    <title>CP-1822</title>
  <subtitle>Remove Dimension Index Sequence requirement for WSI when Dimensions implicitly defined by TILED_FULL</subtitle>
    <chapter>
        <title></title>
        <informaltable rules="all" frame="box">
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Status</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>Final Text</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Date of Last Update</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>2019/01/24</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Person Assigned</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>David Clunie</para><para><link
                    xl:href="mailto:dclunie@dclunie.com"/></para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submitter Name</para></td>
              <td rowspan="1" colspan="1" valign="top"><para>David Clunie</para><para><link
                xl:href="mailto:dclunie@dclunie.com"/></para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submission Date</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>2018/04/19</para></td>
            </tr>
        </informaltable>
        <informaltable rules="all" frame="box">
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Correction Number CP-1822</para></td>
            </tr>
            <tr>
              <td rowspan="1" colspan="1" valign="top"><para>Log Summary: Remove Dimension Index Sequence requirement for WSI when Dimensions implicitly defined by TILED_FULL</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">
                    <para>Name of Standard</para>
                    <para>PS3.3 2018e</para>
                </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Rationale for Correction:</para>
                  <para>It is not necessary to explicitly specify Dimensions by Attribute using the Dimension Index Sequence
                    when the TILED_FULL value of Dimension Organization Type implicitly defines them.</para>
                  <para>Further, having Dimension Index Sequence results in a conditional requirement to include Dimension Index Values in Frame Content Sequence on a per frame basis, if Frame Content Sequence is present (though CP 1713 changed the requirement for Frame Content Sequence from mandatory to optional, so this does not prevent omission of the entire Per-Frame Functional Group Sequence, if there is no other need for it).</para>
          <para>It is proposed that Dimension Index Sequence be made conditional, required if Dimension Organization Type (0020,9311) is absent or not TILED_FULL.</para>
                    <para>It is not considered necessary to provide a canonical explicit description of the dimensions that TILED_FULL implies (i.e., Image Orientation (Slide) and Optical Path Sequence), nor indeed would it be sufficient since the X, Y and Z order specified in C.7.6.17.3 can not be explicitly described as there is no mechanism at present for specifying a selected value of a multi-valued attribute as a single dimension).</para>
                </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">Correction Wording:</td>
            </tr>
        </informaltable>
    </chapter>
    <chapter>
        <title/>

      <informaltable rules="all" frame="box">
        <tr>
          <td>
            <para><emphasis role="italic">Amend DICOM PS3.3 as follows (changes to existing text are bold and <emphasis role="bold"><emphasis role="underline">underlined</emphasis></emphasis> for additions and <emphasis role="bold"><emphasis role="strikethrough">struckthrough</emphasis></emphasis> for removals):</emphasis></para>
          </td>
        </tr>
      </informaltable>
      
        <section label="A.32.8.4" status="4" xml:id="sect_A.32.8.4">
            <title>VL Whole Slide Microscopy Image Functional Group Macros</title>
            <para>...</para>
            <para xml:id="para_c6dd0dd8-a562-406c-a95f-7d307b307f1d">.</para>
            <table frame="box" label="A.32.8-2" rules="all" xml:id="table_A.32.8-2">
                <caption>VL Whole Slide Microscopy Image Functional Group Macros</caption>
                <thead>
                    <tr valign="top">
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_661654ac-dde2-4b5b-b103-22db80f7df54">Functional Group Macro</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_f0df0b22-b551-45da-b912-36013f2dec5a">Section</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_1d9c0115-1d06-4a82-9f8b-b435811646e1">Usage</para>
                        </th>
                    </tr>
