type of entity that is the subject of a specific observation, such as a measurement or qualitative evaluation. E.g., a region of interest, image, series of images, study or patient.</emphasis></emphasis></para>
                            <note>
                                <para><emphasis role="bold"><emphasis role="underline">This is distinct from the Subject Class used in Observation Context, which more generally specifies the entire subject, usually the patient.</emphasis></emphasis></para>
                            </note>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para/>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_130781"><emphasis role="bold"><emphasis role="underline">130781</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Image frame</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">A single two-dimensional pixel plane of an image.</emphasis></emphasis></para>
                            <note>
                                <para><emphasis role="bold"><emphasis role="underline">Usually used in the context of a multi-frame image.</emphasis></emphasis></para>
                            </note>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para/>
                        </td>
                    </tr>
                </tbody>
            </table>
        </section>

        <informaltable rules="all" frame="box">
            <tr>
                <td>
                    <para><emphasis role="italic">PS3.16 for reference unchanged:</emphasis></para>
                </td>
            </tr>
        </informaltable>

        <section label="TID 1006" status="3" xml:id="sect_TID_1006">
            <title>Subject Context</title>
            <para xml:id="para_27b878d4-4efa-4718-aa55-914a4979598e">This Template contains identifying (and optionally descriptive) attributes of the subject of the observation.</para>
            <para xml:id="para_ba8285e8-39fd-4d20-b9bf-a109c4fe411f">Subject context identifies (and optionally) describes the subject of the observation, whether it be a patient (human or animal), a fetus (human or animal), a specimen, or a device.</para>
            <variablelist spacing="compact">
                <varlistentry>
                    <term>
                        <emphasis role="bold">Type:</emphasis>
                    </term>
                    <listitem>
                        <para xml:id="para_6361eadf-47f2-49ca-902e-22d9d79975dd">
                            <emphasis role="bold">Non-Extensible</emphasis>
                        </para>
                    </listitem>
                </varlistentry>
                <varlistentry>
                    <term>
                        <emphasis role="bold">Order:</emphasis>
                    </term>
                    <listitem>
                        <para xml:id="para_ee82aad0-fd60-4924-ada3-f7d198288f08">
                            <emphasis role="bold">Significant</emphasis>
                        </para>
                    </listitem>
                </varlistentry>
                <varlistentry>
                    <term>
                        <emphasis role="bold">Root:</emphasis>
                    </term>
                    <listitem>
                        <para xml:id="para_5467e423-d7a1-455b-b1db-d68d57cef5f9">
                            <emphasis role="bold">No</emphasis>
                        </para>
                    </listitem>
                </varlistentry>
            </variablelist>
            <table frame="box" label="TID 1006" rules="all" xml:id="table_TID_1006">
                <caption>Subject Context</caption>
                <thead>
                    <tr valign="top">
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_3334d125-12b4-486c-9ea6-c2ae8e821f66"/>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_32d44ff9-9f93-4152-b0af-0a6fe651dcd6">NL</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_98cf91d6-481d-489b-89cd-377dd6cf93ea">Rel with Parent</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_8b4a693c-37a1-4047-a4f8-d97b8199ab10">VT</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_9d3a2576-0401-46ad-8b21-7b1a3a26b2be">Concept Name</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_8ef91aa9-4e17-471e-b5c5-24caf9b940eb">VM</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_c83c5c78-1608-4a91-b7dc-ad9579bf702b">Req Type</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_e41e461e-3a79-41ad-a245-c0195f6a056c">Condition</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_fc87e7a1-e795-41e0-b780-fe97274d19c9">Value Set Constraint</para>
                        </th>
                    </tr>
                </thead>
                <tbody>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_41cf745b-46bd-4fce-8523-ee5dfe0d1bdc">1</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_f970a22b-56a2-4e72-9aa9-a67b9e7f30c6"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_b70a40a8-1940-4d20-bf25-d213e7b2141f"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_46306d04-d1ab-45ca-807c-b41389fe6bee">CODE</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_3bc803e6-a5f5-42eb-b4fa-3eb6ee4449f0">EV <link linkend="DCM_121024" xrefstyle="select: nopage">(121024, DCM, "Subject Class")</link>