                </thead>
                <tbody>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_eb389343-b64d-4efe-a32f-71e091518fb8">Pixel Measures</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_ea1313a9-1968-4f04-8cf9-5f7e50434db1">
                                <xref linkend="sect_C.7.6.16.2.1" xrefstyle="select: labelnumber"/>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_b6971130-c9fd-4a1d-b939-8d43782d685c">M - Shall be used as a Shared Functional Group.</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_fa7cff4f-5a24-4b9d-9643-63f2878c0b06">Frame Content</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_927096fd-ffe5-4351-a2e2-0dbf757b07dd">
                                <xref linkend="sect_C.7.6.16.2.2" xrefstyle="select: labelnumber"/>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_fa450bb3-cf94-4482-9690-a8a0953834a5">U - Shall not be used as a Shared Functional Group.</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_06fee48c-e13a-4a73-a2b0-3f7095a05e60">...</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_ac09cd80-0a48-4a55-b984-01ea10623663"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_370b6b45-d0dc-410f-b49e-a39bbe0e4425"/>
                        </td>
                    </tr>
                </tbody>
            </table>
            <note>
                <para xml:id="para_7bfc4a93-3e1d-41dd-80a0-43b1295f3dde">The Plane Position (Slide) and Optical Path Identification Macros are Type C, which allows the Per-Frame Functional Group Sequence (5200,9230) to be entirely omitted in those cases in which there are no other Per-Frame Functional Group Macros with content (i.e., the Frame Content Macro is empty<emphasis role="bold"><emphasis role="underline"> or absent</emphasis></emphasis>).</para>
            </note>
        </section>
        
      <section label="C.7.6.17" status="4" xml:id="sect_C.7.6.17">
          <title>Multi-frame Dimension Module</title>
      <para>...</para>
          <para xml:id="para_e1b5da57-e8e7-4265-ad0d-5b53308cbbc3">
<xref linkend="table_C.7.6.17-1" xrefstyle="select: label"/> specifies the Attributes of the <xref linkend="sect_C.7.6.17" xrefstyle="select: title"/>.</para>
          <table frame="box" label="C.7.6.17-1" rules="all" xml:id="table_C.7.6.17-1">
            <caption>Multi-frame Dimension Module Attributes</caption>
            <thead>
              <tr valign="top">
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_52fedc83-3eba-4eb9-b710-42a6374511bd">Attribute Name</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_bb8b8a4e-884d-4d4a-8f5b-d85cd988f904">Tag</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_a52154a0-b892-41ce-8072-c409e373472c">Type</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_6042c75f-8536-43cf-ada2-e55670bf8a93">Attribute Description</para>
                </th>
              </tr>
            </thead>
            <tbody>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_3c9fc2fc-f7cd-4f03-8913-9309af099d34">Dimension Organization Sequence</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_b142daad-8ba6-43c0-861e-dda13e684c5b">(0020,9221)</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_c93cfbc9-bf03-433a-9ec6-f956e76dc3db">1</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_05cc2719-dd8e-4ca0-869b-6e2f74b3ddde">Sequence that lists the Dimension Organization UIDs referenced by the containing SOP Instance. See <xref linkend="sect_C.7.6.17.2" xrefstyle="select: label"/> for further explanation.</para>
                  <para xml:id="para_c0627d3d-622f-4383-bc91-77f0d75ecbd1">One or more Items shall be included in this Sequence.</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_5da6fb6b-19c6-4eeb-8fea-3feb297deb93">&gt;Dimension Organization UID</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_b4ee974e-add5-4db7-ab7c-3ec1d0593433">(0020,9164)</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_c318f3a5-9039-4037-ad24-dee213cc76af">1</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_a12b4dc4-3b28-4d01-87e1-4b77255cd5f6">Uniquely identifies a set of dimensions referenced within the containing SOP Instance. See <xref linkend="sect_C.7.6.17.2" xrefstyle="select: label"/> for further explanation.</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_2ed4e229-86f0-49eb-a975-82e909a3bda2">Dimension Organization Type</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_057023b5-439a-476b-b653-b891e6760ac0">(0020,9311)</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_eeded002-5f74-4378-87d7-f882f59cab4c">3</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_c658e40b-70bb-4fdd-8d65-b7c387e1b921">Dimension organization of the instance.</para>
                  <variablelist spacing="compact">
                    <title>Defined Terms:</title>
                    <varlistentry>
                      <term>3D</term>
                      <listitem>
                        <para xml:id="para_7f8d5062-be96-4036-9f05-f0bcf74caad8">Spatial Multi-frame image of equally spaced parallel planes (3D volume set)</para>
                      </listitem>
                    </varlistentry>
                    <varlistentry>
                      <term>3D_TEMPORAL</term>
                      <listitem>
                        <para xml:id="para_0d8b0c6d-db9c-4f89-8e1d-234082c4c824">Temporal loop of equally spaced parallel-plane 3D volume sets.</para>
                      </listitem>
                    </varlistentry>
                    <varlistentry>
                      <term>TILED_FULL</term>
                      <listitem>
                        <para xml:id="para_bbbacde5-9240-44c7-aaae-75f23d398fe2">Tiled image in which each frame represents a single tile and the spatial positions of the tiles are implicitly defined as per
                          <xref linkend="sect_C.7.6.17.3" xrefstyle="select: label"/>.</para>
                      </listitem>
                    </varlistentry>
                    <varlistentry>
                      <term>TILED_SPARSE</term>
                      <listitem>
                        <para xml:id="para_ac40ae40-8977-4c43-b515-b181c773e2a3">Tiled image in which each frame represents a single tile and the spatial positions of tiles are explicitly defined by per-frame Functional Group Macro entries.</para>
                      </listitem>
                    </varlistentry>
                  </variablelist>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_9a948690-a5d2-499e-ab49-3354658f32eb">Dimension Index Sequence</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_c1d622aa-159e-4b26-8b77-adffb26e52e2">(0020,9222)</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_fbbb37b9-09b7-4c7a-92c0-183eb09d18be">1<emphasis role="bold"><emphasis role="underline">C</emphasis></emphasis></para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_f2c32ac7-6e56-44f9-9190-09774137ca81">Identifies the Sequence containing the indices used to specify the dimension of the multi-frame object.</para>
                  <para xml:id="para_efd2454b-3e79-4f47-b24f-47d27064bd0d">One or more Items shall be included in this Sequence.</para>
              <para><emphasis role="bold"><emphasis role="underline">Required if Dimension Organization Type (0020,9311) is absent or not TILED_FULL. May be present otherwise.</emphasis></emphasis></para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_1726883b-8c23-4d6d-aaf7-55a208886eb3">&gt;Dimension Index Pointer</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_23c2545b-5e21-4798-9c24-c7962f73e0d1">(0020,9165)</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_79b99d41-fc92-4149-b615-5b4b01dca460">1</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_47caf2ea-2efa-4fa9-b31e-90565d846468">Contains the Data Element Tag that is used to identify the Attribute connected with the index. See <xref linkend="sect_C.7.6.17.1" xrefstyle="select: label"/> for further explanation.</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_6e2e1ef9-ffea-4295-b918-47c477add5e8">&gt;Dimension Index Private Creator</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_eacfd1f5-6ff0-4cad-a407-fdf37eb7c3f9">(0020,9213)</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_6de45546-ad43-4d94-b53e-975163afd60b">1C</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_0aae7b06-c1dd-4d3a-97d7-cda66ecb81fe">Identification of the creator of a group of Private Data Elements.</para>
                  <para xml:id="para_2b1f3578-ff5a-4f1e-a6ae-9483e6b2fdb8">Required if the Dimension Index Pointer (0020,9165) value is the Data Element Tag of a Private Attribute.</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_2401eda0-aeb2-44f2-9a09-8f551992fd25">&gt;Functional Group Pointer</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_87954a21-113b-4324-99dc-14b9948b7143">(0020,9167)</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_1ff96635-7cfb-480e-bd7d-8b4e8728f086">1C</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_14a8cdf9-77a9-465e-9e22-4ae66e1f02cb">Contains the Data Element Tag of the Functional Group Sequence that contains the Attribute that is referenced by the Dimension Index Pointer (0020,9165).</para>