                            </para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_b1c008b1-e0a8-4fd2-92f0-0ad5b28c29aa">1</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_fea4d62f-97f4-4b25-af99-d1847ef9da13">MC</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_3f845b55-1f7c-4467-840e-1040401e5659">IF subject is not the Patient</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_e2615026-1617-4ce2-a980-b1383655dd57">D<xref linkend="sect_CID_271" xrefstyle="select: labelnumber quotedtitle nopage"/>
                            </para>
                            <para xml:id="para_08b64e82-1a40-4373-bc5d-50e7a34e9323">Defaults to <link linkend="DCM_121025" xrefstyle="select: nopage">(121025, DCM, "Patient")</link>
                            </para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_9e3ca4ae-7003-46af-826f-44e616565f0f">2</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_51e4918d-fe2e-4c53-adca-942e85d91468"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_438e597d-ccf5-40b0-b910-f7d5d8c8f75a"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_99b497ce-4434-4d63-877c-42c95f9cfbdf">INCLUDE</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_ebe1db38-9073-473a-b1dc-46337e5a7232">D<xref linkend="sect_TID_1007" xrefstyle="select: labelnumber quotedtitle nopage"/>
                            </para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_96b91d5c-fcfe-457c-902d-1cb2c63b8ba8">1</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_73d4d941-ec6d-4da0-ae77-16a2a5c7a7fd">UC</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_22358fe8-8909-462c-922d-923149b0a498">IFF Row 1 value = <link linkend="DCM_121025" xrefstyle="select: nopage">(121025, DCM, "Patient")</link> or Row 1 is absent</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_dcdb8113-3367-4658-9620-0a697ecba7a5">May be used for human or animal patients</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_0f2c718f-7647-4b9c-9e92-e91266481680">3</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_5b8ba346-d41f-41d0-8e6a-7a1d12b640ef"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_1e436a2d-32c3-41a1-a066-8b868de1ff55"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_3cb0be03-109a-4f51-9747-a80b2a347bbd">INCLUDE</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_3d9bd68e-cb30-4c35-8a0d-4710ed439ddd">D<xref linkend="sect_TID_1008" xrefstyle="select: labelnumber quotedtitle nopage"/>
                            </para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_2352e6a3-47af-48e5-8d51-cd6abffdabfc">1</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_a4e5fa34-0b28-4545-8afc-f870723137bc">UC</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_b21105ea-369a-44ec-8e35-3a46fadd6ccb">IFF Row 1 value = <link linkend="DCM_121026" xrefstyle="select: nopage">(121026, DCM, "Fetus")</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_a948cddc-a3cd-467b-a968-b940e4ff929d">May be used for human or animal fetuses</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_44800471-4ae1-4602-8ac5-ce7997d617f6">4</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_80ab83e4-0992-4e75-9f54-ce701702b6ae"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_783b9c6c-01f7-4fc2-8d27-a07b3e7282e1"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_daeb0fbb-c27f-45b5-9364-ec8aeb906cdf">INCLUDE</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_5c182fe5-b5f2-4482-b6d3-d13cd2a9edfc">D<xref linkend="sect_TID_1009" xrefstyle="select: labelnumber quotedtitle nopage"/>
                            </para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_2e19897b-0b24-4ab6-bcf8-cdc0c76c8551">1</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_eed5fa92-3b41-4383-aab5-d2b875377759">UC</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_85ff6f66-dbf8-4f2e-b81c-fcf5dc288281">IFF Row 1 value = <link linkend="DCM_121027" xrefstyle="select: nopage">(121027, DCM, "Specimen")</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_8d2c3b66-d3fb-4a05-ad66-347e251de092"/>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_cb14cd9f-f7df-4b39-82ad-dd28acd6775b">5</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_6caa33f0-53ba-4e5f-a415-6dca28aa7a4d"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_a7d03fe8-41a1-495e-93cf-3cc6f0f82022"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_d5fbbd4e-3c00-414e-9671-9557269af963">INCLUDE</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_eb9b4c07-affd-4652-93fe-aeced82af923">D<xref linkend="sect_TID_1010" xrefstyle="select: labelnumber quotedtitle nopage"/>
                            </para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_3037d050-816c-4f07-859c-3490237568f7">1</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_b8ee3622-4c5d-4cb2-9c7d-7264713e117a">UC</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_47d7acd6-178e-4216-a033-e4efa099c632">IFF Row 1 value = <link linkend="DCM_121192" xrefstyle="select: nopage">(121192, DCM, "Device Subject")</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_544dfe45-4775-4454-b6f9-fdce8b4c33de"/>
                        </td>
                    </tr>
                </tbody>
            </table>
        </section>