                  <para xml:id="para_818cffac-dbad-40a3-97f3-1abacc220e0d">See <xref linkend="sect_C.7.6.17.1" xrefstyle="select: label"/> for further explanation.</para>
                  <para xml:id="para_04fe7d57-e76e-46ce-b12a-fdf6449feb20">Required if the value of Dimension Index Pointer (0020,9165) is the Data Element Tag of an Attribute that is contained within a Functional Group Sequence.</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_c0e10b3a-cf37-4c42-88de-e4ebb85987ec">&gt;Functional Group Private Creator</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_423d50d4-1f59-47fa-857d-0b5b0425acd0">(0020,9238)</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_71d18b2c-1a59-4afd-a14a-ad22eeea0421">1C</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_09437a5e-a485-4ecf-bbdc-2c68bbad8042">Identification of the creator of a group of Private Data Elements.</para>
                  <para xml:id="para_9d2bb1f6-3bc2-4e82-bcd8-40beb1d76176">Required if the Functional Group Pointer 0020,9167) value is the Data Element Tag of a Private Attribute.</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_fa4969a5-2163-47a4-a169-0377fe85ea96">&gt;Dimension Organization UID</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_ef4b861b-afc0-4d45-b71d-9b2fc74d1fbc">(0020,9164)</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_65124c34-e236-4138-b300-b5f73dccffaa">1</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_a9bab11b-8ac5-41d7-a83f-d852e3969515">Uniquely identifies a set of dimensions referenced within the containing SOP Instance. In particular the dimension described by this Sequence Item is associated with this Dimension Organization UID. See <xref linkend="sect_C.7.6.17.2" xrefstyle="select: label"/> for further explanation.</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_1a297275-f51b-4342-83d1-24279a1150c4">&gt;Dimension Description Label</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_0dc4acfc-565d-41a6-b98b-c3802281c16b">(0020,9421)</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_85bbb392-4042-4268-96f3-ae6bbf9ef0f8">3</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_657bf5db-8f6f-420a-a518-ba7c6f980061">Free text description that explains the meaning of the dimension.</para>
                </td>
              </tr>
            </tbody>
          </table>
          <section label="C.7.6.17.1" status="5" xml:id="sect_C.7.6.17.1">
            <title>Dimension Indices</title>
            <para xml:id="para_ec48b38c-6499-499c-8b5f-415aee99911b">With the Dimension Index Sequence (0020,9222), Data Element Tags are specified that identify the indices used for a particular SOP Instance.</para>
            <para xml:id="para_344a2756-2bbc-4ab9-8afd-036753ea8c71">The actual index values for each frame in a multi-frame header are stored in a single Dimension Index Values Attribute (0020,9157) defined in the Frame Content Functional Group. For each SOP Instance this Attribute has a Value Multiplicity equal to the number of Items in the Sequence. The ordering of the Items in the Sequence defines the ordering in the Dimension Index Values Attribute: Item 1 of the Sequence relates to Value 1, Item 2 to Value 2, etc.</para>
            <para xml:id="para_624da344-1416-423e-a049-3f9ab8d250b3">The Dimension Index Pointer (0020,9165) references a single Attribute that describes the actual values that define the dimension. Each Attribute referenced in the Dimension Index Sequence (0020,9222) will have an index stored in the Dimension Index Values (0020,9157) for each frame. Each index value is an ordinal number starting from 1 and monotonically increasing by 1 within the scope of a Dimension Organization UID (0020,9164). These values are independent of the actual values of the Attribute referenced by the Dimension Index Pointer (0020,9165), i.e., the index values are logical indices, rather than actual indices. Frames assigned the same index shall contain nominally the same value for the referenced Attribute. If the referenced Attribute is not present for some frames, or is present but has no value, then a single index shall be assigned to indicate the lack of the value (i.e., all such frames shall have the same index value, which is different from other
            index values). It is at the discretion of the SOP Instance creator whether the Attribute values are equivalent, and therefore appropriate for assignment to the same index value.</para>