        <section label="TID 1007" status="3" xml:id="sect_TID_1007">
            <title>Subject Context, Patient</title>
            <para xml:id="para_bf468c38-5fec-4320-bd23-8840a4b99ee7">Identifies (and optionally describes) a patient who is the subject.</para>
            <variablelist spacing="compact">
                <varlistentry>
                    <term>
                        <emphasis role="bold">Type:</emphasis>
                    </term>
                    <listitem>
                        <para xml:id="para_60f651c1-5481-4ce9-a8a3-9bcf147ecfe3">
                            <emphasis role="bold">Extensible</emphasis>
                        </para>
                    </listitem>
                </varlistentry>
                <varlistentry>
                    <term>
                        <emphasis role="bold">Order:</emphasis>
                    </term>
                    <listitem>
                        <para xml:id="para_fbe8f43d-9364-4b4f-9949-0b5a3110768b">
                            <emphasis role="bold">Significant</emphasis>
                        </para>
                    </listitem>
                </varlistentry>
                <varlistentry>
                    <term>
                        <emphasis role="bold">Root:</emphasis>
                    </term>
                    <listitem>
                        <para xml:id="para_8f9a9b30-f845-4b70-a55f-f356cf574eea">
                            <emphasis role="bold">No</emphasis>
                        </para>
                    </listitem>
                </varlistentry>
            </variablelist>
            <table frame="box" label="TID 1007" rules="all" xml:id="table_TID_1007">
                <caption>Subject Context, Patient</caption>
                <thead>
                    <tr valign="top">
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_cbfaf626-9734-4d07-a4ac-0aa6958258c5"/>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_3a036833-b7d7-41eb-917b-2ecdcf0d995d">NL</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_901af82f-5f12-455e-96dd-d7660580d658">Rel with Parent</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_5b7de8b7-7530-475a-87e6-8787224443c7">VT</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_aea81ac1-0ded-411c-94bb-f9a48e915cd0">Concept Name</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_9b38dad8-5e65-4e44-9911-505877fa176d">VM</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_80064aea-6c47-48ad-8ebf-a4926020a288">Req Type</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_9af8b701-1cde-485e-9275-6909874ce303">Condition</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_6162a781-0873-4597-824f-bdf777429be7">Value Set Constraint</para>
                        </th>
                    </tr>
                </thead>
                <tbody>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_8019b492-1fd1-44e6-8732-c7f765778080">1</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_3192be22-006e-47b0-9272-80f0d842378d"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_fb168607-e8f0-4e62-8c96-ea881d809fdd"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_e45ac586-76fc-443f-87ec-7ca703334271">UIDREF</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_5ab0ba85-5b97-4079-8945-aac23bdf520e">EV <link linkend="DCM_121028" xrefstyle="select: nopage">(121028, DCM, "Subject UID")</link>
                            </para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_44eec935-d3b5-43ee-95ce-62dec7662df9">1</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_8d64e418-4e74-42cb-8db6-00481b9fd2bf">U</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_bcb64d08-0f26-46fc-b46e-5b47975537db"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_cd49e539-941c-4ec0-9ca6-b4094b8e8ea5">E.g., SOP Instance UID of Detached Patient Instance</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_b62bdc5f-0fa7-45d4-9623-271aecb3e9ce">2</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_f7222674-6cbb-4dcf-8d15-f14be119a27a"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_20adc004-2cb7-4fa8-8ff2-840a37f19dc3"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_67ccdc70-7016-475b-be97-2bc1a3297308">PNAME</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_d5cd837a-602c-4480-86b2-dfc77595374f">EV <link linkend="DCM_121029" xrefstyle="select: nopage">(121029, DCM, "Subject Name")</link>
                            </para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_c7f6005b-ae12-48e9-8334-ec184bd9fe1a">1</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_fe1cd067-95c5-4830-a45a-c070578d1a43">MC</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_8a909eeb-5d29-4f42-9e40-652879a6b837">Required if not inherited.</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_2b970585-8c3b-4ffe-a8f5-8c20ccb87a9e">Defaults to value of Patient's Name (0010,0010) in Patient Module</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_a8bfd02b-567c-436a-b6c6-a5c938dd11b7">3</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_6196223c-988d-4e11-9047-ebd35cb707d1"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_c20cd71e-e02d-4053-b6f3-bc5e5a7f60fa"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_f004ceb4-a948-45bf-b4b9-6036c5ee53bd">CODE</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_14c5a87a-f282-4aa8-9bb7-447651c43256">EV <link linkend="DCM_121030" xrefstyle="select: nopage">(121030, DCM, "Subject ID")</link>
                            </para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_a94a5d71-e7c8-4d36-85f9-5c6192d2a2e9">1</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_e76dc6a8-aa63-47bb-b0ec-28284b3cf7d6">MC</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_ae8ae22f-5f7b-4a18-a071-c76584d9e970">Required if not inherited.</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_be905d69-5b6d-4f67-800e-5d471225b3f5">Defaults to value of Patient ID (0010,0020) in Patient Module</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_501d1342-373c-4dfa-9d30-b27055259fc6">4</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_027507be-2dc1-41af-a9ef-fc489d6bd050"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_898cbf88-ed48-4999-ba4a-b985ffb3842b"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_6c0cfd2b-9789-4b8b-bd7a-42fc04a3103c">DATE</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_13138785-6f10-4b88-86de-0f77ad9f6748">EV <link linkend="DCM_121031" xrefstyle="select: nopage">(121031, DCM, "Subject Birth Date")</link>
                            </para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_8271616d-bffd-4a49-a662-764b8e334557">1</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_2c863b3e-fa6e-4a46-b3fa-34b881e001ac">U</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_3d8d8e18-749b-4d74-a7a2-0ca184bc921d"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_70a0c58a-928c-4404-aed9-a0c44add34ac">Defaults to value of Patient's Birth Date (0010,0030) in Patient Module</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_749f7806-746a-4a26-b47b-1a43c0816ccc">5</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_9ac5e124-794e-42e9-867f-0559625ed8ba"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_fca3d08f-6155-4802-bc66-fddfec099a95"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_9c4bf4dc-2b73-4648-8fbb-90a84c6b524c">CODE</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_a227b0ef-61c8-45b2-922e-d0f2f7e0a56e">EV <link linkend="DCM_121032" xrefstyle="select: nopage">(121032, DCM, "Subject Sex")</link>
                            </para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_5d7e091e-39fe-4e2f-bead-6505c7757610">1</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_901ed7c9-bdc9-4d63-8298-57b0de2716ef">U</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_ae48ab38-fa67-4527-be2c-2681a08fc3b6"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_f2d635a6-832b-4dc3-8273-c452ece78239">Defaults to value equivalent to Patient's Sex (0010,0040) in Patient Module</para>
                            <para xml:id="para_04d6bbb2-818f-489c-84f3-a8b7a08e0061">D<xref linkend="sect_CID_7455" xrefstyle="select: labelnumber quotedtitle nopage"/>
                            </para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_3c7b89a2-e1e9-4c52-a5f4-6eb54fcfdf34">6</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_46576318-be77-4cbc-bd8f-55794271ceca"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_f3aa4002-7226-462c-823c-ae41c3163111"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_7a35b5a6-5ec1-48ea-8b1b-e0453e80affa">NUM</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_2e03c824-c7a7-4cec-9b02-08fe62ec9435">EV <link linkend="DCM_121033" xrefstyle="select: nopage">(121033, DCM, "Subject Age")</link>
                            </para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_bb361d4e-f755-424e-8c07-2258e70d4598">1</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_3a0079da-6041-4ecb-aca5-90072b075a43">U</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_0a38f46b-0c96-42c9-b3d2-69c67edd2094"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_110dea7a-273a-46ca-9f9b-f9bc957c28b6">Defaults to value of Patient's Age (0010,1010) in Patient Study Module</para>
                            <para xml:id="para_70dfdd7b-5815-4ec3-83f0-f7b1789b7146">UNITS = D<xref linkend="sect_CID_7456" xrefstyle="select: labelnumber quotedtitle"/>
                            </para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_655c9b84-a715-44df-9be7-869f30f75930">7</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_1b500f70-cacb-4351-82b1-516314d5db38"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_9a5eee97-d6fc-43c2-a63d-a5136a223f42"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_39d4d93e-e665-45f8-9b96-df513f11e499">CODE</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_3d7b10b5-f314-45b9-be42-a3e9dba5ec1b">EV <link linkend="DCM_121034" xrefstyle="select: nopage">(121034, DCM, "Subject Species")</link>
                            </para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_2950824c-8b38-4e54-9149-34e7d3060d05">1</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_e230982f-4363-477b-9670-548c58049b60">MC</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_40e9cfea-cb92-4eca-aea5-e40782510d01">Required if not inherited.</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_502220dd-3177-4626-8a3c-ea187e44d318">D<xref linkend="sect_CID_7454" xrefstyle="select: labelnumber quotedtitle nopage"/>