            <para xml:id="para_0a6e280e-3d31-42e8-8100-a052c408c655">...</para>
          </section>
          <section label="C.7.6.17.2" status="5" xml:id="sect_C.7.6.17.2">
            <title>Dimension Organization UID</title>
            <para xml:id="para_7b9d94c1-12d5-4f76-b79a-d6f56912270a">The Dimension Organization UID (0020,9164) value identifies a set of dimensions to which an Item of the Dimension Index Sequence (0020,9222) belongs.</para>
            <para xml:id="para_ee63054c-dfdf-476b-b31b-2ffbb0f4064f">When different SOP Instances share the same Dimension Organization UID (0020,9164) for a particular Item of the Dimension Index Sequence (0020,9222), equivalent indices from the corresponding Dimension Index Values (0020,9157) shall have the same meaning across the SOP Instances.</para>
            <para xml:id="para_1a485adb-d2f7-4c94-868c-2ace5d700dd6">This mechanism allows an image creator to explicitly specify that indices are intended to convey identical information across SOP Instances.</para>
            <para xml:id="para_035680e4-5270-4c72-8603-741f6c70343a">...</para>
          </section>
          <section label="C.7.6.17.3" status="5" xml:id="sect_C.7.6.17.3">
            <title>Spatial Location and Optical Path of Tiled Images</title>
            <para xml:id="para_81c46a69-b4dc-428d-822c-2efeee711150">If Dimension Organization Type (0020,9311) is present with a value of TILED_FULL, then the Per-Frame Functional Group Macros that would otherwise describe the spatial location of each tile explicitly (e.g., the X, Y and Z offsets from the origin in the Slide Coordinate System Plane Position (Slide)), and the optical path, may be omitted.</para>
            <para xml:id="para_08fcce55-df5f-49f2-b8a2-978d0468872c">A value of TILED_FULL indicates that the frames across all Instances of a Concatentation, or a single Instance in the absence of a Concatenation, comprise a non-sparse non-overlapping representation of an entire rectangular region, and are sequentially encoded as successive frames in Pixel Data (7FE0,0010) in an implicit order varying:</para>
            <itemizedlist>
              <listitem>
                <para xml:id="para_f0dd20ae-8ade-46f2-ad35-2c50add983d5">first along the row direction from left to right, where the row direction is defined in the Slide Coordinate System by the first three values of Image Orientation (Slide) (0048,0102),</para>
              </listitem>
              <listitem>
                <para xml:id="para_bd60c273-e1a2-4c77-89f4-538768f79537">then along the column direction from top to bottom, where the column direction is defined in the Slide Coordinate System by the second three values of Image Orientation (Slide) (0048,0102),</para>
              </listitem>
              <listitem>
                <para xml:id="para_0ae08229-6360-49ad-a748-9e855f00a350">then along the depth direction from the glass slide towards the coverslip, where the depth direction is defined in the Slide Coordinate System from zero to positive,</para>
              </listitem>
              <listitem>
                <para xml:id="para_e553a9fe-2e1b-42ed-aa5f-a27f3c7c00e6">then along optical paths, where the direction is defined by successive Items of the Optical Path Sequence (0048,0105) in the order in which they are listed in that Sequence.</para>
              </listitem>
            </itemizedlist>
            <para xml:id="para_6d73468e-28c6-4aee-a708-dab0b3a14702">If Dimension Organization Type (0020,9311) is absent or has a value of TILED_SPARSE, then the spatial location of each tile is explicitly encoded using information in the Per-Frame Functional Group Sequence, and the recipient shall not make any assumption about the spatial position or optical path or order of the encoded frames but shall rely on the values of the relevant Per-Frame Functional Group Macro.</para>
          </section>
        </section>
      
      
      <section label="C.7.6.16.2.2" status="6" xml:id="sect_C.7.6.16.2.2">
              <title>Frame Content Macro</title>
              <para xml:id="para_2797cdd6-abe6-4e70-a1f8-c1b965189d4c">
<xref linkend="table_C.7.6.16-3" xrefstyle="select: label"/> specifies the Attributes of the Frame Content Functional Group Macro.</para>
              <para xml:id="para_c05df9f4-9a80-4060-95e7-99b39556fd21">This Functional Group Macro may only be part of the Per-frame Functional Groups Sequence (5200,9230) Attribute.</para>
              <table frame="box" label="C.7.6.16-3" rules="all" xml:id="table_C.7.6.16-3">