                            </para>
                            <para xml:id="para_7bca73e4-ae86-4c63-9d52-855c6bfa2c1b">Defaults to value of Patient Species Code Sequence (0010,2202) in Patient Module, or if absent, <link xl:href="http://snomed.info/id/337915000" xl:show="new">(337915000, SCT, "Homo sapiens")</link>.</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_73319e32-d3b0-42ac-aa41-a02febb46d2e">8</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_adbedfea-63e8-49f5-9db5-32c2bfa77b36"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_46f0fd0a-8eee-48ba-91ca-1903c4efc87d"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_a243c6a0-2c2a-495c-8b3c-0e3809375da6">CODE</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_bc9dcdc7-d19c-4dbf-abdc-943f9257eb42">EV <link linkend="DCM_121035" xrefstyle="select: nopage">(121035, DCM, "Subject Breed")</link>
                            </para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_c8b89d63-c52a-4f92-b994-94495bb150de">1</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_d03f0de1-104d-4dd2-bfcb-fad71641c645">U</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_c5ceed7e-c8b9-4027-b243-96c5ee0aaaad"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_13820a78-c051-4914-8382-08999cd56a2b">Defaults to value of Patient Breed Code Sequence (0010,2293) in Patient Module</para>
                            <para xml:id="para_9b56717c-5a97-4e07-8987-38fec64d94b3">D<xref linkend="sect_CID_7480" xrefstyle="select: labelnumber quotedtitle nopage"/>
                            </para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_244187c3-3e03-4f1c-bced-a23bf7b02956">9</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_199ae9a5-2b85-425f-a4a5-b3e5170e32da"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_c8c2b899-f437-4160-a4ed-6d2b1d0fdea8"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_ae38d44b-07c4-4ca7-a8d6-87dedec09304">CODE</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_001845b6-e368-4f4f-bfa6-92a8cb4ed4c8">EV <link xl:href="http://snomed.info/id/415229000" xl:show="new">(415229000, SCT, "Racial group")</link>
                            </para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_d68fd13c-389e-49f3-8a6b-0ae19e59b592">1</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_2c2aa631-9941-426b-9cca-2f3473987b98">U</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_bd544c37-f3b1-4262-a145-833c83ea61f0"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_2c234fea-fdfd-40c9-93a3-8b07ead0a6fe">Defaults to the coded equivalent of the text value of Ethnic Group (0010,2160) in Patient Module, which is defined as "race or ethnic group"</para>
                            <para xml:id="para_296466ba-c231-41b5-8108-fba2584de6d7">D<xref linkend="sect_CID_6099" xrefstyle="select: labelnumber quotedtitle nopage"/>
                            </para>
                        </td>
                    </tr>
                </tbody>
            </table>
        </section>


        <section label="CID 271" status="3" xml:id="sect_CID_271">
            <title>Observation Subject Class</title>
            <variablelist spacing="compact">
                <varlistentry>
                    <term>
                        <emphasis role="bold">Resources:</emphasis>
                    </term>
                    <listitem>
                        <para xml:id="para_51b5f6c3-7445-4b5d-a307-05e14b787bd1">
                            <emphasis role="bold">
                                <link xl:href="http://dicom.nema.org/medical/dicom/current/output/chtml/part16/sect_CID_271.html" xl:show="new">HTML</link> | <link xl:href="ftp://medical.nema.org/medical/dicom/resources/valuesets/fhir/json/ValueSet-dicom-cid-271-ObservationSubjectClass.json" xl:show="new">FHIR JSON</link> | <link xl:href="ftp://medical.nema.org/medical/dicom/resources/valuesets/fhir/xml/ValueSet-dicom-cid-271-ObservationSubjectClass.xml" xl:show="new">FHIR XML</link> | <link xl:href="ftp://medical.nema.org/medical/dicom/resources/valuesets/ihe/svs/1.2.840.10008.6.1.41.xml" xl:show="new">IHE SVS XML</link>
                            </emphasis>
                        </para>
                    </listitem>
                </varlistentry>
                <varlistentry>
                    <term>
                        <emphasis role="bold">Type:</emphasis>
                    </term>
                    <listitem>
                        <para xml:id="para_8db4e1ef-9fc6-4735-b7e9-81ff384683b2">
                            <emphasis role="bold">Non-Extensible</emphasis>
                        </para>
                    </listitem>
                </varlistentry>
                <varlistentry>
                    <term>
                        <emphasis role="bold">Version:</emphasis>
                    </term>
                    <listitem>
                        <para xml:id="para_20b7cd63-57cc-4a13-a65c-d0313dcc2c43">
                            <emphasis role="bold">20071102</emphasis>
                        </para>
                    </listitem>
                </varlistentry>
                <varlistentry>
                    <term>
                        <emphasis role="bold">UID:</emphasis>
                    </term>
                    <listitem>
                        <para xml:id="para_86161a57-656a-42df-a3aa-e00b5fe87a4d">
                            <emphasis role="bold">1.2.840.10008.6.1.41</emphasis>
                        </para>
                    </listitem>
                </varlistentry>
            </variablelist>
            <table frame="box" label="CID 271" rules="all" xml:id="table_CID_271">
                <caption>Observation Subject Class</caption>
                <thead>
                    <tr valign="top">
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_2067e689-f1d9-455a-a95b-9cca7e7bd387">Coding Scheme Designator</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_fc4c5181-33f0-4397-bf5f-ffe809525600">Code Value</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_2571f60f-a622-461a-9a68-d6fd1b0dab6d">Code Meaning</para>
                        </th>
                    </tr>
                </thead>
                <tbody>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_24a24d40-f656-470d-81d2-c214416485df">DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_cfd89822-6eec-445f-a66b-61d0291e0c76">
                                <link linkend="DCM_121025" xrefstyle="select: nopage">121025</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_76db50c4-2951-415c-b7fa-67b4b19b8aee">Patient</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_77336e8b-c744-42d2-b8d5-b5d27739aa6a">DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_92eaa461-344b-4da9-8a4c-f37fb399aa83">
                                <link linkend="DCM_121026" xrefstyle="select: nopage">121026</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_be21d422-8977-4a5d-9a6e-303fc71f156e">Fetus</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_f717d738-7bd0-4c50-bd1e-9ae268961d1b">DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_5ab3f045-4dcc-4b4a-856b-ea2528583dcf">
                                <link linkend="DCM_121027" xrefstyle="select: nopage">121027</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_a2eaf14f-5c3f-4362-971d-acd68d586a8c">Specimen</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_d2ffc08f-6ab5-46c1-9ca6-b78cd5c59197">DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_386a0801-7b37-4bc8-afa6-9e46dabbfdb8">
                                <link linkend="DCM_121192" xrefstyle="select: nopage">121192</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_06bf0651-79a5-4825-be1c-3d9f26d463a3">Device Subject</para>
                        </td>
                    </tr>
                </tbody>
            </table>
        </section>


        <informaltable rules="all" frame="box">
            <tr>
                <td>
                    <para><emphasis role="italic">Amend PS3.6 as follows (changes to existing text are bold and <emphasis role="bold"><emphasis role="underline">underlined</emphasis></emphasis> for additions and <emphasis role="bold"><emphasis role="strikethrough">struckthrough</emphasis></emphasis> for removals):</emphasis></para>
                </td>
            </tr>
        </informaltable>

        <section label="A" status="1" xml:id="chapter_A">
            <title>Registry of DICOM Unique Identifiers (UIDs) (Normative)</title>
            <table frame="box" label="A-3" rules="all" xml:id="table_A-3">
                <caption>Context Group UID Values</caption>
                <thead>
                    <tr valign="top">
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_3322bcc5-7ea4-44d1-9b17-499a6a5e512c">
                                <emphasis role="bold">Context UID</emphasis>
                            </para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_225c7fe6-91ef-424b-812b-249ecd55f194">
                                <emphasis role="bold">Context Identifier</emphasis>
                            </para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_25349958-e4d1-4bb0-aaa9-33d5cb03df93">
                                <emphasis role="bold">Context Group Name</emphasis>
                            </para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_21c2633a-2b79-49c4-baa8-d930649479d0">
                                <emphasis role="bold">Comment</emphasis>
                            </para>
                        </th>
                    </tr>
                </thead>
                <tbody>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">1.2.840.10008.6.​1.1457</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para>
                                <xref linkend="sect_CID_272" xrefstyle="select: labelnumber"/>
                                <!--<olink targetdoc="PS3.16" targetptr="sect_CID_272" xrefstyle="select: labelnumber"/>-->
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para>
                                <xref linkend="sect_CID_272" xrefstyle="select: title"/>
                                <!--<olink targetdoc="PS3.16" targetptr="sect_CID_272" xrefstyle="select: title"/>-->
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para/>
                        </td>
                    </tr>
                </tbody>
            </table>
        </section>
    </chapter>
</book>