                <caption>Frame Content Macro Attributes</caption>
                <thead>
                  <tr valign="top">
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_524ebe21-f60a-44d2-9648-e16379da4dce">Attribute Name</para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_7fd19267-d36c-4cf5-a83c-2c238303ce38">Tag</para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_f38a031a-bf90-45fb-86b0-4d9633671d73">Type</para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_0467d156-3cf3-43ae-a201-93f7efa5c0d2">Attribute Description</para>
                    </th>
                  </tr>
                </thead>
                <tbody>
                  <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_2cbd10d3-95f9-469c-b795-4aae3e62f9a3">Frame Content Sequence</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                      <para xml:id="para_11f48380-d331-45fe-9341-78ddd8617369">(0020,9111)</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                      <para xml:id="para_3e3dafc7-4788-43af-a588-eff012eb150d">1</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_47f02e7a-77c5-4bf3-9dd4-6176e50f2782">Identifies general characteristics of this frame.</para>
                      <para xml:id="para_8c19b021-9574-49de-9cea-6c45d8e229c5">Only a single Item shall be included in this Sequence.</para>
                    </td>
                  </tr>
                  <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_a24dce14-f92c-4ab9-b15d-88a91fff6f08">...</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                      <para xml:id="para_1430cd99-6dfd-4950-b2b2-ba93a5fa6a34">...</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                      <para xml:id="para_a98dbfa7-d0aa-4f20-b848-168fec91a874">...</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_3ba6b5df-b9f2-4670-9987-687ecfd20f96">...</para>
                    </td>
                  </tr>
                  <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_82637565-18d2-4ae4-8746-11f8aeee41f7">&gt;Dimension Index Values</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                      <para xml:id="para_96e49216-acde-4c00-b63d-07d6854f25ee">(0020,9157)</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                      <para xml:id="para_3b2b3bf8-156f-4459-8c4a-972e14064cc5">1C</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_1f6eaf2e-2d98-4745-be82-1a074d4ab857">Contains the values of the indices defined in the Dimension Index Sequence (0020,9222) for this multi-frame header frame. The number of values is equal to the number of Items of the Dimension Index Sequence and shall be applied in the same order.</para>
                      <para xml:id="para_e1f94cef-b79f-487e-953f-f96a851153bb">See <xref linkend="sect_C.7.6.17.1" xrefstyle="select: label"/> for a description.</para>
                      <note>
                        <para xml:id="para_6949247f-7226-4f36-b966-86c552fbf62e">In <xref linkend="sect_C.7.6.17.1" xrefstyle="select: label"/>, the index values are defined to start from 1 and monotonically increase by 1, within the scope of the Dimension Organization UID (0020,9164).</para>
                      </note>
                      <para xml:id="para_fcdaec69-08fc-476a-8e45-25af692b70f1">Required if the value of Dimension Index Sequence (0020,9222) exists.</para>
                        <note>
                            <para><emphasis role="bold"><emphasis role="underline">For some IODs, such as the VL Whole Slide Microscopy Image IOD, the entire Frame Content Sequence (0020,9111) may be omitted, but if it is present and Dimensions are explicitly defined, then the index values need to be supplied here.</emphasis></emphasis></para>
                        </note>
                    </td>
                  </tr>
                  <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_926d3d20-41aa-4c97-89f2-93aae3df47cd">...</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                      <para xml:id="para_68b9c4b4-5d0c-4596-8791-afd55631ab77">...</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                      <para xml:id="para_dc7381b7-ecbf-48cf-a9d4-44f133f8c84f">...</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_4bd97403-d7b0-4faa-9de7-ed5594914737">...</para>
                    </td>
                  </tr>
                </tbody>
              </table>
            </section>
      
      
        
      
    </chapter>
</book>